| Literature DB >> 32102092 |
Xinzhou Li1,2, Luigi E Perotti1,3, Jessica A Martinez4, Sandra M Duarte-Vogel5, Daniel B Ennis1,2,4, Holden H Wu1,2.
Abstract
PURPOSE: Real-time magnetic resonance imaging (MRI) is a promising alternative to X-ray fluoroscopy for guiding cardiovascular catheterization procedures. Major challenges, however, include the lack of guidewires that are compatible with the MRI environment, not susceptible to radiofrequency-induced heating, and reliably visualized. Preclinical evaluation of new guidewire designs has been conducted at 1.5T. Here we further evaluate the safety (device heating), device visualization, and procedural feasibility of 3T MRI-guided cardiovascular catheterization using a novel MRI-visible glass-fiber epoxy-based guidewire in phantoms and porcine models.Entities:
Mesh:
Substances:
Year: 2020 PMID: 32102092 PMCID: PMC7043930 DOI: 10.1371/journal.pone.0229711
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Fig 6Examples of temperature recordings over time in phantom heating experiments.
Guidewire tip heating (GTH) for stainless steel guidewire only, glass-fiber epoxy-based guidewire only, and glass-fiber epoxy-based guidewire with braided catheter over time during the phantom heating experiments using the gradient echo (GRE, blue line) and turbo spin echo (TSE, red line) sequences. The guidewire tip was placed at position 1 to mimic the descending aorta (DA) location, position 2 to mimic the aortic arch (AA) location, and position 3 to mimic the left ventricle (LV) location. A clear increase in temperature is detected at the tip of the stainless-steel guidewire at positions 1 and 3 during imaging with the TSE sequence. The red zone represents a GTH greater than 2°C, which is not considered procedurally safe. Note that only positive temperature changes above baseline are shown since they correspond to the RF-induced heating.
Imaging parameters.
| Real-Time RF-spoiled GRE MRI Sequence at 3T | |||
|---|---|---|---|
| 5.8 ms– 6 ms / | 385 ms– 400 ms | ||
| 250 mm– 300 mm | 12° | ||
| 1.95 x 1.95 mm2–2.34 x 2.34 mm2 | 10 mm | ||
| 2 / 32 | 128 x 128 | ||
Real-time radiofrequency (RF) spoiled gradient echo (GRE) MRI sequence parameters at 3T. The field of view (FOV) was customized in each experiment to accommodate the subject’s anatomy and left ventricle/aortic arch sizes. TR: repetition time. TE: echo time.
Phantom heating evaluation results.
| Position 1 (DA) | 0.13 | 0.11 | 0.10 |
| Position 2 (AA) | 0.17 | 0.43 | 0.43 |
| Position 3 (LV) | 0.19 | 0.29 | 0.19 |
| Position 1 (DA) | 0.16 | 2.16 | 2.47 |
| Position 2 (AA) | 0.17 | 0.12 | 1.16 |
| Position 3 (LV) | 0.13 | 0.16 | 2.63 |
| Position 1 (DA) | 0.26 | 0.49 | 0.42 |
| Position 2 (AA) | 0.10 | 0.19 | 0.32 |
| Position 3 (LV) | 0.13 | 0.32 | 0.27 |
| Position 1 (DA) | 0.16 | 0.17 | 2.18 |
| Position 2 (AA) | 0.11 | 0.64 | 0.54 |
| Position 3 (LV) | 0.16 | 1.15 | 0.79 |
| Position 1 (DA) | 0.35 | 0.54 | 0.47 |
| Position 2 (AA) | 0.12 | 0.31 | 0.41 |
| Position 3 (LV) | 0.14 | 0.26 | 0.28 |
Maximum increase in temperature (ΔT) at the guidewire tip during the phantom heating experiments at 3T. Results are shown for the MRI-visible glass-fiber epoxy-based guidewire (MR guidewire) and stainless steel guidewire (a, b) and for three different catheters together with the MR guidewire (c-e). The temperatures were measured with the guidewire tip at three positions to mimic the spatial locations for the descending aorta (DA, position 1), the aortic arch (AA, position 2), and the left ventricle (LV, position 3). GRE: gradient echo. bSSFP: balanced steady-state free precession. TSE: turbo spin echo.
In vivo heating evaluation results.
| Maximum ΔT [°C] GRE | Maximum ΔT [°C] TSE | |
|---|---|---|
| 0.17 | 0.15 | |
| 0.16 | 0.20 | |
| 0.26 | 0.25 |
The maximum temperature change (ΔT) close to the glass-fiber epoxy-based MR guidewire tip was measured in vivo at the end of the catheterization procedure in three infarcted porcine models. GRE: gradient echo. TSE: turbo spin echo.
Healthy porcine model experimental results.
| Healthy Porcine Model Experiments | Procedure Time [min:sec] | CNR (blood-signal void) at 3 positions | Signal Void Size (mm) | |||||
|---|---|---|---|---|---|---|---|---|
| Step 2–3 | Step 4 | DA | AA | LV | TL | TW | SW | |
| 10:59 | 2:12 | 72.4 | 93.1 | 117.3 | 12.1 | 5.0 | 2.3 | |
| 1:00 | 0:12 | 31.5 | 37.8 | 37.6 | 11.5 | 5.7 | 2 | |
| 2:05 | 0:49 | 39 | 96.0 | 74.4 | 13.6 | 6.2 | 2.1 | |
| 0:20 | 4:49 | 34.2 | 55.5 | 45.2 | 13.8 | 8.2 | 2.4 | |
| 13:43 | 1:27 | 33.0 | 47.0 | 48.5 | 14 | 7.7 | 2.2 | |
Real-time MR images acquired with a gradient echo (GRE) sequence in five healthy porcine model experiments were analyzed to compute: a) procedure times for step 2–3 and step 4; b) contrast-to-noise ratio (CNR) for guidewire visualization in three positions including the descending aorta (DA), aortic arch (AA), and left ventricle (LV); and c) the MRI signal void size of the glass-fiber epoxy-based guidewire in terms of tip length (TL), tip width (TW), and shaft width (SW).
Infarct porcine model experimental results.
| Infarcted Porcine Model Experiments | Procedure Time [min:sec] | CNR (blood-signal void) at | Signal Void Size (mm) | |||||
|---|---|---|---|---|---|---|---|---|
| Step 2–3 | Step 4 | DA | AA | LV | TL | TW | SW | |
| 4:14 | 0:24 | 32.6 | 46.7 | 54.5 | 12.1 | 5 | 2.3 | |
| switch | 1:14 | |||||||
| 3:41 | 3:03 | 27.5 | 41.2 | 37.6 | 14.8 | 6.8 | 2 | |
| 0:11 | 0:16 | 31.7 | 35.8 | 36.6 | 12.1 | 6.4 | 2 | |
| 1:45 | 0:05 | 33.6 | 41.2 | 40.9 | 12.5 | 6.2 | 2.1 | |
| 1:09 | 0:07 | 31.5 | 38.4 | 39.3 | 12.2 | 5.9 | 2.1 | |
| 3:24 | 0:23 | 47.6 | 64.2 | 74.4 | 13.5 | 6.8 | 2.4 | |
| 0:42 | 1:00 | 61.3 | 66.5 | 62.8 | 12.2 | 6 | 2.1 | |
| 0:42 | 0:17 | 54.3 | 78.2 | 75.5 | 12.8 | 6.2 | 2 | |
Real-time MR images acquired during three successful infarcted porcine model experiments were analyzed to compute: a) procedure times for step 2–3 and step 4; b) contrast-to-noise ratio (CNR) for the glass-fiber epoxy-based MR guidewire visualization in three positions including the descending aorta (DA), aortic arch (AA), and left ventricle (LV); and c) MRI signal void size of the guidewire in terms of tip length (TL), tip width (TW), and shaft width (SW). Multiple catheterization trials were performed in experiments 4 and 5. During experiment 3, the aortic arch was first reached using a metal braided catheter (Expo, Boston Scientific) that was then switched on the wire to a balloon-wedge pressure non-metallic catheter (Arrow, Teleflex).