Verónica García-Vázquez1, Felipe A Calvo1,2,3,4, María J Ledesma-Carbayo5,6, Claudio V Sole1,7, José Calvo-Haro1,8,9, Manuel Desco1,10,11,12, Javier Pascau1,10.
Abstract
In intraoperative electron radiation therapy (IOERT) the energy of the electron beam is selected under the conventional assumption of water-equivalent tissues at the applicator end. However, the treatment field can deviate from the theoretic flat irradiation surface, thus altering dose profiles. This patient-based study explored the feasibility of acquiring intraoperative computed tomography (CT) studies for calculating three-dimensional dose distributions with two factors not included in the conventional assumption, namely the air gap from the applicator end to the irradiation surface and tissue heterogeneity. In addition, dose distributions under the conventional assumption and from preoperative CT studies (both also updated with intraoperative data) were calculated to explore whether there are other alternatives to intraoperative CT studies that can provide similar dose distributions. The IOERT protocol was modified to incorporate the acquisition of intraoperative CT studies before radiation delivery in six patients. Three studies were not valid to calculate dose distributions due to the presence of metal artefacts. For the remaining three cases, the average gamma pass rates between the doses calculated from intraoperative CT studies and those obtained assuming water-equivalent tissues or from preoperative CT studies were 73.4% and 74.0% respectively. The agreement increased when the air gap was included in the conventional assumption (98.1%) or in the preoperative CT images (98.4%). Therefore, this factor was the one mostly influencing the dose distributions of this study. Our experience has shown that intraoperative CT studies are not recommended when the procedure includes the use of shielding discs or surgical retractors unless metal artefacts are removed. IOERT dose distributions calculated under the conventional assumption or from preoperative CT studies may be inaccurate unless the air gap (which depends on the surface irregularities of the irradiated volume and on the applicator pose) is included in the calculations.Entities:
Mesh:
Year: 2020 PMID: 31923183 PMCID: PMC6953834 DOI: 10.1371/journal.pone.0227155
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
IOERT data and CT acquisition parameters.
| IOERT DATA | CT ACQUISITION PARAMETERS | ||||||
|---|---|---|---|---|---|---|---|
| (In all acquisitions, voltage 120 kVp) | |||||||
| 8 | 15° | 6 | 10 | 35 ± 7 | 1.3 x 1.3 x 5.0 | 6 | |
| 125 | 1.3 x 1.3 x 2.0 | ||||||
| 12 | 30° | 8 | 12.5 | 125 | 1.1 x 1.1 x 2.0 | 1 | |
| 1.6 x 1.6 x 2.0 | |||||||
| 5 | 30° | 6 | 10 | 100 | 0.6 x 0.6 x 1.6 | 17 | |
| 127 ± 66 | 1.3 x 1.3 x 3.0 | ||||||
| 5 | 0° | 6 | 10 | 100 | 0.7 x 0.7 x 1.6 | 9 | |
| 114 ± 27 | 1.1 x 1.1 x 5.0 | ||||||
| 10 | 30° | 8 | 12.5 | 196 ± 22 | 0.7 x 0.7 x 1.0 | 1 | |
| 217 ± 7 | 1.4 x 1.4 x 2.0 | ||||||
| 7 | 30° | 9 | 12.5 | 132 ± 62 | 1.2 x 1.2 x 2.0 | 0 | |
| 143 ± 54 | 0.9 x 0.9 x 2.0 | ||||||
aMean ± standard deviation.
bPreoperative CT image.
cIntraoperative CT image.
Fig 1Intraoperative images with metal artefacts.
Volume rendering view (top) and axial view (bottom). (A) Patient 3 (breast cancer, right). (B) Patient 4 (breast cancer, left). (C) Patient 5 (retroperitoneal sarcoma).
Percentage of voxels fulfilling 3D gamma criteria of 3%/3 mm.
Gamma pass rates ≥ 95% highlighted in bold.
| Dose > 10% | Dose > 70% | ||
|---|---|---|---|
| D_intraCT v D_preCTwater | 81.9 | 81.5 | |
| D_intraCT v D_preCT | 81.8 | 81.6 | |
| D_intraCT v D_preCTwater&air | |||
| D_intraCT v D_preCTair | |||
| D_intraCT v D_preCTwater | 90.0 | 92.1 | |
| D_intraCT v D_preCT | 92.3 | 92.2 | |
| D_intraCT v D_preCTwater&air | |||
| D_intraCT v D_preCTair | |||
| D_intraCT v D_preCTwater | 52.6 | 42.0 | |
| D_intraCT v D_preCT | 53.6 | 42.5 | |
| D_intraCT v D_preCTwater&air | |||
| D_intraCT v D_preCTair | 94.5 |
Dose matrices: 183 x 235 x 175, 180 x 156 x 256 and 200 x 179 x 250 for Patient 1, Patient 2 and Patient 6 respectively. Voxel size 1.5 x 1.5 x 1.5 mm.
Average gamma pass rates: 73.4%, 74.0%, 98.1% and 98.4% for D_preCTwater, D_preCT, D_preCTwater&air and D_preCTair respectively.
aAverage gamma pass rate: 91.1%.
bAverage gamma pass rate: 47.3%.
Fig 2Dose distributions calculated using the Monte Carlo algorithm.
(A) Patient 1 (Ewing sarcoma, coronal view). (B) Patient 2 (rhabdomyosarcoma, sagittal view). (C) Patient 6 (chondrosarcoma, axial view). H (head), F (feet), A (anterior), P (posterior), R (right) and L (left).