| Literature DB >> 31660296 |
Hang Shu1,2, Chongxian Yin1,2, Haiyang Zhang3, Ming Liu1,2, Manzhou Zhang1,2, Liying Zhao1,2, Kecheng Chu1,2, Xiaolei Dai1,2, Michael F Moyers4.
Abstract
The Shanghai Advanced Proton Therapy facility (SAPT) is a hospital-based facility that began construction in December of 2014 with commissioning of the first scanned proton beam line starting in October of 2017. Proton beams are extracted from a synchrotron accelerator with energies between 70 and 235 MeV. Beam delivery uses the modulated scanning and energy stacking techniques to produce a maximal scanning area of 40 × 30 cm2 at the iso-center. Prior to clinical use, the beam delivery system was characterized and calibrated following the guidelines of the IEC 62667 medical electronic equipment standard including the spot size in air, spot position, depth dose distributions, and lateral dose profiles, as well as the beam monitor calibrations following the IAEA TRS-398 recommendations with small differences. •The measured dosimetric results showed that the full width at half maximum (FWHM) for the beam spot size in air varied approximately from 6 mm to 13 mm. The dose fall-off (DDF) derived from the measured depth dose in water varied from 4.7 mm at 235 MeV to 0.7 mm at 70 MeV. The homogeneity of the scanned field was better than 2% for various energies as expected.•Furthermore, the beam reproducibility and proportionality delivery accuracy was also stable with the results better than 0.1% and 1% respectively. Finally, the dose monitor calibration factor, its reproducibility and stability were tested. Reproducibility tests exhibited a standard deviation (SD) result of less than 1% during the test period.•All the measured dosimetric parameters showed that the design specifications were well achieved and the results are suitable for being used as a part of the clinical commissioning and quality assurance program for treating patients.Entities:
Keywords: Dosimetry; Proton therapy; Scanned Proton Beam Performance and Calibration; Spot scanning; Technical commissioning
Year: 2019 PMID: 31660296 PMCID: PMC6807371 DOI: 10.1016/j.mex.2019.08.001
Source DB: PubMed Journal: MethodsX ISSN: 2215-0161
Fig. 1Layout of the SAPT scanning system for fixed beam line.
The main parameters of beam delivery system.
| Parameters | Value |
|---|---|
| Energy | 70–235 MeV |
| Energy steps | 94 steps |
| Maximum range | 34 g/cm2 |
| Field size (U × V) | 40 cm × 30 cm |
| SAD (U, V) | 2.87 m, 2.42 m |
| Scanning speed (U, V) | 2 cm/ms, 0.5 cm/ms |
| Spot size at isocenter in air | 6–13 mm (FWHM) |
| Beam intensity | 0.3–3 nA |
| Dose rate | 2 Gy/(L min) |
Fig. 2FWHM of lateral profiles in air of single spot beams at the iso-center plane as a function of energy. The solid curves represent the simulation results while the dots represent the measurements.
Fig. 3The measurement results of spot matrices at 70 MeV, 161.1 MeV and 235 MeV.
Fig. 4a) shows the homogeneity distributions of the standard calibration field with a 60 mm × 60 mm and the spot spacing with 3 mm × 3 mm at 161.1 MeV. Fig. b) and Fig. c) describe the characteristic of the X and Y planes which illustrate that the homogeneity of the delivered dose in the scanning fields was better than 2%.
The homogeneity results of the major axis lateral profiles at 70 MeV, MeV, 161.1 MeV, 179.9 MeV, 202 MeV, 219.2 MeV, and 235 MeV.
| Energy (MeV) | X Penumbra (mm) | X Flatness (mm) | Y Penumbra (mm) | Y Flatness (mm) |
|---|---|---|---|---|
| 235.0 | 5.1 | 1.4% | 4.8 | 1.4% |
| 219.2 | 5.2 | 1.7% | 6.5 | 1.0% |
| 202.0 | 5.4 | 1.0% | 6.2 | 0.9% |
| 179.9 | 5.4 | 1.3% | 5.3 | 0.9% |
| 161.1 | 4.8 | 1.6% | 4.2 | 1.2% |
| 130.1 | 7.3 | 1.5% | 6.0 | 0.8% |
| 70.0 | 9.4 | 1.3% | 8.4 | 0.5% |
Fig. 5Laterally integrated depth dose distributions measured in water with the Peakfinder at 70 MeV, 130.1 MeV, 161.1 MeV, 179.9 MeV, 202 MeV, 219.2 MeV, and 235 MeV.
Parameters derived from the measured depth dose distributions.
| Energy (MeV) | Range (g/cm2) | Entrance-to-peak dose ratio | DDF (mm) |
|---|---|---|---|
| 235.0 | 34.09 | 4.7 | 4.7 |
| 219.2 | 30.17 | 4.9 | 4.2 |
| 202.0 | 26.19 | 5.1 | 3.8 |
| 179.9 | 21.39 | 5.4 | 3.2 |
| 161.1 | 17.67 | 5.6 | 2.6 |
| 130.1 | 12.05 | 5.9 | 1.9 |
| 70.0 | 3.84 | 6.4 | 0.7 |
Fig. 6The beam reproducibility delivery accuracy is better than 0.1% at 235 MeV.
Fig. 7Linear fitting with multiple dose rates of the primary dose monitor.
The result of proportionality of MU delivery at 235 MeV.
| Dose rate (MU/s) | TW34070 /IC output (nC/MU) | Maximum deviation (%) |
|---|---|---|
| 1 | 4.3948 | 0.34 |
| 3 | 4.3975 | 0.29 |
| 5 | 4.3924 | 0.44 |
| 8 | 4.3971 | 0.46 |
| 10 | 4.4004 | 0.19 |
Fig. 8The monitor calibration factor normalized to 161.1 MeV.
Fig. 9Monitor calibration factor measured daily result during the test period.
| Subject Area: | Physics and Astronomy |
| More specific subject area: | Dosimetry |
| Method name: | Scanned Proton Beam Performance and Calibration |
| Name and reference of original method: | IEC-62667 medical standard; |