Elimination of facial nerve stimulation by reimplantation in cochlear implant subjects.

Rolf Battmer1, Joerg Pesch, Timo Stöver, Anke Lesinski-Schiedat, Minoo Lenarz, Thomas Lenarz.   

Abstract

HYPOTHESIS: Perimodiolar intracochlear electrodes with contacts facing towards the modiolus have limited current flow towards the outer wall of the cochlea and therefore, may reduce the occurrence of facial nerve stimulation (FN) in cochlear implant subjects.
BACKGROUND: Facial nerve stimulation is a well-known complication in cochlear implant treatment especially in the group of subjects with otosclerosis. The possible explanation of this side effect is a change of the electrical properties of the otosclerotic bone leading to leakage current and resulting in facial nerve stimulation.
METHODS: Four CI subjects who had been implanted with a Nucleus Mini22 device with a Nucleus Straight electrode between 9 to 12 years ago suffered from severe FN stimulation. Electrode contacts had to be switched off so that they could only use 4, 11, 13, and 15 electrodes of their usual set of 22. The switch off resulted in deteriorating speech understanding over time. Therefore, all subjects were reimplanted with a Nucleus 24R device with a Contour electrode. Preoperatively, the threshold of FN stimulation was obtained on all electrodes subjectively. Intraoperatively, FN stimulation thresholds were measured objectively with both, the old and the new device and were compared. NRT and SRT thresholds were also obtained with the reimplanted device to assure effective electrical stimulation of the auditory nerve.
RESULTS: In all four cases the postoperative fitting demonstrated no FN stimulation on all electrodes up to maximum comfortable level. The insertion of the Contour electrode array was complete in three cases, in one case the array could only be inserted partially similarly to the situation before the reimplantation. Speech perception tests showed a significant improvement in all subjects with the new device.
CONCLUSION: Electrodes with modiolar facing contacts and perimodiolar position like the Nucleus Contour electrode reduce the possibility of facial nerve stimulation significantly due to more focused electrical stimulation.

Entities:  

Mesh:

Year:  2006        PMID: 17006341     DOI: 10.1097/01.mao.0000235374.85739.c6

Source DB:  PubMed          Journal:  Otol Neurotol        ISSN: 1531-7129            Impact factor:   2.311


  10 in total

1.  Intra-Operative Neural Response Telemetry and Acoustic Reflex Assessment using an Advance-In-Stylet Technique and Modiolus-Hugging: A prospective cohort study.

Authors:  Abdulrahman Hagr
Journal:  Sultan Qaboos Univ Med J       Date:  2011-08-15

2.  Can Preoperative CT Scans Be Used to Predict Facial Nerve Stimulation Following CI?

Authors:  Jonathan L Hatch; Habib G Rizk; Michael W Moore; Elizabeth E Camposeo; Shaun A Nguyen; Paul R Lambert; Ted A Meyer; Theodore R McRackan
Journal:  Otol Neurotol       Date:  2017-09       Impact factor: 2.311

3.  The pathologic basis of facial nerve stimulation in otosclerosis and multi-channel cochlear implantation.

Authors:  Mohammad Seyyedi; Barbara S Herrmann; Donald K Eddington; Joseph B Nadol
Journal:  Otol Neurotol       Date:  2013-12       Impact factor: 2.311

4.  Cochlear implantation in cases with incomplete partition type III (X-linked anomaly).

Authors:  Armagan Incesulu; Baki Adapinar; Cem Kecik
Journal:  Eur Arch Otorhinolaryngol       Date:  2008-02-28       Impact factor: 2.503

5.  Facial nerve stimulation after cochlear implantation: our experience.

Authors:  S Berrettini; De A Vito; L Bruschini; S Passetti; F Forli
Journal:  Acta Otorhinolaryngol Ital       Date:  2011-02       Impact factor: 2.124

6.  Hearing Preservation After Cochlear Reimplantation Using Electrocochleography: A Case Report.

Authors:  Amit Walia; Matthew A Shew; Amanda J Ortmann; Craig A Buchman; Jacques A Herzog
Journal:  Laryngoscope       Date:  2021-07-03       Impact factor: 2.970

7.  Speech Recognition Outcomes After Cochlear Reimplantation Surgery.

Authors:  Mariana Reis; Isabelle Boisvert; Valerie Looi; Melville da Cruz
Journal:  Trends Hear       Date:  2017 Jan-Dec       Impact factor: 3.293

Review 8.  Review on cochlear implant electrode array tip fold-over and scalar deviation.

Authors:  Anandhan Dhanasingh; Claude Jolly
Journal:  J Otol       Date:  2019-01-09

9.  A Ten-Year Review of Audiological Performance in Children with Inner Ear Abnormalities after Cochlear Implantation in Singapore.

Authors:  Sok Yan Tay; Rosslyn Anicete; Kun Kiaang Henry Tan
Journal:  Int J Otolaryngol       Date:  2019-12-01

10.  Postoperative radiological assessment of the mastoid facial canal in cochlear implant patients in correlation with facial nerve stimulation.

Authors:  Iris Burck; Rania A Helal; Nagy N N Naguib; Nour-Eldin A Nour-Eldin; Jan-Erik Scholtz; Simon Martin; Martin Leinung; Silke Helbig; Timo Stöver; Annette Lehn; Thomas J Vogl
Journal:  Eur Radiol       Date:  2021-07-05       Impact factor: 5.315

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.