RF induced energy for partially implanted catheters: a computational study.

Elena Lucano, Micaela Liberti, Tom Lloyd, Francesca Apollonio, Steve Wedan, Wolfgang Kainz, Leonardo M Angelone.   

Abstract

Magnetic Resonance Imaging (MRI) is a radiological imaging technique widely used in clinical practice. MRI has been proposed to guide the catheters for interventional procedures, such as cardiac ablation. However, there are risks associated with this procedure, such as RF-induced heating of tissue near the catheters. The aim of this study is to develop a quantitative RF-safety method for patients with partially implanted leads at 64 MHz. RF-induced heating is related to the electric field incident along the catheter, which in turns depends on several variables, including the position of the RF feeding sources and the orientation of the polarization, which are however often unknown. This study evaluates the electric field profile along the lead trajectory using simulations with an anatomical human model landmarked at the heart. The energy absorbed in the volume near the tip of ageneric partially implanted lead was computed for all source positions and field orientation. The results showed that varying source positions and field orientation may result in changes of up to 18% for the E-field magnitude and up to 60% for the 10g-averaged specific absorption rate (SAR) in the volume surrounding the tip of the lead.

Entities:  

Mesh:

Year:  2016        PMID: 28268553      PMCID: PMC5809151          DOI: 10.1109/EMBC.2016.7590934

Source DB:  PubMed          Journal:  Conf Proc IEEE Eng Med Biol Soc        ISSN: 1557-170X


  4 in total

1.  Assessing the Electromagnetic Fields Generated By a Radiofrequency MRI Body Coil at 64 MHz: Defeaturing Versus Accuracy.

Authors:  Elena Lucano; Micaela Liberti; Gonzalo G Mendoza; Tom Lloyd; Maria Ida Iacono; Francesca Apollonio; Steve Wedan; Wolfgang Kainz; Leonardo M Angelone
Journal:  IEEE Trans Biomed Eng       Date:  2015-12-17       Impact factor: 4.538

Review 2.  Radiofrequency safety for interventional MRI procedures.

Authors:  Ergin Atalar
Journal:  Acad Radiol       Date:  2005-09       Impact factor: 3.173

3.  Calculation of MRI-induced heating of an implanted medical lead wire with an electric field transfer function.

Authors:  Sung-Min Park; Rungkiet Kamondetdacha; John A Nyenhuis
Journal:  J Magn Reson Imaging       Date:  2007-11       Impact factor: 4.813

4.  Development of a new generation of high-resolution anatomical models for medical device evaluation: the Virtual Population 3.0.

Authors:  Marie-Christine Gosselin; Esra Neufeld; Heidi Moser; Eveline Huber; Silvia Farcito; Livia Gerber; Maria Jedensjö; Isabel Hilber; Fabienne Di Gennaro; Bryn Lloyd; Emilio Cherubini; Dominik Szczerba; Wolfgang Kainz; Niels Kuster
Journal:  Phys Med Biol       Date:  2014-08-21       Impact factor: 3.609

  4 in total
  1 in total

1.  Real-time 3T MRI-guided cardiovascular catheterization in a porcine model using a glass-fiber epoxy-based guidewire.

Authors:  Xinzhou Li; Luigi E Perotti; Jessica A Martinez; Sandra M Duarte-Vogel; Daniel B Ennis; Holden H Wu
Journal:  PLoS One       Date:  2020-02-26       Impact factor: 3.240

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.