Impact of High-Dose Intravenous Vitamin C for Treatment of Sepsis on Point-of-Care Blood Glucose Readings.

Anna Peyton Howell1, Jenna L Parrett2, Daniel R Malcom3.   

Abstract

BACKGROUND: Intravenous vitamin C therapy has been associated with reduced mortality in patients with sepsis. Of potential concern with this therapy are falsely elevated point-of-care (POC) blood glucose values vs laboratory analyzed (LA) readings. The purpose of this study was to compare POC and LA blood glucose measurements in patients receiving intravenous vitamin C therapy.
METHODS: All adults (≥18 years old) admitted from January 2017 to December 2018 who received at least two doses of intravenous vitamin C and had at least one paired blood glucose collection were eligible for inclusion. The primary endpoint was the accuracy in paired blood glucose values determined using the International Organization for Standardization (ISO) 15197:2013 criteria. Paired values were assessed for clinical impact using the Parkes consensus error grid analysis. A subgroup analysis was conducted to determine the impact of impaired renal function on outcomes.
RESULTS: Fourteen patients were included for analysis with 46 paired blood glucose levels. Compliance with ISO15197:2013 criteria was met in 34 (73.9%) paired values, which did not meet the minimum criteria for accuracy. Subgroup analysis showed that the paired values from patients with impaired renal function did not meet the minimum requirements for compliance, while those from patients without impaired renal function did. The Parkes error grid showed that the variation in POC measurements likely had minimal clinical impact.
CONCLUSIONS: Our study suggests that most patients receiving vitamin C for sepsis may still be monitored at POC with the glucose meter used in our study with minimal clinical impact.

Entities:  

Keywords:  blood glucose; critical illness; hexokinase; point-of-care systems; sodium-coupled vitamin C transporters

Year:  2019        PMID: 31766883      PMCID: PMC8256056          DOI: 10.1177/1932296819889638

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  38 in total

1.  Effects of drugs on glucose measurements with handheld glucose meters and a portable glucose analyzer.

Authors:  Z Tang; X Du; R F Louie; G J Kost
Journal:  Am J Clin Pathol       Date:  2000-01       Impact factor: 2.493

2.  A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose.

Authors:  J L Parkes; S L Slatin; S Pardo; B H Ginsberg
Journal:  Diabetes Care       Date:  2000-08       Impact factor: 19.112

3.  Accuracy of glycemic measurements in the critically ill.

Authors:  Linda Slater-MacLean; George Cembrowski; Dat Chin; Carol Shalapay; Tanya Binette; Kathy Hegadoren; Christine Newburn-Cook
Journal:  Diabetes Technol Ther       Date:  2008-06       Impact factor: 6.118

4.  Response.

Authors:  Paul E Marik; Racquel Rivera; Michael H Hooper; Vikramjit Khangoora; John Catravas
Journal:  Chest       Date:  2017-08       Impact factor: 9.410

5.  Accuracy of point-of-care blood glucose measurements in critically ill patients in shock.

Authors:  Carlo Jan Pati-An Garingarao; Myrna Buenaluz-Sedurante; Cecilia Alegado Jimeno
Journal:  J Diabetes Sci Technol       Date:  2014-06-08

Review 6.  Vitamin C in Burn Resuscitation.

Authors:  Julie A Rizzo; Matthew P Rowan; Ian R Driscoll; Kevin K Chung; Bruce C Friedman
Journal:  Crit Care Clin       Date:  2016-08-02       Impact factor: 3.598

7.  Performance of strip-based glucose meters and cassette-based blood gas analyzer for monitoring glucose levels in a surgical intensive care setting.

Authors:  Helena Claerhout; Martine De Prins; Dieter Mesotten; Greet Van den Berghe; Chantal Mathieu; Johan Van Eldere; Florent Vanstapel
Journal:  Clin Chem Lab Med       Date:  2016-01       Impact factor: 3.694

8.  Vitamin C Administration to the Critically Ill: A Systematic Review and Meta-Analysis.

Authors:  Pascal L Langlois; William Manzanares; Neill K J Adhikari; François Lamontagne; Christian Stoppe; Aileen Hill; Daren K Heyland
Journal:  JPEN J Parenter Enteral Nutr       Date:  2018-11-19       Impact factor: 4.016

9.  Bedside capillary blood glucose measurements in critically ill patients: influence of catecholamine therapy.

Authors:  M Fekih Hassen; S Ayed; R Gharbi; H Ben Sik Ali; S Marghli; S Elatrous
Journal:  Diabetes Res Clin Pract       Date:  2009-10-23       Impact factor: 5.602

10.  Differences in glucose values obtained from point-of-care glucose meters and laboratory analysis in critically ill patients.

Authors:  Anjannette Cook; Delyla Laughlin; Margery Moore; Doreen North; Kathleen Wilkins; Gay Wong; Allyson Wallace-Scroggs; Lisa Halvorsen
Journal:  Am J Crit Care       Date:  2009-01       Impact factor: 2.228

View more
  3 in total

1.  Interferences With CGM Systems: Practical Relevance?

Authors:  Lutz Heinemann
Journal:  J Diabetes Sci Technol       Date:  2021-12-15

2.  Never a rose without a prick: pseudohyperglycemia when administering high-dose intravenous vitamin C.

Authors:  Yongli Zhang; Wei Huang
Journal:  Crit Care       Date:  2020-05-24       Impact factor: 9.097

3.  Pharmacokinetic Evaluation of Intravenous Vitamin C: A Classic Pharmacokinetic Study.

Authors:  Ping Chen; Greg Reed; Joyce Jiang; Yaohui Wang; Jean Sunega; Ruochen Dong; Yan Ma; Anna Esparham; Ryan Ferrell; Mark Levine; Jeanne Drisko; Qi Chen
Journal:  Clin Pharmacokinet       Date:  2022-06-25       Impact factor: 5.577

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.