Validation of reflex testing rules and establishment of a new workflow for body fluid cell analysis using a Sysmex XN-550 automatic hematology analyzer.

Y-U Cho1, E You1, S Jang1, C-J Park1.   

Abstract

INTRODUCTION: We developed and validated reflex testing rules for the microscopic examination (ME) of body fluids (BFs) on the Sysmex XN-550 (Sysmex Corporation) instrument.
METHODS: We assessed the detection limits, precision, linearity, and carryover. To develop the reflex testing rules (derivation arm), we tested 515 samples and then validated the rules using another 507 samples (validation arm).
RESULTS: All analytical performances were acceptable, and the carryover was negligible. There was agreement between the automated count and ME of red blood cells (r = .98) and total nucleated cells (TNCs) (r = .98), as well as the differential counts of neutrophils (r = .90) and lymphocytes (r = .84). We developed reflex testing rules: TNCs <10/μL, cell 2/cell 1 ratio ≤0.7, HF-BF cells >7.9/100 white blood cells, LY-X ≥85 or LY-Y ≥90, and eosinophils >2.5%. In the validation arm, implementation of the rules resulted in 126 rule-negative samples (24.9%) that were well correlated between the 2 methods. We propose a new workflow for BF cell analysis based on automated counting.
CONCLUSION: The Sysmex XN-550 can be a suitable alternative to ME for BF cell analysis, especially for screening samples and subsequent automatic reporting under the rational use of laboratory-specific rules.
© 2018 John Wiley & Sons Ltd.

Entities:  

Keywords:  Sysmex XN-550; body fluid; reflex testing rules; validation; workflow

Mesh:

Year:  2018        PMID: 29314650     DOI: 10.1111/ijlh.12774

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  3 in total

1.  A comparison of the analysis of 3 types of body fluids using the XN-350 hematology analyzer versus light microscopy assessment.

Authors:  Jiwon Lee; Younggeun Cho; Han-Sung Kim; Hee Jung Kang; Miyoung Kim; Young Kyung Lee
Journal:  Medicine (Baltimore)       Date:  2021-03-19       Impact factor: 1.817

2.  Definition of the Immune Parameters Related to COVID-19 Severity.

Authors:  Sarah Birindelli; Maciej S Tarkowski; Marcello Gallucci; Marco Schiuma; Alice Covizzi; Przemysław Lewkowicz; Elena Aloisio; Felicia Stefania Falvella; Alberto Dolci; Agostino Riva; Massimo Galli; Mauro Panteghini
Journal:  Front Immunol       Date:  2022-03-18       Impact factor: 7.561

3.  Performance Evaluation of Body Fluid Cellular Analysis Using the Beckman Coulter UniCel DxH 800, Sysmex XN-350, and UF-5000 Automated Cellular Analyzers.

Authors:  Jooyoung Cho; Joowon Oh; Sang Guk Lee; You Hee Lee; Jaewoo Song; Jeong Ho Kim
Journal:  Ann Lab Med       Date:  2020-03       Impact factor: 3.464

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.