Preliminary evaluation of UF-5000 Body Fluid Mode for automated cerebrospinal fluid cell counting.

Michela Seghezzi1, Barbara Manenti1, Giulia Previtali1, Maria Grazia Alessio1, Paola Dominoni1, Sabrina Buoro2.   

Abstract

BACKGROUND: Cellular analysis of cerebrospinal fluid (CSF) provides important diagnostic information in various medical conditions. The aim of this study was to evaluate the application of Sysmex UF-5000 body fluid mode in cytometric analysis of CSF compared to Light Microscopic (LM).
METHODS: Eighty-one consecutive CSF samples were analyzed by UF-5000 body fluid mode and by LM. The study also included the evaluation of: limit of Blank (LoB), limit of Detection (LoD), limit of Quantitation (LoQ), carryover and linearity.
RESULTS: For total nucleated cells (TNC-UF) and white blood cells (WBC-UF) LoB, LoD and LoQ were 1×106cells/L, 1.8×106cells/L and 1.9×106cells/L respectively. For red blood cells (RBC) LoB was 2×106cells/L, LoD was 3.5×106cells/L and LoQ was 14×106cells/L respectively. Linearity was excellent, carryover was negligible. The agreement between UF-5000 body fluid mode parameters and manual cell counts was good in all CSF samples with bias ranged between -0.5 and 25.1×106cells/L. The ROC curve analysis showed an area under curve of 0.99 for both TNC-UF and WBC-UF parameters.
CONCLUSIONS: The UF-5000 body fluid mode offers rapid and accurate counts in clinically relevant concentration ranges, replacing the LM for most samples. However, in samples with abnormal cell counts or with abnormal scattergram the need for microscopic review remains.
Copyright © 2017 Elsevier B.V. All rights reserved.

Keywords:  Automated cell counting; Body fluids; Cerebrospinal fluid; Sysmex UF-5000

Mesh:

Year:  2017        PMID: 28843601     DOI: 10.1016/j.cca.2017.08.026

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  3 in total

1.  A performance comparison of the fully automated urine particle analyzer UF-5000 with UF-1000i and Gram staining in predicting bacterial growth patterns in women with uncomplicated urinary tract infections.

Authors:  Stephen Shei-Dei Yang; Chun-Chun Yang; Yi-Sheng Chen; Shang-Jen Chang
Journal:  BMC Urol       Date:  2021-02-12       Impact factor: 2.264

2.  Performance Evaluation of Body Fluid Cellular Analysis Using the Beckman Coulter UniCel DxH 800, Sysmex XN-350, and UF-5000 Automated Cellular Analyzers.

Authors:  Jooyoung Cho; Joowon Oh; Sang Guk Lee; You Hee Lee; Jaewoo Song; Jeong Ho Kim
Journal:  Ann Lab Med       Date:  2020-03       Impact factor: 3.464

3.  UF-5000 flow cytometer: A new technology to support microbiologists' interpretation of suspected urinary tract infections.

Authors:  Roberto Ippoliti; Isabella Allievi; Andrea Rocchetti
Journal:  Microbiologyopen       Date:  2020-01-06       Impact factor: 3.139

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.