Influence of cortical bone and implant design in the primary stability of dental implants measured by two different devices of resonance frequency analysis: An in vitro study.

David Chávarri-Prado1, Aritza Brizuela-Velasco2, Markel Diéguez-Pereira1, Esteban Pérez-Pevida2, Antonio Jiménez-Garrudo2, Iratxe Viteri-Agustín3, Alejandro Estrada-Martínez1, Oier Montalbán-Vadillo2.   

Abstract

BACKGROUND: This study aimed to evaluate the effect of the implant design and the presence of cortical bone in the primary stability, as well as analyze the differences between the stability measurements obtained by two different resonance frequency analysis (RFA) devices.
MATERIAL AND METHODS: A total of 80 Klockner implants of two different models [40 Essential Cone implants (group A) and 40 Vega implants (group B)] were used. The implants were placed in two polyurethane blocks that simulated the mechanical properties of the maxillary bone. One block featured a layer of cortical bone that was absent from the other block. The primary stability of all implants was measured by insertion torque and RFA using two different devices: Penguin RFA and Osstell IDX.
RESULTS: Primary stability was superior in the cortical bone in both torque and RFA. In the block containing cortical bone, group A implants obtained a greater insertion torque than did group B. The insertion torque was lesser in the bone lacking cortex. Regarding the ISQ of the implants, group A presented higher values in the block with cortical bone, but the values were lower in the block without cortical bone. There were no significant differences between the values obtained from the Osstell IDX and Penguin RFA.
CONCLUSIONS: The presence of cortical bone positively influences the primary stability of dental implants. The design of the implant also has a statistically significant influence on implant primary stability, although the impact depends on whether there is coronal cerclage or not. There were no statistically significant differences in the implant stability measurements obtained by two different devices. Key words:Implant stability, resonance frequency analysis, torque, osstell, penguin, cortical. Copyright:
© 2020 Medicina Oral S.L.

Entities:  

Year:  2020        PMID: 32190194      PMCID: PMC7071536          DOI: 10.4317/jced.56014

Source DB:  PubMed          Journal:  J Clin Exp Dent        ISSN: 1989-5488


  38 in total

1.  Differences in implant stability associated with various methods of preparation of the implant bed: an in vitro study.

Authors:  Su-Jin Ahn; Richard Leesungbok; Suk-Won Lee; Young-Ku Heo; Kyung Lhi Kang
Journal:  J Prosthet Dent       Date:  2012-06       Impact factor: 3.426

2.  Early detection of alterations in the resonance frequency assessment of oral implant stability on various bone types: a clinical study.

Authors:  Metin Sençimen; Aydın Gülses; Jülide Ozen; Cem Dergin; Kemal Murat Okçu; Simel Ayyıldız; Hasan Ayberk Altuğ
Journal:  J Oral Implantol       Date:  2010-03-16       Impact factor: 1.779

3.  Comparison of dental implant stabilities by impact response and resonance frequencies using artificial bone.

Authors:  Dae-Seung Kim; Woo-Jin Lee; Soon-Chul Choi; Sam-Sun Lee; Min-Suk Heo; Kyung-Hoe Huh; Tae-Il Kim; Won-Jin Yi
Journal:  Med Eng Phys       Date:  2013-12-28       Impact factor: 2.242

4.  Influence of factors related to implant stability detected by wireless resonance frequency analysis device.

Authors:  K Ohta; M Takechi; M Minami; H Shigeishi; M Hiraoka; M Nishimura; N Kamata
Journal:  J Oral Rehabil       Date:  2009-11-30       Impact factor: 3.837

5.  Relationship of three-dimensional bone-to-implant contact to primary implant stability and peri-implant bone strain in immediate loading: microcomputed tomographic and in vitro analyses.

Authors:  Jui-Ting Hsu; Heng-Li Huang; Chih-Han Chang; Ming-Tzu Tsai; Wei-Ching Hung; Lih-Jyh Fuh
Journal:  Int J Oral Maxillofac Implants       Date:  2013 Mar-Apr       Impact factor: 2.804

6.  Biomechanical evaluation of subcrestal dental implants with different bone anchorages.

Authors:  Bruno Salles Sotto-Maior; Camila de Andrade Lima; Plínio Mendes Senna; Germana de Villa Camargos; Altair Antoninha Del Bel Cury
Journal:  Braz Oral Res       Date:  2014-08-04

7.  Mechanical state assessment of the implant-bone continuum: a better understanding of the Periotest method.

Authors:  J Tricio; P Laohapand; D van Steenberghe; M Quirynen; I Naert
Journal:  Int J Oral Maxillofac Implants       Date:  1995 Jan-Feb       Impact factor: 2.804

8.  Osseointegrated titanium implants. Requirements for ensuring a long-lasting, direct bone-to-implant anchorage in man.

Authors:  T Albrektsson; P I Brånemark; H A Hansson; J Lindström
Journal:  Acta Orthop Scand       Date:  1981

9.  The effect of the thread depth on the mechanical properties of the dental implant.

Authors:  Sun-Young Lee; Sung-Jun Kim; Hyun-Wook An; Hyun-Seung Kim; Dong-Guk Ha; Kyung-Ho Ryo; Kwang-Bum Park
Journal:  J Adv Prosthodont       Date:  2015-04-23       Impact factor: 1.904

10.  The effects of bone density and crestal cortical bone thickness on micromotion and peri-implant bone strain distribution in an immediately loaded implant: a nonlinear finite element analysis.

Authors:  Tsutomu Sugiura; Kazuhiko Yamamoto; Satoshi Horita; Kazuhiro Murakami; Sadami Tsutsumi; Tadaaki Kirita
Journal:  J Periodontal Implant Sci       Date:  2016-06-28       Impact factor: 2.614

View more
  3 in total

1.  In vitro evaluation of the influence of bone cortical thickness on the primary stability of conventional- and short-sized implants.

Authors:  Luiz-Antônio-Borelli Barros; Caio-Fossalussa da Silva; Germana-de Villa Camargos; Elcio Marcantonio; Guilherme-José-Pimentel-Lopes de Oliveira; Luiz-Antônio-Borelli Barros-Filho
Journal:  J Clin Exp Dent       Date:  2022-02-01

2.  Effect of macro-design in the primary stability of short and extra-short implants using resonance frequency analysis. An ex vivo study.

Authors:  Ramón Silva; Pablo Villalón; Felipe Cáceres
Journal:  J Oral Biol Craniofac Res       Date:  2020-08-29

3.  Evaluation of the primary stability in dental implants placed in low density bone with a new drilling technique, Osseodensification: an in vitro study.

Authors:  J Barberá-Millán; C Larrazábal-Morón; J-J Enciso-Ripoll; E Pérez-Pevida; D Chávarri-Prado; M-D Gómez-Adrián
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2021-05-01
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.