Literature DB >> 12616189

Comparative study of cochlear damage with three perimodiolar electrode designs.

Adrien A Eshraghi1, Nathaniel W Yang, Thomas J Balkany.   

Abstract

OBJECTIVE: To describe intracochlear insertion trauma caused by three perimodiolar cochlear implant electrodes. STUDY
DESIGN: Descriptive histological study of 15 human cadaver temporal bones.
METHODS: Fifteen cadaver temporal bones underwent surface preparation and were implanted with one of the following perimodiolar electrode arrays: Combi 40+PM (MedEl Corporation), HiFocus II (Advanced Bionics Corporation), or Contour (Cochlear Corporation). A cryosectioning technique was used to study horizontal sections at 200 microm intervals with the electrode in place. Image-enhanced videofluoroscopy and computer-assisted morphometrics were used to assess the mechanism of insertion trauma and to determine electrode position within the modiolus.
RESULTS: Histological examination revealed varying degrees of damage to the spiral ligament, basilar membrane, and osseous spiral lamina. Using a novel grading system for electrode trauma, there was no statistically significant difference among the three electrodes. A literature search of histological studies of a commonly used "standard" electrode showed damage equal to or greater than that seen in the current study.
CONCLUSIONS: Insertion trauma caused by periomodiolar electrodes occurs to an acceptable degree. Refinement of electrodes based on mechanisms of trauma may be able to further reduce damage.

Entities:  

Mesh:

Year:  2003        PMID: 12616189     DOI: 10.1097/00005537-200303000-00005

Source DB:  PubMed          Journal:  Laryngoscope        ISSN: 0023-852X            Impact factor:   3.325


  49 in total

1.  Intra-Operative Neural Response Telemetry and Acoustic Reflex Assessment using an Advance-In-Stylet Technique and Modiolus-Hugging: A prospective cohort study.

Authors:  Abdulrahman Hagr
Journal:  Sultan Qaboos Univ Med J       Date:  2011-08-15

2.  Rotation of the osseous spiral lamina from the hook region along the basal turn of the cochlea: results of a magnetic resonance image anatomical study using high-resolution DRIVE sequences.

Authors:  Daren Gibson; Michael B Gluth; Andy Whyte; Marcus D Atlas
Journal:  Surg Radiol Anat       Date:  2011-11-26       Impact factor: 1.246

3.  In vitro and in vivo pharmacokinetic study of a dexamethasone-releasing silicone for cochlear implants.

Authors:  Ya Liu; Claude Jolly; Susanne Braun; Thomas Stark; Elias Scherer; Stefan K Plontke; Jan Kiefer
Journal:  Eur Arch Otorhinolaryngol       Date:  2015-08-29       Impact factor: 2.503

4.  Anatomy of the round window and hook region of the cochlea with implications for cochlear implantation and other endocochlear surgical procedures.

Authors:  Peter M M C Li; Haobing Wang; Clarinda Northrop; Saumil N Merchant; Joseph B Nadol
Journal:  Otol Neurotol       Date:  2007-08       Impact factor: 2.311

Review 5.  Soft cochlear implantation: rationale for the surgical approach.

Authors:  David R Friedland; Christina Runge-Samuelson
Journal:  Trends Amplif       Date:  2009-06

6.  Evaluation of a new slim lateral wall electrode for cochlear implantation: an imaging study in human temporal bones.

Authors:  Aarno Dietz; Matti Iso-Mustajärvi; Sini Sipari; Jyrki Tervaniemi; Dzemal Gazibegovic
Journal:  Eur Arch Otorhinolaryngol       Date:  2018-05-24       Impact factor: 2.503

7.  Automated insertion of preformed cochlear implant electrodes: evaluation of curling behaviour and insertion forces on an artificial cochlear model.

Authors:  Thomas S Rau; Andreas Hussong; Martin Leinung; Thomas Lenarz; Omid Majdani
Journal:  Int J Comput Assist Radiol Surg       Date:  2009-04-10       Impact factor: 2.924

8.  Considerations for design of future cochlear implant electrode arrays: electrode array stiffness, size, and depth of insertion.

Authors:  Stephen J Rebscher; Alexander Hetherington; Ben Bonham; Peter Wardrop; David Whinney; Patricia A Leake
Journal:  J Rehabil Res Dev       Date:  2008

Review 9.  Cochlear Implants and Children with Vestibular Impairments.

Authors:  Sharon L Cushing; Blake C Papsin
Journal:  Semin Hear       Date:  2018-07-20

10.  Temporal Bone Histopathology of First-Generation Cochlear Implant Electrode Translocation.

Authors:  Akira Ishiyama; Gail Ishiyama; Ivan A Lopez; Fred H Linthicum
Journal:  Otol Neurotol       Date:  2019-07       Impact factor: 2.311

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.