Callistus M Nguyen1,2, Alan H Baydush2, James D Ververs2, Scott Isom3, Charles M Able4, Michael T Munley1,2.
Abstract
The purpose of this study was to determine if medical linear accelerators (linac) produced by the same manufacturer exhibit operational consistency within their subsystems and components. Two linacs that were commissioned together and installed at the same facility were monitored. Each machine delivered a daily robust quality assurance (QA) irradiation. Linacs and their components operate consistently, but have different operational parameter levels even when produced by the same manufacturer and commissioned in series. These findings have implications on the feasibility of true clinical beam matching.Entities:
Keywords: Medical linear accelerator; Performance; Quality assurance; Statistical process control
Year: 2018 PMID: 32095575 PMCID: PMC7033762 DOI: 10.1016/j.tipsro.2018.05.001
Source DB: PubMed Journal: Tech Innov Patient Support Radiat Oncol ISSN: 2405-6324
Statistics for BGM-PWM performance parameters in two linacs.
| Performance parameter | Range | P-value |
|---|---|---|
| Buncher Radial Current (A) – BRC | −0.110 to 0.125 | <0.00001 |
| Buncher Transverse Current (A) – BTC | 0.065 to 0.115 | <0.00001 |
| Angle Radial Current (A) – ARC | −0.064 to 0.006 | <0.00001 |
| Angle Transverse Current (A) – ATC | −0.007 to 0.011 | <0.00001 |
| Position Radial Current (A) – PRC | −0.847 to 0.273 | <0.00001 |
| Position Transverse Current (A) – PTC | 0.296 to 0.759 | <0.00001 |
Fig. 1Graphical analysis of the BGM-PWM subnode. (a) Median values for the BGM-PWM parameters of the two linacs in this study. The median values show that each linac performance parameter is distinct. (b) The control limits of the “I” chart for BGM-PWM parameters of two linacs. The control limits show that each linac has its own operating window.
Statistics for motion control performance parameters in two linacs.
| Performance parameter | Range | P-value |
|---|---|---|
| Gantry speed 1 | 4.994 to 4.993 | <0.00001 |
| Gantry speed 2 | 5.679 to 5.672 | <0.00001 |
| MLC bank A speed 1 | 1.8287 to 1.8283 | <0.00001 |
| MLC bank A speed 2 | 2.400 to 2.400 | 0.006 |
| MLC bank B speed 1 | 2.417 to 2.417 | 0.01 |
| MLC bank B speed 2 | 2.279 to 2.417 | <0.00001 |
Fig. 2Graphical analysis of Motion Control parameters. (a) The ratio of MLC speeds in Bank A of two linacs. The spread of the data shows that these parameters are different from each other. (b) Median values for Motion Control parameters of the two linacs in this study. The median values show that each linac performance parameter is distinct. (c) The control limits of the “I” chart for the Motion Control parameters of two linacs. The control limits show that each linac has its own operating window.