Jonas D Fontenot1, Hassaan Alkhatib, Jeffrey A Garrett, Andrew R Jensen, Steven P McCullough, Arthur J Olch, Brent C Parker, Ching-Chong J Yang, Lynne A Fairobent.
Abstract
Entities:
Mesh:
Year: 2014 PMID: 24423852 PMCID: PMC5711227 DOI: 10.1120/jacmp.v15i1.4528
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Recommended minimum practices for commissioning and QA of an IGRT system
|
| |
|---|---|
|
| |
| Customer acceptance procedures | |
| TPS integration | |
| OIS integration | |
| Establish routine QA baselines | |
| QA documentation | |
|
| |
|
|
|
|
| |
|
|
|
| Imaging‐treatment isocenter coincidence (SRS only) | 1 mm |
| Positioning/repositioning (SRS only) | 1 mm |
| Imaging‐treatment isocenter coincidence (SBRT only) | 2 mm |
| Positioning/repositioning (SBRT only) | 2 mm |
|
| |
| Imaging‐treatment isocenter coincidence (non‐SRS/SBRT) | 2 mm |
| Positioning/repositioning (non‐SRS/SBRT) | 2 mm |
|
| |
| Image scaling | 2 mm |
|
| |
| Imaging dose | |
| 2D MV |
|
| 2D kV (static imaging mode) |
|
| 2D kV (fluoroscopy mode) |
|
| All 3D imaging modes |
|
| Image quality | |
| 2D (spatial resolution, contrast) | Baseline value |
| 3D (uniformity, spatial resolution, contrast) | |
|
| |
| Verify / Reestablish QA baselines (as appropriate) | |
; .