Chengcheng Zhu1, Lizhen Cao2,3, Zhaoying Wen4,5, Sinyeob Ahn6, Esther Raithel7, Christoph Forman7, Michael Hope2, David Saloner2.
Abstract
BACKGROUND: 3D non-contrast high-resolution black-blood cardiovascular magnetic resonance (CMR) (DANTE-SPACE) has been used for surveillance of abdominal aortic aneurysm (AAA) and validated against computed tomography (CT) angiography. However, it requires a long scan time of more than 7 min. We sought to develop an accelerated sequence applying compressed sensing (CS-DANTE-SPACE) and validate it in AAA patients undergoing surveillance.Entities:
Keywords: Abdominal aortic aneurysm, black blood CMR; Compressed sensing; Patient surveillance
Mesh:
Year: 2019 PMID: 31660983 PMCID: PMC6816154 DOI: 10.1186/s12968-019-0571-2
Source DB: PubMed Journal: J Cardiovasc Magn Reson ISSN: 1097-6647 Impact factor: 5.364
Patient demographics
| All ( | |
|---|---|
| Age, mean ± SD | 73 ± 6 |
| Male, n, % | 38 (100) |
| Height (cm) | 176 ± 6 |
| Weight (Kg) | 82.9 ± 18.8 |
| Body Mass Index | 27.0 ± 6.1 |
| Hypertension a, n, % | 28 (73.6) |
| DM b, n, % | 13 (34.2) |
| Smoking, n, % | 22 (57.9) |
| CAD c, n, % | 15 (39.5) |
a: Hypertension is defined as resting blood pressure > 140/90 mmHg;
b: Diabetes Mellitus;
c: Coronary Artery Disease
Fig. 1Optimization of reconstruction parameters for compressed sensing (CS) CS-DANTE-SPACE. A regularization factor of 0.002 had best image quality and 10 iteration times were sufficient
Qualitative and quantitative image quality assessment of DANTE-SPACE and CS-DANTE-SPACE. Results from reviewer 1 are shown. ILT = intraluminal thrombus
| DANTE- | CS-DANTE | ICC | Bias (LOA) | CV (100%) | ||
|---|---|---|---|---|---|---|
| Maximal Diameter (cm) | 4.85 ± 0.87 | 4.84 ± 0.88 | 0.55 | 0.997 | 0.01(− 0.13, 0.14) | 1.4 |
| ILT/wall Area (cm2) | 9.55 ± 5.72 | 9.62 ± 5.81 | 0.46 | 0.995 | −0.07(−1.22, 1.10) | 6.2 |
| Lumen Area (cm2) | 7.62 ± 4.46 | 7.64 ± 4.41 | 0.65 | 0.996 | −0.02(− 0.77,0.72) | 5.0 |
| ILT signal ratio | 1.05 ± 0.29 | 1.03 ± 0.32 | 0.35 | 0.851 | 0.02(−0.30,0.35) | 16.0 |
| Contrast Ratio | 2.50 ± 0.91 | 2.95 ± 0.98 | < 0.001 | NA | NA | NA |
| Sharpness-inner (mm−1) | 0.46 ± 0.53 | 0.52 ± 0.57 | 0.04 | NA | NA | NA |
| Sharpness-outer (mm−1) | 0.42 ± 0.83 | 0.47 ± 0.64 | 0.02 | NA | NA | NA |
| Sharpness-average (mm−1) | 0.44 ± 0.55 | 0.50 ± 0.48 | 0.002 | NA | NA | NA |
| Image quality score | 3.22 ± 0.58 | 3.24 ± 0.62 | 0.766 | N | NA | NA |
Fig. 2Bland-Altman plots for measurements from DANTE-SPACE and CS-DANTE-SPACE images (data from Reader 1 is shown). ILT = intraluminal thrombus
Inter-reader agreement for quantitative measurements and qualitative image quality on DANTE-SPACE
| DANTE-SPACE | DANTE-SPACE | ICC | Bias (LOA) | CV | ||
|---|---|---|---|---|---|---|
| Maximal Diameter (cm) | 4.85 ± 0.87 | 4.84 ± 0.87 | 0.92 | 0.985 | 0.01(−0.30,0.30) | 3.2 |
| ILT/wall Area (cm2) | 9.55 ± 5.72 | 9.44 ± 5.78 | 0.44 | 0.983 | 0.11(−1.90, 2.12) | 10.8 |
| Lumen Area (cm2) | 7.62 ± 4.46 | 7.53 ± 4.52 | 0.31 | 0.990 | 0.09(−1.05, 1.22) | 7.6 |
| Contrast Ratio | 2.50 ± 0.91 | 2.44 ± 0.94 | 0.18 | 0.944 | 0.06(−0.54,0.66) | 12.4 |
| ILT signal ratio | 1.05 ± 0.29 | 0.98 ± 0.26 | < 0.001 | 0.884 | 0.06(−0.16,0.30) | 11.6 |
| Image quality score | 3.22 ± 0.58 | 2.84 ± 0.65 | < 0.001 | 0.577 | 0.38(−0.73,1.49) | 18.7 |
Inter-reader agreement for quantitative measurements and qualitative image quality on CS-DANTE-SPACE
| CS-SPACE | CS-SPACE | ICC | Bias (LOA) | CV | ||
|---|---|---|---|---|---|---|
| Maximal Diameter (cm) | 4.84 ± 0.88 | 4.83 ± 0.92 | 0.76 | 0.985 | 0.01(−0.30,0.31) | 3.2 |
| ILT/wall Area (cm2) | 9.62 ± 5.81 | 9.39 ± 5.72 | 0.06 | 0.989 | 0.23(−1.38,1.83) | 8.6 |
| Lumen Area (cm2) | 7.64 ± 4.41 | 7.72 ± 4.49 | 0.31 | 0.993 | −0.08(−1.13,0.97) | 7.0 |
| Contrast Ratio | 2.95 ± 0.98 | 2.89 ± 0.93 | 0.50 | 0.944 | 0.06(−1.16,1.28) | 21.4 |
| ILT signal ratio | 1.03 ± 0.32 | 0.95 ± 0.29 | 0.01 | 0.815 | 0.08(−0.28,0.42) | 17.8 |
| Image quality score | 3.24 ± 0.62 | 3.06 ± 0.71 | 0.05 | 0.559 | 0.18(−1.05,1.41) | 20.0 |
Fig. 3Bland-Altman plots showing inter-reader agreement of DANTE-SPACE measurements
Fig. 4Bland-Altman plots showing inter-reader agreement of CS-DANTE-SPACE measurements
Fig. 5A patient with a 4.7 cm aneurysm. CS-DANTE-SPACE achieved comparable image quality with DANTE-SPACE, but with a 41% scan time reduction. Reference CTA 2 month before CMR is shown on the right. Red arrows show the level of maximal diameter. Heterogeneous intraluminal thrombus (ILT) was demonstrated in both sequences. A sharper image was noted for CS-DANTE-SPACE, possibly due to the reduced motion artifacts because of shorter scan time
Fig. 6A patient with a 4.7 cm aneurysm. CS-DANTE-SPACE achieved comparable image quality with DANTE-SPACE, but with a 41% scan time reduction. CS-DANTE-SPACE has slightly higher sharpness and contrast ratio than DANTE-SPACE. Red arrows show the level of maximal diameter. Bright (fresh) intraluminal thrombus (ILT) was demonstrated in both sequences