Cecilio Velasco-Barona1, Claudia Corredor-Ortega1, Alejandra Mendez-Leon1, Nadia L Casillas-Chavarín1, Daniel Valdepeña-López Velarde1, Guadalupe Cervantes-Coste1, Daniel Malacara-Hernández2,3, Roberto Gonzalez-Salinas3,4.
Abstract
Prospective, randomized, comparative, and controlled study to estimate the association between angle κ distance and higher-order aberrations (HOAs) with postoperative visual acuity after presbyopia-correcting IOL implantation. Forty-three eyes from 43 patients were included and randomly assigned in two groups for either AT LISA tri 839MP or Acrysof IQ PanOptix IOL implantation. The OPD-Scan III analyzer was utilized to assess the angle κ distance and higher-order aberration (HOAs). Twenty-three eyes were in the Acrysof IQ PanOptix group and 20 patients in the AT LISA tri 839MP group. The uncorrected distance visual acuity (UDVA) for the PanOptix group was 0.092 ± 0.10, whereas for AT LISA tri was 0.050 ± 0.06 (P=0.229). The uncorrected intermediate visual acuity (UIVA) for the PanOptix group was 0.173 ± 0.18, whereas for AT LISA tri, it was 0.182 ± 0.11 (P=0.669). Uncorrected near visual acuity (UNVA) was 0.068 ± 0.04 and 0.085 ± 0.07, respectively (P=0.221). Also, correlation coefficient between HOAs and the Strehl ratio for each group were -0.768 (P < 0.0001) and -0.863 (P=0.0001). Patients implanted with both trifocal IOLs showed excellent postoperative visual performance at all distances at the six-month follow-up visit. No association was found between angle κ distance and postoperative visual acuity regardless of the angle κ magnitude or the two trifocal IOLs inner optical diameter. Also, internal aberrations demonstrated a significant inverse correlation with the Strehl ratio for both trifocal IOLs.Entities:
Year: 2019 PMID: 31885892 PMCID: PMC6900936 DOI: 10.1155/2019/7018937
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Figure 1Pupillary diameter display and diagram of κ angle distance. (a) Comparison of pupillary diameter under mesopic and photopic conditions. (b) Diagram of κ angle formed by the visual axis and the pupillary axis. (c) Graphic representation of κ angle, visual axis, and pupillary axis, showing the center of the visual axis (green cross, representing the center of the reflection points), corneal center (violet dot in diagram, similar to the anatomic center), and pupillary center (blue dot, representing the center of the circle). The radial distance between the green cross and the violet dot represents angle alpha (α). The radial distance between the blue dot and the green cross represents angle κ (κ). The + sign represents the positive angle; and the—sign represents the negative angle.
Figure 2HOAs measurements using an OPD-Scan III analyzer obtained at the six-month follow-up visit.
Trifocal IOL features [10].
| Feature | Acrysof IQ PanOptix® | AT LISA tri 839MP® |
|---|---|---|
| Technology | Trifocal | Trifocal |
| Diffractive zone (mm) | 4.5 | 6.0 |
| Central zone | Diffractive | Diffractive |
| Optic type | Nonapodized | Nonapodized |
| Optic diameter (mm) | 6.0/4.5 mm diffractive region | 6.0/4.3 mm trifocal/4.3 to 6.0 mm bifocal |
| Near add power (D) | +3.25 | +3.33 |
| Intermediate IOL power (D) | +2.17 | +1.66 |
| Asphericity ( | −0.10 | −0.18 |
| IOL color | Yellow | Clear |
| Inner ring optical diameter | 1.164 mm | 1.04 mm |
Figure 3Inner ring optical diameter of the trifocal IOLs included in the study: (a) Acrysof IQ PanOptix® and (b) the AT LISA tri 839MP®.
Preoperative measurements.
| Parameter | Value | Acrysof IQ PanOptix | AT LISA tri 839MP | Difference between means |
|
|---|---|---|---|---|---|
| Age | Mean ± SD | 67.52 ± 8.22 | 65.29 ± 8.12 | 2.23 ± 2.7 | 0.423 |
| Range | 57–81 | 52–80 | |||
| UDVA (logMAR) | Mean ± SD | 0.324 ± 0.38 | 0.360 ± 0.42 | 0.03 ± 0.02 | 0.251 |
| Range | 0.09–2.00 | 1.1–2.00 | |||
| Spherical equivalent (D) | Mean ± SD | 0.24 ± 2.1 | 0.34 ± 3.2 | 0.01 ± 0.01 | 0.224 |
| Range | −6.25 to +4.50 | −7.50 to +3.50 | |||
| Corneal astigmatism (D) | Mean ± SD | 0.43 ± 031 | 0.51 ± 0.23 | 0.04 ± 0.10 | 0.683 |
| Range | 0–1.00 | 0.25–0.1.00 | |||
| Steep keratometry (D) | Mean ± SD | 44.15 ± 1.30 | 43.80 ± 1.29 | 0.36 ± 0.70 | 0.210 |
| Range | 41.7–46.5 | 42.0–46.3 | |||
| Flat keratometry (D) | Mean ± SD | 43.50 ± 1.23 | 43.20 ± 1.47 | 0.39 ± 0.63 | 0.152 |
| Range | 40.0–45.3 | 41.2–45.1 |
UDVA = uncorrected distance visual acuity. Mann–Whitney test.
Mean visual acuity and angle κ values per group at the six-month follow-up visit.
| Parameter | Value | Acrysof IQ PanOptix | AT LISA tri 839MP | 95% CI |
|
|---|---|---|---|---|---|
| UDVA (logMAR) | Mean ± (SD) | 0.092 ± 0.10 | 0.050 ± 0.06 | 0.04, 0.13 | 0.229 |
| Range | 0–0.39 | 0–0.27 | |||
| UIVA (logMAR) | Mean ± (SD) | 0.173 ± 0.18 | 0.182 ± 0.11 | 0.08, 0.13 | 0.669 |
| Range | 0–0.91 | 0–0.39 | |||
| UNVA (logMAR) | Mean ± (SD) | 0.068 ± 0.04 | 0.085 ± 0.07 | 0.08, 0.17 | 0.221 |
| Range | 0–0.09 | 0–0.30 | |||
| Angle | Mean ± (SD) | 0.337 ± 0.15 | 0.278 ± 0.13 | −0.24, 0.11 | 0.093 |
| Range | 0.10–0.62 | 0.02–0.64 |
UDVA = uncorrected distance visual acuity; UIVA = uncorrected intermediate visual acuity; UNVA = uncorrected near visual acuity. Mann–Whitney test.
Comparison of the postoperative total and internal aberrations per group.
| Value | Acrysof IQ PanOptix | AT LISA tri 839MP | Difference between medians |
| |
|---|---|---|---|---|---|
| Total aberrations | |||||
| TILT ( | Mean ± SD | 0.291 ± 0.22 | 0.467 ± 0.45 | 0.004 | 0.387 |
| Range | 0.01–1.05 | 0.07–1.54 | |||
| HOAs ( | Mean ± SD | 0.381 ± 0.21 | 0.485 ± 0.26 | 0.111 | 0.195 |
| Range | 0.08–0.86 | 0.10–1.00 | |||
| COMA ( | Mean ± SD | 0.133 ± 0.11 | 0.247 ± 0.23 | 0.031 | 0.164 |
| Range | 0.02–0.58 | 0.05–0.65 | |||
| TREFOIL ( | Mean ± SD | 0.289 ± 0.18 | 0.255 ± 0.56 | 0.081 | 0.073 |
| Range | 0.02–0.76 | 0.15–2.17 | |||
| SPHERE ( | Mean ± SD | 0.045 ± 0.04 | 0.326 ± 0.40 | 0.098 | 0.075 |
| Range | 0.00–0.17 | 0.00–1.00 | |||
|
| |||||
| Internal aberration | |||||
| TILT ( | Mean ± SD | 0.440 ± 0.39 | 1.11 ± 2.18 | 0.015 | 0.401 |
| Range | 0.05–1.50 | 0.07–9.30 | |||
| HOAs ( | Mean ± SD | 0.435 ± 0.67 | 0.831 ± 1.16 | 0.547 | 0.065 |
| Range | 0.10–3.46 | 0.17–4.0 | |||
| COMA ( | Mean ± SD | 0.183 ± 0.20 | 0.443 ± 0.42 | 0.086 | 0.071 |
| Range | 0.01–0.97 | 0.05–1.2 | |||
| TREFOIL ( | Mean ± SD | 0.289 ± 0.18 | 0.653 ± 1.15 | 0.054 | 0.256 |
| Range | 0.05–2.13 | 0.15–2.17 | |||
| SPHERE ( | Mean ± SD | 0.140 ± 0.12 | 0.524 ± 0.65 | 0.019 | 0.509 |
| Range | 0.02–0.57 | 0.00–2.29 | |||
Mann–Whitney test.
Correlation between angle κ distance and UDVA, UNVA, and UIVA.
|
| 95% CI |
|
| |
|---|---|---|---|---|
| Acrysof IQ PanOptix ( | ||||
| UDVA (logMAR) | −0.127 | −0.52, −0.31 | 0.016 | 0.573 |
| UIVA (logMAR) | −0.279 | −0.62, 0.16 | 0.077 | 0.208 |
| UNVA (logMAR) | −0.095 | −0.49, 0.33 | −0.009 | 0.671 |
| AT LISA tri 839MP ( | ||||
| UDVA (logMAR) | −0.432 | −0.39, 0.87 | 0.187 | 0.284 |
| UIVA (logMAR) | −0.360 | −0.84, 0.46 | 0.130 | 0.380 |
| UNVA (logMAR) | −0.452 | −0.87, 0.36 | −0.206 | 0.258 |
UDVA = uncorrected distance visual acuity; UNVA = uncorrected near visual acuity; UIVA = uncorrected intermediate visual acuity. Pearson correlation coefficient (r).
The correlation coefficient (r) between angle κ distance and internal aberration.
|
| 95% CI |
|
| |
|---|---|---|---|---|
| Acrysof IQ PanOptix | ||||
| Total HOAs (D) | 0.371 | −0.05, 0.68 | 0.138 | 0.088 |
| Internal aberration (D) | 0.304 | −0.13, 0.64 | 0.092 | 0.168 |
| AT LISA tri 839MP | ||||
| Total HOAs (D) | 0.173 | −0.27, 0.56 | 0.030 | 0.226 |
| Internal aberration (D) | 0.240 | −0.21, 0.60 | 0.57 | 0.146 |
Pearson correlation coefficient (r).