Accuracy of washout rate analysis for thallium-201 single-photon emission computed tomography myocardial perfusion imaging using cadmium zinc telluride detectors: A phantom study.

Masaru Ishihara1, Masahisa Onoguchi2, Takayuki Shibutani2.   

Abstract

OBJECTIVE: The aim of this study was to assess the accuracy of washout rate (WOR) analysis for thallium-201 chloride (201Tl) single-photon emission computed tomography myocardial perfusion imaging data acquired using cadmium zinc telluride detectors and a myocardial phantom.
METHODS: A myocardial phantom was injected with 10.5 MBq 201Tl, and 10-min acquisitions were performed at 0, 24, 46, and 62 h to accommodate natural radioactive decay over time. Global myocardial WOR (global-WOR) and regional WOR (regional-WOR, left anterior descending artery [LAD], right coronary artery [RCA], and left circumflex artery [LCX]) were analyzed between 0 and 24 h (infarction model), 0 and 46 h (ischemia model), and 0 and 62 h (normal model), respectively. We compared the calculated radioactive decay-rate as a reference standard and phantom imaging WOR (phantom-WOR).
RESULTS: Decay-rate versus phantom-WOR were 20.4% vs. 20.8% (global-WOR), 21.3% (LAD), 21.2% (RCA), and 19.7% (LCX) for the infarction model; 35.4% vs. 35.6% (global-WOR), 35.5% (LAD), 36.2% (RCA), and 35.2% (LCX) for the ischemia model; and 44.5% vs. 45.1% (global-WOR), 45.4% (LAD), 44.7% (RCA), and 43.5% (LCX) for the normal model.
CONCLUSION: WOR analysis for 201Tl single-photon emission computed tomography myocardial perfusion imaging using cadmium zinc telluride detectors is a reliable analysis method.
© The Author(s) 2020.

Entities:  

Keywords:  Thallium-201 chloride; cadmium zinc telluride detectors; single-photon emission computed tomography; washout rate

Year:  2020        PMID: 31976070      PMCID: PMC6958653          DOI: 10.1177/2048004019900600

Source DB:  PubMed          Journal:  JRSM Cardiovasc Dis        ISSN: 2048-0040


Introduction

Thallium-201 chloride (201Tl) is widely used as a single-photon emission computed tomography (SPECT) myocardial perfusion imaging (MPI) agent. 201Tl SPECT allows for noninvasive evaluation of coronary artery disease (CAD).[1,2] In 201Tl SPECT MPI, regional hypoperfusion can be evaluated using short-axis, vertical long-axis, and horizontal long-axis images. However, global hypoperfusion in myocardial ischemia (balanced ischemia and diffuse abnormal flow reserve) might be underestimated.[3,4] Washout rate (WOR) analysis for 201Tl SPECT MPI can help detect abnormalities in patients with apparently normal MPI.[5-7] For clinical use, it is important to confirm the accuracy of WOR analysis for 201Tl SPECT MPI. The aim of this study was to assess the accuracy of WOR analysis for 201Tl SPECT MPI data acquired using cadmium zinc telluride (CZT) detectors and a myocardial phantom.

Methods

A scanner with CZT detectors (D-SPECT cardiac scanner; Spectrum Dynamics, Caesarea, Israel) and a myocardial phantom (RH-2; Kyoto-Kagaku, Kyoto, Japan) were used. The phantom (Figure 1) was injected with 10.5 MBq 201Tl, and 10-min acquisitions were performed at 0, 24, 46, and 62 h to accommodate natural radioactive decay over time. Natural radioactive decay over time was used to calculate radioactivity, and the WOR for the abnormal model was <40%.[8] For the calculation analysis of the basic WOR, the first myocardial counts (stress imaging in clinical cases) and decreased second myocardial counts, due to physiological washout over time (redistribution imaging in clinical cases), were used. In our phantom study, myocardial counts at 24, 46, and 62 h were measured due to natural decay.
Figure 1.

Photograph showing the myocardial phantom and thallium-201 chloride (201Tl) injection site (arrow) in the myocardial area (187 mL).

Photograph showing the myocardial phantom and thallium-201 chloride (201Tl) injection site (arrow) in the myocardial area (187 mL). Imaging was conducted using an energy peak and energy window width of 71 keV ± 15%. For all acquisitions, the raw projection data were reconstructed using an iterative maximum likelihood expectation maximization algorithm (ML-EM) using a variant of the ordered-subset expectation maximization algorithm (OS-EM) using three and four iterations and 32 subsets.[9] The WORs were analyzed between 0 and 24 h (infarction model), 0 and 46 h (ischemia model), and 0 and 62 h (normal model) (Figure 2) using the scanner’s software (D-SPECT WOR Software, Spectrum Dynamics, Caesarea, Israel). D-SPECT WOR software decay correction was not employed during WOR analysis in this study, and natural decay was used to calculate the WOR because decay correction underestimates the WOR.
Figure 2.

Models of 201Tl for various levels of myocardial blood flow: infarction model (N0–N24), ischemia model (N0–N46), and normal model (N0–N62).

Models of 201Tl for various levels of myocardial blood flow: infarction model (N0–N24), ischemia model (N0–N46), and normal model (N0–N62). The formula for calculating radioactivity from the images at 0, 24, 46, and 62 h was where N = radioactivity at time t, N0 = initial radioactivity, t = time (h), T = 72.97 h. The formula for decay-rate in the infarction model, ischemia model, and normal model was The same formula was used to calculate the WOR in all models. The initial (N0) radioactivity in the myocardial phantom was 10.5 MBq. The 24 h (N24) radioactivity of the myocardial phantom was 8.4 MBq (= N0[1/2]24.00/72.97), 46 h (N46) radioactivity of the myocardial phantom was 6.8 MBq (= N0[1/2]46.00/72.97), and 62 h (N62) radioactivity of the myocardial phantom was 5.8 MBq (= N0[1/2]62.00/72.97). The 0–24 h decay-rate was 20.4% (= [[N0–N24]/N0] × 100), 0–46 h decay-rate was 35.4% (= [[N0–N46]/N0] × 100), and 0–62 h decay-rate was 44.5% (= [[N0–N62]/N0] × 100).[8,10] Differences between the decay-rate as a reference standard and the phantom imaging WOR (phantom-WOR) were compared. Phantom-WORs were used for the global myocardial WOR (global-WOR) and coronary artery regions (regional-WOR). All WORs were automatically calculated using the D-SPECT WOR software. Regional-WORs were divided into three main territories: left anterior descending artery (LAD), right coronary artery (RCA), and left circumflex artery (LCX). The left ventricle was divided into 17 segments using the American Heart Association model[11] with the LAD supplying segments 1, 2, 7, 8, 13, 14, and 17; RCA supplying segments 3, 4, 9, 10, and 15; and LCX supplying segments 5, 6, 11, 12, and 16. Statistical analyses were performed using spreadsheet software (Microsoft Excel 2016, Microsoft Corp., Redmond WA, USA). The error was obtained as the absolute error. It is shown as follows:

Results

The reconstructed myocardial counts were 2.44 million at 0 h, 1.92 million at 24 h, 1.54 million at 46 h, and 1.33 million at 62 h. Decay-rate, phantom-WOR, and absolute error for each model are shown in Tables 1 and 2. Absolute error ranged from 0.2% to 0.6% for global-WOR, and from 0.1% to 1.0% for regional-WOR.
Table 1.

Global-WOR values.

Infarction modelIschemia modelNormal model
Decay-rate (%)20.435.444.5
Phantom-WOR (%)20.835.645.1
Absolute error (%)0.40.20.6

Decay-rate: calculated natural radioactive decay rate; Phantom-WOR: phantom imaging washout rate

Table 2.

Regional-WOR values.

Infarction model
Ischemia model
Normal model
LADRCALCXLADRCALCXLADRCALCX
Decay-rate (%)20.435.444.5
Phantom-WOR (%)21.321.219.735.536.235.245.444.743.5
Absolute error (%)0.90.80.70.10.80.20.90.21.0

LAD: left anterior descending artery, RCA: right coronary artery, LCX: left circumferential artery, Decay-rate: calculated natural radioactive decay rate; Phantom-WOR: phantom imaging washout rate

Global-WOR values. Decay-rate: calculated natural radioactive decay rate; Phantom-WOR: phantom imaging washout rate Regional-WOR values. LAD: left anterior descending artery, RCA: right coronary artery, LCX: left circumferential artery, Decay-rate: calculated natural radioactive decay rate; Phantom-WOR: phantom imaging washout rate Phantom-WOR images are shown in Figure 3.
Figure 3.

Phantom-WOR analysis images of 201Tl SPECT MPI: infarction model (a), ischemia model (b), and normal model (c).

Phantom-WOR analysis images of 201Tl SPECT MPI: infarction model (a), ischemia model (b), and normal model (c).

Discussion

Recently, dedicated cardiac systems with CZT detectors were introduced in Japan. The improvements in efficiency and resolution allow low-dose imaging protocols, reduce the acquisition time, and permit quantification of the myocardial perfusion reserve using dynamic SPECT images.[12-14] 201Tl is widely used for SPECT MPI in many Asian countries. 201Tl SPECT MPI WOR is also a valuable tool when diagnosing CAD.[7,15] There are methods of calculating WOR from planar and SPECT images.[15-17] In this study, we used CZT detectors and a myocardial phantom and assessed the accuracy of WOR analysis for 201Tl SPECT MPI. We observed the natural radioactive decay that occurs over time and created an infarction model, ischemia model, and normal model. In our previous study,[18] 201Tl SPECT MPI required a counts of ≥1.2 million. Therefore, the myocardial counts at 0 h were ≥2.20 million, and at 62 h (used for the normal model) were calculated to be ≥1.20 million, with an acquisition time set to 10-min in this study (to record natural radioactive decay). In clinical cases, the decrease in reproducibility can be due to repositioning effects between the first and second acquisitions. A previous study showed that reproducibility was unaffected by the myocardial phantom and clinical cases in several instances.[19] In the present study, inaccuracy of misalignment was controlled for by keeping the CZT detectors and the myocardial phantom immobile over acquisitions performed at 24, 46, and 62 h. We found an absolute error of decay-rate and phantom-WOR for both global-WOR and regional-WOR to be ≤1.0%, with excellent accuracy of the WOR analysis results. This study elucidated the accuracy of the WOR using the new CZT system. In the future, it is expected to have clinical application in the diagnosis of disease, especially multivessel disease and slightly abnormal myocardial perfusion.

Conclusion

The WOR analysis for 201Tl SPECT MPI using CZT detectors is a reliable analysis method.
  19 in total

Review 1.  Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. A statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association.

Authors:  Manuel D Cerqueira; Neil J Weissman; Vasken Dilsizian; Alice K Jacobs; Sanjiv Kaul; Warren K Laskey; Dudley J Pennell; John A Rumberger; Thomas Ryan; Mario S Verani
Journal:  Circulation       Date:  2002-01-29       Impact factor: 29.690

2.  Negative washout rate of myocardial thallium-201--a specific marker for high grade coronary artery narrowing.

Authors:  M Yamada; T Chikamori; Y Doi; Y Yonezawa; H Seo; T Maeda; T Ozawa
Journal:  Jpn Circ J       Date:  1992-10

3.  Low washout rate during stress thallium-201 myocardial scintigraphy. Transient left ventricular dysfunction in a patient with grant splenomegaly due to polycytemia vera.

Authors:  Naoyuki Sata; Yasuhiro Tanaka; Katsunori Toufuku; Katsurou Kashima; Kenkichi Miyahara
Journal:  J Cardiol       Date:  2005-07       Impact factor: 3.159

4.  Underestimation of extent of ischemia by gated SPECT myocardial perfusion imaging in patients with left main coronary artery disease.

Authors:  Daniel S Berman; Xingping Kang; Piotr J Slomka; James Gerlach; Ling de Yang; Sean W Hayes; John D Friedman; Louise E J Thomson; Guido Germano
Journal:  J Nucl Cardiol       Date:  2007-07       Impact factor: 5.952

5.  Improved noninvasive assessment of coronary artery disease by quantitative analysis of regional stress myocardial distribution and washout of thallium-201.

Authors:  J Maddahi; E V Garcia; D S Berman; A Waxman; H J Swan; J Forrester
Journal:  Circulation       Date:  1981-11       Impact factor: 29.690

6.  Comparison of myocardial perfusion imaging using thallium-201 between a new cadmium-zinc-telluride cardiac camera and a conventional SPECT camera.

Authors:  Bernard Songy; David Lussato; Mohamed Guernou; Mathieu Queneau; Ricardo Geronazzo
Journal:  Clin Nucl Med       Date:  2011-09       Impact factor: 7.794

7.  High-speed myocardial perfusion imaging initial clinical comparison with conventional dual detector anger camera imaging.

Authors:  Tali Sharir; Simona Ben-Haim; Konstantine Merzon; Vitali Prochorov; Dalia Dickman; Shlomo Ben-Haim; Daniel S Berman
Journal:  JACC Cardiovasc Imaging       Date:  2008-03

8.  Increased (99m)Tc-sestamibi washout reflects impaired myocardial contractile and relaxation reserve during dobutamine stress due to mitochondrial dysfunction in dilated cardiomyopathy patients.

Authors:  Daisuke Hayashi; Satoru Ohshima; Satoshi Isobe; Xian Wu Cheng; Kazumasa Unno; Hidehito Funahashi; Norihiro Shinoda; Takahiro Okumura; Akihiro Hirashiki; Katsuhiko Kato; Toyoaki Murohara
Journal:  J Am Coll Cardiol       Date:  2013-03-13       Impact factor: 24.094

9.  Intra-patient reproducibility of myocardial SPECT imaging with 201Tl.

Authors:  Barbra E Backus; Frederik A Verburg; R Leo Romijn; Mark W Konijnenberg; Freek J Beekman; J Fred Verzijlbergen
Journal:  J Nucl Cardiol       Date:  2009-01-20       Impact factor: 5.952

Review 10.  Standardization of 123I-meta-iodobenzylguanidine myocardial sympathetic activity imaging: phantom calibration and clinical applications.

Authors:  Kenichi Nakajima; Derk O Verschure; Koichi Okuda; Hein J Verberne
Journal:  Clin Transl Imaging       Date:  2017-05-04
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.