| Literature DB >> 31976070 |
Masaru Ishihara1, Masahisa Onoguchi2, Takayuki Shibutani2.
Abstract
OBJECTIVE: The aim of this study was to assess the accuracy of washout rate (WOR) analysis for thallium-201 chloride (201Tl) single-photon emission computed tomography myocardial perfusion imaging data acquired using cadmium zinc telluride detectors and a myocardial phantom.Entities:
Keywords: Thallium-201 chloride; cadmium zinc telluride detectors; single-photon emission computed tomography; washout rate
Year: 2020 PMID: 31976070 PMCID: PMC6958653 DOI: 10.1177/2048004019900600
Source DB: PubMed Journal: JRSM Cardiovasc Dis ISSN: 2048-0040
Figure 1.Photograph showing the myocardial phantom and thallium-201 chloride (201Tl) injection site (arrow) in the myocardial area (187 mL).
Figure 2.Models of 201Tl for various levels of myocardial blood flow: infarction model (N0–N24), ischemia model (N0–N46), and normal model (N0–N62).
Global-WOR values.
| Infarction model | Ischemia model | Normal model | |
|---|---|---|---|
| Decay-rate (%) | 20.4 | 35.4 | 44.5 |
| Phantom-WOR (%) | 20.8 | 35.6 | 45.1 |
| Absolute error (%) | 0.4 | 0.2 | 0.6 |
Decay-rate: calculated natural radioactive decay rate; Phantom-WOR: phantom imaging washout rate
Regional-WOR values.
| Infarction model | Ischemia model | Normal model | |||||||
|---|---|---|---|---|---|---|---|---|---|
| LAD | RCA | LCX | LAD | RCA | LCX | LAD | RCA | LCX | |
| Decay-rate (%) | 20.4 | 35.4 | 44.5 | ||||||
| Phantom-WOR (%) | 21.3 | 21.2 | 19.7 | 35.5 | 36.2 | 35.2 | 45.4 | 44.7 | 43.5 |
| Absolute error (%) | 0.9 | 0.8 | 0.7 | 0.1 | 0.8 | 0.2 | 0.9 | 0.2 | 1.0 |
LAD: left anterior descending artery, RCA: right coronary artery, LCX: left circumferential artery, Decay-rate: calculated natural radioactive decay rate; Phantom-WOR: phantom imaging washout rate
Figure 3.Phantom-WOR analysis images of 201Tl SPECT MPI: infarction model (a), ischemia model (b), and normal model (c).