Hossein Aslian1, Tomas Kron2, Troy Watts3, Cagla Akalanli2, Nicholas Hardcastle2, Peta Lonski2, Atousa Montaseri2, Barry Hay2, James Korte2, Kemal Berk2, Francesco Longo1,4, Mara Severgnini5.
Abstract
PURPOSE: This study focused on determining risks from stereotactic radiotherapy using flattening filter-free (FFF) beams for patients with cardiac implantable electronic device (CIEDs). Two strategies were employed: a) a retrospective analysis of patients with CIEDs who underwent stereotactic radiosurgery (SRS)/SBRT at the Peter MacCallum Cancer Centre between 2014 and 2018 and b) an experimental study on the impact of FFF beams on CIEDs.Entities:
Keywords: cardiac implantable electronic device (CIED); flattening filter-free (FFF) beams; stereotactic body radiotherapy (SBRT); volumetric-modulated arc therapy (VMAT)
Mesh:
Year: 2020 PMID: 32277741 PMCID: PMC7324704 DOI: 10.1002/acm2.12873
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Fig. 1The isolation circuitry between the load and the oscilloscope was used to monitor the CIED during irradiation. A star load configuration with the electrodes, except the case, was tied to a common node through 250‐Ω resistors. The common node is tied back to the case via a 10‐Ω resistor.
Fig. 2(Top) The experimental setup; (bottom left) The solid slab phantom; (bottom right) top view of the acrylic slab designed to position the CIEDs into six specific holes. Four holes were designed to be 3 cm away from the planning target volume (PTV), and two holes were designed to be placed partially inside the PTV.
Pulse repetition frequency (Hz) for dose rate settings in TrueBeam in FFF mode.
| Energy | Repetition rate (MU/min) | Pulse repetition frequency (Hz) |
|---|---|---|
| 6 MV | 1400 | 360 |
| 1200 | 310 | |
| 1000 | 260 | |
| 800 | 206/207 | |
| 600 | 155 | |
| 400 | 206/207 | |
| 10 MV | 2400 | 360 |
| 2000 | 300 | |
| 1600 | 240 | |
| 1200 | 180 | |
| 800 | 120 | |
| 400 | 120 |
MU/min, Monitor unit per minute; Hz, Hertz.
Rep‐rate tests (RRTs).
| RRT #1 | RRT #2 | |
|---|---|---|
| Site | – | – |
| Delivery technique | Open field irradiation at isocenter | Open field irradiation at isocenter |
| Energy & Beam mode | 6 MV‐FFF | 10 MV‐FFF |
| Dose/Fx | From 0.5 Gy up to 1.5 Gy | From 0.5 Gy up to 2 Gy |
| The maximum rep‐rate |
600 MU/min 800 MU/min 1000 MU/min 1200 MU/min 1400 MU/min |
400 MU/min 800 MU/min 1200 MU/min 1600 MU/min 2000 MU/min 2400 MU/min |
| Position of integrated circuits to PTV |
Inside |
Inside |
| CIED # |
4 CIEDs (Inside)
ICD (Medtronic)
ICD (St. Jude Medical)
CRT‐D (St. Jude Medical)
CRT‐D (Medtronic) |
4 CIEDs (Inside)
ICD (St. Jude Medical)
ICD (Medtronic)
CRT‐D (St. Jude Medical)
CRT‐D (Medtronic) |
CIED, Cardiac implantable electronic device; CRT‐D, Cardiac resynchronization therapy defibrillator; FFF beam, Flattening‑filter‑free beam; ICD, Implantable cardioverter defibrillator; MU/min, Monitor unit per minute; PTV, Planning target volume; RRT, Repetition rate test.
Clinical treatment plans (CTPs).
| CTP #1 | CTP #2 | CTP #3 | CTP #4 | |
|---|---|---|---|---|
| Site | Lung and Chest | Lung and Chest | Lung and Chest | Lung and Chest |
| Delivery technique | VMAT‐SBRT | VMAT‐SBRT | VMAT‐ SBRT | 3DCRT‐ SBRT |
| Energy & Beam mode | 10 MV‐FFF | 6 MV‐FFF | 10 MV‐FFF | 6 MV‐FFF |
| Dose/Fx | 28 Gy/1 Fx | 28 Gy/1 Fx | 48 Gy/4 Fx | 48 Gy/4 Fx |
| The maximum rep‐rate | 2400 MU/min | 1400 MU/min | 2400 MU/min | 1400 MU/min |
| Position of integrated circuits to PTV |
Partially inside Outside: 3 cm away |
Partially inside Outside: 3 cm away |
Partially inside Outside: 3 cm away |
Partially inside Outside: 3 cm away |
| CIED # |
2 CIEDs (Partially inside)
ICD (Medtronic)
ICD 4 CIEDs (Outside)
CRT‐D (Medtronic)
ICD (St. Jude Medical)
ICD (Medtronic
ICD (St. Jude Medical) |
2 CIEDs (Partially inside)
ICD (Medtronic)
ICD (St. Jude Medical)
3 CIEDs (Outside)
ICD (Medtronic)
ICD (St. Jude Medical)
CRT‐D (Medtronic) |
2 CIEDs (Partially inside)
ICD (St. Jude Medical)
ICD (Medtronic)
4 CIEDs (Outside)
ICD (St. Jude Medical)
ICD (Medtronic)
CRT‐D (St. Jude Medical)
ICD (Medtronic) |
2 CIEDs (Partially inside)
ICD (St. Jude Medical)
ICD Medtronic)
3 CIEDs (Outside)
ICD (St. Jude Medical)
ICD (Medtronic)
CRT‐D (St. Jude Medical) |
CTP Clinical treatment plan; VMAT‐SBRT, Volumetric‐modulated arc therapy‐Stereotactic body radiotherapy; 3DCRT‐SBRT, Three‐dimensional conformal radiation therapy‐ Stereotactic body radiotherapy; PTV, Planning target volume; MU/min, Monitor unit per minute; CIED, Cardiac implantable electronic device; ICD, Implantable cardioverter defibrillator; CRT‐D, Cardiac resynchronization therapy defibrillator.
Fig. 3A retrospective analysis of patients with CIEDs undergoing radiotherapy at PeterMac between 2014 and 2018; CF‐3DCRT: Conventionally fractionated three‐dimensional conformal radiotherapy, CF‐IMRT: Conventionally fractionated intensity modulated radiotherapy, CF‐VMAT: Conventionally fractionated volumetric modulated arc therapy, SBRT/SRS: Stereotactic body radiotherapy/Stereotactic radiosurgery, SXRT: Superficial x ray radiation therapy.
Summary of the clinical treatment plans and their related features.
| Characteristics | Study population | SBRT/SRS with FFF | SBRT/SRS with FF |
|---|---|---|---|
| No. | 28 | 13 | 15 |
| Patient age, median (range), y | 80 (57–93) | – | – |
| Type of CIED, No. | |||
| PM | 23 | 11 | 12 |
| ICD | 4 | 1 | 3 |
| CRT‐D | 1 | 1 | 0 |
| Site of irradiation, No. | |||
| Lung and chest | 17 | 11 | 5 |
| Brain | 2 | 0 | 2 |
| Liver | 2 | 1 | 1 |
| Kidney | 5 | 0 | 5 |
| Spine | 1 | 0 | 1 |
| Pelvis | 1 | 1 | 1 |
| Treatment plan specification photon energy | |||
| 6 MV | 17 | 2 | 15 |
| 10 MV | 11 | 11 | 0 |
| Number of fraction, median (range) | 3 (1–5) | – | – |
| Dose/fraction, median (range), Gy | 29 (28–54) | ||
| Rep‐rate (MU/Min) | |||
| 2400 (MU/Min) | 11 | 11 | 0 |
| 1400 (MU/Min) | 0 | 2 | 0 |
| 600 (MU/Min) | 15 | 0 | 15 |
| Distance between CIED and closest treatment field edge | |||
| less than 5 cm | 1 | 1 | 0 |
| 5–10 cm | 4 | 0 | 4 |
| More than 10 cm | 23 | 10 | 13 |
| SBRT delivery technique | |||
| VMAT | 11 | 4 | 7 |
| 3DCRT | 17 | 4 | 13 |
CIED, Cardiac implantable electronic device; PM, Pacemaker; ICD, Implantable cardioverter defibrillator; CRT‐D, Cardiac resynchronization therapy; SBRT, Stereotactic body radiotherapy; SRS, Stereotactic radiosurgery; FFF beam, Flattening‑filter‑free beam; FF beams flattened beams.
Fig. 4Histograms show estimated dose to CIEDs from FF and FFF beams. The median (range) calculated dose to CIEDs was 0.11 (0–1.5) Gy for FFF beams and was 0.22 (0.01–1.86) for FF beams.
Fig. 5Results of the rep‐rate test for 6‐MV FFF (up) and 10‐MV FFF (down) to determine when inappropriate sensing (either over‐sensing or inhibition) occurs.
Types of CIED errors.
| CIED # | CTP # | Position of electronic circuit of CIED to PTV | Inappropriate sensing during irradiation | Interrogation after irradiation | Total calculated dose to | |
|---|---|---|---|---|---|---|
|
the part of device inside the field including integrated circuits (Gy) |
the part of device outside the field (Gy) | |||||
| 9 | 1 | Partially inside | No | Reprogramming of pacing rate |
Dmax: 26.1 Dmean: 16.2 |
Dmax: 17.8 Dmean: 5.6 |
| 10 | 1 | Partially inside | Yes | Reprogramming of pacing rate |
Dmax: 28 Dmean: 16.9 |
Dmax: 18.7 Dmean: 2.8 |
| 20 | 3 | Partially inside | Yes | A decrease in pulse amplitude after delivering the last fraction |
Dmax: 46.4 Dmean: 28.8 |
Dmax: 30.7 Dmean: 9.6 |
| 21 | 3 | Partially inside | No | Reprogramming of ventricular pacing threshold |
Dmax: 42.7 Dmean: 28.4 |
Dmax: 29.2 Dmean: 4.8 |
|
Total calculated dose to the CIED (Gy) | ||||||
| 13 | 1 | Outside | Yes | Change in pacing mode |
Dmax: 0.89 Gy Dmean: 0.66 Gy | |
| 24 | 3 | Outside | No | Change in pacing mode |
Dmax: 0.96 Gy Dmean: 0.85 Gy | |
CTP, Clinical treatment plan; CIED, Cardiac implantable electronic device; PTV, Planning target volume; Dmax, Maximum dose; Dmean Mean dose.
Fig. 6The CIED electrograms demonstrates transient inappropriate sensing during the delivery of CTP #1.