Literature DB >> 32617079

Retrospective study on performance of constancy check device in Linac beam monitoring using Statistical Process Control.

Bipasha Pal1, Angshuman Pal2, Suresh Das1, Soura Palit1, Papai Sarkar1, Subhayan Mondal1, Suman Mallik1, Jyotirup Goswami1, Sayan Das1, Arijit Sen1, Monidipa Mondol1.   

Abstract

AIM: To examine the application of Statistical Process Control (SPC) and Ishikawa diagrams for retrospective evaluation of machine Quality Assurance (QA) performance in radiotherapy.
BACKGROUND: SPC is a popular method for supplementing the performance of QA techniques in healthcare. This work investigates the applicability of SPC techniques and Ishikawa charts in machine QA.
MATERIALS AND METHODS: SPC has been applied to recommend QA limits on the particular beam parameters using the QUICKCHECK webline QA portable constancy check device for 6 MV and 10 MV flattened photon beams from the Elekta Versa HD linear accelerator (Linac). Four machine QA parameters - beam flatness, beam symmetry along gun target direction and left-right direction, and beam quality factor (BQF) - were selected for retrospective analysis. Shewhart charts, Exponentially Weighted Moving Average (EWMA) charts and Cumulative Sum (CUSUM) charts were obtained for each parameter. The root causes for a failure in machine QA were broken down into an Ishikawa diagram enabling the user to identify the root cause of error and rectify the problem subsequently.
RESULTS: Shewhart charts and EWMA charts applied could detect loss in control in one variable in the 6 MV beams and in all four variables in 10 MV beams. CUSUM charts detected offsets in the readings. The Ishikawa chart exhaustively included the possible errors that lead to loss of control.
CONCLUSION: SPC is proven to be effective for detection of loss in control in machine QA. The Ishikawa chart provides the set of probable root causes of machine error useful while troubleshooting.
© 2019 Greater Poland Cancer Centre. Published by Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Ishikawa diagram; Quality assurance; Radiotherapy; Shewhart chart; Statistical process control

Year:  2019        PMID: 32617079      PMCID: PMC7315042          DOI: 10.1016/j.rpor.2019.12.004

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  22 in total

1.  Phantom dosimetry for conformal stereotactic radiotherapy with a head and neck localizer frame.

Authors:  B P Ravindra; L Fairclough; S M Jaywant
Journal:  Phys Med Biol       Date:  2001-07       Impact factor: 3.609

2.  Statistical process control for radiotherapy quality assurance.

Authors:  Todd Pawlicki; Matthew Whitaker; Arthur L Boyer
Journal:  Med Phys       Date:  2005-09       Impact factor: 4.071

3.  Quality in radiation oncology.

Authors:  Todd Pawlicki; Arno J Mundt
Journal:  Med Phys       Date:  2007-05       Impact factor: 4.071

4.  A comprehensive analysis of the IMRT dose delivery process using statistical process control (SPC).

Authors:  Karine Gérard; Jean-Pierre Grandhaye; Vincent Marchesi; Hanna Kafrouni; François Husson; Pierre Aletti
Journal:  Med Phys       Date:  2009-04       Impact factor: 4.071

5.  Statistical process control for IMRT dosimetric verification.

Authors:  Stephen L Breen; Douglas J Moseley; Beibei Zhang; Michael B Sharpe
Journal:  Med Phys       Date:  2008-10       Impact factor: 4.071

6.  Applying Lean-Six-Sigma Methodology in radiotherapy: Lessons learned by the breast daily repositioning case.

Authors:  Pietro Mancosu; Giorgia Nicolini; Giulia Goretti; Fiorenza De Rose; Davide Franceschini; Chiara Ferrari; Giacomo Reggiori; Stefano Tomatis; Marta Scorsetti
Journal:  Radiother Oncol       Date:  2018-03-06       Impact factor: 6.280

7.  Monitoring Daily QA 3 constancy for routine quality assurance on linear accelerators.

Authors:  Diana Binny; Craig M Lancaster; Tanya Kairn; Jamie V Trapp; Scott B Crowe
Journal:  Phys Med       Date:  2016-11-10       Impact factor: 2.685

Review 8.  Quality assurance in Health Services: the paradigm of radiotherapy.

Authors:  Aristoula Papakostidi; Maria Tolia; Nikolaos Tsoukalas
Journal:  J BUON       Date:  2014 Jan-Mar       Impact factor: 2.533

9.  The next step in patient-specific QA: 3D dose verification of conformal and intensity-modulated RT based on EPID dosimetry and Monte Carlo dose calculations.

Authors:  Wouter van Elmpt; Sebastiaan Nijsten; Ben Mijnheer; André Dekker; Philippe Lambin
Journal:  Radiother Oncol       Date:  2007-12-03       Impact factor: 6.280

10.  A multi-institutional evaluation of machine performance check system on treatment beam output and symmetry using statistical process control.

Authors:  Diana Binny; Trent Aland; Ben R Archibald-Heeren; Jamie V Trapp; Tanya Kairn; Scott B Crowe
Journal:  J Appl Clin Med Phys       Date:  2019-02-20       Impact factor: 2.102

View more