| Literature DB >> 31650728 |
Jooyoung Cho1,2, Joowon Oh1,3, Sang Guk Lee1, You Hee Lee1, Jaewoo Song1, Jeong Ho Kim4.
Abstract
BACKGROUND: Automated cellular analyzers are expected to improve the analytical performance in body fluid (BF) analysis. We evaluated the analytical performance of three automated cellular analyzers and established optimum reflex analysis guidelines.Entities:
Keywords: Analytical performance; Automated cellular analyzer; Body fluid; Differential cell counting; UF-5000; UniCel DxH 800; XN-350
Year: 2020 PMID: 31650728 PMCID: PMC6822009 DOI: 10.3343/alm.2020.40.2.122
Source DB: PubMed Journal: Ann Lab Med ISSN: 2234-3806 Impact factor: 3.464
TNC and RBC precision, linearity, and detection capabilities using three automated analyzers
| Within-device SD (CV%) | AMR | Detection capability | |||||||
|---|---|---|---|---|---|---|---|---|---|
| Low | Mid | High | Claimed | Validated | LoB | LoD | LoQ | ||
| UniCel DxH 800 | TNC × 109/L | 0.125 ± 0.012 (9.3) | 1.382 ± 0.060 (4.4) | 65.676 ± 0.791 (1.2) | 0.020–89.000 | 9.501–38.005 | 0.012 | 0.018 | 0.037 |
| RBC × 1012/L | 0.012 ± 0.001 (4.3) | 1.741 ± 0.014 (0.8) | 5.843 ± 0.137 (2.4) | 0.001–6.200 | 1.499–5.996 | < 0.001 | < 0.001 | > 0.005 | |
| XN-350 | TNC × 109/L | 0.083 ± 0.005 (5.6) | - | 0.324 ± 0.008 (2.4) | 0.003–10.000 | 2.629–10.517 | < 0.001 | 0.003 | 0.027 |
| RBC × 1012/L | 0.026 ± 0.002 (8.0) | - | 0.077 ± 0.003 (3.3) | 0.002–5.000 | 1.250–5.001 | < 0.001 | < 0.001 | 0.003 | |
| UF-5000 | TNC × 109/L | 0.038 ± 0.003 (6.7) | - | 0.796 ± 0.017 (2.1) | 0.001–10.000 | 2.354–9.416 | < 0.001 | 0.002 | 0.025 |
| RBC × 1012/L | < 0.001 (5.4) | - | < 0.001 (3.2) | 0.000–0.010 | 0.002–0.009 | < 0.001 | < 0.001 | < 0.001 | |
UniCel DxH 800 (Beckman Coulter, Brea, CA, USA), XN-350 (Sysmex, Kobe, Japan), and UF-5000 (Sysmex, Kobe, Japan).
Abbreviations: TNC, total nucleated cell; RBC, red blood cell; AMR, analytical measurement range; LoB, limit of blank; LoD, limit of detection; LoQ, limit of quantification.
Passing-Bablok regression and Bland-Altman analysis for TNC and RBC counts, PMN (%), and MN (%) in CSF (N=88) and non-CSF (N=454) samples compared with the manual method
| CSF samples (N = 88) | Non-CSF samples (N = 454) | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| Slope (95% CI) | Intercept (95% CI) | %bias of the slope | Bland-Altman bias (95% CI) | Slope (95% CI) | Intercept (95% CI) | %bias of the slope | Bland-Altman bias (95% CI) | ||
| UniCel DxH 800 | TNC × 109/L | 0.89 (0.76–1.22) | 0.002 (-0.017–0.019) | 10.8 | -0.01 (-0.033–0.013) | 0.93 (0.91–0.95) | 0.023 (0.017–0.030) | -6.6 | -0.067 (-0.154–0.020) |
| RBC × 1012/L | 1.41 (1.10–2.14) | 0.045 (0.024–0.070) | 40.9 | 0.174 (0.066–0.283) | 1.13 (1.10–1.15) | 0.164 (0.141–0.193) | 12.8 | 0.77 (0.582–0.959) | |
| PMN (%) | 0.63 (0.47–1.01) | 3.19 (0.58–7.83) | -036.9 | -7.8 (-18.15–2.55) | 0.72 (0.65–0.79) | 11.29 (8.96–14.31) | -028.1 | 3.3 (0.64–5.96) | |
| MN (%) | 0.64 (0.41–8.88) | 19.19 (-792.13–36.95) | -036.1 | -9.92 (-24.62–4.79) | 0.75 (0.65–0.86) | 1.49 (-6.75–8.19) | -024.9 | -16.05 (-19.31–-12.78) | |
| XN-350 | TNC × 109/L | 1.12 (0.98–1.23) | 0.001 (0.001–0.002) | 12.1 | 0.015 (-0.006–0.035) | 1.11 (1.09–1.14) | 0.003 (0.001–0.007) | 11.2 | 0.105 (0.029–0.180) |
| RBC × 1012/L | 0.96 (0.86–1.10) | 0.1 (0.099–0.103) | -4.2 | 0.139 (0.048–0.229) | 1.05 (1.03–1.08) | 0.1 (0.092–0.119) | 5.3 | 0.455 (0.268–0.644) | |
| PMN (%) | 0.97 (0.90–1.47) | 0.93 (-1.23–2.60) | -2.9 | 1.33 (-2.67–5.33) | 0.91 (0.89–0.93) | 3.63 (2.82–4.50) | -8.6 | 1.93 (0.95–2.91) | |
| MN (%) | 0.97 (0.88–1.47) | 1.96 (-47.48–9.78) | -2.9 | -1.35 (-5.35–2.64) | 0.91 (0.88–0.93) | 7.03 (5.01–9.74) | -9.5 | 1.9 (0.29–3.52) | |
| UF-5000 | TNC × 109/L | 1.08 (1.00–1.20) | 0.001 (0.001–0.002) | 33 | 0.009 (-0.004–0.022) | 0.97 (0.93–1.00) | 0.011 (0.005–0.016) | -3.2 | -0.199 (-0.431–0.032) |
| RBC × 1012/L | 1.11 (1.05–1.21) | < 0.001 | 11.2 | < 0.001 | 1.1 (1.08–1.12) | < 0.001 | 10.4 | < 0.001 | |
| PMN (%) | 0.86 (0.66–1.10) | 1.18 (-0.30–4.06) | -13.9 | -2.27 (-5.88–1.35) | 0.84 (0.81–0.88) | 6.92 (5.71–8.57) | -16.1 | 5.03 (3.46–6.59) | |
| MN (%) | 0.86 (0.65–0.98) | 12.78 (1.13–33.21) | -14.1 | 2.16 (-1.49–5.81) | 0.82 (0.78–0.85) | 12.61 (9.71–15.77) | -18.5 | -1.12 (-2.98–0.75) | |
Abbreviations: TNC, total nucleated cell; RBC, red blood cell; PMN, polymorphonuclear cell; MN, mononuclear cell; CSF, cerebrospinal fluid; CI, confidence interval.
Passing-Bablok regression and Bland-Altman analysis for differential cell counts using XN-350 compared with manual differential counting (N=803)
| Slope (95% CI) | Intercept (95% CI) | %bias of the slope | Bland-Altman bias (95% CI) | |
|---|---|---|---|---|
| PMN | 0.96 (0.95–0.98) | 4.31 (3.94–4.75) | -3.6 | 4.60 (4.10–5.11) |
| Neutrophil | 0.96 (0.94–9.97) | 4.51 (4.20–4.92) | -4.2 | 4.59 (4.10–5.08) |
| Eosinophil | 2.50 (1.29–∞) | -2.10 (-∞–-0.80) | 150.1 | 0.71 (0.53–0.89) |
| MN | 0.94 (0.92–0.95) | 2.95 (1.59–4.30) | -6.2 | -0.31 (-1.08–0.47) |
| Lymphocyte | 1.13 (1.09–1.16) | 5.04 (4.16–6.03) | 12.7 | 9.44 (8.75–10.13) |
| Monocyte | 0.74 (0.71–0.76) | -0.01 (-0.70–0.69) | -26.5 | -9.77 (-10.61–-8.93) |
| HF-BF | 1.60 (1.28–2.30) | 0.70 (-0.80–1.79) | 60.2 | 8.77 (6.99–10.56) |
Abbreviations: CI, confidence interval; PMN, polymorphonuclear cell; MN, mononuclear cell; HF-BF, high fluorescence-body fluid cell.
Diagnostic performance of high eosinophil (>10%) using XN-350 at various eosinophil (%) cut-off values (mean and 95% confidence intervals)
| Cut-off | Sensitivity (%) | Specificity (%) | PPV (%) | NPV (%) | Accuracy (%) | LR+ | LR- |
|---|---|---|---|---|---|---|---|
| ≥ 1.0 | 99.0 (94.4–99.8) | 71.9 (69.3–74.3) | 21.6 (20.1–23.2) | 99.9 (99.2–99.9) | 73.9 (68.0–79.7) | 3.52 (3.22–3.86) | 0.01 (0.00–0.08) |
| ≥ 1.7 | 99.0 (94.4–99.8) | 80.5 (78.3–82.7) | 28.5 (26.2–30.9) | 99.9 (99.3–99.9) | 81.9 (75.2–88.5) | 5.09 (4.54–5.71) | 0.01 (0.00–0.01) |
| ≥ 3.1 | 97.9 (92.8–99.4) | 90.2 (88.5–91.8) | 44.0 (39.8–48.2) | 99.8 (99.3–99.9) | 90.8 (83.1–98.5) | 10.03 (8.46–11.92) | 0.02 (0.01–0.02) |
| ≥ 3.8 | 95.9 (89.9–98.4) | 92.5 (90.9–93.8) | 50.0 (45.0–55.0) | 99.7 (99.1–99.9) | 92.7 (84.8–99.9) | 12.77 (10.47–15.60) | 0.04 (0.01–0.10) |
| ≥ 5.2 | 90.7 (83.3–95.0) | 95.5 (94.2–96.5) | 61.1 (54.7–67.2) | 99.3 (98.6–99.6) | 95.1 (86.8–99.9) | 20.07 (15.43–26.12) | 0.10 (0.05–0.17) |
| ≥ 8.3 | 80.4 (71.4–87.1) | 98.0 (97.0–98.6) | 75.7 (67.6–82.3) | 98.5 (97.7–99.0) | 96.6 (88.1–99.9) | 39.85 (26.78–59.33) | 0.20 (0.13–0.29) |
| ≥ 10.1 | 70.1 (60.4–78.3) | 98.7 (97.9–99.2) | 81.0 (72.0–87.6) | 97.7 (96.9–98.3) | 96.5 (87.9–99.9) | 54.29 (32.97–89.40) | 0.30 (0.22–0.40) |
| ≥ 13.3 | 51.5 (43.7–63.2) | 99.5 (98.8–99.8) | 92.6 (79.3–97.2) | 96.1 (95.7–97.1) | 95.9 (87.4–99.9) | 110.71 (46.86–245.97) | 0.47 (0.37–0.57) |
| ≥ 16.1 | 39.2 (30.1–49.1) | 99.7 (99.2–99.9) | 90.5 (77.6–96.3) | 95.4 (94.7–96.1) | 95.1 (86.5–99.9) | 121.35 (45.98–320.40) | 0.61 (0.51–0.70) |
Abbreviations: PPV, positive predictive value; NPV, negative predictive value; LR+, positive likelihood ratio; LR-, negative likelihood ratio.
Fig. 1ROC curve analysis of HF-BFs obtained using the Sysmex XN-350 and other cells obtained by manual counting compared with positive results (positive for malignancy and atypical cells) determined by cytology. Lining cells and monocytes/histiocytes were classified together as MNs. Macrophages, lining cells, such as mesothelial cells, atypical cells, and malignant cells were counted as “other cells.”
Abbreviations: HF-BFs, high fluorescence-body fluid cells; TPF, true positive fraction; FPF, false positive fraction; AUC, area under the curve; CI, confidence interval; MNs, mononuclear cells.
Diagnostic performance according to various HF-BFs/100 WBCs cut-off values obtained using XN-350 for detection of potential malignancy (mean and 95% confidence intervals)
| Cut-off | Sensitivity (%) | Specificity (%) | PPV (%) | NPV (%) | Accuracy (%) | LR+ | LR- |
|---|---|---|---|---|---|---|---|
| > 0.7 | 95.1 (91.2–97.3) | 19.6 (16.4–23.2) | 31.4 (30.2–32.5) | 91.1 (84.6–95.1) | 40.6 (35.6–45.4) | 1.18 (1.12–1.25) | 0.25 (0.14–0.46) |
| ≥ 2.6 | 81.8 (75.9–86.5) | 36.3 (32.3–40.5) | 33.2 (31.1–35.2) | 83.7 (79.1–87.6) | 49.0 (45.7–52.3) | 1.28 (1.17–1.40) | 0.50 (0.37–0.68) |
| ≥ 4.3 | 71.4 (64.9–77.2) | 45.1 (40.9–49.3) | 33.5 (30.9–36.0) | 80.3 (76.3–83.8) | 52.4 (49.7–55.1) | 1.30 (1.15–1.46) | 0.63 (0.49–0.79) |
| ≥ 5.9 | 67.0 (60.3–73.1) | 52.7 (48.4–56.9) | 35.4 (32.4–38.4) | 80.5 (77.0–83.6) | 56.7 (54.0–59.3) | 1.42 (1.24–1.61) | 0.63 (0.50–0.77) |
| ≥ 8.7 | 52.2 (45.4–59.0) | 63.7 (59.5–67.7) | 35.7 (31.8–39.8) | 77.5 (74.7–80.2) | 60.5 (57.5–63.5) | 1.44 (1.20–1.70) | 0.75 (0.64–0.87) |
| ≥ 12.3 | 42.4 (35.8–49.2) | 72.1 (68.1–75.7) | 37.0 (32.1–41.9) | 76.4 (74.0–78.7) | 63.8 (60.2–67.4) | 1.52 (1.22–1.87) | 0.80 (0.70–0.90) |
| ≥ 15.8 | 33.0 (26.9–39.7) | 78.1 (74.4–81.5) | 36.9 (31.1–42.9) | 75.1 (73.1–77.1) | 65.6 (61.4–69.7) | 1.51 (1.17–1.94) | 0.86 (0.77–0.95) |
| ≥ 20.3 | 22.7 (17.4–28.9) | 84.8 (81.5–87.6) | 36.6 (29.4–44.3) | 73.9 (72.3–75.5) | 67.5 (62.6–72.4) | 1.49 (1.07–2.05) | 0.91 (0.83–0.99) |
| ≥ 48.2 | 11.8 (8.1–17.0) | 97.1 (95.3–98.3) | 61.6 (46.1–74.9) | 74.0 (73.0–75.1) | 73.4 (67.2–79.6) | 4.15 (2.24–7.66) | 0.91 (0.85–0.95) |
| ≥ 54.8 | 11.3 (7.7–16.4) | 97.9 (96.3–98.8) | 67.7 (50.9–80.8) | 74.1 (73.1–75.1) | 73.8 (67.5–80.1) | 5.42 (2.73–10.76) | 0.91 (0.85–0.95) |
| ≥ 83.4 | 6.9 (4.2–11.2) | 99.2 (97.8–99.6) | 77.8 (55.5–88.5) | 73.4 (72.6–74.1) | 73.5 (66.9–79.9) | 9.07 (2.75–19.14) | 0.94 (0.90–0.97) |
Abbreviations: see Table 4.