Assessment of clinically relevant dose distributions in pelvic IOERT using Gafchromic EBT3 films.

Filipa Costa1, Sandra Sarmento2, Olga Sousa3.   

Abstract

PURPOSE: In IOERT a single dose of radiation is delivered to the tumour site during surgery. Manual dose calculations are used and the irradiation target volume, electron energy and applicator are decided on site by the radiation oncologist. This work assesses the effect that irregular and curved surfaces, typical of pelvic IOERT, may have on the expected dose distribution.
METHODS: The feasibility of using Gafchromic EBT3 films and a slab phantom to obtain 2D dose distributions was investigated. Different set-ups were tested by comparison with water tank measurements, applying the gamma function analysis with 2% and 2 mm criteria. The validated set-up was then used to obtain reference dose distributions, which were converted to colour-coded graphical representations. Phantoms with step-like and curved surfaces were created to simulate typical pelvic IOERT irradiation surfaces, and the dose distributions were obtained and compared with the reference distributions.
RESULTS: Good agreement with water tank measurements was obtained for all applicators below 2 mm, using the chosen setup in reference conditions. In non-reference conditions, the presence of a step-like surface creates an adjacent hotspot, followed by a quick reduction of the dose in depth. With curved surfaces, the dose distribution is shifted forward, becoming curved and deeper, but when the applicator is larger than the hole, hotspots are also observed.
CONCLUSIONS: The shape of the irradiation surfaces alters the dose distribution. Visualization of these effects is important to assess target coverage and interpret in vivo measurements in pelvic IOERT.
Copyright © 2015 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Dose distributions; Gafchromic EBT3 films; Pelvic IOERT; Rectal cancer

Mesh:

Year:  2015        PMID: 26078013     DOI: 10.1016/j.ejmp.2015.05.013

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  4 in total

1.  In vivo dosimetry using Gafchromic films during pelvic intraoperative electron radiation therapy (IOERT).

Authors:  Filipa Costa; Sandra Sarmento; Dora Gomes; Helena Magalhães; Rosário Arrais; Graciete Moreira; Maria Fátima Cruz; José Pedro Silva; Lúcio Santos; Olga Sousa
Journal:  Br J Radiol       Date:  2016-05-18       Impact factor: 3.039

2.  Surface scanning for 3D dose calculation in intraoperative electron radiation therapy.

Authors:  Verónica García-Vázquez; Begoña Sesé-Lucio; Felipe A Calvo; Juan J Vaquero; Manuel Desco; Javier Pascau
Journal:  Radiat Oncol       Date:  2018-12-07       Impact factor: 3.481

3.  Intraoperative computed tomography imaging for dose calculation in intraoperative electron radiation therapy: Initial clinical observations.

Authors:  Verónica García-Vázquez; Felipe A Calvo; María J Ledesma-Carbayo; Claudio V Sole; José Calvo-Haro; Manuel Desco; Javier Pascau
Journal:  PLoS One       Date:  2020-01-10       Impact factor: 3.240

4.  Treatment Planning in Intraoperative Radiation Therapy (IORT): Where Should We Go?

Authors:  Carlo Cavedon; Renzo Mazzarotto
Journal:  Cancers (Basel)       Date:  2022-07-20       Impact factor: 6.575

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.