Literature DB >> 28419185

A New Automated Technology for Cerebrospinal Fluid Cell Counts: Comparison of Accuracy and Clinical Impact of GloCyte, Sysmex XN, and Manual Methods.

Linda M Sandhaus1, Christine A Dillman1, Warren P Hinkle2, Julie M MacKenzie3, George Hong3.   

Abstract

OBJECTIVES: The purpose of the study was to compare the performance of GloCyte (Advanced Instruments, Norwood, MA), a new semiautomated instrument for cerebrospinal fluid cell counting, with the manual hemocytometer method and the automated Sysmex XN (Sysmex, Kobe, Japan) body fluid mode. The clinical impact of replacing the manual method with either automated method was determined.
METHODS: Fifty-seven samples from 38 patients were analyzed by all three methods. Pearson correlation and Passing-Bablok regression were used to compare methods. Cytospin smears were reviewed on all samples, and clinical histories were obtained.
RESULTS: There was a strong linear relationship between the manual and automated methods for WBC counts ( R  = 0.988 for GloCyte; R  = 0.980 for Sysmex XN). Positive bias was absent or negligible for WBC counts less than 30/μL. GloCyte and manual RBC counts were equivalent. There were no samples for which replacement of manual WBC counts by automated counts would have changed the diagnosis. Both automated methods showed improved precision for WBC counts compared with the manual method.
CONCLUSIONS: Replacing manual WBC counts by GloCyte or Sysmex XN WBC counts would improve consistency of results without compromising diagnostic accuracy. © American Society for Clinical Pathology, 2017. All rights reserved. For permissions, please e-mail: journals.permissions@oup.com

Entities:  

Keywords:  Automated CSF cell counts; Body fluid cell counts; Cerebrospinal fluid cell counts; GloCyte; Sysmex XN; Total nucleated cells

Mesh:

Year:  2017        PMID: 28419185     DOI: 10.1093/ajcp/aqx026

Source DB:  PubMed          Journal:  Am J Clin Pathol        ISSN: 0002-9173            Impact factor:   2.493


  4 in total

1.  The World Health Organization's Essential Diagnostics List: Diagnostics for neurologic disorders.

Authors:  Greer Waldrop; Teddy G Goetz; Omar K Siddiqi; Igor J Koralnik; Hiral Shah; Kiran T Thakur
Journal:  Neurology       Date:  2019-10-08       Impact factor: 9.910

Review 2.  Technological features of blast identification in the cerebrospinal fluid: A systematic review of flow cytometry and laboratory haematology methods.

Authors:  John L Frater; Cara Lunn Shirai; Jonathan R Brestoff
Journal:  Int J Lab Hematol       Date:  2022-07-04       Impact factor: 3.450

3.  On-Chip Cell Staining and Counting Platform for the Rapid Detection of Blood Cells in Cerebrospinal Fluid.

Authors:  Yujin Lee; Byeongyeon Kim; Sungyoung Choi
Journal:  Sensors (Basel)       Date:  2018-04-07       Impact factor: 3.576

4.  Performance Evaluation of Body Fluid Cellular Analysis Using the Beckman Coulter UniCel DxH 800, Sysmex XN-350, and UF-5000 Automated Cellular Analyzers.

Authors:  Jooyoung Cho; Joowon Oh; Sang Guk Lee; You Hee Lee; Jaewoo Song; Jeong Ho Kim
Journal:  Ann Lab Med       Date:  2020-03       Impact factor: 3.464

  4 in total