Literature DB >> 32278270

Portable pulsed xenon ultraviolet light disinfection in a teaching hospital animal laboratory in China.

Jing-Jing Li1, Shan-Ni Wang2, Jiao-Jiao Qiao2, Li-Hua Chen3, Yu Li4, Yong Wu3, Yan-Xia Ding2, Mei-Mei Wang2, Yun Tian2, Yun-Bo Liu5, Chen Yan6, Chen Zhang7, Chang-Qing Gao8.   

Abstract

An animal laboratory in a teaching hospital is a possible cause of cross infection. We aimed to assess the infection control in our animal laboratory and evaluate the disinfectant effects of a portable pulsed xenon ultraviolet (PX-UV) machine. Samples were taken from the surface of research tables, other high touch places, such as doorknobs, weighing scales, and handles of trolleys, and from air in the barrier system pre- and post-manual cleaning and post-PX-UV disinfection. The bacteria types were identified. We found that routine manual cleaning significantly reduced bacterial colony form unit (CFU)/cm2 (P = .02), and the median of CFU/cm2 reduced from 0.5 pre-cleaning to zero post-cleaning. PX-UV disinfection also significantly reduced residual bacterial counts (P = .002), with the highest counts 10 pre-PX-UV disinfection and 1 afterwards. Without manual cleaning, PX-UV disinfected surfaces significantly (P < .001), median count 6 pre-PX-UV disinfection and zero afterwards. PX-UV significantly reduced bacterial colony counts in the air with the median count falling from 6 to zero (P < .001). Some of the 21 species of pathogens we identified in the current study are pathogenic, resistant to antibiotics, and able to cause nosocomial infections and zoonosis. PX-UV reduced counts of most of the pathogens. PX-UV is an effective agent against these pathogens.
Copyright © 2020 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Animal laboratory; Disinfection; Pulsed xenon ultraviolet light

Mesh:

Substances:

Year:  2020        PMID: 32278270      PMCID: PMC7136868          DOI: 10.1016/j.jphotobiol.2020.111869

Source DB:  PubMed          Journal:  J Photochem Photobiol B        ISSN: 1011-1344            Impact factor:   6.252


Introductions

Medical researchers and life scientists are often involved in animal studies. In China, staff in medical schools and teaching hospitals are the main resources for medical and life sciences research. Hence, hospitals may have one or more animal laboratory. Our animal laboratory is on a hospital campus where the majority of investigators are Masters or PhD students from three affiliated Xiang Ya hospital, as well as from the medical school. As a result, there is a risk of cross infection by transmitting pathogens between a hospital clinic and the animal laboratory, and triggering a disaster like SARS, the severe acute respiratory syndrome that originated in China and spread to many countries in 2003 [1]. Pulsed xenon ultraviolet (PX-UV) is proven as an effective tool for disinfection of pathogens in various hospital settings, including wards [[2], [3], [4], [5], [6]], operating rooms [7], surgical sites [8,9], nursing rooms [10], human milk feeding rooms [11], and burns units [12]; and this disinfectant effect also applies to clinical laboratories and blood sampling rooms in China [13,14]. However, the effectiveness of PX-UV in disinfection of a hospital animal laboratory is unstudied. Hence, we aimed to evaluate the usefulness of PX-UV in this setting in China. Our animal laboratory is specific pathogen free (SPF), which means that animals kept in the barrier system should be free of particular pathogens, such as those able to cause infections and zoonoses, and those that will interfere with research. Pathogens from outside the barrier system are not allowed to be transferred in according to the regulations. Thus, we did not introduce any bacteria from outside the system, but measured the pathogens that exist in the two rooms in the animal laboratory that are used for surgical operations and administering test agents.

Methods and Materials

Sampling

Sampling Sites

This study was conducted in the animal laboratory on the campus of the Third Xiang Ya Hospital, which has 2200 beds. The size of each room is 7 m × 7 m. Research tables are placed along the wall as shown in Fig. 1 . We measured pathogens in the air and on the surface of the research tables, and other high touch places such as doorknobs, weighing scales, and handles of trolleys.
Fig. 1

The sampling sites for surfaces and air in the two rooms. Pathogens in the air and on the surface of the research tables, doorknobs, weighing scales, and handles of trolleys were measured.

The sampling sites for surfaces and air in the two rooms. Pathogens in the air and on the surface of the research tables, doorknobs, weighing scales, and handles of trolleys were measured. The sampling sites for surfaces and air are indicated in Fig. 1. In each room we sampled 7 surface sites and 5 air sites for pathogens, with each site one meter away from the wall. The tables are 0.9 m high.

Routine Manual Cleaning

At the end of each day, after removing all unnecessary items from the tables and the room, the laboratory staff cleans the research tables using rags wetted with 0.5% peracetic acid and the air in the room is disinfected by spraying 1% peracetic acid.

Sampling Time

Samples were taken at the end of the working day, at about 7 pm, when the facility was about to close. The disinfectant effect of PX-UV was evaluated in two ways. First, after routine manual cleaning of the research tables, samples were taken at three time points: pre- and post-manual cleaning of the surfaces as well as post-PX-UV disinfection. Second, an evaluation was performed without manual cleaning, for which samples were taken pre- and post-PX-UV disinfection. The comparison of colony counts pre- and post-manual cleaning and post-PX-UV on surfaces was repeated 4 times with 7 sampling sites; the comparison of colony counts pre- and post-PX-UV disinfection in the air was repeated 4 times with 10 sampling sites. The comparison of colony counts pre- and post-PX-UV without manual cleaning on surfaces was repeated twice with 7 sampling sites. We did not repeat the last comparison four times because the results were consistent with our other findings.

Sampling Methods

For air sampling, 64 cm2 tryptic soy agar plates were open for 30 mins to collect pathogens falling from air into the plates. The control plates, one in each room, were open and covered immediately. Other procedures were the same in both test and control plates. For surface sampling, a sterilized specification board with a 5 cm × 5 cm window (Nanjing Bizheng Biological Technology Co., Ltd.) was used to define the sampling area. A total of 100 cm2 were sampled for each table. For the weighing scales and doorknobs, the whole surface of the plate containing the materials for measuring and the whole doorknob were sampled. Sterilized sponges moistened with saline were used to scrub the surface up and down five times in the specification window. The sponge was then placed in a 50 mL conical tube containing 10 mL sterile saline for testing. The tubes were then placed in a mixer for two minutes, from which 100 μL was cultured for 48 h at 36 ± 1 °C. Sterilized gloves were worn to prevent contamination during sampling.

Culture Count Results

The bacterial colonies in each plate were counted and the types of bacteria were classified with matrix-assisted laser desorption/ionization time of flight mass spectrometry (MicroflexLT/SH. BRUKER, Germany). For the surface study, the colony form unit (CFU)/cm2 is the number of colonies counted in each plate. For the air pathogen study, results were reported as total CFU per plate. When CFU was over 100 per plate, results were recorded as 100.

Device

The PX-UV machine MX-3600 (Xi'An Fukang Air Purification Equipment and Engineering Co. Ltd., Xi'an, China) was initially assessed for its efficacy and we found that it was set at a wavelength of 100 nm to 400 nm, a frequency of 3 Hz, and a duration of work of 6 mins. The disinfectant efficacy of the machine was 100% for Staphylococcus aureus and Escherichia coli. Thus, these parameters were used for this study. The machine was deployed on the floor close to the table in the center of the room. When working, the top of the machine's UV light lamp is 1.1 m high, well above the table surfaces (0.9 m). Hence, all the table surfaces and air was exposed to UV light.

Data Analysis

Data were analyzed by performing a Wilcoxon rank sum test, using a software program called MedCalc, Version 17 (MedCalc Software, Ostend, Belgium), to compare the difference between the data pre- and post-disinfection. Statistical significance was set at P < .05.

Results

Effect of Manual Cleaning and PX-UV Disinfection on Surfaces

As demonstrated in Table 1A and Fig. 2A, routine manual cleaning significantly reduced bacterial colony counts (P = .02). The median of the colony count dropped from 0.5 before cleaning to zero after cleaning. PX-UV disinfection significantly reduced residual bacterial counts (P = .002), with the highest count being 10 before PX-UV disinfection and 1 afterwards (Table 1B and Fig. 2B). PX-UV disinfects surfaces without manual cleaning (P < .001), with a median count of 6 before PX-UV disinfection reducing to zero afterwards (Table 1C and Fig. 2 C).
Table 1

Comparison of colony counts pre- and post-manual cleaning and post-PX-UV disinfection on surfaces and air.

A
Pre- manual cleaning
Post- manual cleaning

No. samplesMedian (95% CV)Lowest - highest valuesNo. samplesMedian (95% CV)Lowest - highest valuesP
560.5 (0.00–0.69)0.00–100.00560 (0.00–1.00)0.00–10.000.02



Fig. 2

Effect of manual cleaning and pulsed xenon ultraviolet light (PX-UV) on surfaces and air. Routine manual cleaning significantly reduces bacterial colony counts (CFU) (P = .02), with the median of the CFU reduced from 0.5 before cleaning to zero after cleaning (A). PX-UV disinfection significantly reduces residual bacterial counts (P = .002), with the highest count being 10 before PX-UV disinfection and 1 afterwards (B). PX-UV disinfects the surface without manual cleaning (P < .001), with the median count of 6 before PX-UV disinfection reducing to zero afterwards (C). PX-UV disinfection significantly reduces bacterial colony counts in the air, with a median count 6 reducing to zero (P < .001) (D).

Comparison of colony counts pre- and post-manual cleaning and post-PX-UV disinfection on surfaces and air. Effect of manual cleaning and pulsed xenon ultraviolet light (PX-UV) on surfaces and air. Routine manual cleaning significantly reduces bacterial colony counts (CFU) (P = .02), with the median of the CFU reduced from 0.5 before cleaning to zero after cleaning (A). PX-UV disinfection significantly reduces residual bacterial counts (P = .002), with the highest count being 10 before PX-UV disinfection and 1 afterwards (B). PX-UV disinfects the surface without manual cleaning (P < .001), with the median count of 6 before PX-UV disinfection reducing to zero afterwards (C). PX-UV disinfection significantly reduces bacterial colony counts in the air, with a median count 6 reducing to zero (P < .001) (D).

Effect of PX-UV Disinfection on the Air

As shown in Table 1D and Fig. 2D, PX-UV disinfection significantly reduced bacterial colony counts in the air, with a median count 6 reducing to zero (P < .001). All controls were bacteria free.

Pathogens Identified

As shown in Table 2 , 16 species of pathogens were identified from the surfaces of the research tables, weighing scales, handles of trolleys, and doorknobs. Ten species decreased, one remained unchanged, and five increased after manual cleaning. Due to technical reasons, pathogen identification was not performed after the first test when only one colony was found in each of the air and surface cultures after PX-UV disinfection. In the other tests, no pathogen was detected after PX-UV disinfection.
Table 2

Comparison of the colony counts and bacteria kinds before and after annual cleaning and after PX-UV disinfection on surfaces.

Name of the bacteria identified on surface of the table, balance, handle of trolley and doorknobStainingCounts identified before manual cleaningCounts identified after manual cleaningCounts identified after PX-UVCharacteristics
Aerococcus viridansG+300Generally acquired in hospital environment and pathogenic, can infect newborns[21,22]
Bacillus flexusG+200
Brevibacillus centrosporusG+400
Brevibacillus centrosporusG+020
Comamonas kerstersiiG−400
Corynebacterium glutamicumG+200
Corynebacterium stationisG+100
Jeotgalicoccus halotoleransG+010
Lactobacillus pantherisG+200
Lactobacillus paracasei ssp. paracaseiG+110
Proteus mirabilisG−200Causing urinary tract infection, nephrolith and cystic calculus, sepsis[16,17]
Proteus vulgarisG−0100
Staphylococcus nepalensisG+3400Zoonotic potential[18]
Staphylococcus sciuri ssp. sciuriG+150pathogenic[19]
Staphylococcus succinus ssp. succinusG+0170
Staphylococcus xylosusG+600Pathogenic and resistance to different kinds of antibiotics[20]
Comparison of the colony counts and bacteria kinds before and after annual cleaning and after PX-UV disinfection on surfaces. Table 3 shows that 11 species of bacteria were identified in the air. Four decreased and six increased after manual cleaning. As was the case for the surface study, PX-UV disinfected most of the pathogens. Among these 11 species of pathogens, six were the same pathogens as found on the surfaces, and five were different.
Table 3

Comparison of the colony counts and bacteria kinds before and after annual cleaning and after PX-UV disinfection in air.

Name of the bacteria identified in the airStainingCounts identified before manual cleaningCounts identified after manual cleaningCounts identified after PX-UVCharacteristics
Brevibacillus centrosporusG+020
Flavobacterium gelidilacusG−100
Hydrogenophaga pseudoflavaG−030
Lactobacillus brevisG+200
Lactobacillus paracasei ssp. paracaseiG+3100
Microbacterium lacticumG+100
Proteus mirabilisG−060Causing urinary tract infection, nephrolith and cystic calculus, sepsis[16,17]
Staphylococcus nepalensisG+420Zoonotic potential[18]
Staphylococcus sciuri ssp. camaticusG+010
Staphylococcus sciuri ssp. sciuriG+0110Pathogenic and resistance to different kinds of antibiotics[19]
Staphylococcus xylosusG+360Weak pathogenic but has strong resistance to antibiotics
Comparison of the colony counts and bacteria kinds before and after annual cleaning and after PX-UV disinfection in air.

Discussion

In this study, we found that routine manual cleaning significantly reduces bacterial colony counts, but bacteria remain. PX-UV effectively disinfects the surfaces of research tables, weighing scales, doorknobs, handles of trolleys, and simultaneously the air in the room, with or without manual cleaning in a short time. Some of the bacteria identified in the barrier system are pathogenic and are able to cause nosocomial infection. Careful cleaning and disinfection of environmental surfaces in hospital is essential for effective infection control. An animal laboratory should be cleaned and disinfected regularly. Traditionally, the surfaces and the air of an animal laboratory are cleaned and disinfected by spraying peracetic acid or hydrogen peroxide, or by ultraviolet radiation. These methods, however, have intrinsic shortcomings. Spraying peracetic acid or hydrogen peroxide may corrode the surfaces of the racks of the individual ventilation cages (IVC) system, and those of the tables, which are often made of stainless steel. The detergents may also affect animals if the IVC, the chamber for raising animals, draws air from inside the room. The effectiveness of manual cleaning and disinfection is variable [5,7]. Inherent human error often results in incomplete cleaning, as demonstrated by our study. Continuous ultraviolet radiation takes about 30 min to disinfect target surfaces and air [15]. But this is too long when the surfaces and air must be disinfected quickly. In such cases, PX-UV disinfection has its advantages as shown in our study, with its fast (achieving satisfactory results in 6 min) and effective disinfection. A number of pathogens can infect both animals and humans causing zoonosis, such as plague, Lyme disease, rabies, anthrax, tuberculosis, and epidemic hemorrhagic fever. In recent years, emerging zoonoses such as SARS, Ebola, and COVID-2019 have received worldwide attention because of major outbreaks [1,16,17]. Minor outbreaks of emerging and re-emerging zoonoses occur frequently in China [18]. Zoonotic outbreaks occurred in students and teachers because of using infected animals during teaching [19]. In our study, we found that several types of infectious bacteria exist on the surfaces and air inside the barrier system. Among them, proteus mirabilis can cause urinary tract infection, nephrolith and cystic calculus, sepsis [20,21]; staphylococcus nepalensis has zoonotic potential [22]; staphylococcus sciuri ssp sciuri [23], and staphylococcus xylosus [24] are pathogenic and resistant to different antibiotics; and aerococcus viridans can cause serious infection [25,26]. The existence of these bacteria in the barrier system suggests a risk of cross infection with the hospital environment. Strict control of the transmission of infectious pathogens in an animal laboratory is important, not only for the health of laboratory animals [27], which helps achieve reliable and reproducible research results, but also for the control of nosocomial infection. PX-UV disinfection is an optimal choice to achieve this with its advantages of fast disinfection and satisfactory efficacy without damaging equipment and apparatus. In our study, the discrepancy between the bacterial species on the surfaces and in the air, as well as the increase in bacteria counts and species after manual cleaning, is possibly caused by contamination from the cloth used by the cleaner, or contamination from researchers as some students are still working while the cleaner is cleaning and sampling. As we only identified a few bacteria in each plate, we did not test their resistance to antibiotics. In the barrier system of an animal laboratory, bacteria are being exchanged with the hospital environment. The bacteria are pathogenic, resistant to antibiotics, and have potential to cause cross-infection, resulting in nosocomial infections and zoonosis. PX-UV disinfection is an effective agent to disinfect the system.

Author Contributions

C.Q.G., Y.B.L., C.Y., and C.Z., conceptualized the study and wrote the manuscript. J.J.L, S.N. W., L.H.C., Y.L., Y.W., Y.X.D, M.M.W., J.J.Q. and Y.T. conducted the experiments. All authors reviewed the manuscript.

Ethics Approval

Not required.

Declaration of Competing Interest

None.
  26 in total

1.  Evaluation of a pulsed xenon ultraviolet light device for reduction of pathogens with biofilm-forming ability and impact on environmental bioburden in clinical laboratories.

Authors:  Li-Hua Chen; Yu Li; Yong Qi; Shan-Ni Wang; Chang-Qing Gao; Yong Wu
Journal:  Photodiagnosis Photodyn Ther       Date:  2019-08-24       Impact factor: 3.631

2.  Survey of multidrug resistance integrative mobilizable elements SGI1 and PGI1 in Proteus mirabilis in humans and dogs in France, 2010-13.

Authors:  Eliette Schultz; Marisa Haenni; Laurent Mereghetti; Eliane Siebor; Catherine Neuwirth; Jean-Yves Madec; Axel Cloeckaert; Benoît Doublet
Journal:  J Antimicrob Chemother       Date:  2015-06-11       Impact factor: 5.790

3.  Emergence of Zaire Ebola virus disease in Guinea.

Authors:  Sylvain Baize; Delphine Pannetier; Lisa Oestereich; Toni Rieger; Lamine Koivogui; N'Faly Magassouba; Barrè Soropogui; Mamadou Saliou Sow; Sakoba Keïta; Hilde De Clerck; Amanda Tiffany; Gemma Dominguez; Mathieu Loua; Alexis Traoré; Moussa Kolié; Emmanuel Roland Malano; Emmanuel Heleze; Anne Bocquin; Stephane Mély; Hervé Raoul; Valérie Caro; Dániel Cadar; Martin Gabriel; Meike Pahlmann; Dennis Tappe; Jonas Schmidt-Chanasit; Benido Impouma; Abdoul Karim Diallo; Pierre Formenty; Michel Van Herp; Stephan Günther
Journal:  N Engl J Med       Date:  2014-04-16       Impact factor: 91.245

4.  Environmental effectiveness of pulsed-xenon light in the operating room.

Authors:  Sarah Simmons; Charles Dale; James Holt; Deborah G Passey; Mark Stibich
Journal:  Am J Infect Control       Date:  2018-04-13       Impact factor: 2.918

5.  The oral microbiota of domestic cats harbors a wide variety of Staphylococcus species with zoonotic potential.

Authors:  Ciro César Rossi; Ingrid da Silva Dias; Igor Mansur Muniz; Walter Lilenbaum; Marcia Giambiagi-deMarval
Journal:  Vet Microbiol       Date:  2017-01-25       Impact factor: 3.293

Review 6.  Microbiota and reproducibility of rodent models.

Authors:  Craig L Franklin; Aaron C Ericsson
Journal:  Lab Anim (NY)       Date:  2017-03-22       Impact factor: 12.625

7.  Is the pulsed xenon ultraviolet light no-touch disinfection system effective on methicillin-resistant Staphylococcus aureus in the absence of manual cleaning?

Authors:  Chetan Jinadatha; Frank C Villamaria; Marcos I Restrepo; Nagaraja Ganachari-Mallappa; I-Chia Liao; Eileen M Stock; Laurel A Copeland; John E Zeber
Journal:  Am J Infect Control       Date:  2015-05-23       Impact factor: 2.918

8.  Pulsed xenon ultraviolet and non-thermal atmospheric plasma treatments are effective for the disinfection of air in hospital blood sampling rooms.

Authors:  Shan-Ni Wang; Jing-Jing Li; Ying-Xin Liu; Zhi Lin; Jiao-Jiao Qiao; Li-Hua Chen; Yu Li; Yong Wu; Mei-Mei Wang; Yun-Bo Liu; Chen Yan; Zhi-Heng Chen; Chang-Qing Gao
Journal:  Photodiagnosis Photodyn Ther       Date:  2019-05-29       Impact factor: 3.631

Review 9.  Major emerging and re-emerging zoonoses in China: a matter of global health and socioeconomic development for 1.3 billion.

Authors:  Quan Liu; Lili Cao; Xing-Quan Zhu
Journal:  Int J Infect Dis       Date:  2014-05-21       Impact factor: 3.623

Review 10.  A review of studies on animal reservoirs of the SARS coronavirus.

Authors:  Zhengli Shi; Zhihong Hu
Journal:  Virus Res       Date:  2007-04-23       Impact factor: 3.303

View more
  6 in total

1.  Comparison of the Antimicrobial Effect of Zataria Multiflora Essence and Deconex Surface on Microbial Load of Emergency Ambulances.

Authors:  Ali Jadidi; Behrooz Irannejad; Maryam Mohsenifard; Ali Khanmohamadi Hezave; Seyed Mohamad Aghaepour
Journal:  Inquiry       Date:  2022 Jan-Dec       Impact factor: 2.099

2.  A Practical Assessment of the Disinfectant Efficacy of UV Light with and without Ozone Using a Novel Transfer Hatch in a Research Animal Facility.

Authors:  Jiao-Jiao Qiao; Jing-Jing Li; Chun-Hui Li; Yong Qi; Li-Yu Chen; Shan-Ni Wang; Paul E Honess; Yun-Bo Liu; Chen Zhang; Qing-Xia Liu; Bin Yi; Chang-Qing Gao
Journal:  J Am Assoc Lab Anim Sci       Date:  2022-04-07       Impact factor: 1.706

Review 3.  Comprehensive review of guidelines to practice prosthodontic and implant procedures during COVID-19 pandemic.

Authors:  Gunjan Pruthi; Hari Parkash; Vijaya Bharathi P; Radhika Jain; Arpit Gupta; Sachin Rai
Journal:  J Oral Biol Craniofac Res       Date:  2020-10-17

Review 4.  Hospital surface disinfection using ultraviolet germicidal irradiation technology: A review.

Authors:  Robert Scott; Lovleen Tina Joshi; Conor McGinn
Journal:  Healthc Technol Lett       Date:  2022-05-28

Review 5.  Potential of pulsed light technology for control of SARS-CoV-2 in hospital environments.

Authors:  Julie Jean; María Isabel Rodríguez-López; Eric Jubinville; Estrella Núñez-Delicado; Vicente M Gómez-López
Journal:  J Photochem Photobiol B       Date:  2020-12-28       Impact factor: 6.252

Review 6.  Environmental Disinfection of a Dental Clinic during the Covid-19 Pandemic: A Narrative Insight.

Authors:  Antonio Scarano; Francesco Inchingolo; Felice Lorusso
Journal:  Biomed Res Int       Date:  2020-10-28       Impact factor: 3.411

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.