Literature DB >> 30140564

RELIABILITY AND VALIDITY OF THE HALO DIGITAL GONIOMETER FOR SHOULDER RANGE OF MOTION IN HEALTHY SUBJECTS.

Sarah Correll1, Jennifer Field2, Heather Hutchinson3, Gabby Mickevicius4, Amber Fitzsimmons5, Betty Smoot5.   

Abstract

BACKGROUND: Range of motion (ROM) of the shoulder is an integral component of assessment of musculoskeletal shoulder impairments. ROM is typically measured using a universal goniometer (UG). The UG has demonstrated good intra and inter-rater reliability for measuring shoulder ROM, although limitations exist. In recent years, alternative measurement devices such as smartphone applications and digital goniometers have been introduced, potentially addressing some of the shortcomings of the UG. Limited research is available on the validity and reliability of these alternative devices, including the laser-guided digital goniometer, in measuring shoulder ROM.Purpose: The purpose of this study was to investigate the intra- and inter-rater reliability and concurrent validity of a laser-guided digital goniometer (HALO) for measuring active shoulder ROM.
METHODS: A convenience sample of healthy volunteers was recruited. To be eligible, participants were required to be between 18 and 75 years of age and able to actively move at least one shoulder into 90 ° of glenohumeral abduction. Self-report of previous significant shoulder injury; previous shoulder surgery; current bilateral shoulder pain; current neck or upper back pain; or referred pain into the upper extremity were exclusion criteria. Active shoulder flexion, abduction, internal rotation, and external rotation were measured for each shoulder. Two evaluators measured each motion twice with each device (HALO and the UG) per shoulder. The intra-class correlation coefficient (ICC) for reliability and validity/agreement between devices was calculated using a two-way mixed model with a 95% confidence interval.
RESULTS: Data were analyzed for 75 shoulders from 41 participants (seven participants had only one shoulder evaluated). Intra-rater reliability ICCs are between 0.82 and 0.91 for the HALO, and 0.83 to 0.95 for the UG. Inter-rater reliability for the HALO was 0.89 to 0.98 and for the UG was 0.90 to 0.98. The ICCs for agreement, comparing the HALO digital goniometer to the UG ranged from 0.79 to 0.99.
CONCLUSION: This study provides evidence that the HALO digital goniometer can be a reliable and valid tool for measuring shoulder ROM in individuals with healthy shoulders. However, the two devices should not be used interchangeably to evaluate a single individual's change over time for any motion. LEVEL OF EVIDENCE: Diagnostic Study (clinical measurement), Level 2b.

Entities:  

Keywords:  Clinimetrics; goniometry; reliability; shoulder; validity

Year:  2018        PMID: 30140564      PMCID: PMC6088125     

Source DB:  PubMed          Journal:  Int J Sports Phys Ther        ISSN: 2159-2896


  13 in total

1.  The reliability and concurrent validity of shoulder mobility measurements using a digital inclinometer and goniometer: a technical report.

Authors:  Morey J Kolber; William J Hanney
Journal:  Int J Sports Phys Ther       Date:  2012-06

2.  Sample size and optimal designs for reliability studies.

Authors:  S D Walter; M Eliasziw; A Donner
Journal:  Stat Med       Date:  1998-01-15       Impact factor: 2.373

3.  Validity and reliability of smartphone magnetometer-based goniometer evaluation of shoulder abduction--A pilot study.

Authors:  Linda B Johnson; Sean Sumner; Tina Duong; Posu Yan; Ruzena Bajcsy; R Ted Abresch; Evan de Bie; Jay J Han
Journal:  Man Ther       Date:  2015-03-17

4.  A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research.

Authors:  Terry K Koo; Mae Y Li
Journal:  J Chiropr Med       Date:  2016-03-31

5.  Clinical measurement of range of motion. Review of goniometry emphasizing reliability and validity.

Authors:  R L Gajdosik; R W Bohannon
Journal:  Phys Ther       Date:  1987-12

6.  Glenohumeral joint rotation range of motion in competitive swimmers.

Authors:  Bryan L Riemann; Joe Witt; George J Davies
Journal:  J Sports Sci       Date:  2011-07-22       Impact factor: 3.337

7.  Reliability and validity of goniometric iPhone applications for the assessment of active shoulder external rotation.

Authors:  Katy Mitchell; Simran Bakshi Gutierrez; Stacy Sutton; Stephanie Morton; Andrea Morgenthaler
Journal:  Physiother Theory Pract       Date:  2014-03-21       Impact factor: 2.279

8.  Axial rotation gravity goniometer. A simple design of instrument and a controlled reliability study.

Authors:  W Laupattarakasem; W Sirichativapee; W Kowsuwon; S Sribunditkul; C Suibnugarn
Journal:  Clin Orthop Relat Res       Date:  1990-02       Impact factor: 4.176

9.  A Comparison of Elbow Range of Motion Measurements: Smartphone-Based Digital Photography Versus Goniometric Measurements.

Authors:  Megan A Meislin; Eric R Wagner; Alexander Y Shin
Journal:  J Hand Surg Am       Date:  2016-02-13       Impact factor: 2.230

10.  Concurrent validity and reliability of the Simple Goniometer iPhone app compared with the Universal Goniometer.

Authors:  Anne Jones; Rebecca Sealey; Michael Crowe; Susan Gordon
Journal:  Physiother Theory Pract       Date:  2014-03-25       Impact factor: 2.279

View more
  10 in total

1.  Validation of a novel range of motion assessment tool for the cervical spine: the HALO© digital goniometer.

Authors:  Ashley R Wilson-Smith; Saimurooban Muralidaran; Monish Maharaj; Matthew H Pelletier; Peter Beshara; Prashanth Rao; Louise M Pearce; Tian Wang; Ralph J Mobbs; William R Walsh
Journal:  J Spine Surg       Date:  2022-03

2.  Effect of Task-Specific Training on Trunk Control and Balance in Patients with Subacute Stroke.

Authors:  Mohamed E Khallaf
Journal:  Neurol Res Int       Date:  2020-11-17

3.  Effects of SNAG mobilization combined with a self-SNAG home-exercise for the treatment of cervicogenic headache: a pilot study.

Authors:  Jean-Philippe Paquin; Yannick Tousignant-Laflamme; Jean-Pierre Dumas
Journal:  J Man Manip Ther       Date:  2021-02-05

4.  Evaluation of a New Rotator Cuff Trainer Based on Oscillating Hydraulic Damping.

Authors:  Yinghao Wang; Jianfeng Wu; Hongchun Yang; Zhichuan Tang; Guozhong Chai
Journal:  Healthcare (Basel)       Date:  2020-01-28

5.  Validity and Reliability of a New Optoelectronic System for Measuring Active Range of Motion of Upper Limb Joints in Asymptomatic and Symptomatic Subjects.

Authors:  Rodrigo Martín-San Agustín; Jose A García-Vidal; German Cánovas-Ambit; Aurelio Arenas-Della Vecchia; Manuel López-Nicolás; Francesc Medina-Mirapeix
Journal:  J Clin Med       Date:  2019-11-02       Impact factor: 4.241

6.  Effects of a 12-week resistance and aerobic exercise program on muscular strength and quality of life in breast cancer survivors: Study protocol for the EFICAN randomized controlled trial.

Authors:  Alberto Soriano-Maldonado; Álvaro Carrera-Ruiz; David M Díez-Fernández; Alba Esteban-Simón; Mercedes Maldonado-Quesada; Nuria Moreno-Poza; María Del Mar García-Martínez; Celia Alcaraz-García; Rosa Vázquez-Sousa; Herminia Moreno-Martos; Antonio Toro-de-Federico; Nur Hachem-Salas; Eva Artés-Rodríguez; Manuel A Rodríguez-Pérez; Antonio J Casimiro-Andújar
Journal:  Medicine (Baltimore)       Date:  2019-11       Impact factor: 1.817

Review 7.  The Reliability of the Microsoft Kinect and Ambulatory Sensor-Based Motion Tracking Devices to Measure Shoulder Range-of-Motion: A Systematic Review and Meta-Analysis.

Authors:  Peter Beshara; David B Anderson; Matthew Pelletier; William R Walsh
Journal:  Sensors (Basel)       Date:  2021-12-08       Impact factor: 3.576

8.  Validity and reliability of a smartphone application for self-measurement of active shoulder range of motion in a standing position among healthy adults.

Authors:  Hiroki Shimizu; Takanobu Saito; Chika Kouno; Kanako Shimoura; Rika Kawabe; Yuki Shinohara; Kohei Mukaiyama; Chen Changyu; Masaya Kato; Momoko Nagai-Tanima; Tomoki Aoyama
Journal:  JSES Int       Date:  2022-05-05

9.  Reliability of a Smartphone Compared With an Inertial Sensor to Measure Shoulder Mobility: Cross-Sectional Study.

Authors:  Cristina Roldán-Jiménez; Jaime Martin-Martin; Antonio I Cuesta-Vargas
Journal:  JMIR Mhealth Uhealth       Date:  2019-09-06       Impact factor: 4.773

10.  Roller Massage: Comparing the Immediate Post-Treatment Effects Between an Instructional Video and a Self-Preferred Program Using Two Different Density-Type Roller Balls.

Authors:  Scott W Cheatham; Kyle R Stull; Wendy N Batts; Tony Ambler-Wright
Journal:  J Hum Kinet       Date:  2020-01-31       Impact factor: 2.193

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.