Differences in glucose values obtained from point-of-care glucose meters and laboratory analysis in critically ill patients.

Anjannette Cook1, Delyla Laughlin, Margery Moore, Doreen North, Kathleen Wilkins, Gay Wong, Allyson Wallace-Scroggs, Lisa Halvorsen.   

Abstract

BACKGROUND: Blood for glucose analysis is often obtained interchangeably from indwelling catheters and fingersticks.
OBJECTIVES: To determine the level of agreement between glucose values obtained by laboratory analysis and with a point-of-care device for blood from 2 different sources: fingerstick and a central venous catheter.
METHODS: A method-comparison design was used. Point-of-care values for blood from fingersticks and catheters were compared with laboratory values for blood from catheters in a convenience sample of 67 critically ill patients. The effects of hematocrit level and finger edema on differences in glucose values between the 2 methods were also evaluated. A t test was used to determine differences in glucose values obtained via the 2 methods. Differences and limits of agreement were also calculated.
RESULTS: Laboratory glucose values for blood from a catheter differed significantly from point-of-care values for blood from the catheter (t(1,66) = -9.18; P < .001) and from a fingerstick (t(1,66) = 6.53; P < .001). Glucose values for the 2 methods differed by 20 mg/dL or more for 1 of 6 patients (15%) for catheter samples and for 1 of 5 (21%) for fingerstick samples. Point-of-care glucose values for fingerstick and catheter samples did not differ (P = .98). Hematocrit level significantly explained the difference in glucose values between the 2 methods for both catheter (R(2) = 0.288; P < .001) and fingerstick (R(2) = 0.280; P = .02) samples.
CONCLUSIONS: Use of a commonly used point-of-care device when precise glucose values are needed may lead to faulty treatment decisions.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19116407     DOI: 10.4037/ajcc2009626

Source DB:  PubMed          Journal:  Am J Crit Care        ISSN: 1062-3264            Impact factor:   2.228


  11 in total

1.  Impact of High-Dose Intravenous Vitamin C for Treatment of Sepsis on Point-of-Care Blood Glucose Readings.

Authors:  Anna Peyton Howell; Jenna L Parrett; Daniel R Malcom
Journal:  J Diabetes Sci Technol       Date:  2019-11-25

Review 2.  Accuracy of point-of-care glucose measurements.

Authors:  Annette Rebel; Mark A Rice; Brenda G Fahy
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

Review 3.  Glycemic control in the burn intensive care unit: focus on the role of anemia in glucose measurement.

Authors:  Elizabeth A Mann; Alejandra G Mora; Heather F Pidcoke; Steven E Wolf; Charles E Wade
Journal:  J Diabetes Sci Technol       Date:  2009-11-01

4.  Accuracy of point-of-care blood glucose measurements in critically ill patients in shock.

Authors:  Carlo Jan Pati-An Garingarao; Myrna Buenaluz-Sedurante; Cecilia Alegado Jimeno
Journal:  J Diabetes Sci Technol       Date:  2014-06-08

5.  Estimates of total analytical error in consumer and hospital glucose meters contributed by hematocrit, maltose, and ascorbate.

Authors:  Martha E Lyon; Jeffrey A DuBois; Gordon H Fick; Andrew W Lyon
Journal:  J Diabetes Sci Technol       Date:  2010-11-01

Review 6.  Monitoring nutrition and glucose in acute brain injury.

Authors:  Neeraj Badjatia; Paul Vespa
Journal:  Neurocrit Care       Date:  2014-12       Impact factor: 3.210

Review 7.  Accuracy of blood-glucose measurements using glucose meters and arterial blood gas analyzers in critically ill adult patients: systematic review.

Authors:  Shigeaki Inoue; Moritoki Egi; Joji Kotani; Kiyoshi Morita
Journal:  Crit Care       Date:  2013-03-18       Impact factor: 9.097

8.  Association between glycemic variability and mortality: How robust is the evidence?

Authors:  John Victor Peter; Raj Kumar Mani
Journal:  Indian J Crit Care Med       Date:  2014-05

9.  Glucose Meter Use in the Intensive Care Unit: Much Ado About Something.

Authors:  S Karon Brad
Journal:  EJIFCC       Date:  2014-09-04

10.  Sialochemical analysis in polytraumatized patients in intensive care units.

Authors:  Maria Heloisa Madruga Chaves; Amanda Rebeca da Silveira Wolf; Kelly Aline Lima Nascimento; Danielle Nawcki; Gabriele Muller Feustel; Patricia Vida Cassi Bettega; Sergio Aparecido Ignacio; João Armando Brancher; Luana Alves Tannous; Renata Iani Werneck; Paulo Henrique Couto Souza; Marlene Maria Tourais de Barros; Aline Cristina Batista Rodrigues Johann
Journal:  PLoS One       Date:  2019-10-03       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.