Mohammad Rawashdeh1, Charbel Saade2, Maha Zaitoun1, Mostafa Abdelrahman1, Patrick Brennan3, Haytham Alewaidat1, Mark F McEntee4.
Abstract
The aim of this study was to determine diagnostic reference levels (DRLs) for cardiac computed tomography (CCT) in Jordan. Volume computed tomography dose index (CTDIvol ) and dose-length product (DLP) were collected from 228 CCTs performed at seven Jordanian hospitals specialized in cardiac CT. DRLs for cardiac CT were defined at the 75th percentile of CTDIvol and DLP. CTDIvol and DLP were collected from 30 successive cardiac CT in each center except for one center (18 scans). The 75th percentile of the CTDIvol and the DLP of the centers calculated from mixed retrospective and prospective gated modes were 47.74 milligray (mGy) and 1035 mGy/cm, respectively. This study demonstrated wide dose variations among the surveyed hospitals for cardiac CT scans; there was a 5.1-fold difference between the highest and lowest median DLP with a range of 223.2-1146.7 mGy/cm. Differences were associated with variations in the mAs and kVp. This study confirmed large variability in CTDIvol and DLP for cardiac CT scans; variation was associated with acquisition protocols and highlights the need for dose optimization. DRLs are proposed for CCT; there remains substantial potential for optimization of cardiac CT examinations for adults in Jordan.Entities:
Keywords: CCT; CTDIvol; DLP; DRL; Jordan
Mesh:
Year: 2019 PMID: 31469229 PMCID: PMC6807208 DOI: 10.1002/acm2.12711
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Information about the scanners in each hospital.
| Hospital | Manufacturer | Year of manufacture | CT model type | Year of installation | Gantry rotation | Capacity (slice No.) |
|---|---|---|---|---|---|---|
| A | Philips | 2008 | Brilliance ICT scan 256 | 2011 | 360 | 256 |
| B | Siemens | 2014 | Somatom force | 2016 | 360 | 384 |
| C | Philips | 2010 | Philips multislices 64p/s | 2010 | 360 | 65 |
| D | Siemens | 2010 | Multislices 128 | 2011 | 360 | 128 |
| E | Siemens | 2006 | Somatom sensation 64 | 2006 | 360 | 64 |
| F | GE | 2011 | Optima 99 multislice | 2012 | 360 | 180 |
| G | Siemens | 2008 | Somatom sensation 64 dual Source | 2008 | 360 | 128 |
Patients and equipment's characteristics and cardiac computed tomography (CT) protocols (n = 228).
| Characteristics | n (%) or mean (IQR) |
|---|---|
| Gender | |
| Male | 139 (61.0%) |
| Female | 89 (39.0%) |
| Age (years) | 50.72 (41–61) |
| Weight (kg) | 70.1 (66.2–80.0) |
| Body mass index (kg/m2) | 30 (28–32) |
| Scan mode | |
| PGM | 60 (26.3%) |
| RGM | 168 (73.7%) |
| CT model (slice) | |
| Philips | 59 (25.9%) |
| Siemens | 138 (60.5%) |
| GE Optima 99 | 31 (13.6%) |
Diagnostic reference levels (DRLs) for mixed, prospective gating mode (PGM), and respective gating modes (RGM) cardiac computed tomography (CT) scans in Jordan.
| Scan type | Mixed modes | PGM (n = 60) | RGM (n = 168) |
|---|---|---|---|
| CTDIvol | |||
| Minimum | 2.00 | 2.00 | 6.60 |
| 25th | 13.71 | 4.50 | 21.00 |
| Median | 31.93 | 7.84 | 40.42 |
| 75th | 47.74 | 33.37 | 64.54 |
| Maximum | 86.64 | 33.80 | 86.64 |
| DLP | |||
| Minimum | 27.60 | 27.60 | 216.00 |
| 25th | 329.58 | 151.80 | 431.97 |
| Median | 727.00 | 626.60 | 888.30 |
| 75th | 1035.00 | 692.95 | 1141.50 |
| Maximum | 2865.00 | 740.00 | 2865.00 |
Diagnostic reference levels (DRLs) for cardiac computed tomography (CT) scans performed with Mixed Modes per center.
| Center | 1 (n = 30) | 2 (n = 45) | 3 (n = 29) | 4 (n = 45) | 5 (n = 18) | 6 (n = 31) | 7 (n = 30) |
|---|---|---|---|---|---|---|---|
| CTDIvol | |||||||
| Minimum | 7.84 | 24.06 | 21.0 | 4.50 | 13.55 | 2.63 | 2.0 |
| 25th | 10.78 | 38.85 | 31.93 | 4.50 | 13.55 | 64.54 | 5.80 |
| Median | 12.69 | 43.89 | 64.54 | 9.90 | 18.37 | 64.54 | 19.59 |
| 75th | 19.15 | 47.87 | 79.43 | 33.80 | 22.30 | 64.54 | 21.70 |
| Maximum | 44.73 | 79.43 | 86.64 | 33.80 | 22.30 | 64.56 | 21.9 |
| DLP | |||||||
| Minimum | 146.90 | 385.00 | 583.9 | 151.80 | 329.58 | 35.50 | 27.60 |
| 25th | 189.10 | 807.40 | 831.20 | 618.10 | 329.58 | 1126.20 | 140.00 |
| Median | 223.20 | 888.30 | 1035.00 | 653.60 | 431.97 | 1146.70 | 251.05 |
| 75th | 583.90 | 934.60 | 1435.00 | 726.50 | 1267.07 | 1236.80 | 293.00 |
| Maximum | 933.40 | 1441.90 | 1549.00 | 740.00 | 1267.07 | 1356.10 | 2865.00 |
Cardiac computed tomography (CT) diagnostic reference levels (DRLs) in Jordan compared with other international cardiac CT DRLs.
| Iran | France | Italy | Netherlands | Japan | Switzerland | KSA | Jordan (current study) | |
|---|---|---|---|---|---|---|---|---|
| CTDIvol | ||||||||
| Mixed | 66.5 | – | 61 | – | – | 50 | – | 47.74 |
| PGM | – | 26 | – | – | – | – | 29 | 33.37 |
| RGM | – | 44 | – | – | – | – | 43 | 64.54 |
| DLP | ||||||||
| Mixed | 1073 | – | 1208 | 671 | 1510 | 1000 | – | 1035.0 |
| PGM | – | 370 | – | – | – | – | 343 | 692.95 |
| RGM | – | 870 | – | – | – | – | 808 | 1141.50 |
Predictive of CTDIvol, and dose–length product (DLP) from exposure factors using stepwise regression factors for Mixed Modes, prospective gated mode (PGM), and retrospective ECG‐gating modes (RGM).
| R2 |
| |
|---|---|---|
| Mixed modes | ||
| CTDIvol | ||
| mAs | 0.414 | ≤0.001 |
| kVp | 0.496 | ≤0.001 |
| Number of slices | 0.530 | ≤0.001 |
| DLP | ||
| mAs | 0.243 | ≤0.001 |
| kVp | 0.335 | ≤0.001 |
| Number of slices | 0.364 | ≤0.001 |
| PGM | ||
| CTDIvol | ||
| mAs | 0.646 | ≤0.001 |
| DLP | ||
| kVp | 0.585 | ≤0.001 |
| Number of slices | 0.711 | ≤0.001 |
| mAs | 0.820 | ≤0.001 |
| RGM | ||
| CTDIvol | ||
| kVp | 0.342 | ≤0.001 |
| Pitch | 0.506 | ≤0.001 |
| mAs | 0.600 | ≤0.001 |
| Number of slices | 0.625 | ≤0.001 |
| Slices thickness | 0.635 | ≤0.001 |
| DLP | ||
| mAs | 0.167 | ≤0.001 |
| kVp | 0.232 | ≤0.001 |
| Number of slices | 0.268 | ≤0.001 |