Manasvini Sharma1, Suma Ganesh1, Shailja Tibrewal1, Shalinder Sabharwal2, Neha Sachdeva3, Mohd Adil1, Jyotsana Chaudhary1, Zeeshan Siddiqui2.
Abstract
Purpose: To evaluate the accuracy of Spot photoscreener (PS) as a noncycloplegic photorefractor in detecting amblyopia risk factors (ARFs) in preschool children in an Indian eye clinic setting. Also, to derive appropriate cutoff values for screening to obtain maximum sensitivity and specificity of the device in detecting ARF.Entities:
Keywords: Amblyopia; cyclorefraction; photoscreener
Mesh:
Year: 2020 PMID: 32057013 PMCID: PMC7043179 DOI: 10.4103/ijo.IJO_701_19
Source DB: PubMed Journal: Indian J Ophthalmol ISSN: 0301-4738 Impact factor: 1.848
Figure 1Flow of the process of inclusion and examination of children in the study
2 × 2 contingency table of PS vs clinical examination for detection of ARFs
| Clinic examination including CR | Total | |||
|---|---|---|---|---|
| Positive | Negative | |||
| PS result | ||||
| Positive | 106 (48.40%) | 29 (13.24%) | 135 (61.64%) | <.0001 |
| Negative | 18 (8.22%) | 66 (30.14%) | 84 (38.36%) | |
| Total | 124 (56.62%) | 95 (43.38%) | 219 (100.00%) | |
PS=Photoscreener, ARFs=Amblyopia risk factors, CR=Cyclorefraction
Sensitivity, specificity, positive predictive value, and negative predictive value of PS for individual ARFs
| Sensitivity | Specificity | Positive Predictive Value | Negative Predictive Value | |
|---|---|---|---|---|
| Myopia | 73.02% | 94.40% | 68.66% | 95.42% |
| Astigmatism | 57.86% | 84.90% | 64.29% | 81.09% |
| Hypermetropia | 36.07% | 97.08% | 66.67% | 90.37% |
| Strabismus | 42.22% | 91.95% | 57.58% | 86.02% |
| Anisometropia | 53.85% | 92.72% | 31.82% | 96.95% |
PS=Photoscreener, ARFs=Amblyopia risk factors
Comparison of individual refractive risk factors between CR and PS
| Mean of the Difference (CR-PS) | SD | ||
|---|---|---|---|
| CR SE-PS SE | 0.06 | 2.82 | 0.676 |
| CR Myopia-PS Myopia | -2.23 | 3.61 | 0.000 |
| CR Hypermetropia-PS Hypermetropia | 0.26 | 1.95 | 0.021 |
| CR Astigmatism-PS Astigmatism | 0.14 | 1.19 | 0.054 |
| CR Anisometropia-PS Anisometropia | -0.14 | 1.06 | 0.059 |
CR=Cyclorefraction, PS=Photoscreener, SD=Standard deviation
Figure 2Bland–Altman assessment of agreement between the cyclorefraction and photoscreener for spherical equivalent (a), hypermetropia (b), myopia (c), astigmatism (d), and anisometropia (e). The dashed line depicts the mean deviation and the solid lines depict the 95% limits of agreement
Figure 3ROC curve analysis for detection of refractive amblyopia risk factors (a) Hypermetropia, (b) myopia, (c) astigmatism, and (d) anisometropia