Literature DB >> 19909872

Incidence and predictors of drug-eluting stent fracture in human coronary artery a pathologic analysis.

Gaku Nakazawa1, Aloke V Finn, Marc Vorpahl, Elena Ladich, Robert Kutys, Isidora Balazs, Frank D Kolodgie, Renu Virmani.   

Abstract

OBJECTIVES: The aim of this study was to perform pathologic assessment on stent fracture.
BACKGROUND: Clinically, stent fracture has been reported in 1% to 2% of patients after drug-eluting stent (DES) implantation.
METHODS: High-contrast film-based radiographs of 177 consecutive lesions from the CVPath DES autopsy registry were reviewed. Stent fracture was graded as I (single-strut fracture), II (> or =2 struts), III (> or =2 struts with deformation), IV (with transection without gap), and V (with transection causing gap in stent segment). The incidence of adverse pathologic findings (thrombosis and restenosis) was assessed histologically.
RESULTS: Stent fracture was documented in 51 lesions (29%; grade I = 10, II = 14, III = 12, IV = 6, and V = 9). Lesions with stent fracture had longer duration after implantation (172 days [interquartile range (IQR) 31 to 630 days] vs. 44 days [IQR 7 to 270 days], p = 0.004), a higher rate of Cypher (Cordis Corp., Miami Lakes, Florida) stent usage (63% vs. 36%, p = 0.001), longer stent length (30.0 mm [IQR 22.0 to 40.0 mm] vs. 20.0 mm [IQR 14.0 to 27.3 mm], p < 0.0001), and a higher rate of overlapping stents (45% vs. 22%, p = 0.003). Although fracture with grade I to IV did not have significant impact on the occurrence of adverse pathologic findings such as thrombosis and restenosis, 67% of the grade V fracture lesions were associated with adverse pathologic findings at fracture sites. Longer stent length, use of Cypher, and longer duration of implant were identified as independent risk factors of stent fracture by logistic regression analysis.
CONCLUSIONS: The incidence of stent fracture was 29% lesions at autopsy, which is much higher than clinically reported. A high rate of adverse pathologic findings was observed in lesions with grade V stent fracture, whereas fracture with grade I to IV did not have a significant impact on the pathological outcome.

Entities:  

Mesh:

Year:  2009        PMID: 19909872     DOI: 10.1016/j.jacc.2009.05.075

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  52 in total

1.  Prevalence and prognosis of coronary stent gap detected by multi-detector CT: a follow-up study.

Authors:  Xinghua Zhang; Li Yang; Haiyue Ju; Fan Zhang; Jian Wu; Bin He; Yundai Chen
Journal:  Eur Radiol       Date:  2012-04-27       Impact factor: 5.315

2.  Circumferential stent fracture: novel detection and treatment with the use of StentBoost.

Authors:  Raghu T Ramegowda; Srinivas B Chikkaswamy; Ashalatha Bharatha; Jayashree Radhakrishna; Geetha B Krishnanaik; Manjunath C Nanjappa; Arunkumar Panneerselvam
Journal:  Tex Heart Inst J       Date:  2012

3.  Recurrent stent fracture in the right coronary artery.

Authors:  Adnan K Chhatriwalla; Samuel Unzek; Samir R Kapadia
Journal:  Clin Cardiol       Date:  2010-11-11       Impact factor: 2.882

4.  Pathology of drug-eluting versus bare-metal stents in saphenous vein bypass graft lesions.

Authors:  Saami K Yazdani; Andrew Farb; Masataka Nakano; Marc Vorpahl; Elena Ladich; Aloke V Finn; Frank D Kolodgie; Renu Virmani
Journal:  JACC Cardiovasc Interv       Date:  2012-06       Impact factor: 11.195

5.  Stent fracture is associated with a higher mortality in patients with type-2 diabetes treated by implantation of a second-generation drug-eluting stent.

Authors:  Zhen Ge; Zhi-Zhong Liu; Jing Kan; Jun-Jie Zhang; Sui-Ji Li; Nai-Liang Tian; Fei Ye; Xue-Song Qian; Song Yang; Meng-Xuan Chen; Tanveer S Rab; Shao-Liang Chen
Journal:  Int J Cardiovasc Imaging       Date:  2017-06-19       Impact factor: 2.357

Review 6.  The importance of the endothelium in atherothrombosis and coronary stenting.

Authors:  Fumiyuki Otsuka; Aloke V Finn; Saami K Yazdani; Masataka Nakano; Frank D Kolodgie; Renu Virmani
Journal:  Nat Rev Cardiol       Date:  2012-05-22       Impact factor: 32.419

7.  Impact of stent mis-sizing and mis-positioning on coronary fluid wall shear and intramural stress.

Authors:  Henry Y Chen; Bon-Kwon Koo; Deepak L Bhatt; Ghassan S Kassab
Journal:  J Appl Physiol (1985)       Date:  2013-05-30

Review 8.  Are we aware of stent fracture?

Authors:  R D Acar; M Bulut; M Akcakoyun
Journal:  Herz       Date:  2013-10-25       Impact factor: 1.443

9.  Late stent fracture - A potential role of left ventricular dilatation.

Authors:  Ahmed Al Mamary; Gilberto Dariol; Massimo Napodano
Journal:  J Saudi Heart Assoc       Date:  2014-02-10

10.  Validating Fatigue Safety Factor Calculation Methods for Cardiovascular Stents.

Authors: 
Journal:  J Biomech Eng       Date:  2018-06-01       Impact factor: 2.097