Comparison of three different phantoms used for Winston-Lutz test with Artiscan software.

Hubert Szweda1, Kinga Graczyk2, Dawid Radomiak1, Krzysztof Matuszewski1, Bartosz Pawałowski1.   

Abstract

BACKGROUND: One of the most important test in every quality assurances process of medical linear accelerators is the Winston-Lutz test, allowing an evaluation of the treatment isocentre in the light of uncertainty of the position of the collimator, the gantry and the couch. AIM: The purpose of this work was analysis of the results of the Winston-Lutz test performed with three different phantoms for two different accelerators.
MATERIALS AND METHODS: Measurements were performed on two Varian machines: TrueBeam equipped with aS1200 EPID and TrueBeam equipped with aS1000 EPID. During the study three different phantoms dedicated for verification of the radiation isocentre were used: PTW Isoball, AQUILAB Isocentre Phantom and Varian Isocentre Cube. Analysis of the DICOM images was performed in Artiscan software.
RESULTS: For TrueBeam with as1200 EPID, gantry MV isocentre was about 0.18 mm larger for Varian Isocentre Cube than for two other phantoms used in this study. The largest variability of this parameter was observed for the couch. The results differed to 1.16 mm. For TrueBeam with as1000 EPID, results for collimator isocentre with PTW Isoball phantom were about 0.10 mm larger than for two other phantoms. For the gantry, results obtained with Varian Isocentre Cube were 0.21 mm larger.
CONCLUSION: The obtained results for all three phantoms are within the accepted tolerance range. The largest differences were observed for treatment couch, which may be related to the phantom mobility during couch movement.
© 2020 Published by Elsevier B.V. on behalf of Greater Poland Cancer Centre.

Entities:  

Keywords:  Artiscan; IGRT; QA; Winston-Lutz test

Year:  2020        PMID: 32214910      PMCID: PMC7083885          DOI: 10.1016/j.rpor.2020.03.003

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  10 in total

1.  Quality assurance for image-guided radiation therapy utilizing CT-based technologies: a report of the AAPM TG-179.

Authors:  Jean-Pierre Bissonnette; Peter A Balter; Lei Dong; Katja M Langen; D Michael Lovelock; Moyed Miften; Douglas J Moseley; Jean Pouliot; Jan-Jakob Sonke; Sua Yoo
Journal:  Med Phys       Date:  2012-04       Impact factor: 4.071

2.  Relations between dose cumulated in organs at risk and treatment based on different image-guidance strategies of cervical cancer.

Authors:  Agata Jodda; Tomasz Piotrowski; Bartosz Urbański; Julian Malicki
Journal:  Phys Med       Date:  2019-01-11       Impact factor: 2.685

3.  Evaluation of cone-beam computed tomography image quality assurance for Vero4DRT system.

Authors:  Hideharu Miura; Shuichi Ozawa; Masahiro Hayata; Shintarou Tsuda; Tsubasa Enosaki; Kiyoshi Yamada; Yasushi Nagata
Journal:  Rep Pract Oncol Radiother       Date:  2017-04-25

4.  An end-to-end postal audit test to examine the coincidence between the imaging isocenter and treatment beam isocenter of the IGRT linac system for Japan Clinical Oncology Group (JCOG) clinical trials.

Authors:  Yu Kumazaki; Shuichi Ozawa; Mitsuhiro Nakamura; Satoshi Kito; Toshiyuki Minemura; Hidenobu Tachibana; Teiji Nishio; Satoshi Ishikura; Yasumasa Nishimura
Journal:  Phys Med       Date:  2018-09-11       Impact factor: 2.685

5.  [Winston-lutz test and acquisition of flexmap using rotational irradiation].

Authors:  Toshikazu Imae; Akihiro Haga; Naoya Saotome; Shigeharu Takenaka; Yukari Okano; Katutake Sasaki; Makoto Nedu; Shigeki Saegusa; Takashi Shiraki; Keiichi Yano; Keiichi Nakagawa; Kuni Ohtomo
Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi       Date:  2014-04

Review 6.  Design principles and clinical possibilities with a new generation of radiation therapy equipment. A review.

Authors:  A Brahme
Journal:  Acta Oncol       Date:  1987       Impact factor: 4.089

7.  A closer look at the conventional Winston-Lutz test: Analysis in terms of dose.

Authors:  Juan-Francisco Calvo-Ortega; Sandra Moragues-Femenía; Coral Laosa-Bello; Sol San José-Maderuelo; Joan Casals-Farran
Journal:  Rep Pract Oncol Radiother       Date:  2019-07-18

8.  Evaluation of image-guidance strategies for prostate cancer.

Authors:  T Piotrowski; K Kaczmarek; T Bajon; A Ryczkowski; A Jodda; J Kaźmierska
Journal:  Technol Cancer Res Treat       Date:  2013-08-31

Review 9.  Isocenter verification for linac-based stereotactic radiation therapy: review of principles and techniques.

Authors:  Pejman Rowshanfarzad; Mahsheed Sabet; Daryl J O'Connor; Peter B Greer
Journal:  J Appl Clin Med Phys       Date:  2011-11-15       Impact factor: 2.102

10.  AAPM Medical Physics Practice Guideline 2.a: Commissioning and quality assurance of X-ray-based image-guided radiotherapy systems.

Authors:  Jonas D Fontenot; Hassaan Alkhatib; Jeffrey A Garrett; Andrew R Jensen; Steven P McCullough; Arthur J Olch; Brent C Parker; Ching-Chong J Yang; Lynne A Fairobent
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.