Comparison of dental implant stabilities by impact response and resonance frequencies using artificial bone.

Dae-Seung Kim1, Woo-Jin Lee1, Soon-Chul Choi2, Sam-Sun Lee2, Min-Suk Heo2, Kyung-Hoe Huh2, Tae-Il Kim3, Won-Jin Yi4.   

Abstract

PURPOSE: We compared implant stability as determined by the peak frequency from the impact response with the implant stability quotient (ISQ) by resonance frequency analysis (RFA) in various artificial bone conditions. The clinical bone conditions were simulated using an artificial bone material with different cortical thicknesses and trabecular densities.
MATERIALS AND METHODS: The artificial bone material was solid, rigid polyurethane. The polyurethane foam of 0.8g/cm(3) density was used for the cortical bone layer, and that of 0.08, 0.16, 0.24, 0.32, and 0.48g/cm(3) densities for the trabecular bone layer. The cortical bone material of 4 different thicknesses (1.4, 1.6, 1.8, and 2.0mm) was attached to the trabecular bone with varying density. Two types of dental implants (10 and 13mm lengths of 4.0mm diameter) were placed into the artificial bone blocks. An inductive sensor was used to measure the vibration caused by tapping the adapter-implant assembly. The peak frequency of the power spectrum of the impact response was used as the criterion for implant stability. The ISQ value was also measured for the same conditions.
RESULTS: The stability, as measured by peak frequency (SPF) and ISQ value, increased as the trabecular density and the cortical density increased in linear regression analysis. The SPF and ISQ values were highly correlated with each other when the trabecular bone density and cortical bone thickness changed (Pearson correlation=0.90, p<0.01). The linear regression of the SPF with the cortical bone thickness showed higher goodness of fit (R(2) measure) than the ISQ value with the cortical bone thickness. The SPF could differentiate implantation conditions as many as the ISQ value when the trabecular bone density and the cortical density changed. However, the ISQ value was not consistent with the general stability tendency in some conditions.
CONCLUSION: The SPF showed better consistency and differentiability with implant stability than the ISQ value by resonance frequency analysis in the various implantation conditions.
Copyright © 2013 IPEM. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Impact response; Implant stability; Implant stability quotient; Peak frequency from power spectrum; Resonance frequency analysis

Mesh:

Substances:

Year:  2013        PMID: 24378382     DOI: 10.1016/j.medengphy.2013.12.004

Source DB:  PubMed          Journal:  Med Eng Phys        ISSN: 1350-4533            Impact factor:   2.242


  4 in total

1.  Analysis of the influence of implant shape on primary stability using the correlation of multiple methods.

Authors:  Mariana Lima da Costa Valente; Denise Tornavoi de Castro; Antonio Carlos Shimano; César Penazzo Lepri; Andréa Cândido dos Reis
Journal:  Clin Oral Investig       Date:  2015-02-14       Impact factor: 3.573

2.  Influence of cortical bone and implant design in the primary stability of dental implants measured by two different devices of resonance frequency analysis: An in vitro study.

Authors:  David Chávarri-Prado; Aritza Brizuela-Velasco; Markel Diéguez-Pereira; Esteban Pérez-Pevida; Antonio Jiménez-Garrudo; Iratxe Viteri-Agustín; Alejandro Estrada-Martínez; Oier Montalbán-Vadillo
Journal:  J Clin Exp Dent       Date:  2020-03-01

3.  Evaluation of the primary stability in dental implants placed in low density bone with a new drilling technique, Osseodensification: an in vitro study.

Authors:  J Barberá-Millán; C Larrazábal-Morón; J-J Enciso-Ripoll; E Pérez-Pevida; D Chávarri-Prado; M-D Gómez-Adrián
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2021-05-01

4.  The relationship between dental implant stability and trabecular bone structure using cone-beam computed tomography.

Authors:  Se-Ryong Kang; Sung-Chul Bok; Soon-Chul Choi; Sam-Sun Lee; Min-Suk Heo; Kyung-Hoe Huh; Tae-Il Kim; Won-Jin Yi
Journal:  J Periodontal Implant Sci       Date:  2016-04-26       Impact factor: 2.614

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.