Literature DB >> 34761000

Normal limits for oscillometric bronchodilator responses and relationships with clinical factors.

Kanika Jetmalani1, Nathan J Brown1,2, Chantale Boustany1,3, Brett G Toelle1,4, Guy B Marks1,5,6, Michael J Abramson7, David P Johns8, Alan L James9,10,11, Michael Hunter9,12, Arthur W Musk9,10, Norbert Berend1, Claude S Farah1,13, David G Chapman1,14, Cindy Thamrin1, Gregory G King1,4.   

Abstract

INTRODUCTION: We aimed to determine normal thresholds for positive bronchodilator responses for oscillometry in an Australian general population sample aged ≥40 years, to guide clinical interpretation. We also examined relationships between bronchodilator responses and respiratory symptoms, asthma diagnosis, smoking and baseline lung function.
METHODS: Subjects recruited from Sydney, Melbourne and Busselton, Australia, underwent measurements of spirometry, resistance (R rs6 ) and reactance (X rs6 ) at 6 Hz, before and after inhalation of salbutamol 200 μg. Respiratory symptoms and/or medication use, asthma diagnosis, and smoking were recorded. Threshold bronchodilator responses were defined as the fifth percentile of decrease in R rs6 and 95th percentile increase in X rs6 in a healthy subgroup.
RESULTS: Of 1318 participants, 1145 (570 female) were analysed. The lower threshold for ΔR rs6 was -1.38 cmH2O·s·L-1 (-30.0% or -1.42 Z-scores) and upper threshold for ΔX rs6 was 0.57 cmH2O·s·L-1 (1.36 Z-scores). Respiratory symptoms and/or medication use, asthma diagnosis, and smoking all predicted bronchodilator response, as did baseline oscillometry and spirometry. When categorised into clinically relevant groups according to those predictors, ΔX rs6 was more sensitive than spirometry in smokers without current asthma or chronic obstructive pulmonary disease (COPD), ∼20% having a positive response. Using absolute or Z-score change provided similar prevalences of responsiveness, except in COPD, in which responsiveness measured by absolute change was twice that for Z-score. DISCUSSION: This study describes normative thresholds for bronchodilator responses in oscillometry parameters, including intra-breath parameters, as determined by absolute, relative and Z-score changes. Positive bronchodilator response by oscillometry correlated with clinical factors and baseline function, which may inform the clinical interpretation of oscillometry.
Copyright ©The authors 2021.

Entities:  

Year:  2021        PMID: 34761000      PMCID: PMC8573235          DOI: 10.1183/23120541.00439-2021

Source DB:  PubMed          Journal:  ERJ Open Res        ISSN: 2312-0541


Introduction

Reversibility of airflow obstruction in response to a bronchodilator is a hallmark of asthma and is commonly seen in patients with chronic obstructive pulmonary disease (COPD) [1]. In patients with asthma, bronchodilator responsiveness is useful for diagnosis and is associated with a greater risk of future adverse events [2]. In patients with COPD, bronchodilator responsiveness is only modestly repeatable [1, 3] and correlates with: forced expiratory volume in 1 s (FEV1) improvement following oral corticosteroids [4]; with response to inhaled corticosteroid/long-acting bronchodilator combination treatment (albeit inconsistently) [5, 6]; and with lower risk of hospitalisation and mortality in patients with severe COPD [7]. Although spirometry is the gold standard for bronchodilator testing, it likely misses clinically important responses. For example, reduced hyperinflation to a short-acting bronchodilator is greater with more severe COPD, whereas the FEV1 response is less [8]. Respiratory system impedance measured by oscillometry (also known as the forced oscillation technique (FOT)) is a sensitive way to measure bronchodilator responses. There are currently insufficient data on oscillometric responses to bronchodilators in healthy subjects [9] and consequently, little data on its prevalence in disease are available. Oscillometry involves low amplitude pressure oscillations applied to the mouth during tidal breathing. This is an important physiological difference compared with spirometry in which the forced manoeuvre of spirometry may induce airway collapse and expiratory flow limitation, or complete closure. This likely explains the poor correlation between spirometric and oscillometric bronchodilator responses [10]. The resistance of the respiratory system (Rrs) reflects airway calibre, while respiratory system reactance (Xrs) reflects predominantly the oscillatory stiffness of the respiratory system. Oscillatory stiffness is thus a dynamic stiffness (as opposed to static) and, as such, is sensitive to heterogeneously distributed airway narrowing and closure [11-13]. The literature suggests that Rrs and Xrs are more sensitive than spirometry in detecting bronchodilator responses in asthma [14-16], smokers without COPD [17] and in COPD [18], which also correlate with changes in ventilation distribution seen on lung imaging [19, 20]. However, further data on cut-off values are needed to inform clinical interpretation. There is a single published study that includes sufficient participants to allow the reliable estimation of cut-off values. This study involved five different devices used in healthy populations across four countries [21]. While this facilitated applicability of the derived cut-offs, there were some measurement differences between devices, which complicated the interpretation of the values. Thus, we aimed to examine bronchodilator responses using a single device to derive cut-off values for bronchodilator responsiveness, and determine its relationship with respiratory symptoms, asthma diagnosis, smoking and spirometric airflow obstruction in a general population sample over the age of 40 years. We hypothesised that oscillometry would be more sensitive than spirometry in detecting bronchodilator responses, and that bronchodilator responsiveness measured by oscillometry was related to symptoms, asthma diagnosis, smoking and spirometric airflow obstruction. This study was a retrospective analysis of the data obtained in the Australian arm of the Burden of Obstructive Lung Diseases (BOLD) study, which was part of a multi-centre study of the prevalence of spirometrically determined airflow obstruction [22, 23]. Parts of this analysis have been published previously in abstract form [24] and oscillometry data from the Busselton site have been published [25].

Methods

Study population

Subjects 40 years or older were recruited randomly from three cities across Australia (Sydney in New South Wales, Melbourne in Victoria and Busselton in Western Australia). Details on how the study populations were sampled are in the supplementary material. Study operations and testing methods were consistent across sites, which were part of the global BOLD project [23], with added local tests and questionnaires. The study was approved by the Human Research Ethics Committee of the University of Sydney (ref. 12–2006/9724). All subjects gave informed, written consent.

Study design

This was a cross-sectional study of oscillometric bronchodilator responses in the BOLD cohort in which the limits of responsiveness in healthy subjects were defined, then applied to the remainder of the cohort to determine the prevalence of increased bronchodilator responsiveness and its relationships with symptoms, doctor diagnosis of asthma and smoking. All subjects underwent oscillometry and spirometry measurements, in that order, at baseline and 15 min after the administration of 200 μg salbutamol administered by metered-dose inhaler through a spacer. Participants had all respiratory medications withheld on the day of testing; with short-acting bronchodilator inhalers for ≥6 h before testing. Respiratory symptoms, medication use and smoking history were obtained using the BOLD core questionnaire. To define cut-off values we identified a healthy group (HealthyAsym) as those with no respiratory symptoms or inhaled medications use in the past year, no doctor diagnosis ever of either asthma or COPD ever, currently not smoking and having less than 10 pack-years past smoking. Spirometry was not used to define this group as we wanted to remain consistent with definitions used in studies of normal spirometry, such as the Global Lung Function Initiative (GLI). Normative equations for oscillometry parameters were developed from this healthy group, based on previously published methodology [25], from which Z-scores for bronchodilator responsiveness were calculated.

Oscillometry

The oscillometry device was built in-house and has been described previously [26]. A detailed description is presented in the supplementary material. In brief, a multi-frequency pressure oscillation (6, 11 and 19 Hz) was imposed at the mouth. After establishing stable tidal breathing, a single 60-s recording was acquired with subjects supporting their own cheeks. Only the impedance parameters at 6 Hz were analysed in this study. Quality control procedures were applied as previously described [27]. Resistance was expressed as the mean across the entire recording (Rrs), and also separately for inspiration only (Rrs). Similarly, reactance was expressed as the mean (Xrs) and Xrs. Mean Xrs minus mean expiratory Xrs was calculated as an index of expiratory flow limitation (EFLi) [28]. Z-scores were calculated for each pre-bronchodilator and post-bronchodilator measurement, for each participant, based on age- and sex-specific expected mean and standard deviation values in the HealthyAsym group.

Spirometry

Spirometry was performed according to American Thoracic Society/European Respiratory Society Task Force criteria [29], using an EasyOne Plus hand-held spirometer (NDD Medical Technologies, Andover, MA, USA). All spirograms were reviewed by one study investigator (D.P. Johns) who assigned a standardised quality score. The highest recorded FEV1 and forced vital capacity (FVC) from acceptable trials were used in the analysis. Prediction equations of the GLI [30] were used.

Bronchodilator responses

The bronchodilator responses (Δ) were calculated as post-bronchodilator values minus baseline (pre-bronchodilator) values, and expressed as absolute change, proportional (%) change from baseline and as changes in Z-scores (derived from the HealthyAsym group). As very extreme values of relative ΔXrs occur, even for transformed data, absolute and Z-score changes only were used. A negative ΔRrs indicated decrease (the expected response), therefore the lower limit of normal (LLN) was defined as the fifth percentile of bronchodilator response in the HealthyAsym group. A positive ΔXrs indicated an increase (the expected response), therefore the upper limit of normal (ULN) was defined as the 95th percentile. Positive bronchodilator responsiveness in either FEV1 or FVC was defined as an increase of ≥12% and ≥200 mL.

Statistical analyses

The data were analysed using SPSS software (IBM, Armonk, NY, USA; version 21). Paired t-tests and the Wilcoxon signed-rank test were used to compare baseline and post-bronchodilator lung function as appropriate. Natural logarithm and exponential transformations were used to normalise Rrs and Xrs distributions, respectively, in the healthy group. Multiple linear regressions were used to define normative equations as performed previously [25]. Spearman correlations were used to evaluate the relationships between bronchodilator responses and potential anthropometric predictors (age, sex, height and body mass index (BMI)), baseline lung function, and clinical predictors (respiratory symptoms, asthma diagnosis and smoking history).

Results

Of 1318 subjects, 163 were excluded due to incomplete data, 10 had highly disparate Rrs and Xrs values indicating artefact, leaving 1145 (86.9%) with complete questionnaire and technically satisfactory pre-bronchodilator and post-bronchodilator FOT and spirometry data. The anthropometric characteristics of the entire cohort are described in table 1. The anthropometric characteristics of the healthy subgroup of the entire cohort are also shown. A smoking history of ≥10 pack-years was reported by 27.9% (320 out of 1145), while 10.4% (119 out of 1145) had obstructed baseline spirometry and 7.1% (81 out of 1145) had obstructed post-bronchodilator spirometry. Positive spirometric bronchodilator responses occurred in 6.6% (75 out of 1145). The post-bronchodilator spirometry and all oscillometric parameters, were all significantly different compared with baseline (figure 1).
TABLE 1

Anthropometric characteristics of the entire cohort and the healthy subgroup reporting no respiratory symptoms (HealthyAsymp)

Entire cohort# HealthyAsym
Subjects 1145577
Age, years 61.3±12.460.6±12.5
Male/female 570/575244/333
Height, m 1.68±0.091.67±0.09
Smoking, pack-years 0.0 (0.0–12.0)0 (0.0–0.75)
BMI, kg·m−2 27.8±4.626.8±4.0
FEV1, Z-score −0.24±1.030.1±0.9
FEV1/FVC, Z-score −0.46±0.96−0.3±0.8
Rrs6, cmH2O·s·L−1 3.29 (2.65–4.15)3.10 (2.53, 3.78)
Xrs 6 , cmH2O·s·L−1 −0.45 (−0.83–  −0.22)−0.39 (−0.66– −0.18)
EFLi, cmH2O·s·L−1 −0.12 (−0.27–0.89)−0.15 (−0.27– −0.02)

Data are presented as mean±sd or median (interquartile range), unless otherwise stated. BMI: body mass index; FEV1: forced expiratory volume in 1 s; FVC: forced vital capacity; Rrs: resistance of the respiratory system measured at 6 Hz; Xrs: reactance of the respiratory system measured at 6 Hz; EFLi: expiratory flow limitation index. #: includes HealthyAsym subjects.

FIGURE 1

The bronchodilator-induced changes in a) resistance and b) reactance parameters of the respiratory system, for each of the clinically defined groups. BDR: bronchodilator response; Δ: post-bronchodilator minus pre-bronchodilator change; Rrs: resistance of the respiratory system measured at 6 Hz; insp: inspiratory; Xrs: reactance of the respiratory system measured at 6 Hz; EFLi: expiratory flow limitation index; Asym; asymptomatic: Symp: symptomatic.

Anthropometric characteristics of the entire cohort and the healthy subgroup reporting no respiratory symptoms (HealthyAsymp) Data are presented as mean±sd or median (interquartile range), unless otherwise stated. BMI: body mass index; FEV1: forced expiratory volume in 1 s; FVC: forced vital capacity; Rrs: resistance of the respiratory system measured at 6 Hz; Xrs: reactance of the respiratory system measured at 6 Hz; EFLi: expiratory flow limitation index. #: includes HealthyAsym subjects. The bronchodilator-induced changes in a) resistance and b) reactance parameters of the respiratory system, for each of the clinically defined groups. BDR: bronchodilator response; Δ: post-bronchodilator minus pre-bronchodilator change; Rrs: resistance of the respiratory system measured at 6 Hz; insp: inspiratory; Xrs: reactance of the respiratory system measured at 6 Hz; EFLi: expiratory flow limitation index; Asym; asymptomatic: Symp: symptomatic.

Bronchodilator responses in healthy subjects

There were 577 subjects in the HealthyAsym group (table 1). Their baseline and post-bronchodilator spirometry and oscillometry parameters are shown in table E1. This shows that there were minimal but statistically significant bronchodilator-associated changes in FEV1, FVC and FEV1/FVC ratio, and all oscillometry parameters. The normative equations for oscillometry that were derived from this group to determine Z-scores, are listed in table E2 and the normative thresholds for bronchodilator responses are given in table 2. The LLN of ΔRrs parameters and ULN for ΔXrs parameters are also provided in table 2. The bronchodilator responses in all other subjects (table 2) were significantly different for all oscillometry and spirometry parameters, compared with the HealthyAsym group.
TABLE 2

Bronchodilator responses for resistance and reactance of the respiratory system and spirometry parameters in the healthy asymptomatic (HealthyAsym) group and remainder

Parameter HealthyAsym (n=577) LLN or ULN Remainder (n=568)
ΔRrs6, cmH2O·s·L−1 −0.23 (−0.57–0.04)LLN −1.38−0.36 (−0.81– −0.01)
ΔRrs6, % −6.4 (−17.2–1.3)LLN −30.0−11.3 (−20.4– −0.2)
ΔRrs6, Z-score −0.31 (−0.73–0.03)LLN −1.42−1.1 (−2.4– −0.3)
ΔRrs6,insp, cmH2O·s·L−1−0.31 (−0.67– −0.04)LLN −1.38−0.41 (−0.93– −0.11)
ΔRrs6,insp, % −8.5 (−20.7–1.3)LLN −42.9−12.7 (−25.6– −0.2)
ΔXrs6, cmH2O·s·L−1 0.09±0.26ULN 0.570.28±0.57
ΔXrs6, Z-score 0.25±0.67ULN 1.360.5±0.8
Δ Xrs6,insp, cmH2O·s·L−1 0.12±0.23ULN 0.530.22±0.40
ΔEFLi, cmH2O·s·L−1 0.039±0.281LLN −0.37−0.099±0.592
ΔFEV1, mL 76±119106±135
ΔFVC, mL 34±17639±208

Data are presented as median (interquartile range) or mean±sd, unless otherwise stated. Responses for resistance are defined as the fifth percentiles and upper limits of normal (ULNs) for reactance are defined as the 95th percentiles. LLN: lower limit of normal; Δ: post-bronchodilator minus pre-bronchodilator change; Rrs6: resistance of the respiratory system measured at 6 Hz; insp: inspiratory; Xrs: reactance of the respiratory system measured at 6 Hz; EFLi: expiratory flow limitation index; FEV1: forced expiratory volume in 1 s; FVC: forced vital capacity. p<0.002 for HealthyAsym compared with remainder all Rrs parameters (Mann–Whitney); p<0.001 HealthyAsym compared with remainder for all Xrs parameters (t-test).

Bronchodilator responses for resistance and reactance of the respiratory system and spirometry parameters in the healthy asymptomatic (HealthyAsym) group and remainder Data are presented as median (interquartile range) or mean±sd, unless otherwise stated. Responses for resistance are defined as the fifth percentiles and upper limits of normal (ULNs) for reactance are defined as the 95th percentiles. LLN: lower limit of normal; Δ: post-bronchodilator minus pre-bronchodilator change; Rrs6: resistance of the respiratory system measured at 6 Hz; insp: inspiratory; Xrs: reactance of the respiratory system measured at 6 Hz; EFLi: expiratory flow limitation index; FEV1: forced expiratory volume in 1 s; FVC: forced vital capacity. p<0.002 for HealthyAsym compared with remainder all Rrs parameters (Mann–Whitney); p<0.001 HealthyAsym compared with remainder for all Xrs parameters (t-test).

Predictors of bronchodilator responses

The anthropometric (sex, height and BMI), clinical (symptoms, asthma diagnosis and smoking history) and baseline lung function (Rrs, Xrs and spirometry Z-scores) predictors of bronchodilator responses for the entire cohort are shown in table E3. Male patients had larger ΔXrs Z-score and ΔFVC, but changes were very small and clinically insignificant. Higher BMI was associated with larger ΔRrs, ΔRrs, all ΔXrs parameters and ΔFVC. Asthma diagnosis and respiratory symptoms were associated with larger bronchodilator responses in all but one of the spirometry and oscillometry parameters (ΔEFLi). Smoking history was associated with larger bronchodilator responses in spirometry and all oscillometry parameters. Greater impairment of baseline oscillometry and spirometry parameters predicted greater bronchodilator responses in all parameters except for FEV1/FVC not predicting ΔFVC (table E3). In the HealthyAsym group, anthropometric parameters were not predictive of bronchodilator responses (tables E4 and E5). However, baseline Rrs and Xrs Z-scores predicted all ΔRrs and ΔXrs parameters (except ΔEFLi and ΔRrs (%) in men). Baseline FEV1 and FEV1/FVC Z-scores predicted ΔFEV1 and all ΔRrs parameters in women, while only FEV1/FVC Z-scores were predictive in men.

Clinical context of bronchodilator responses

Given the above associations with asthma diagnosis, symptoms and/or medication use, smoking history and airflow obstruction, the prevalence of bronchodilator responses was determined for clinical groups based on those parameters, to provide clinical context. Consequently, five mutually exclusive groups were defined: 1) symptomatic nonsmokers (HealthySymp), defined as no asthma diagnosis but reported respiratory symptoms in the last year and <10 pack-years smoking; 2) asymptomatic smokers (SmokersAsym), with no respiratory symptoms in the last year and no asthma diagnosis but ≥10 pack-years smoking history and FEV1/FVC ratio Z-score ≥LLN; 3) symptomatic smokers (SmokersSymp), the same as SmokersAsym but reporting respiratory symptoms; 4) nonsmokers with current asthma, defined as any past asthma diagnosis and reporting respiratory symptoms and/or respiratory medication use in the last year, and <10 pack-years smoking; 5) smokers with fixed airflow obstruction (“COPD”), defined as ≥10 pack-years smoking history and post-bronchodilator FEV1/FVC ratio Z-score <−1.645). Therefore, COPD subjects could have a doctor diagnosis of asthma (present in 20 out of 46 subjects) and subjects with asthma could have an abnormal FEV1/FVC ratio (32 and 20 out of 122 had reduced Z-scores pre-bronchodilator and post-bronchodilator, respectively). The definition of these groups is shown in figure 2.
FIGURE 2

Disposition into clinical groups. See Results section for definition of groups. Asym: asymptomatic; Symp: symptomatic.

Disposition into clinical groups. See Results section for definition of groups. Asym: asymptomatic; Symp: symptomatic. The anthropometric characteristics of each group are shown in table E6, their median bronchodilator changes are shown in table E7 and the prevalence of abnormal baseline function in table E8. The percentage of positive bronchodilator responsiveness in each group is shown in table 3 and in figure 1. The proportions of positive bronchodilator responsiveness were compared with that in the HealthyAsym group. The HealthySymp group were similar. SmokersAsym had more positive bronchodilator responsiveness in ΔXrs (absolute change) and ΔEFLi, but not in spirometry; SmokersSymp had more positive bronchodilator responses in all of the ΔXrs parameters and in ΔFEV1. In the asthma and COPD groups, bronchodilator responsiveness was increased for almost all parameters (except ΔRrs % and ΔRrs Z-score in COPD). The asthma and SmokersSymp groups appeared similar in terms of responsiveness in both oscillometry and spirometry. Notably, 109 out of 122 (89.3%) of people with asthma reported wheeze in the previous 12 months. Bronchodilator responsiveness was greatest in COPD, in ΔXrs expressed as absolute change or ΔEFLi, of whom approximately half exhibited positive responses. Positive ΔFEV1 responsiveness was also demonstrated in about a third of this group.
TABLE 3

Percentage of positive bronchodilator responses for resistance and reactance of the respiratory system, and spirometry parameters, for each of the clinical groups

HealthyAsym (n=577) HealthySymp (n=126) SmokersAsym (n=159) SmokersSymp (n=115) Asthma (n=122) COPD (n=46)
ΔRrs6 (abs) 4.94.08.27.812.3*17.4#
ΔRrs6 (%) 5.06.38.811.313.9*10.9
ΔRrs6 (Z-score) 4.56.39.412.2*13.1*10.9
ΔRrs6,insp (abs) 4.97.18.29.615.6*21.7*
ΔRrs6,insp (%) 4.96.38.811.313.9*10.9
ΔXrs6 (abs) 4.97.911.9*17.4*19.7*47.8
ΔXrs6 (Z-score) 5.06.37.517.4*14.8*26.1+
Δ Xrs6,insp (abs) 4.77.97.520.0*§18.9*37.0+
ΔEFLi (abs) 4.99.513.2*19.1*13.9*45.7
ΔFEV1 (%) 2.30.81.99.6#17.2+30.4ƒ
ΔFVC (%) 1.40.80.65.29.8+30.4

Asym: asymptomatic; Symp: symptomatic; Δ: post-bronchodilator minus pre-bronchodilator change; Rrs6: resistance of the respiratory system measured at 6 Hz; abs: absolute; insp: inspiratory; Xrs: reactance of the respiratory system measured at 6 Hz; EFLi: expiratory flow limitation index; FEV1: forced expiratory volume in 1 s; FVC: forced vital capacity. *: p<0.05 compared with HealthyAsym only (Z-test with Bonferroni correction); #: p<0.05 compared with HealthyAsym and HealthySymp; ¶: p<0.05 compared with all other groups.

+: p<0.05 compared with HealthyAsym, HealthySymp and SmokersAsym; §: p<0.05 compared with SmokersAsym; ƒ: p<0.05 compared with HealthyAsym, HealthySymp, SmokersAsym and asthma.

Percentage of positive bronchodilator responses for resistance and reactance of the respiratory system, and spirometry parameters, for each of the clinical groups Asym: asymptomatic; Symp: symptomatic; Δ: post-bronchodilator minus pre-bronchodilator change; Rrs6: resistance of the respiratory system measured at 6 Hz; abs: absolute; insp: inspiratory; Xrs: reactance of the respiratory system measured at 6 Hz; EFLi: expiratory flow limitation index; FEV1: forced expiratory volume in 1 s; FVC: forced vital capacity. *: p<0.05 compared with HealthyAsym only (Z-test with Bonferroni correction); #: p<0.05 compared with HealthyAsym and HealthySymp; ¶: p<0.05 compared with all other groups. +: p<0.05 compared with HealthyAsym, HealthySymp and SmokersAsym; §: p<0.05 compared with SmokersAsym; ƒ: p<0.05 compared with HealthyAsym, HealthySymp, SmokersAsym and asthma. Except in the COPD group, the prevalence of responsiveness was similar for each of the ΔRrs and ΔXrs parameters. In COPD, absolute changes resulted in more positive responders than either relative or Z-score changes (p<0.01, chi-squared test), likely due to their more severe baseline obstruction. Also, responsiveness was generally more frequent for Xrs than for Rrs parameters in smoking and disease groups, particularly in COPD (figure 1 and table 3). Bronchodilator responsiveness was more prevalent for ΔXrs parameters, compared with ΔFEV1, in the symptomatic and smoking groups, but not in the asthma or COPD groups (table 4).
TABLE 4

Unadjusted p-values from comparisons of the proportion of bronchodilator responsiveness within groups, measured by change in (Δ) reactance of the respiratory system measured at 6 Hz (Xrs) parameters versus change in forced expiratory volume in 1 s, in healthy but symptomatic (Symp), smoking, asthmatic and COPD groups

HealthySymp (n=126) SmokersAsym (n=159) SmokersSymp (n=115) Asthma (n=122) COPD (n=46)
ΔXrs6 (abs) 0.012*0.001*0.1080.6780.077
ΔXrs6 (Z-score) 0.0390.0350.0780.6480.774
ΔXrs6,insp (abs) 0.012*0.0350.023*0.8240.549
ΔEFLi (abs) 0.003*0.0003*0.0430.5560.118

Asym: asymptomatic; abs: absolute; insp: inspiratory; EFLi: expiratory flow limitation index. Comparisons were by McNemar tests. *: significant p-values after Benjamini–Hochberg adjustment for multiple comparisons, using α=0.10 (10% false discovery rate).

Unadjusted p-values from comparisons of the proportion of bronchodilator responsiveness within groups, measured by change in (Δ) reactance of the respiratory system measured at 6 Hz (Xrs) parameters versus change in forced expiratory volume in 1 s, in healthy but symptomatic (Symp), smoking, asthmatic and COPD groups Asym: asymptomatic; abs: absolute; insp: inspiratory; EFLi: expiratory flow limitation index. Comparisons were by McNemar tests. *: significant p-values after Benjamini–Hochberg adjustment for multiple comparisons, using α=0.10 (10% false discovery rate).

Discussion

In summary, we have defined lower and upper limits of bronchodilator responsiveness for Rrs and Xrs parameters, respectively, from a healthy, community cohort, which included intra-breath parameters and changes in Z-scores [9], to allow us to compare bronchodilator responsiveness measured by oscillometry to responsiveness measured by spirometry. Our thresholds for bronchodilator responsiveness measured by oscillometry were almost identical to published values [21] and oscillometry was more sensitive than spirometry in symptomatic nonsmokers and in both symptomatic and asymptomatic smokers who had normal spirometry. We also found bronchodilator responsiveness was related to asthma diagnosis, respiratory symptoms, smoking history and baseline airway function (both spirometry and oscillometry). In five clinical groups defined according to those factors, oscillometric bronchodilator responsiveness quantified as absolute, percentage and Z-score changes, produced similar outcomes, except in COPD for which responses measured as absolute changes were about twice as frequent as measured by relative or Z-score change. This was likely due to the dependence of bronchodilator responsiveness on baseline values. Only one of several published studies on bronchodilator responses in healthy subjects [21, 31–33] had sufficient numbers to reliably define upper and lower limits of bronchodilator responses [9, 21], although not for intra-breath parameters and not expressed as changes in Z-scores, as we produced. These normal limits were derived from pooled measurements from five different devices, in a slightly younger population than the present study. Despite these differences, our lower limit of bronchodilator response for Rrs (−1.38 cmH2O·s·L−1 or −30.0%) and upper limit for Xrs (0.57 cmH2O·s·L−1) are practically identical to that reported in the study by Oostveen et al. [21] (Rrs –1.28 cmH2O·s·L−1 or −31.5% and Xrs 0.47 cmH2O·s·L−1). This suggests that these cut-points based on absolute and relative change in mean Rrs and absolute change in Xrs are fairly robust given their consistency across different populations and devices. Although increasing BMI predicted lower bronchodilator response in Rrs in healthy men, it was marginal (table E5) and did not predict responses in women. Correction for BMI was therefore not performed in determining normal responsiveness. Bronchodilator responsiveness of all oscillometry parameters (and for ΔFEV1 and ΔFVC) correlated with baseline values for Rrs, Xrs and spirometry. This suggests that bronchodilator responses should be expressed as relative change [34]. However, we did not use relative change for Xrs parameters because of the very large values and highly skewed distribution that resulted from values that were close to zero. However, use of the absolute, relative or Z-score changes in Rrs or Xrs did not affect the prevalence of bronchodilator responders in groups other than COPD, therefore these cut-offs could be used interchangeably. In the COPD group, when ΔXrs was expressed as absolute change, positive bronchodilator responsiveness was about twice that responsiveness expressed as Z-score change. This is likely due to the strong dependence of ΔXrs, expressed as absolute change, on baseline values. This may be explained by Xrs being sensitive to airway closure and heterogeneous severe narrowing that is common in COPD [11–13, 35, 36]. Even small decreases in bronchoconstriction could alleviate airway severe narrowing and closure, but would have a large effect in improving Xrs. Thus, bronchodilator responses should arguably be expressed as either relative or Z-score change, given their dependence on baseline function. However, it is also possible that the high prevalence of bronchodilator responsiveness measured by ΔXrs as absolute change may be clinically meaningful; this requires further study. Showing the prevalence of bronchodilator responsiveness by groups was to provide clinical context. Disease or diagnostic labels are arbitrary and our primary aim was not to investigate disease or diagnostic label differences. Nevertheless, there were some clinically interesting observations. The asthma group was large (10.7%) and, despite being a community sample with normal baseline function, oscillometric bronchodilator responsiveness was common in nearly 20%. In the COPD group, bronchodilator responsiveness was present in nearly half according to either ΔXrs or spirometry. This could be related to asthma diagnoses in 20 out of 46 in addition to their smoking and, arguably, they may justifiably have an asthma/COPD overlap label. Although BMI predicted ΔXrs responses in the whole cohort (table E3), it may be explained by its association with symptoms and smoking since HealthySymp and both smoking groups were heavier than the reference group (tables E4 and E5). Despite their normal spirometry, 18.2% of asymptomatic smokers and 31.3% of symptomatic smokers had abnormal baseline oscillometry (defined as abnormality in either Rrs or Xrs). Furthermore, their bronchodilator responsiveness in Xrs parameters (up to 20%) was about twice that of spirometry. There may be value in exploring potential clinical correlations in symptomatic smokers, given the evidence of their increased morbidity [37]. Because of the greater bronchodilator responsiveness in oscillometry in the COPD and asthma groups, in particular in Xrs, the concordance between the bronchodilator responses measured by oscillometry and spirometry was only fair (κ=0.21–0.40) while being poor or slight in the other groups. This is consistent with published studies [10, 21, 38, 39]. In COPD, this difference could possibly be due to oscillometry being more sensitive to smoking-related lung damage than spirometry. The basis of the differences is unknown but may reflect the different manoeuvres used during spirometry and oscillometry measurements. In a tertiary asthma clinic study [15], κ between spirometry and Xrs was 0.45, perhaps due to a wider range of lung function impairment. However, oscillometric bronchodilator responses related more strongly to asthma control than spirometry [15]. Discordant bronchodilator responsiveness between oscillometry and spirometry could be clinically important, i.e. complementary information, but this also needs further study. A potential limitation of this study was the participant ages of 40 years or older, which meant that these findings could not be applied to younger subjects. The oscillometry device was a proprietary device and, therefore, not used clinically elsewhere, and our use of 6 Hz and 19 Hz is nonstandard (usually 5 Hz and 19 Hz). Theoretically, any differences in measurements associated with different frequencies are likely to be insignificant. We have also compared our device with three other commercial oscillometry devices and showed that, while Rrs was comparable between devices, there was greater variability in Xrs [40]. These findings are consistent with other studies comparing those same devices [21, 41] but it should also be noted that there are greater disparities in Xrs measurements between other oscillometry devices [41]. Nevertheless, any small differences in measurements between devices are unlikely to affect bronchodilator responses given the within-session changes being measured. The COPD group was small (46 out of 1145) and approximately half also had an asthma diagnosis. Given the many COPD phenotypes, our findings would not be generalisable to COPD or to other smoking-related airways diseases. In conclusion, we have defined normative values for bronchodilator responses for oscillometry parameters, in a large, well characterised healthy population sample of 40 years and older. These thresholds are potentially useful to inform interpretation of oscillometry in airway disease. Bronchodilator responsiveness was associated with respiratory symptoms, asthma diagnosis and smoking history. ΔXrs may be a particularly sensitive measure of airway dysfunction in smokers with normal spirometry. For the purposes of clinical interpretation, given the strong dependence of bronchodilator response on baseline oscillometric or spirometric function, it may be more appropriate to express responses as either relative or Z-score change. The clinical significance of oscillometric bronchodilator responsiveness, particularly in relation to disease phenotypes and treatable traits, needs further study, given that the fair concordance with spirometry suggests potential complementarity. Please note: supplementary material is not edited by the Editorial Office, and is uploaded as it has been supplied by the author. Supplementary material 00439-2021.SUPPLEMENT Figure E1 00439-2021.FigureE1
  41 in total

1.  Lung function and symptom improvement with fluticasone propionate/salmeterol and ipratropium bromide/albuterol in COPD: response by beta-agonist reversibility.

Authors:  Eugene R Bleecker; Amanda Emmett; Glenn Crater; Katharine Knobil; Christopher Kalberg
Journal:  Pulm Pharmacol Ther       Date:  2008-05-01       Impact factor: 3.410

2.  Clinical Significance of Symptoms in Smokers with Preserved Pulmonary Function.

Authors:  Prescott G Woodruff; R Graham Barr; Eugene Bleecker; Stephanie A Christenson; David Couper; Jeffrey L Curtis; Natalia A Gouskova; Nadia N Hansel; Eric A Hoffman; Richard E Kanner; Eric Kleerup; Stephen C Lazarus; Fernando J Martinez; Robert Paine; Stephen Rennard; Donald P Tashkin; MeiLan K Han
Journal:  N Engl J Med       Date:  2016-05-12       Impact factor: 91.245

3.  Procedures to improve the repeatability of forced oscillation measurements in school-aged children.

Authors:  Paul D Robinson; Martin Turner; Nathan J Brown; Cheryl Salome; Norbert Berend; Guy B Marks; Greg G King
Journal:  Respir Physiol Neurobiol       Date:  2011-02-12       Impact factor: 1.931

4.  Relationship between heterogeneous changes in airway morphometry and lung resistance and elastance.

Authors:  K R Lutchen; H Gillis
Journal:  J Appl Physiol (1985)       Date:  1997-10

5.  Effect of stochastic heterogeneity on lung impedance during acute bronchoconstriction: a model analysis.

Authors:  C W Thorpe; J H Bates
Journal:  J Appl Physiol (1985)       Date:  1997-05

6.  Triple Inhaled Therapy at Two Glucocorticoid Doses in Moderate-to-Very-Severe COPD.

Authors:  Klaus F Rabe; Fernando J Martinez; Gary T Ferguson; Chen Wang; Dave Singh; Jadwiga A Wedzicha; Roopa Trivedi; Earl St Rose; Shaila Ballal; Julie McLaren; Patrick Darken; Magnus Aurivillius; Colin Reisner; Paul Dorinsky
Journal:  N Engl J Med       Date:  2020-06-24       Impact factor: 91.245

7.  Lung hyperinflation and its reversibility in patients with airway obstruction of varying severity.

Authors:  Athavudh Deesomchok; Katherine A Webb; Lutz Forkert; Yuk-Miu Lam; Dror Ofir; Dennis Jensen; Denis E O'Donnell
Journal:  COPD       Date:  2010-12       Impact factor: 2.409

8.  A comparison of plethysmography, spirometry and oscillometry for assessing the pulmonary effects of inhaled ipratropium bromide in healthy subjects and patients with asthma.

Authors:  C M Houghton; A A Woodcock; D Singh
Journal:  Br J Clin Pharmacol       Date:  2005-02       Impact factor: 4.335

9.  International variation in the prevalence of COPD (the BOLD Study): a population-based prevalence study.

Authors:  A Sonia Buist; Mary Ann McBurnie; William M Vollmer; Suzanne Gillespie; Peter Burney; David M Mannino; Ana M B Menezes; Sean D Sullivan; Todd A Lee; Kevin B Weiss; Robert L Jensen; Guy B Marks; Amund Gulsvik; Ewa Nizankowska-Mogilnicka
Journal:  Lancet       Date:  2007-09-01       Impact factor: 79.321

10.  Airflow reversibility and long-term outcomes in patients with COPD without comorbidities.

Authors:  José M Marín; Maria Ciudad; Virginia Moya; Santiago Carrizo; Salvador Bello; Barbara Piras; Bartolomé R Celli; Marc Miravitlles
Journal:  Respir Med       Date:  2014-05-20       Impact factor: 3.415

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.