Bioelectronic neural interfaces can fail in vivo due to water penetration and corrosion of the packaging technology used to protect sensitive portions of the device. Although anisotropic conductive adhesive (ACA) is gaining popularity in the neural interface community to connect fabricated electrode arrays with back-end packages, the durability of ACA in chronic implants is largely unknown. We have designed a platform that uses an aggressive reactive-accelerated aging (RAA) environment to rapidly assess the ability of ACA and silicone-rubber encapsulation to maintain electrical integrity in vitro. All RAA experiments were performed at 77°C, for 24 days, and with 10 to 20 mM H2O2, which approximates a 1 year implantation. Results from these experiments showed that ACA rapidly fails (i.e., 2 to 4 days RAA) due to water absorption through the silicone encapsulant. Electrical impedance spectroscopy (EIS) confirmed water penetration through the package and the resulting corrosion of the sensitive metallic components.

Entities:  

Mesh:

Substances:

Year:  2019        PMID: 31946694      PMCID: PMC8059779          DOI: 10.1109/EMBC.2019.8856692

Source DB:  PubMed          Journal:  Annu Int Conf IEEE Eng Med Biol Soc        ISSN: 2375-7477


  12 in total

1.  Chronic neural recording using silicon-substrate microelectrode arrays implanted in cerebral cortex.

Authors:  Rio J Vetter; Justin C Williams; Jamille F Hetke; Elizabeth A Nunamaker; Daryl R Kipke
Journal:  IEEE Trans Biomed Eng       Date:  2004-06       Impact factor: 4.538

2.  Investigation on the hermeticity of an implantable package with 32 feedthroughs for neural prosthetic applications.

Authors:  Lorenzo Bisoni; Matthias Mueller; Paul Cvancara; Caterina Carboni; Roberto Puddu; Luigi Raffo; Massimo Barbaro; Thomas Stieglitz
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2016-08

3.  Pressure-volume-temperature data for oxygen.

Authors:  C H MEYERS
Journal:  J Res Natl Bur Stand (1934)       Date:  1948

4.  Investigations on different epoxies for electrical insulation of microflex structures.

Authors:  P Cvancara; S Lauser; L Rudmann; T Stieglitz
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2016-08

5.  Rapid evaluation of the durability of cortical neural implants using accelerated aging with reactive oxygen species.

Authors:  Pavel Takmakov; Kiersten Ruda; K Scott Phillips; Irada S Isayeva; Victor Krauthamer; Cristin G Welle
Journal:  J Neural Eng       Date:  2015-01-28       Impact factor: 5.379

6.  Chronic in vivo stability assessment of carbon fiber microelectrode arrays.

Authors:  Paras R Patel; Huanan Zhang; Matthew T Robbins; Justin B Nofar; Shaun P Marshall; Michael J Kobylarek; Takashi D Y Kozai; Nicholas A Kotov; Cynthia A Chestek
Journal:  J Neural Eng       Date:  2016-10-05       Impact factor: 5.379

7.  Long-term stability of neural prosthetic control signals from silicon cortical arrays in rhesus macaque motor cortex.

Authors:  Cynthia A Chestek; Vikash Gilja; Paul Nuyujukian; Justin D Foster; Joline M Fan; Matthew T Kaufman; Mark M Churchland; Zuley Rivera-Alvidrez; John P Cunningham; Stephen I Ryu; Krishna V Shenoy
Journal:  J Neural Eng       Date:  2011-07-20       Impact factor: 5.379

8.  Neural control of cursor trajectory and click by a human with tetraplegia 1000 days after implant of an intracortical microelectrode array.

Authors:  J D Simeral; S-P Kim; M J Black; J P Donoghue; L R Hochberg
Journal:  J Neural Eng       Date:  2011-03-24       Impact factor: 5.379

9.  Accelerated aging for testing polymeric biomaterials and medical devices.

Authors:  D W L Hukins; A Mahomed; S N Kukureka
Journal:  Med Eng Phys       Date:  2008-08-09       Impact factor: 2.242

10.  in vitro Reactive-Accelerated-Aging (RAA) Assessment of Tissue-Engineered Electronic Nerve Interfaces (TEENI).

Authors:  Cary A Kuliasha; Jack W Judy
Journal:  Annu Int Conf IEEE Eng Med Biol Soc       Date:  2018-07
View more
  3 in total

1.  Microgaskets for High-Channel-Density Reconnectable Implantable Packaging.

Authors:  Paritosh Rustogi; Jack W Judy
Journal:  J Microelectromech Syst       Date:  2022-03-22       Impact factor: 2.829

2.  Electrical Isolation Performance of Microgasket Technology for Implant Packaging.

Authors:  Paritosh Rustogi; Jack W Judy
Journal:  Electron Compon Technol Conf       Date:  2020-08-05

3.  Bonding methods for chip integration with Parylene devices.

Authors:  James Yoo; Ellis Meng
Journal:  J Micromech Microeng       Date:  2021-02-19       Impact factor: 2.282

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.