Literature DB >> 30241749

An end-to-end postal audit test to examine the coincidence between the imaging isocenter and treatment beam isocenter of the IGRT linac system for Japan Clinical Oncology Group (JCOG) clinical trials.

Yu Kumazaki1, Shuichi Ozawa2, Mitsuhiro Nakamura3, Satoshi Kito4, Toshiyuki Minemura5, Hidenobu Tachibana6, Teiji Nishio7, Satoshi Ishikura8, Yasumasa Nishimura9.   

Abstract

PURPOSE: The aim of this study was to develop an end-to-end postal audit test to examine the coincidence between the imaging isocenter and treatment beam isocenter of the image guided radiotherapy (IGRT) linac system for Japan Clinical Oncology Group (JCOG) trials, as a part of IGRT credentialing of institutions participating in JCOG trials.
METHODS: We developed an end-to-end postal audit test to verify radiation positional errors associated with IGRT techniques. This test is intended for simulating a clinical IGRT flow and uses a static cubic phantom measuring 15 × 15 × 15 cm3 and weighing approximately 3.4 kg. The phantom has four gold fiducial markers and a spherical dummy target for setup, with known shift values from the phantom center. Two pairs of Gafchromic RTQA2 films were inserted 5 mm from the phantom's anterior-posterior and right-left surfaces. Radiation positional errors at the isocenter were determined by analyzing the center of the radiation field on the films and the known shift values of the dummy target. The test was performed on 47 IGRT devices at 35 institutions.
RESULTS: Radiation positional errors were within acceptance levels (1 mm/1°) for 42 IGRT devices (89.4%) in the first check. Median time to complete IGRT credentialing was 11.5 days. This audit method was applicable for any radiotherapy machine with an IGRT device.
CONCLUSIONS: A postal audit test to verify radiation positional errors for JCOG trials was successfully developed. In the postal audit, all but one institution passed this credentialing item within two trials.
Copyright © 2018 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Keywords:  Clinical trial; Credentialing; IGRT; Postal audit

Mesh:

Year:  2018        PMID: 30241749     DOI: 10.1016/j.ejmp.2018.08.010

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  2 in total

1.  An overview of the medical-physics-related verification system for radiotherapy multicenter clinical trials by the Medical Physics Working Group in the Japan Clinical Oncology Group-Radiation Therapy Study Group.

Authors:  Teiji Nishio; Mitsuhiro Nakamura; Hiroyuki Okamoto; Satoshi Kito; Toshiyuki Minemura; Shuichi Ozawa; Yu Kumazaki; Masayori Ishikawa; Naoki Tohyama; Masahiko Kurooka; Takeo Nakashima; Hidetoshi Shimizu; Ryusuke Suzuki; Satoshi Ishikura; Yasumasa Nishimura
Journal:  J Radiat Res       Date:  2020-11-16       Impact factor: 2.724

2.  Comparison of three different phantoms used for Winston-Lutz test with Artiscan software.

Authors:  Hubert Szweda; Kinga Graczyk; Dawid Radomiak; Krzysztof Matuszewski; Bartosz Pawałowski
Journal:  Rep Pract Oncol Radiother       Date:  2020-03-04
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.