Correct visualization of the vascular lumen is impaired in standard computed tomography (CT) because of blooming artifacts, increase of apparent size, induced by metallic stents and vascular calcifications. Recently, due to the introduction of photon-counting detectors in the X-ray imaging field, a new prototype spectral photon-counting CT (SPCCT) based on a modified clinical CT system has been tested in a feasibility study for improving vascular lumen delineation and visualization of coronary stent architecture. Coronary stents of different metal composition were deployed inside plastic tubes containing hydroxyapatite spheres to simulate vascular calcifications and in the abdominal aorta of one New Zealand White (NZW) rabbit. Imaging was performed with an SPCCT prototype, a dual-energy CT system, and a conventional 64-channel CT system (B64). We found the apparent widths of the stents significantly smaller on SPCCT than on the other two systems in vitro (p < 0.01), thus closer to the true size. Consequently, the intra-stent lumen was significantly larger on SPCCT (p < 0.01). In conclusion, owing to the increased spatial resolution of SPCCT, improved lumen visualization and delineation of stent metallic mesh is possible compared to dual-energy and conventional CT.Entities:
Mesh:
Substances:
Year: 2019 PMID: 31882698 PMCID: PMC6934567 DOI: 10.1038/s41598-019-56427-6
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Figure 1Representative conventional HU images acquired on the B64, the IQon, and the SPCCT of the three stents in vitro in water. Transverse views of the stents at similar locations in the absence (first row) and presence of calcification (second row). The better spatial resolution of the SPPCT system results in an improved visualization of the stent metallic struts. On the SPCCT images the stents can be visually separated from the calcification, while this separation is not possible on the B64 and IQon images due to larger detector size. A 5 mm scale bar is shown on the transverse views.
Figure 2Longitudinal views of the entire length of the platinum coated stent (Promus Premier) are presented in the top row and 3D Volume Rendering are presented in the bottom row. The improved quality of SPCCT images allows clear visualization of the metallic mesh of the stent and its deformation due to the presence of the calcification insert (arrows indicate the calcification), while this is not possible for IQon and B64.
Figure 3Representative conventional HU images and Pt-specific K-edge images acquired on the SPCCT in the calcification containing regions of the three different stents. The platinum coated Promus Premier stent is clearly visible on the Pt specific K-edge images while all other attenuation sources such as the calcification and the intra-luminal iodinated contrast agent are absent from these images. The clear visualization of the deformation of the metallic mesh due to the calcification is to be noted. A 5 mm scale bar is shown on each image.
In vitro measured apparent strut width and intra-stent lumen diameter from SPCCT, IQon, and B64.
| Coroflex Blue Neo | Promus Premier | Tsunami Gold | ||
|---|---|---|---|---|
| SPCCT apparent strut width (mm) | Conventional HU | 0.69 ± 0.002 | 0.71 ± 0.002 | 0.69 ± 0.004 |
| Pt K-edge images | — | 0.71 ± 0.007 | — | |
| IQon apparent strut width (mm) | Conventional HU | 0.98 ± 0.06 | 0.99 ± 0.05 | 1.01 ± 0.16 |
| B64 apparent strut width (mm) | Conventional HU | 1.03 ± 0.004 | 1.00 ± 0.003 | 1.04 ± 0.009 |
| SPCCT intra-stent lumen (mm) | Conventional HU | 2.44 ± 0.01 | 2.28 ± 0.01 | 2.71 ± 0.04 |
| IQon intra-stent lumen (mm) | Conventional HU | 2.04 ± 0.08 | 2.21 ± 0.06 | 1.94 ± 0.21 |
| B64 intra-stent lumen (mm) | Conventional HU | 1.96 ± 0.006 | 1.9 ± 0.003 | 2.18 ± 0.04 |
Figure 4Representative in vivo images of the three coronary stents placed in the abdominal aorta of a NZW rabbit. Conventional HU images are presented on the two left columns of the figure. Importantly, the Pt-specific K-edge images of SPCCT (right column) enable visualization of the Pt coated stent while all other sources of attenuation are removed.
In vivo measured apparent strut width and intra-stent lumen diameter from SPCCT and B64.
| Coroflex Blue Neo | Promus Premier | Tsunami Gold | ||
|---|---|---|---|---|
| SPCCT apparent strut width (mm) | Conventional HU | 0.79 ± 0.006 | 0.74 ± 0.003 | 0.75 ± 0.004 |
| Pt K-edge images | — | 0.70 ± 0.012 | — | |
| B64 apparent strut width (mm) | Conventional HU | 1.13 ± 0.009 | 1.11 ± 0.005 | 1.06 ± 0.008 |
| SPCCT intra-stent lumen (mm) | Conventional HU | 2.06 ± 0.006 | 1.85 ± 0.004 | 1.81 ± 0.003 |
| B64 intra-stent lumen (mm) | Conventional HU | 1.73 ± 0.008 | 1.60 ± 0.004 | 1.50 ± 0.008 |
Detector characteristics.
| Pixel Size | 500 µm pitch, flip-chip to sensor |
| Observable count rate | >10 Mcps/pixel, Paralyzable |
| Input referred noise | <350 |
| Leakage compensation | Static 200 nA Dynamic 60/600 nA 3 dB at 10 kHz |
| Thresholds | 5 |
| Energy range | >160 keV |
| Energy resolution | 0.5 keV/LSB |
| Frame rate | >10 kHz, zero dead-time |
Description of the three types of coronary stents used in the study.
| Stent 1 | Stent 2 | Stent 3 | |
|---|---|---|---|
| Commercial Name | Coroflex Blue Neo (Braun, Melsungen AG, Germany) | Promus Premier Monorail (Boston Scientific, France) | Tsunami Gold (Terumo Europe NV) |
| Composition | Cobalt-Chromium | Platinum (Pt)-Chromium | Stainless steel with gold markers |
| Strut thickness (mm) | 0.081 | 0.060 | 0.080 |
| Length (mm) | 19 | 24 | 10 |