Literature DB >> 7615316

Mechanical state assessment of the implant-bone continuum: a better understanding of the Periotest method.

J Tricio1, P Laohapand, D van Steenberghe, M Quirynen, I Naert.   

Abstract

The aim of this study was to obtain a better understanding of the Periotest method when used to detect subclinical mobility of osseointegrated implants. Four hundred two screw-shaped implants were tested with the Periotest device at the time of abutment connection. Several factors, including jaw location, implant and abutment length, and gender, were related to Periotest values (PTVs). Implants located in the anterior region of the mandible showed the lowest mean PTV (-3.2). The influence of abutment and implant length upon PTVs could be detected in the maxilla. In the mandible, only abutment length had influence on PTVs. Women showed higher mean Periotest scores in the maxilla compared with men. This difference was not found in the mandible. The Periotest method, its clinical limitations, and the meaning of a given PTV are also discussed.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7615316

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  7 in total

1.  Evaluating systemic administration effect of propranolol on osseointegration around titanium implants: A histomorphometric study in dogs.

Authors:  Mohammad Tavakoli; Mohammad Jafari Farshami; Nakisa Torabinia; Jaber Yaghini; Samaneh Shams
Journal:  Dent Res J (Isfahan)       Date:  2022-04-27

2.  The effect of various thread designs on the initial stability of taper implants.

Authors:  Ju-Hee Park; Young-Jun Lim; Myung-Joo Kim; Ho-Beom Kwon
Journal:  J Adv Prosthodont       Date:  2009-03-31       Impact factor: 1.904

3.  An alternative treatment for hopeless implants: Coronal resection.

Authors:  H Burak Kutlu; Samir Goyushov; Tolga F Tözüm
Journal:  J Oral Biol Craniofac Res       Date:  2020-03-13

4.  Influence of cortical bone and implant design in the primary stability of dental implants measured by two different devices of resonance frequency analysis: An in vitro study.

Authors:  David Chávarri-Prado; Aritza Brizuela-Velasco; Markel Diéguez-Pereira; Esteban Pérez-Pevida; Antonio Jiménez-Garrudo; Iratxe Viteri-Agustín; Alejandro Estrada-Martínez; Oier Montalbán-Vadillo
Journal:  J Clin Exp Dent       Date:  2020-03-01

5.  A retrospective study on related factors affecting the survival rate of dental implants.

Authors:  Hee-Won Jang; Jeong-Kyung Kang; Ki Lee; Yong-Sang Lee; Pil-Kyoo Park
Journal:  J Adv Prosthodont       Date:  2011-12-28       Impact factor: 1.904

6.  Implant Stability Changes for Nonsubmerged and Submerged Protocols for a Single Implant Mandibular Overdenture Using Ball Attachment.

Authors:  Ahmed Salah; Karim Foda; Mohamed Farouk Abdalla; Marwa Abdel Aal; Amr Naguib; Nouran Abdel Nabi
Journal:  Int J Dent       Date:  2021-09-16

7.  Factors Affecting the Survival Rate of Dental Implants: A Retrospective Study.

Authors:  Sonal Raikar; Pratim Talukdar; Sarala Kumari; Sangram Kumar Panda; Vinni Mary Oommen; Arvind Prasad
Journal:  J Int Soc Prev Community Dent       Date:  2017-12-29
  7 in total