Establishment of diagnostic reference levels in cardiac computed tomography.

Mohammad Rawashdeh1, Charbel Saade2, Maha Zaitoun1, Mostafa Abdelrahman1, Patrick Brennan3, Haytham Alewaidat1, Mark F McEntee4.   

Abstract

The aim of this study was to determine diagnostic reference levels (DRLs) for cardiac computed tomography (CCT) in Jordan. Volume computed tomography dose index (CTDIvol ) and dose-length product (DLP) were collected from 228 CCTs performed at seven Jordanian hospitals specialized in cardiac CT. DRLs for cardiac CT were defined at the 75th percentile of CTDIvol and DLP. CTDIvol and DLP were collected from 30 successive cardiac CT in each center except for one center (18 scans). The 75th percentile of the CTDIvol and the DLP of the centers calculated from mixed retrospective and prospective gated modes were 47.74 milligray (mGy) and 1035 mGy/cm, respectively. This study demonstrated wide dose variations among the surveyed hospitals for cardiac CT scans; there was a 5.1-fold difference between the highest and lowest median DLP with a range of 223.2-1146.7 mGy/cm. Differences were associated with variations in the mAs and kVp. This study confirmed large variability in CTDIvol and DLP for cardiac CT scans; variation was associated with acquisition protocols and highlights the need for dose optimization. DRLs are proposed for CCT; there remains substantial potential for optimization of cardiac CT examinations for adults in Jordan.
© 2019 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  CCT; CTDIvol; DLP; DRL; Jordan

Mesh:

Year:  2019        PMID: 31469229      PMCID: PMC6807208          DOI: 10.1002/acm2.12711

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


INTRODUCTION

Radiation doses differ between hospitals for standard‐sized patients undergoing the same examination.1 These dose variations are “primarily attributable to local choices regarding technical parameters, rather than patient, institution, or machine characteristics.” Computed tomography (CT) has a large range of scanning options that lead to a large variation in patient radiation doses. Several factors affect radiation dose to patients undergoing cardiac computed tomography (CCT) procedures.2 Concerns have been raised about the radiation exposure of CCT. Recently, prospective gated mode (PGM) was developed to reduce radiation dose. PGM, also known as “step‐and‐shoot” or “sequential scan mode,”3 has been developed as an alternative mode to standard helical (spiral) scanning with retrospective electrocardiographic gating with the aim of decreasing radiation dose during CCT to patients with stable heart rates. Previous work by other authors has identified key contributors to variations in dose as radiographic settings, equipment factors, the level of quality assurance in place, radiographer training, radiographer experience, as well as patient body habitus.4, 5 Dose optimization requires identification of which factors are the greatest contributor to variations in dose. Once the contribution of factors is established, operators can consider corrective action in a cost‐effective manner.4 In order to achieve optimization, guidance on appropriate levels of patient exposure is required. Basic safety standards have been provided by the International Commission on Radiological Protection (ICRP), World Health Organization (WHO), and International Atomic Energy Agency (IAEA) in order to optimize the protection of patients during all radiological procedures, including CT.6 These include a recommendation for DRLs7 or guidance levels to be put in place, monitored, and used to improved radiological procedures.8 DRLs enhance patient protection and allow comparisons of the radiation dose of different CT systems, and procedures. DRLs offer a framework where dose levels from different hospitals can be compared and when DRLs are exceeded by a department, corrective strategies can be taken. DRLs are not dose limits, but rather guidelines. Where they are regularly exceeded, corrective action should be taken. DRLs have been effective in reducing radiation dose, with radiation levels being reduced by 50% in the UK since their adoption.9, 10 As dose variation for the same examination can reach up to 23‐fold between centers for non‐CT X‐ray examinations2 and 13‐fold for CT examinations.11 Strategies such as DRLs are required to reduce dose variation between centers. The method of establishing DRLs starts with a determination of radiation dose levels delivered for specific examinations in several hospitals in an individual country (or state). These data are then used to compute examination‐specific DRLs for that country, state or region, usually in terms of the 75th percentile of the dose distribution. Due to the differing ethnicities, procedures, and equipment across different countries, it is advisable that each country determine their own DRLs. While the requirement for DRLs have been described in legislative documents in Jordan and internationally, and their implementation for general examinations is seen in Europe and the US, national DRL values do not exist for cardiac CT examinations in Jordan. Cardiac CT examinations are among the highest dose examinations, with patients undergoing chest CT having doses ranging from 4 to 18 mSv but those undergoing cardiac CT receiving 5–32 for a coronary CT angiogram, and 12–18 mSv for a 64‐slice coronary CT Angiography (CTA) with tube modulation. However, with good technique, a prospectively triggered coronary CTA patient can receive 2–4 mSv.12 The dose for cardiac CT in Jordan is currently not known, so we cannot compare our performance internally or internationally. The current work aims to address this gap.

MATERIALS AND METHODS

In 2018, a retrospective survey was performed across seven Jordanian hospitals that routinely perform CCT. Data from 228 cardiac CT scans were collected over a 3‐month period from May to August. These cardiac CT scans were performed on a wide range of scanners with different number of slices from three manufacturers (GE, Siemens, and Philips). These are representative of CT scanners being used across Jordan. According to the Jordanian regulations, CT scanners undergo a quality control program that includes daily, monthly, quarterly, and annual checking of the assessment of the radiation dose.13 Dose in CTDIvol and DLP was provided in 30 successive cardiac CT scans in each center except for one, which provided 18 scans. There survey respondents used two scanning modes, prospective electrocardiographic (ECG) gating modes (PGM) where the scanner monitors the patient's ECG and retrospective ECG‐gating modes (RGM) where the patient is scanned continuously, while only certain portions of the scan are used to reconstruct the image. Participants were all adults between 17 to 75 yr and all genders were included. In line with a methodology previously published,9, 14 protocol and scan sequence details were grouped in the survey (Table 1). A clinical coordinator with CT experience was appointed in each center to administer and receive the questionnaires. The survey focused on CTA performed for the assessment of coronary artery disease. CTA included one single scan per examination. Calcium scoring, evaluation bypass graft patency and preparation of trans catheter aortic valve implantation were excluded.
Table 1

Information about the scanners in each hospital.

HospitalManufacturerYear of manufactureCT model typeYear of installationGantry rotationCapacity (slice No.)
APhilips2008Brilliance ICT scan 2562011360256
BSiemens2014Somatom force2016360384
CPhilips2010Philips multislices 64p/s201036065
DSiemens2010Multislices 1282011360128
ESiemens2006Somatom sensation 64200636064
FGE2011Optima 99 multislice2012360180
GSiemens2008Somatom sensation 64 dual Source2008360128
Information about the scanners in each hospital. To standardize weight for the sampled population and in line with international recommendations, the survey included only those patients who weighed between 60 and 80 kg.9

Radiation dose data

On modern CT scanners CTDIvol and DLP are provided for every sequence and examination. However, on older equipment it was necessary to calculate the dose using exposure and procedural data. The patient dose data, CTDIvol and DLP values, were extracted from the picture archiving and communication system (PACS). A summary of the two relevant dose parameters is given below along with methods for calculating these factors. CTDIvol is defined as weighted CTDI (CTDIw) divided by CT pitch and provide an estimation for average phantom dose for a complete spiral CT scan.15

Weighted CT Dose Index — CTDIw

This describes the radiation dose delivered per unit cranio–caudal (z) axis thickness. Significant previous work has provided methodologies so that the baseline dose value CTDIw can be calculated for specific examinations and from this baseline value other important dose metrics can be calculated4, 7, 16, 17, 18, 19, 20, 21, 22 and these are explained below. Using typical CTDIw values calculated from a dosimetry phantom for each CT scanner model, with specific exposure factors, the IMPACT group14 has provided a calculator from which CTDIw can be calculated for any sequence, for any model using a range of exposure factors. This calculation is shown in Formula 1.where ( CTDIw)scanner, phantom, kV, NxT is a coefficient based on the normalized dose [CTDIw for a specific scanner at a particular kV, slice number (N), and thickness (T)]; (FNxT)scanner are dose coefficients for all other slice number and thicknesses; C is the mAs or effective mAs if pitch correction was included

Volumetric CT dose index ‐ CTDIvol

Since CTDIw does not consider whether axial slices are contiguous, noncontiguous, or overlapping, a “pitch” correction must be added which provides a more representative volume CTDI or CTDIvol (Formula 2).where pitch factor is the distance the table moves in the z axis divided by the slice number and each slice thickness. CTDIw is calculated from Formula 1.

Dose Length Product (DLP)

The dose measurements above do not consider the total length of the patient who has been irradiated during each examination sequence. This is calculated using Formula 3.where CTDIvol is calculated from Formula 2.

Data analysis

The minimum, maximum, and the first, second, and third quartiles were calculated for CTDIvol, and DLP using Statistical Package for the Social Sciences (SPSS) software (SPSS for Windows, version 22.0, SPSS Inc.). Stepwise regressions were performed to find which exposure factors were associated with high dose and the level of contribution of each factor. The following factors were included: kVp, mAs, pitch, slice thickness, and number of slices. In addition, international DRLs were compared to the current national DRL level established in the current work.

RESULTS

Seven hospitals (four public, one university, and two private) participated in the study. Recordings from a total of 228 CCT examinations were collected, of which 60 CCT were performed using PGM and 168 using RGM. Sixty‐one percent of patients were males and 39% were females and the mean weight for both of them was 70.1 kg (Min = 60; Max = 80, SD = 8.93). Data were collected from seven scanners, of which two and five centers acquired image data in 256 and 128 slices. This represents approximately 31% of cardiac CT units in Jordan and exceeds the sample sizes used previously to establish reference levels in other countries.5, 7 A sample size calculation indicated a difference <5% in dose would be detected at 0.8 power. Patient and equipment characteristics and cardiac CT protocols are shown in Table 2. A summary of DRLs per scan mode (PGM, RGM, and Mixed Modes) and per center is shown Table 3.
Table 2

Patients and equipment's characteristics and cardiac computed tomography (CT) protocols (n = 228).

Characteristicsn (%) or mean (IQR)
Gender
Male139 (61.0%)
Female89 (39.0%)
Age (years)50.72 (41–61)
Weight (kg)70.1 (66.2–80.0)
Body mass index (kg/m2)30 (28–32)
Scan mode
PGM60 (26.3%)
RGM168 (73.7%)
CT model (slice)
Philips59 (25.9%)
Siemens138 (60.5%)
GE Optima 9931 (13.6%)
Table 3

Diagnostic reference levels (DRLs) for mixed, prospective gating mode (PGM), and respective gating modes (RGM) cardiac computed tomography (CT) scans in Jordan.

Scan typeMixed modesPGM (n = 60)RGM (n = 168)
CTDIvol
Minimum2.002.006.60
25th13.714.5021.00
Median31.937.8440.42
75th47.7433.3764.54
Maximum86.6433.8086.64
DLP
Minimum27.6027.60216.00
25th329.58151.80431.97
Median727.00626.60888.30
75th1035.00692.951141.50
Maximum2865.00740.002865.00
Patients and equipment's characteristics and cardiac computed tomography (CT) protocols (n = 228). Diagnostic reference levels (DRLs) for mixed, prospective gating mode (PGM), and respective gating modes (RGM) cardiac computed tomography (CT) scans in Jordan. The mean CTDIvol and DLP per CCT examination were calculated for each site and used to compare doses across CCT centers (Table 4). Wide variations were found between hospitals surveyed, with 1‐ to 5.1‐fold differences in mean CTDIvol and DLP reported for the examinations surveyed.
Table 4

Diagnostic reference levels (DRLs) for cardiac computed tomography (CT) scans performed with Mixed Modes per center.

Center1 (n = 30)2 (n = 45)3 (n = 29)4 (n = 45)5 (n = 18)6 (n = 31)7 (n = 30)
CTDIvol
Minimum7.8424.0621.04.5013.552.632.0
25th10.7838.8531.934.5013.5564.545.80
Median12.6943.8964.549.9018.3764.5419.59
75th19.1547.8779.4333.8022.3064.5421.70
Maximum44.7379.4386.6433.8022.3064.5621.9
DLP
Minimum146.90385.00583.9151.80329.5835.5027.60
25th189.10807.40831.20618.10329.581126.20140.00
Median223.20888.301035.00653.60431.971146.70251.05
75th583.90934.601435.00726.501267.071236.80293.00
Maximum933.401441.901549.00740.001267.071356.102865.00
Diagnostic reference levels (DRLs) for cardiac computed tomography (CT) scans performed with Mixed Modes per center. Comparisons of the findings with DRLs published from other countries are shown in Table 5.18‐24
Table 5

Cardiac computed tomography (CT) diagnostic reference levels (DRLs) in Jordan compared with other international cardiac CT DRLs.

Iran18 France19 Italy23 Netherlands20 Japan21 Switzerland24 KSA22 Jordan (current study)
CTDIvol
Mixed66.5615047.74
PGM262933.37
RGM444364.54
DLP
Mixed10731208671151010001035.0
PGM370343692.95
RGM8708081141.50
Cardiac computed tomography (CT) diagnostic reference levels (DRLs) in Jordan compared with other international cardiac CT DRLs. Multiple regression analysis for the Mixed Modes suggested that mAs, kVp, and number of slices were more accurately predictive of CTDIvol than any individual variable alone, with R2 of 0.530 (P ≤ 0.001). The results of the regression also showed that combination of mAs, kVp, and number of slices could significantly predicate DLP. All factors had a significant positive predictor value, with a comparatively low R2 of 0.364 (P ≤ 0.001). The equations werewhere A is mAs, B is kVp, and C is the number of slices.where A is mAs, B is kVp, and C is the number of slices. In the PGM mode, the results from the multiple linear regressions demonstrated that mAs was more accurately predictive of CTDIvol than any other variable, with R2 of 0.646 (P ≤ 0.001), while for the case of the DLP, kVp, number of slices, and mAs could significantly predicate DLP, with R2 of 0.820 (P ≤ 0.001), which is significantly higher (Table 6).where A is mAswhere A is kVp, B is the number of slices, and C is mAs
Table 6

Predictive of CTDIvol, and dose–length product (DLP) from exposure factors using stepwise regression factors for Mixed Modes, prospective gated mode (PGM), and retrospective ECG‐gating modes (RGM).

R2 P‐value
Mixed modes
CTDIvol
mAs0.414≤0.001
kVp0.496≤0.001
Number of slices0.530≤0.001
DLP
mAs0.243≤0.001
kVp0.335≤0.001
Number of slices0.364≤0.001
PGM
CTDIvol
mAs0.646≤0.001
DLP
kVp0.585≤0.001
Number of slices0.711≤0.001
mAs0.820≤0.001
RGM
CTDIvol
kVp0.342≤0.001
Pitch0.506≤0.001
mAs0.600≤0.001
Number of slices0.625≤0.001
Slices thickness0.635≤0.001
DLP
mAs0.167≤0.001
kVp0.232≤0.001
Number of slices0.268≤0.001
Predictive of CTDIvol, and dose–length product (DLP) from exposure factors using stepwise regression factors for Mixed Modes, prospective gated mode (PGM), and retrospective ECG‐gating modes (RGM). For the RGM mode, the results from the multiple linear regressions demonstrated that kVp, pitch, mAs, number of slices, and slice thickness were the predictive factors of CTDIvol, with R2 of 0.635 (P ≤ 0.001), while mAs, kVp, and number of slices were the predictive factors of DLP with R2 of 0.268 (P ≤ 0.001), which is relatively lower as demonstrated in the following two formulas.where A is kVp, B is pitch, C is mAs, D is the number of slices, and E is the slice thickness.where A is mAs, B is kVp, and C is the number of slices.

DISCUSSION

DRLs were first proposed by the International Commission on Radiological Protection in 1991. They are defined as “dose levels in radio diagnostic practices for typical examinations for patient groups or standard phantoms for broadly defined groups of equipment.”9, 25, 26 Patient radiation doses that exceed established DRLs should be investigated to identify potential reasons for higher dose and to allow better management of the radiation dose of similar procedures in the future.21, 27 With the significant amount of studies that have been conducted to establish the DRL levels in other countries and in the different CT scans,4, 7, 16, 17, 18, 19, 20, 21, 22 this is the first study to establish DRLs in CCT in Jordan. Research to date has demonstrated a maximum potential dose reduction in CT scan between 60% and 80%.28, 29, 30, 31 However, it is important to appropriately balance the need to achieve low radiation dose with the likelihood of creating useful diagnostic CT images. Low radiation exposure for a certain patient during CCT scan may result high image noise; however, it needs to be acknowledged that while high radiation exposure may increase image quality and reduce image noise, this does not automatically mean additional diagnostic information.32 The current work highlighting wide dose variations for cardiac CT scans were shown across hospitals. With a range of 223.2–1146.7 mGy cm, the highest mean DLP was 5.1 times higher than the median value. These differences were primarily attributable to local choices regarding technical parameters, rather than patient, institution, or machine characteristics. Changes in CTDIvol were associated with variations in the mAs, kVp pitch, and slice thickness. Multiple regression analysis predicted that low DLP was most dependent on mAs, kVp, mode of scan, and number of slices. The current study reports that PGM allowed a significant dose reduction with CTDIvol of 60.4% compared with the RGM mode. These wide dose variations between modes emphasize the need to analyze the CTDIvol and DLP individually and, therefore, establish DRL for each mode.22 The current work did not compare the diagnostic performance of PGM with RGM. However, the reduction in radiation dose with PGM scanning was larger than the effect of other radiation dose‐reduction techniques. The 75th DLP in scanned average‐sized patients was only 692.95 mGy cm. Among those centers using PGM, the lowest median DLP was 223.2 mGy cm in Centre A, this dose was in contrast with 75th DLP of RGM 1146.60 mGy cm. This finding supports the use of PGM as an effective tool in comprehensive radiation dose‐reduction technique. The current work shows that a reduction of 26.3% in the DLP in CT scanning within participating centers is achievable with PGM, this is a larger decrease in dose than reported in other works.2, 19, 22 Compared with nationwide surveys from other countries, Jordanian CCT centers generally appear to employ higher doses than those countries previously surveyed; therefore, there is a large potential for optimization of CCT examinations for adults in Jordan. Variation in radiation dose shown in the current work highlights the need for the adoption of DRLs that radiologists or radiographers need to optimize their CCT protocols and that interest in dose optimization must be improved. The work also demonstrates the need for periodic reassessment of DRLs at short‐time intervals. Clinical audits should also identify further causal agents, eliminate unjustified examinations, and optimize procedures. Further work should investigate size‐specific dose estimate (SSDE). SSDE accounts for patient parameters establishment of DRLs and removes the requirement to limit the average of the sample to between 60 and 80 kg. Additional studies should to be conducted on the SSDE application to DLP so that scan length can be considered for in the equation of patient dose. There are several noteworthy limitations to our study. First, as our analysis was retrospective, we could not obtain information on several parameters that possibly influence the radiation dose such as beam collimation, rotation time, and patients' diameter. Second, we did not assess the CT scan indications; hence, the parameters of these CT scans may not represent the routine protocols of the respective institutions. Third, since our analysis was conducted on only seven institutions, bias could have been introduced, even though the CT scans we obtained were from various centers geographically dispersed throughout Jordan. This study demonstrates large variability in CTDIvol and DLP during CCT examinations in Jordan and highlights the need for doses to be reduced. PGM is clearly an effective dose‐reduction technique for cardiac CT examinations and the use of this mode should be encouraged. Local DRL results should be communicated back to each CT center to encourage optimization of scan parameters and develop more proactive comparisons with national DRL and other CT centers.

CONFLICT OF INTEREST

There is no conflict of interest declared in this article.
  24 in total

1.  Diagnostic reference level of computed tomography (CT) in Japan.

Authors:  Yasuhiro Fukushima; Yoshito Tsushima; Hiroyuki Takei; Ayako Taketomi-Takahashi; Hidenori Otake; Keigo Endo
Journal:  Radiat Prot Dosimetry       Date:  2011-12-05       Impact factor: 0.972

2.  Establishment of diagnostic reference levels in cardiac CT in France: a need for patient dose optimisation.

Authors:  F Mafalanka; C Etard; J L Rehel; D Pesenti-Rossi; F Amrar-Vennier; N Baron; L Christiaens; R Convers-Domart; D Defez; P Douek; V Gaxotte; J L Georges; S Leygnac; P Ou; J L Sablayrolles; C Salvat; E Schouman-Claeys; M Sirol; B Aubert
Journal:  Radiat Prot Dosimetry       Date:  2014-10-23       Impact factor: 0.972

3.  ESTABLISHING DIAGNOSTIC REFERENCE LEVELS FOR CARDIAC COMPUTED TOMOGRAPHY ANGIOGRAPHY IN SAUDI ARABIA.

Authors:  Ali B Alhailiy; Peter L Kench; Mark F McEntee; Patrick C Brennan; Elaine A Ryan
Journal:  Radiat Prot Dosimetry       Date:  2018-10-01       Impact factor: 0.972

4.  Reference dose levels for patients undergoing common diagnostic X-ray examinations in Irish hospitals.

Authors:  D A Johnston; P C Brennan
Journal:  Br J Radiol       Date:  2000-04       Impact factor: 3.039

5.  ESTIMATION OF CARDIAC CT ANGIOGRAPHY RADIATION DOSE TOWARD THE ESTABLISHMENT OF NATIONAL DIAGNOSTIC REFERENCE LEVEL FOR CCTA IN IRAN.

Authors:  Seyed Mohammad Bagher Hosseini Nasab; Ali Shabestani-Monfared; Mohammad Reza Deevband; Reza Paydar; Mehrdad Nabahati
Journal:  Radiat Prot Dosimetry       Date:  2017-05-01       Impact factor: 0.972

Review 6.  Radiation dose and safety in cardiac computed tomography.

Authors:  Thomas C Gerber; Birgit Kantor; Cynthia H McCollough
Journal:  Cardiol Clin       Date:  2009-11       Impact factor: 2.213

7.  Dose reduction in computed tomography by individualized scan protocols.

Authors:  L Jangland; E Sanner; J Persliden
Journal:  Acta Radiol       Date:  2004-05       Impact factor: 1.990

8.  Initial evaluation of coronary images from 320-detector row computed tomography.

Authors:  Frank J Rybicki; Hansel J Otero; Michael L Steigner; Gabriel Vorobiof; Leelakrishna Nallamshetty; Dimitrios Mitsouras; Hale Ersoy; Richard T Mather; Philip F Judy; Tianxi Cai; Karl Coyner; Kurt Schultz; Amanda G Whitmore; Marcelo F Di Carli
Journal:  Int J Cardiovasc Imaging       Date:  2008-03-27       Impact factor: 2.357

9.  Database of normalised computed tomography dose index for retrospective CT dosimetry.

Authors:  Eunah Lee; Stephanie Lamart; Mark P Little; Choonsik Lee
Journal:  J Radiol Prot       Date:  2014-04-14       Impact factor: 1.394

10.  Estimated radiation dose associated with cardiac CT angiography.

Authors:  Jörg Hausleiter; Tanja Meyer; Franziska Hermann; Martin Hadamitzky; Markus Krebs; Thomas C Gerber; Cynthia McCollough; Stefan Martinoff; Adnan Kastrati; Albert Schömig; Stephan Achenbach
Journal:  JAMA       Date:  2009-02-04       Impact factor: 56.272