Instrument accuracy and reproducibility in measurements of pulmonary function.

Robert L Jensen1, John G Teeter, Richard D England, Heather J White, Eve H Pickering, Robert O Crapo.   

Abstract

BACKGROUND: The objective of the study was to quantify the accuracy and reproducibility of five commercially available pulmonary function test (PFT) instruments (Collins CPL [Ferraris Respiratory; Louisville, CO]; Morgan Transflow Test PFT System [Morgan Scientific; Haverhill, MA]; SensorMedics Vmax 22D [VIASYS Healthcare; Yorba Linda, CA]; Jaeger USA Masterscreen Diffusion TP [VIASYS Healthcare]; and Medical Graphics Profiler DX System [Medical Graphics Corp; St. Paul, MN]) that are associated with spirometry and the measurement of pulmonary diffusing capacity.
METHODS: In a multifactor, single-center, repeated-measures, full factorial 90-day study, a pulmonary waveform generator and a single-breath simulator of diffusing capacity of the lung for carbon monoxide (Dlco) were used to perform simulations of FVC and Dlco maneuvers. Accuracy was assessed as the difference between the observed and simulated values. Reproducibility was determined as the coefficient of variation of all measurements made during the study.
RESULTS: All instruments demonstrated a high degree of accuracy in the measurement of FVC and FEV(1). Overall, the accuracies associated with the measurement of peak flow, forced expiratory flow, mid-expiratory phase, and diffusing capacity were generally lower and more variable among the instruments tested. The coefficients of variation of Dlco measurements over 90 days were higher than those observed for spirometry.
CONCLUSIONS: This study demonstrates the feasibility of assessing the accuracy and reproducibility of modern PFT instruments using simulation testing. Our results provide an assessment of the component of PFT accuracy and reproducibility that is due to instrumentation alone.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17573502     DOI: 10.1378/chest.06-1998

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  11 in total

1.  Technical performance analysis of different types of spirometers.

Authors:  Zhongping Wu; Ruibo Huang; Liping Zhong; Yi Gao; Jinping Zheng
Journal:  BMC Pulm Med       Date:  2022-01-05       Impact factor: 3.317

2.  Using hyperpolarized 129Xe MRI to quantify regional gas transfer in idiopathic pulmonary fibrosis.

Authors:  Jennifer M Wang; Scott H Robertson; Ziyi Wang; Mu He; Rohan S Virgincar; Geoffry M Schrank; Rose Marie Smigla; Thomas G O'Riordan; John Sundy; Lukas Ebner; Craig R Rackley; Page McAdams; Bastiaan Driehuys
Journal:  Thorax       Date:  2017-08-31       Impact factor: 9.139

3.  Simulation system for a rebreathing technique to measure multiple cardiopulmonary function parameters.

Authors:  Cuneyt Yilmaz; William W Chance; Robert L Johnson; Connie C W Hsia
Journal:  Chest       Date:  2009-05       Impact factor: 9.410

Review 4.  Alzheimer's disease drug development: old problems require new priorities.

Authors:  Robert E Becker; Nigel H Greig
Journal:  CNS Neurol Disord Drug Targets       Date:  2008-12       Impact factor: 4.388

5.  Intersession variability in single-breath diffusing capacity in diabetics without overt lung disease.

Authors:  Michael B Drummond; Pamela F Schwartz; William T Duggan; John G Teeter; Richard J Riese; Richard C Ahrens; Robert O Crapo; Richard D England; Neil R Macintyre; Robert L Jensen; Robert A Wise
Journal:  Am J Respir Crit Care Med       Date:  2008-05-08       Impact factor: 21.405

6.  Sequential changes of serum KL-6 predict the progression of interstitial lung disease.

Authors:  Ying Jiang; Qun Luo; Qian Han; Junting Huang; Yonger Ou; Miao Chen; Yu Wen; Silas Sethiel Mosha; Kuimiao Deng; Rongchang Chen
Journal:  J Thorac Dis       Date:  2018-08       Impact factor: 2.895

Review 7.  New Developments in Imaging Idiopathic Pulmonary Fibrosis With Hyperpolarized Xenon Magnetic Resonance Imaging.

Authors:  Joseph G Mammarappallil; Leith Rankine; Jim M Wild; Bastiaan Driehuys
Journal:  J Thorac Imaging       Date:  2019-03       Impact factor: 3.000

8.  Two-year pulmonary safety and efficacy of inhaled human insulin (Exubera) in adult patients with type 2 diabetes.

Authors:  Julio Rosenstock; William T Cefalu; Priscilla A Hollander; Andre Belanger; Freddy G Eliaschewitz; Jorge L Gross; Solomon S Klioze; Lisa B St Aubin; Howard Foyt; Masayo Ogawa; William T Duggan
Journal:  Diabetes Care       Date:  2008-06-05       Impact factor: 19.112

9.  Comparison of oscillometry devices using active mechanical test loads.

Authors:  Ronald J Dandurand; Jean-Pierre Lavoie; Larry C Lands; Zoltán Hantos
Journal:  ERJ Open Res       Date:  2019-12-23

10.  Accurate Spirometry with Integrated Barometric Sensors in Face-Worn Garments.

Authors:  Bo Zhou; Alejandro Baucells Costa; Paul Lukowicz
Journal:  Sensors (Basel)       Date:  2020-07-29       Impact factor: 3.576

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.