| Literature DB >> 31886158 |
Ronald J Dandurand1,2, Jean-Pierre Lavoie3, Larry C Lands1, Zoltán Hantos4.
Abstract
Noninvasiveness, low cooperation demand and the potential for detailed physiological characterisation have promoted the use of oscillometry in the assessment of lung function. However, concerns have been raised about the comparability of measurement outcomes delivered by the different oscillometry devices. The present study compares the performances of oscillometers in the measurement of mechanical test loads with and without simulated breathing. Six devices (five were commercially available and one was custom made) were tested with mechanical test loads combining resistors (R), gas compliances (C) and a tube inertance (L), to mimic respiratory resistance (R rs) and reactance (X rs) spectra encountered in clinical practice. A ventilator was used to simulate breathing at tidal volumes of 300 and 700 mL at frequencies of 30 and 15 min-1, respectively. Measurements were evaluated in terms of R, C, L, resonance frequency (f res), reactance area (AX ) and resistance change between 5 and 20 or 19 Hz (R 5-20(19)). Increasing test loads caused progressive deviations in R rs and X rs from calculated values at various degrees in the different oscillometers. While mean values of R rs were recovered acceptably, some devices exhibited serious distortions in the frequency dependences of R rs and X rs, leading to large errors in C, L, f res, AX and R 5-20(19). The results were largely independent of the simulated breathing. Simplistic calibration procedures and mouthpiece corrections, in addition to unknown instrumental and signal processing factors, may be responsible for the large differences in oscillometry measures. Rigorous testing and ongoing harmonisation efforts are necessary to better exploit the diagnostic and scientific potential of oscillometry.Entities:
Year: 2019 PMID: 31886158 PMCID: PMC6926364 DOI: 10.1183/23120541.00160-2019
Source DB: PubMed Journal: ERJ Open Res ISSN: 2312-0541
FIGURE 1Schematic arrangement of the mechanical test loads.
Mechanical test loads (M1–M6) as combinations of resistors and compliances (the same inertance tube was used in all test loads) and ventilation parameters
| 700 | 700 | 700 | 300 | 300 | 300 | |
| 15 | 15 | 15 | 30 | 30 | 30 | |
#: measured at 50 mL·s−1; ¶: measured at the mean flow of the ventilator setting.
FIGURE 2Representative tracing of pressure, flow and volume (blue) signals with “intra-breath” resistance (black) and reactance (red) computed at an 11-Hz oscillation frequency in test load M2.
Types and frequency contents of the oscillation signals
| 5, 11, 13, 17, 19, 23, 29, 31, 37 | Pseudo-random, relative primes | 16 s | |
| 5, 11, 13, 17, 19, 23, 29, 31, 37 | Pseudo-random, relative primes | 16 s | |
| 5, 10, 15, 20, 25, 35 | Recurrent impulses | 30 s | |
| 5, 10, 15, 20, 25, 30, 35 | Recurrent impulses | 60 s | |
| 5, 10, 15, 20, 25, 30, 35 | Pseudo-random | 60 s | |
| 4, 6, 8, …, 32, 34, 36 | Pseudo-random | 8 s | |
| 5, 7, 11, 13, 17, 19, 23, 29, 31, 37 | Pseudo-random, relative primes | 30 s | |
| 5, 11, 19 | Pseudo-random, relative primes | 5 breaths |
FIGURE 3Impedance measurements in the mechanical test loads a) M1 and b) M6.
FIGURE 4Values of a) resistance (R), b) compliance (C), c) inertance (L) and d) the fitting error (F) obtained from model fitting to impedance data measured in the mechanical test loads M1–M6, with the different devices and modes. Horizontal lines indicate values from fitting to calculated impedances. #: C>0.06 L·hPa−1.
FIGURE 5Values of a) resonance frequency (fres), b) reactance area (A) and c) the frequency dependence of resistance (R5–20(19)) calculated from impedance data measured in the mechanical test loads M1–M6 with the different devices and modes. Horizontal lines indicate values obtained from the calculated impedance data. Zero crossing of reactance occurred below (#) or above the measured frequency range (¶).