Caner Kara1, Ikbal Seza Petricli1.
Abstract
OBJECTIVES: This study was designed to assess the accuracy of photorefraction and wavefront-based autorefraction in children under 3 years of age.Entities:
Keywords: Accuracy; amblyopia; child; ocular; refraction; refractive errors; refractometry
Year: 2021 PMID: 35005493 PMCID: PMC8651029 DOI: 10.14744/bej.2021.09825
Source DB: PubMed Journal: Beyoglu Eye J ISSN: 2459-1777
The mean refractive values obtained with the devices studied and cycloplegic retinoscopy
| SE (D) Mean±SD (Range) | J0 (D) Mean±SD (Range) | J45 (D)Mean±SD (Range) | |
|---|---|---|---|
| Spot Vision | |||
| Non-cycloplegic | 0.43±0.75 (-1.75 to 3.50) | 0.49±0.48 (-0.62 to 2.24) | 0.02±0.25 (-0.87 to 0.95) |
| Cycloplegic | 1.80±1.25 (-1.37 to 5.38) | 0.53±0.42 (-0.72 to 1.94) | 0.00±0.25 (-0.80 to 0.80) |
| SureSight | |||
| Non-cycloplegic | 0.90±0.86 (-2.37 to 3.63) | 0.50±0.45 (-0.63 to 1.50) | 0.04±0.21 (-0.93 to 0.83) |
| Cycloplegic | 1.55±1.4 (-1.87 to 5.00) | 0.42±0.41 (-0.70 to 1.50) | 0.04±0.19 (-0.86 to 1.06) |
| Retinoscopy | |||
| Cycloplegic | 1.53±1.34 (-2.00 to 5.25) | 0.44±0.4 (-0.5 to 1.75) | 0.00±0.09 (-0.98 to 0.87) |
D: Diopter; SE: Spherical equivalent; J0: Jackson cross-cylinder power at axis 90° and 180°; J45: Jackson cross-cylinder power at axis 45° and 135°.
Comparisons between the measurements obtained with the devices evaluated and cycloplegic retinoscopy
| SE (D) Mean (95% LOA) | J0 (D) Mean (95% LOA) | J45 (D) Mean (95% LOA) | |||
|---|---|---|---|---|---|
| Spot Vision | vs. | Retinoscopy | -1.10 (-3.30 to 1.30) | 0.05 (-0.63 to 0.72) | 0.02 (-0.48 to 0.51) |
| Non-cycloplegic | Cycloplegic | ||||
| Spot Vision | vs. | Retinoscopy | 0.27 (-0.94 to 1.47) | 0.09 (-0.56 to 0.74) | 0.00 (-0.51 to 0.50) |
| Cycloplegic | Cycloplegic | ||||
| SureSight | vs. | Retinoscopy | -0.63 (-3.30 to 2.10) | 0.06 (-0.47 to 0.59) | 0.04 (-0.36 to 0.44) |
| Non-cycloplegic | Cycloplegic | ||||
| SureSight | vs. | Retinoscopy | 0.01 (-1.10 to 1.10) | -0.03 (-0.48 to 0.43) | 0.03 (-0.32 to 0.38) |
| Cycloplegic | Cycloplegic |
significant at p<0.001,
significant at p<0.05; D: Diopter; SE: Spherical equivalent, J0: Jackson cross-cylinder power at axis 90° and 180°; J45: Jackson cross-cylinder power at axis 45° and 135°; LOA: Limit of agreement.
Figure 1Bland-Altman plots showing agreement between non-cycloplegic device measurements and cycloplegic retinoscopy measurements for spherical equivalent (SE) and Jackson cross-cylinder power values (J0 and J45).
Figure 2Frequency distribution of differences in the spherical equivalent (SE) values obtained with both devices with and without cycloplegia and the mean SE values obtained with cycloplegic retinoscopy.
Figure 3Correlation between the mean spherical equivalent (SE) values obtained with cycloplegic retinoscopy and the difference in SE values obtained with both devices without cycloplegia and cycloplegic retinoscopy.
Figure 4Bland-Altman plots showing the agreement between cycloplegic measurements of spherical equivalent (SE) and Jackson cross-cylinder power values (J0 and J45) obtained with both devices and cycloplegic retinoscopy.