Sustained Long-Term Outcomes With Closed-Loop Spinal Cord Stimulation: 12-Month Results of the Prospective, Multicenter, Open-Label Avalon Study.

Marc Russo1, Charles Brooker2,3, Michael J Cousins2,3, Nathan Taylor2,3, Tillman Boesel4, Richard Sullivan5, Lewis Holford2,3, Erin Hanson6, Gerrit Eduard Gmel6, Nastaran Hesam Shariati6, Lawrence Poree7, John Parker6,8.   

Abstract

BACKGROUND: Spinal cord stimulation (SCS) activates the dorsal column fibers using electrical stimuli. Current SCS systems function in fixed-output mode, delivering the same stimulus regardless of spinal cord (SC) activation.
OBJECTIVE: To present long-term outcomes of a novel closed-loop SCS system that aims to maintain the SC activation near a set target level and within a therapeutic window for each patient. SC activation is measured through the evoked compound action potential (ECAP) generated by each stimulus pulse.
METHODS: Fifty patients with lower back and/or leg pain who were successfully trialed received a permanent system (Evoke; Saluda Medical, Sydney, Australia). Ratings of pain (visual analog scale), quality of life, function, sleep, and medication use were collected at baseline and at each visit. SC activation levels were reported in summary statistics. The therapeutic window for each individual patient was defined as the range of ECAP amplitudes between sensation threshold and uncomfortably strong stimulation.
RESULTS: At 12 mo, the proportion of patients with ≥50% relief was 76.9% (back), 79.3% (leg), and 81.4% (overall), and the proportion with ≥80% pain relief was 56.4% (back), 58.6% (leg), and 53.5% (overall). Patients spent a median of 84.9% of their time with stimulation in their therapeutic window, and 68.8% (22/32) eliminated or reduced their opioid intake. Statistically significant improvements in secondary outcomes were observed.
CONCLUSION: The majority of patients experienced more than 80% pain relief with stable SC activation, as measured by ECAP amplitude at 12 mo, providing evidence for the long-term effectiveness of the Evoke closed-loop SCS system. © Congress of Neurological Surgeons 2020.

Entities:  

Keywords:  Action potentials; Back pain; Chronic pain; Electric stimulation; Feedback; Pain management; Spinal cord stimulation

Mesh:

Year:  2020        PMID: 32023344      PMCID: PMC8184296          DOI: 10.1093/neuros/nyaa003

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


Brief Pain Inventory closed-loop stimulator external closed-loop stimulator evoked compound action potential EuroQol instrument failed back surgery syndrome failed neck surgery syndrome morphine milligram equivalents Oswestry Disability Index Pittsburgh Sleep Quality Index serious adverse event spinal cord spinal cord stimulation standard deviation standard error of the mean therapeutic window visual analog scale Spinal cord stimulation (SCS) is an established therapy that has been used to provide relief from chronic neuropathic pain for over 50 yr.[1] Although the mechanisms of action of SCS are not fully understood, a vast amount of evidence suggests that it acts by activating the dorsal column fibers, which, in turn, inhibit pain processing in the dorsal horn.[2] All currently available commercial SCS systems still operate in a fixed-output stimulation configuration (eg, fixed frequency, fixed pulse width, fixed amplitude, and fixed pulse train) regardless of the neural activation of the spinal cord (SC) fibers. These fixed-output (or open-loop) systems do not account for the large variation of electrical field strength and, subsequently, volume of tissue stimulated that occurs because of changes in the distance between the electrode and SC from normal physiological activity (eg, breathing and heartbeat) and movement.[3-5] Changes in stimulation strength lead to variable activation of the SC and, thus, variable inhibition of pain processing pathways. Each suprathreshold stimulus elicits an evoked compound action potential (ECAP), which represents the sum of all single-fiber action potentials generated by the stimulus. The ECAP is therefore a direct measure of SC activation and provides information on the fibers elicited by SCS, contributing to the therapeutic effect of stimulation.[6-8] In order to determine whether maintaining stable SC activation has a beneficial outcome on pain relief, a prospective, multicenter, single-arm study was designed to demonstrate the safety and performance of a new closed-loop SCS system that maintains stable SC activation via continuous ECAP measurement. This closed-loop system adjusts the stimulus current after each pulse to maintain the resulting ECAP amplitude near a target amplitude. Details regarding the patients, study design, device, and results through 6 mo have been previously published.[9] Here, the evidence for closed-loop SCS is expanded upon by including results through the 12-mo follow-up and additional patients enrolled in the study. An elective extension of follow-up to 24 mo was offered to all patients; these results will be published once all patients have completed the study.

METHODS

Patients

Patients were consented at 5 clinical sites in Australia from August 2015 to April 2017. Key inclusion criteria included the following: patients diagnosed with chronic, intractable pain (visual analog scale [VAS] ≥6 cm for the past week) that was refractory to conservative therapy for at least 3 mo and having maintained their prescription pain medication dosage stable for at least 1 mo. Key exclusion criteria included the following: having a contraindication to SCS, a condition that was likely to interfere with study conduct or treatment outcome evaluation, and involvement in litigation involving their pain condition. The study was limited to patients with chronic back and/or leg pain but not restricted to patients with previous back surgery. The study was designed to show that at least 90% of patients could be programmed with closed-loop stimulation with a power of 0.85.

Study Design and Data Collection

The Avalon study was a prospective, multicenter, single-arm study approved by local ethics committees prior to patient enrollment. All patients provided written consent to participate in the study. The protocol was publicly registered at Australian New Zealand Clinical Trials Registry. Baseline assessments included ratings of pain (100-mm VAS),[10] impact of pain (Brief Pain Inventory [BPI])[11] function (Oswestry Disability Index [ODI]),[12] sleep (Pittsburgh Sleep Quality Index [PSQI]),[13] quality of life (EuroQol instrument [EQ-5D-5L]),[14] and medication usage. Assessments were repeated at the end of the trial period and at 1 (data not shown), 3, 6, and 12 mo postimplantation. Adverse events were assessed throughout the study.

Device and Implantation

The Evoke SCS system (Saluda Medical, Sydney, Australia) consists of a rechargeable external closed-loop stimulator (eCLS), implantable closed-loop stimulator (CLS), two 12-contact percutaneous leads, and all necessary surgical and supporting tools. For the trial procedure, patients were implanted with trial leads in the epidural space over the dorsal columns at the thoracic vertebral level associated with pain for an average of 7 d to evaluate its effect. If the trial was deemed successful (≥40% reduction in pain in any segmental VAS score with baseline VAS ≥ 6 cm), patients were given the option to receive the implanted system. Evoke is the first closed-loop SCS system to have the ability to measure human SC activation in real time. The recording electrodes for measuring the ECAP can be any 2 nonstimulating electrodes on either lead, and either or both leads may be used for stimulation. The ECAP elicited by the stimulation is sensed, sampled, and processed by the stimulator to measure its amplitude (Figure 1). The amplitude is then used to drive a feedback loop that adjusts the stimulus current at each stimulus pulse to maintain near-constant SC activation. For each patient, programming involved first optimizing the stimulation location and then identifying appropriate electrodes to optimize ECAP measurement to assess SC activation. The level of SC activation, in other words the ECAP amplitude, is obtained by filtering the measured response resulting in a single value measured in microvolts.[9]
FIGURE 1.

Schematic representation of an ECAP. ECAPs have a well-defined shape with 3 peaks: 2 positive and 1 negative, labeled P1, N1, and P2, in order of appearance. In simple terms, they are the combined electrical field generated by the action potentials of stimulated fibers at the recording site. The first P1 peak stems from capacitive coupling between the inside and outside of the fibers and is caused by the incoming action potential. The N1 and P2 peaks result from ionic flow (sodium [Na+] and potassium [K+]) in and out of the fibers that form the well-known action potential. As the measurement is done outside of the fibers, the polarity of the peaks is reversed compared with intracellular recordings often depicted in textbooks. ECAP, evoked compound action potential.

Schematic representation of an ECAP. ECAPs have a well-defined shape with 3 peaks: 2 positive and 1 negative, labeled P1, N1, and P2, in order of appearance. In simple terms, they are the combined electrical field generated by the action potentials of stimulated fibers at the recording site. The first P1 peak stems from capacitive coupling between the inside and outside of the fibers and is caused by the incoming action potential. The N1 and P2 peaks result from ionic flow (sodium [Na+] and potassium [K+]) in and out of the fibers that form the well-known action potential. As the measurement is done outside of the fibers, the polarity of the peaks is reversed compared with intracellular recordings often depicted in textbooks. ECAP, evoked compound action potential. Prior work established that ECAP amplitude correlates with stimulation sensation intensity in a linear fashion.[15] In other words, a higher ECAP amplitude results in an equal or stronger perceived stimulus sensation (never weaker). Therefore, the patient's therapeutic window is defined as the range of ECAP amplitudes lying between the stimulus perception threshold and the maximum level of stimulation (discomfort threshold). This definition is consistent with the literature,[16,17] but creates a relationship between SC activation and stimulation perception. The implant automatically records and stores the amplitudes of the ECAPs elicited by each stimulus pulse. To investigate the patients’ SC activation patterns and therapy use, the ECAP amplitudes for the patients’ preferred (or most used) program during the 1 wk prior to the scheduled visits were extracted and analyzed.

Data Management and Analysis

Throughout the study, standard data management procedures were observed, and monitoring was carried out periodically by an independent clinical research organization to ensure data quality. Along with raw scores and percent change from baseline, VAS data were also analyzed as responders (≥50% pain relief) and high responders (≥80% pain relief). Paired t-tests with an alpha of 0.05 were used to test that mean change from baseline was different from 0 using SAS Enterprise Guide 7.1. Efficacy, medication, and device data are presented for the permanently implanted patients only; for the sake of transparency, we present safety data for all enrolled patients.

RESULTS

Patients, Demographics, and Baseline Characteristics

A total of 70 patients underwent a trial procedure. Of these, 68 (97.1%) completed the end-of-trial assessments and were evaluable. Of the 68 patients, 56 (82.4%) with assessment data had a reduction of 40% or more from baseline in their overall VAS rating; of those, 48 patients elected to proceed with a permanent implant. Two additional patients with a segmental VAS reduction of 40% or more proceeded with a permanent implant as per the protocol inclusion criterion (Figure 2). Fifty subjects were implanted (71.4% of those trialed), which is in line with the literature (reported rates are between 41.4% and 86.4%[18-21]). On average, patients had a permanent implant for 19.8 mo (range, 0.5-25.8 mo) at the time of this report. One subject died of cardiac arrest between the 3- and 6-mo visit, but the occurrence was determined to not be related to the device or procedure. One patient was withdrawn by the investigator for noncompliance with study requirements between the 6- and 12-mo visits. All patients had completed the 12-mo follow-up at the time of this report.
FIGURE 2.

Patient progression through the study.

Patient progression through the study. Demographics and baseline characteristics for the cohort of implanted patients are presented in Table 1. Persistent or recurrent pain following spinal surgery was the main diagnosis across the cohort (56.0%), and the lower back was the most commonly reported primary pain area (78.0%). Three patients have had previous experience with SCS.
TABLE 1.

Baseline Demographics and Characteristics for Permanently Implanted Patients

Implanted patients N = 50
Age (yr) at enrollment
 Mean (SD)56.7 (12.2)
Gender, n (%)
 Male23 (46.0)
 Female27 (54.0)
Primary diagnosis, n (%)
 FBSS/FNSS28 (56.0)
 Radiculopathy9 (18.0)
 Other[a]13 (26.0)
Primary region of pain, n (%)
 Lower back39 (78.0)
 Leg8 (16.0)
 Foot3 (6.0)
Prior history of SCS, n (%)
 Yes3 (6.0)
 No47 (94.0)
Duration (yr) of pain
 Mean (SD)15.0 (11.0)

FBSS, failed back surgery syndrome; FNSS, failed neck surgery syndrome; SCS, spinal cord stimulation; SD, standard deviation.

aOther diagnoses: discogenic back (or lower back) pain/internal disc disruption (n = 5), lumbar spondylosis (n = 4), lumbar degenerative disease (n = 1), neuropathic pain/neuropathic low back pain post trauma (n = 1), peripheral neuropathy (n = 1), and sciatica and gluteal tendinopathy (n = 1).

Baseline Demographics and Characteristics for Permanently Implanted Patients FBSS, failed back surgery syndrome; FNSS, failed neck surgery syndrome; SCS, spinal cord stimulation; SD, standard deviation. aOther diagnoses: discogenic back (or lower back) pain/internal disc disruption (n = 5), lumbar spondylosis (n = 4), lumbar degenerative disease (n = 1), neuropathic pain/neuropathic low back pain post trauma (n = 1), peripheral neuropathy (n = 1), and sciatica and gluteal tendinopathy (n = 1).

Pain Relief Outcomes

Across all permanently implanted patients, mean rating of back pain was 81.3 mm (±1.4) at baseline (n = 46 patients). After 12 mo of treatment, back pain rating was reduced by 57.9 mm (±4.2, P < .0001) to 22.7 (±4.1), a mean percent reduction of 72.0% (±5.0). At 12 mo, 76.9% of patients were back pain responders (≥50% pain reduction), with 56.4% being classified as high responders (≥80% pain reduction). Individual patient responses for back pain at 12 mo are shown in Figure 3.
FIGURE 3.

Individual patient responses for back pain VAS reduction at 12 mo. VAS, visual analog scale.

Individual patient responses for back pain VAS reduction at 12 mo. VAS, visual analog scale. Mean rating of leg pain for permanently implanted patients was 77.7 mm (±1.8) at baseline (n = 35 patients). Leg pain was reduced to 21.1 mm (±4.7) after 12 mo of treatment, a statistically signficant mean reduction of 56.6 mm (±5.1, P < .0001) and a mean percent reduction of 72.1% (±6.1). The proportion of patients who were leg pain responders at 12 mo was 79.3%, and 58.6% of patients were high responders. Mean rating of overall pain for permanently implanted patients was 81.3 mm (±1.6) at baseline (n = 50 patients). Overall pain was reduced to 21.0 mm (±3.4) after 12 mo of treatment, a statistically signficant mean reduction of 59.8 mm (±3.8, P < .001) and a mean percent reduction of 73.6% (±4.3). The proportion of patients who were overall pain responders at 12 mo was 81.4%, and 53.5% of patients were high responders. Back, leg, and overall pain outcomes are presented in Table 2 and Figures 4 and 5.
TABLE 2.

Summary of Back, Leg, and Overall Pain Visual Analog Scale Scores Over Time for Permanently Implanted Patients

Baseline3 mo6 mo12 mo
Back pain
 N46414239
 Mean raw VAS score, mm (SEM)81.3 (1.4)24.9 (3.4)22.6 (3.7)22.7 (4.1)
 Mean percent improvement in VAS scores (SEM)69.2 (4.3)72.6 (4.3)72.0 (5.0)
 Mean improvement in VAS scores, mm (SEM, P value)56.7 (3.7, P < .0001)58.4 (3.5, P < .0001)57.9 (4.2, P < .0001)
 Proportion of patients responding at ≥ 50% improvement, %75.676.276.9
 Proportion of patients responding at ≥ 80% improvement, %53.750.056.4
Leg pain
 N35323229
 Mean raw VAS score, mm (SEM)77.7 (1.8)18.6 (3.4)15.7 (3.2)21.1 (4.7)
 Mean percent improvement in VAS scores (SEM)75.7 (4.3)78.9 (4.6)72.1 (6.1)
 Mean improvement in VAS scores, mm (SEM, P value)59.6 (3.9, P < .0001)61.7 (3.9, P < .0001)56.6 (5.1, P < .0001)
 Proportion of patients responding at ≥ 50% improvement, %87.584.479.3
 Proportion of patients responding at ≥ 80% improvement, %53.162.558.6
Overall pain
 N50454643
 Mean raw VAS score, mm (SEM)81.3 (1.6)22.8 (2.9)22.6 (3.5)21.0 (3.4)
 Mean percent improvement in VAS scores (SEM)71.2 (4.0)71.7 (4.5)73.6 (4.3)
 Mean improvement in VAS scores, mm (SEM, P value)58.6 (3.3, P < .0001)58.7 (3.7, P < .0001)59.8 (3.8, P < .0001)
 Proportion of patients responding at ≥ 50% improvement, %80.078.381.4
 Proportion of patients responding at ≥ 80% improvement, %42.252.253.5

SEM, standard error of the mean; VAS, visual analog scale.

The change from baseline summary presented in the table is based on subjects providing data both at baseline and at follow-up. All available data at the individual timepoints are presented for completeness.

FIGURE 4.

Mean VAS ratings over time for A, low back pain, B, leg pain, and C, overall pain for permanently implanted patients. *P < .0001. Error bars represent the standard error of the mean. VAS, visual analog scale.

FIGURE 5.

Responder and high responder rates for overall pain over time.

Mean VAS ratings over time for A, low back pain, B, leg pain, and C, overall pain for permanently implanted patients. *P < .0001. Error bars represent the standard error of the mean. VAS, visual analog scale. Responder and high responder rates for overall pain over time. Summary of Back, Leg, and Overall Pain Visual Analog Scale Scores Over Time for Permanently Implanted Patients SEM, standard error of the mean; VAS, visual analog scale. The change from baseline summary presented in the table is based on subjects providing data both at baseline and at follow-up. All available data at the individual timepoints are presented for completeness.

Secondary Outcomes

BPI, EQ-5D-5 L, ODI, and PSQI for permanently implanted patients all showed statistically significant improvements at 12 mo compared with baseline. Outcomes at 12 mo are described below, and outcomes at other timepoints can be found in Table 3 and Figures 6 and 7.
TABLE 3.

Summary of Secondary Outcomes Over Time for Permanently Implanted Patients

Baseline3 mo6 mo12 mo
BPI
 N50454643
 Mean severity score (SEM)6.8 (0.2)3.3 (0.3)3.2 (0.3)3.1 (0.3)
 Mean change from baseline, severity score (SEM, P value)3.5 (0.3, P < .0001)3.6 (0.3, P < .0001)3.6 (0.3, P < .0001)
 Mean interference score (SEM)7.1 (0.2)3.5 (0.4)3.4 (0.4)3.2 (0.4)
 Mean change from baseline, interference score (SEM, P value)3.5 (0.4, P < .0001)3.6 (0.4, P < .0001)3.8 (0.4, P < .0001)
EQ-5D-5L
 N50454643
 Mean EQ index score (SEM)0.404 (0.030)0.637 (0.029)0.688 (0.029)0.633 (0.034)
 Mean change from baseline – EQ index score (SEM, P value)0.233 (0.036, P < .0001)0.278 (0.034, P < .0001)0.214 (0.032, P < .0001)
 Minimally important difference from baseline (≥0.074)80.0% (36/45)87.0% (40/46)88.4% (38/43)
 Mean EQ Health VAS score (SEM)53.3 (2.9)73.2 (2.5)75.2 (3.0)71.4 (3.3)
 Mean change from baseline – EQ Health VAS (SEM, P value)18.4 (3.7, P < .0001)20.7 (4.1, P < .0001)17.4 (4.1, P = .0001)
ODI
 N50444443
 Mean ODI score (SEM)52.3 (1.7)34.6 (2.1)31.7 (2.3)31.2 (2.5)
 Mean change from baseline – final score (SEM, P value)17.1 (1.8, P < .0001)20.1 (2.1, P < .0001)20.3 (2.1, P < .0001)
 Minimum detectable change from baseline (≥10%)70.5% (31/44)70.5% (31/44)76.7% (33/43)
PSQI
 N50454642
 Mean PSQI score (SEM)12.0 (0.6)8.4 (0.7)8.3 (0.7)8.6 (0.8)
 Mean change from baseline – global score (SEM, P value)3.5 (0.7, P < .0001)3.6 (0.7, P < .0001)3.1 (0.7, P < .0001)

BPI, Brief Pain Inventory; EQ-5D-5 L, EuroQol instrument; PSQI, Pittsburgh Sleep Quality Index; ODI, Oswestry Disability; SEM, standard error of the mean; VAS, visual analog scale.

The change from baseline summary presented in the table is based on patients providing data both at baseline and at follow-up. All available data at the individual timepoints are presented for completeness.

FIGURE 6.

Proportion of patients reporting each level of disability on the Oswestry Disability Index over time for permanently implanted patients.

FIGURE 7.

A, Mean daily use (MME/d) at baseline and 12 mo for all patients on opioids at baseline. B, Proportion of patients (who were taking opioids at baseline) reducing or eliminating opioid medications over time. C, Mean daily use (MME/d) at baseline and 12 mo of all patients on high-dose opioids (>50 MME/d) at baseline. MME, morphine milligram equivalent.

Proportion of patients reporting each level of disability on the Oswestry Disability Index over time for permanently implanted patients. A, Mean daily use (MME/d) at baseline and 12 mo for all patients on opioids at baseline. B, Proportion of patients (who were taking opioids at baseline) reducing or eliminating opioid medications over time. C, Mean daily use (MME/d) at baseline and 12 mo of all patients on high-dose opioids (>50 MME/d) at baseline. MME, morphine milligram equivalent. Summary of Secondary Outcomes Over Time for Permanently Implanted Patients BPI, Brief Pain Inventory; EQ-5D-5 L, EuroQol instrument; PSQI, Pittsburgh Sleep Quality Index; ODI, Oswestry Disability; SEM, standard error of the mean; VAS, visual analog scale. The change from baseline summary presented in the table is based on patients providing data both at baseline and at follow-up. All available data at the individual timepoints are presented for completeness. The mean BPI severity score was more than halved over the 12-mo follow-up period, with a mean change of 3.6 (±0.3, P < .0001). The mean BPI interference score decreased by 3.8 (±0.4, P < .0001) from a baseline of 7.1 (±0.2). The mean EQ-5D-5L index score increased significantly from baseline to 12 mo by 0.214 (±0.032, P < .0001); similarly, the EQ-5D-5L Health VAS score increased significantly by 17.4 (±4.1, P = .0001). At 12 mo, 88.4% of patients experienced at least a minimally important difference (≥0.074) in EQ-5D-5L.[22] The mean ODI score decreased by 20.3 (±2.1, P < .0001). This resulted in a large shift of the patient population toward lower disability, with 74.4% of implanted patients being only minimally or moderately disabled at 12 mo, compared with 18.0% at baseline (Figure 6). Sleep quality, as measured by the mean PSQI score, was also improved from 12 (±0.6) at baseline to 8.6 (±0.8) at 12 mo, with 22 patients (54.2%) having a clinically meaningful PSQI change at 12 mo (≥3 point reduction) compared with baseline.[23] Remarkably, opioid use was almost halved over the course of the study. At baseline, 76.0% (38/50) of permanently implanted patients were on opioids with an average daily dose of 62.9 morphine milligram equivalents ([MME]/d), which decreased throughout the study to 32.3 MME/d after 12 mo of treatment (Figure 7A showing actual mg vs time). In addition, 68.8% (22/32) of permanently implanted patients who were on opioids at baseline eliminated or reduced MMEs at the 12-mo visit (Figure 7B). More notably, 14 patients on high-dose opioid therapy (>50 MME) at baseline had a significant decrease in opioid usage from 133.5 MME/d down to 66.8 MME/d at 12 mo (Figure 7C). Finally, patients were asked to rate their overall satisfaction with the stimulator using a 5-point scale (with options ranging from “very unsatisfied” to “very satisfied”). At 12 mo, 88.4% of patients reported they were “satisfied” or “very satisfied” with their treatment.

ECAP Amplitude Data

Closed-loop stimulation was programmable in all patients at 12 mo. For each patient, we calculated the most frequent ECAP amplitudes for their preferred program; the median value across patients ranged from 20.3 to 28.5 μV across the 3-, 6-, and 12-mo visits without clear upward or downward trend. Time spent within and below the therapeutic window across scheduled visits is displayed in Figure 8. At the 3-, 6-, and 12-mo follow-up visits, patients were in the therapeutic window a median of between 83.1% and 96.7% of the time. Patients spent a median of 2.9% to 6.8% of the total time below the therapeutic window and a median of 0% to 0.1% of the time above the therapeutic window across 3-, 6-, and 12-mo visits. Because of the non-normal distribution of the population, median values were used to present the therapeutic window and ECAP amplitude data.
FIGURE 8.

Median percent stimuli below, within, and above the therapeutic window from the 3- to 12-mo visits. TW, therapeutic window. The reader will note that, unlike the average values, median values do not add up to 100%.

Median percent stimuli below, within, and above the therapeutic window from the 3- to 12-mo visits. TW, therapeutic window. The reader will note that, unlike the average values, median values do not add up to 100%.

Safety Outcomes

At the time of writing, 3 (4.3%) study/device-related serious adverse events (SAEs) have been reported (see Table 4). One patient developed an allergic reaction to titanium after the implant, another patient experienced severe new low back pain during the trial period, and the third experienced postoperative wound dehiscence following the implant procedure because of poor skin integrity. All 3 SAEs resolved with treatment.
TABLE 4.

Serious Adverse Event Rates

Serious adverse event ratesEventsPatients
Nn/N (%)
Study-related SAEs33/70 (4.2)
Unanticipated adverse device effects00/70 (0)

SAE, serious adverse event.

Serious Adverse Event Rates SAE, serious adverse event. No unanticipated adverse events were recorded, and the type, rate, nature, and severity of adverse events that have occurred in the Avalon study were consistent and comparable with other SCS device studies and to reported adverse events in the literature.[24]

Programming Burden

During the course of the study, patients could return to the clinic at any time to get support with the device or ask for reprogramming. These visits were unscheduled programming visits and showed a clear downward trend as seen in Figure 9. After an initial optimization period, reprogramming was necessary only in a subset of patients and became increasingly rare. These data suggest that, in the long term, patients may need no more than 1 programming session per year on average.
FIGURE 9.

Number of unscheduled programming visits per patient per month between each study time point up to 12 mo.

Number of unscheduled programming visits per patient per month between each study time point up to 12 mo.

DISCUSSION

During the development of the Evoke SCS system (Saluda Medical), it was hypothesized that controlling the amount of neural activation and maintaining the SC activation within an individual's therapeutic range could have profound benefits for long-term efficacy. In this study, we demonstrate that ECAP measurements are possible long term and remain stable over a 12-mo period. Closed-loop SCS is able to maintain SC activation within a therapeutic window for more than 80% of the time. This helps to avoid discomfort and overstimulation in patients as well as possible understimulation leading to suboptimal therapeutic delivery.[16,25] Over the past decade, a large amount of effort has been spent on developing novel waveforms for SCS in an attempt to optimize the therapy.[20,21] However, no existing device, independent of the waveform, has an objective measure of the effect on the SC. It is likely that stimulation outside the therapeutic window is one cause for loss of efficacy, the main contributor to the 20% to 25% explant rate over a 5-yr period reported in the literature.[26-29] Although the maintenance of efficacy over several years still needs to be investigated, the stable 12-mo outcome data are promising and could indicate that closed-loop SCS can decrease the explant rate of SCS associated with loss of efficacy. The study has demonstrated sustained high rates of VAS pain reduction through 12 mo, with mean percent reduction of 72% or greater in back, leg, and overall pain at 12 mo. Responder and high responder rates were profound. More than 76% of patients achieved at least 50% pain relief, and more than 53% of patients achieved at least 80% pain relief in the back, legs, and overall pain at 12 mo. These are the highest levels of pain relief reported at 1-yr follow-up for any SCS system to date. Functional disability, quality of life, and sleep-related patient-reported outcomes also showed significant improvements over time. After 12 mo of treatment, the proportion of patients reporting a severe or worse disability on the ODI was reduced by two-thirds (82%-26%). Because it takes time for patients’ lives to normalize after improving their pain relief, it is likely that a stable therapy will induce further improvments in quality of life over time. Therefore, these scores will be monitored in this study as the patients progress through their 24-mo visit. It is well known that excessive opioid use has a wide-ranging impact on both the patient's well-being as well as their family, and the economic burden on the healthcare system is a matter of public health concern.[30] The opioid reduction over the course of the study is therefore particularly promising, as finding alternative long-term treatments for pain is a major focus of the international community, and closed-loop SCS could become a valuable tool in helping patients reduce their opioid intake and further improve their quality of life. In addition to offering an objective measure of SC activation and enabling closed-loop SCS, ECAP measurements have the potential to inform about the health of the SC and the effects of treatment over time. The objective measurements of SC activation and other neurophysiological properties have the potential to hold substantial clinical utility. These data may, in the future, be used to help the diagnosis and treatment of chronic pain and could provide an avenue for more effective, individualized, mechanism-based treatments. This may be extended to other treatments that impact neural activation, such as some types of pain medications (eg, anticonvulsants and opioids), in which measurement of ECAPs may be used to titrate/optimize dosing and measure the interaction between these medications and SCS. Research is currently underway investigating the mechanisms of action of SCS and the effect of various stimulus paradigms on SC activation.

CONCLUSION

The 12-mo results from the Avalon study show the highest degree of pain relief recorded for an SCS system to date. We postulate that the stable level of SC activation is the main factor contributing to achieving this profound level of pain relief. To further test this hypothesis, the Avalon study was extended to a follow-up of 24 mo for consenting patients. Additionally, the Evoke SCS system (Saluda Medical) is currently being evaluated in a randomized, controlled, double-blind study in the United States, comparing the safety and efficacy of open-loop SCS to closed-loop SCS utilizing ECAP measurements.

Disclosures

This study was funded by Saluda Medical Pty Ltd Dr Russo has served as a consultant for Abbott, Boston Scientific, Mainstay Medical, Medtronic Inc, Nevro Corp, Presidio Medical, Saluda Medical Pty Ltd, and Stimwave. Dr Brooker has served as a consultant for Sequirus and as an investigator on sponsored research for Abbott, Medtronic Inc, Phosphagenics, RRMedsciences, Saluda Medical Pty Ltd, and Spinal Modulation. Dr Cousins has served as a consultant for Saluda Medical. Mr Sullivan has served as a consultant for Abbott, Medtronic, Nevro, and Seqirus. Ms Hanson, Dr Gmel, Dr Shariati, and Dr Parker are employees of Saluda Medical. Dr Poree has served as a consultant for Medtronic, Sollis, and Circuit Tx; a consultant with stock options for Nalu; a consultant and medical monitor for Saluda Medical; a principal investigator for the Abbott DRG study; a past consultant for StimWave; and a past principal investigator and shareholder for Spinal Modulation.
  29 in total

1.  Changes in axonal physiology and morphology after chronic compressive injury of the rat thoracic spinal cord.

Authors:  R Nashmi; M G Fehlings
Journal:  Neuroscience       Date:  2001       Impact factor: 3.590

2.  EuroQol--a new facility for the measurement of health-related quality of life.

Authors: 
Journal:  Health Policy       Date:  1990-12       Impact factor: 2.980

3.  Changing body position alters the location of the spinal cord within the vertebral canal: a magnetic resonance imaging study.

Authors:  M R B Ranger; G J Irwin; K M Bunbury; J M Peutrell
Journal:  Br J Anaesth       Date:  2008-10-19       Impact factor: 9.166

4.  Spinal Cord Stimulator Explantation: Motives for Removal of Surgically Placed Paddle Systems.

Authors:  Derrick A Dupré; Nestor Tomycz; Donald Whiting; Michael Oh
Journal:  Pain Pract       Date:  2017-11-20       Impact factor: 3.183

Review 5.  Spinal cord stimulation: mechanisms of action.

Authors:  John C Oakley; Joshua P Prager
Journal:  Spine (Phila Pa 1976)       Date:  2002-11-15       Impact factor: 3.468

6.  Conduction properties of epidurally recorded spinal cord potentials following lower limb stimulation in man.

Authors:  J P Halonen; S J Jones; M A Edgar; A O Ransford
Journal:  Electroencephalogr Clin Neurophysiol       Date:  1989 May-Jun

7.  The validation of visual analogue scales as ratio scale measures for chronic and experimental pain.

Authors:  Donald D Price; Patricia A McGrath; Amir Rafii; Barbara Buckingham
Journal:  Pain       Date:  1983-09       Impact factor: 6.961

8.  The Pittsburgh Sleep Quality Index: a new instrument for psychiatric practice and research.

Authors:  D J Buysse; C F Reynolds; T H Monk; S R Berman; D J Kupfer
Journal:  Psychiatry Res       Date:  1989-05       Impact factor: 3.222

9.  Success Using Neuromodulation With BURST (SUNBURST) Study: Results From a Prospective, Randomized Controlled Trial Using a Novel Burst Waveform.

Authors:  Timothy Deer; Konstantin V Slavin; Kasra Amirdelfan; Richard B North; Allen W Burton; Thomas L Yearwood; Ed Tavel; Peter Staats; Steven Falowski; Jason Pope; Rafael Justiz; Alain Y Fabi; Alexander Taghva; Richard Paicius; Timothy Houden; Derron Wilson
Journal:  Neuromodulation       Date:  2017-09-29

Review 10.  Improving patient experience with spinal cord stimulation: implications of position-related changes in neurostimulation.

Authors:  Edgar Ross; David Abejón
Journal:  Neuromodulation       Date:  2011-12-02
View more
  7 in total

1.  A fully implantable wireless bidirectional neuromodulation system for mice.

Authors:  Jason P Wright; Ibrahim T Mughrabi; Jason Wong; Jose Mathew; Naveen Jayaprakash; Christine Crosfield; Eric H Chang; Sangeeta S Chavan; Kevin J Tracey; Valentin A Pavlov; Yousef Al-Abed; Theodoros P Zanos; Stavros Zanos; Timir Datta-Chaudhuri
Journal:  Biosens Bioelectron       Date:  2021-12-11       Impact factor: 12.545

2.  Research design considerations for randomized controlled trials of spinal cord stimulation for pain: Initiative on Methods, Measurement, and Pain Assessment in Clinical Trials/Institute of Neuromodulation/International Neuromodulation Society recommendations.

Authors:  Nathaniel Katz; Robert H Dworkin; Richard North; Simon Thomson; Sam Eldabe; Salim M Hayek; Brian H Kopell; John Markman; Ali Rezai; Rod S Taylor; Dennis C Turk; Eric Buchser; Howard Fields; Gregory Fiore; McKenzie Ferguson; Jennifer Gewandter; Chris Hilker; Roshini Jain; Angela Leitner; John Loeser; Ewan McNicol; Turo Nurmikko; Jane Shipley; Rahul Singh; Andrea Trescot; Robert van Dongen; Lalit Venkatesan
Journal:  Pain       Date:  2021-07-01       Impact factor: 6.961

3.  The Effect of Spinal Cord Stimulation Frequency on the Neural Response and Perceived Sensation in Patients With Chronic Pain.

Authors:  Gerrit Eduard Gmel; Rosana Santos Escapa; John L Parker; Dave Mugan; Adnan Al-Kaisy; Stefano Palmisani
Journal:  Front Neurosci       Date:  2021-01-21       Impact factor: 4.677

4.  Successful application of burst spinal cord stimulation for refractory upper limb pain: a case series.

Authors:  Kuen Su Lee; Yoo Kyung Jang; Gene Hyun Park; In Jae Jun; Jae Chul Koh
Journal:  J Int Med Res       Date:  2021-03       Impact factor: 1.671

Review 5.  A Call to Action Toward Optimizing the Electrical Dose Received by Neural Targets in Spinal Cord Stimulation Therapy for Neuropathic Pain.

Authors:  Krishnan Chakravarthy; Rajiv Reddy; Adnan Al-Kaisy; Thomas Yearwood; Jay Grider
Journal:  J Pain Res       Date:  2021-09-07       Impact factor: 2.832

Review 6.  Non-Spinal Neuromodulation of the Lumbar Medial Branch Nerve for Chronic Axial Low Back Pain: A Narrative Review.

Authors:  Vinicius Tieppo Francio; Benjamin D Westerhaus; Adam Rupp; Dawood Sayed
Journal:  Front Pain Res (Lausanne)       Date:  2022-02-25

Review 7.  Advances in Pain Medicine: a Review of New Technologies.

Authors:  Natalie Strand; Maloney J; Vinicius Tieppo Francio; Murphy M; Michal Turkiewicz; Antonios El Helou; Maita M; Covington S; Singh N; Peck J; Wie C
Journal:  Curr Pain Headache Rep       Date:  2022-07-29
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.