| Literature DB >> 31436183 |
Sara Ortiz-Toquero1, Guadalupe Rodriguez1, Victoria de Juan2, Raul Martin3.
Abstract
Purpose: Compare the agreement between the finally fitted back optic zone radius (BOZR) of a spherical gas permeable (GP) contact lense (CL) with those proposed by different guidelines currently available to fit GP CLs in keratoconus.Entities:
Keywords: Contact lens; fitting; gas permeable; guidelines; keratoconus
Mesh:
Year: 2019 PMID: 31436183 PMCID: PMC6727727 DOI: 10.4103/ijo.IJO_1538_18
Source DB: PubMed Journal: Indian J Ophthalmol ISSN: 0301-4738 Impact factor: 1.848
Description of guidelines used in the study
| Guideline | Description |
|---|---|
| Guideline #1 | Suggested by the APEX software CL fitting (APEX, version 1.1.0.6, developed by Hecht Contactlinsen in association with Oculus, which displays a simulated fluorescein pattern of the specified GP design to aid the fitting procedure).[13] Recommend 3-touch fitting philosophy. |
| Guideline #2 | BOZR = (BOZRAPEX*0.88)+0.77 Improvement of BOZR proposed by APEX software CL fitting.[13] Recommend 3-touch fitting philosophy. |
| Guideline #3 | BOZR=Horizontal K (mm) - 0.10 Recommended by Conoptica-Hecht Contactlinsen, (Germany) to fit KAKC lens.[19] Recommend 3-touch fitting philosophy |
| Guideline #4 | If corneal astigmatism < - 3.75 D: BOZR=Kf (D) - 0.61 x Astigmatism If corneal astigmatism-4.00 to-7.50 D: BOZR=Kf (D) - 0.50 x Astigmatism If corneal astigmatism > - 7.50 D: BOZR=Kf (D) - 0.35 x Astigmatism *Calculated to diameter of 9.40 mm Proposed by Centre of Contact Lens Research (University of Waterloo, Canada).[15] Recommend 3-touch fitting philosophy |
| Guideline #5 | If Kf<7.00 mm: BOZR=0.211*Kf (mm) + 5.904 If Kf 7 to 8 mm: BOZR=0.465*Kf (mm) + 4.16 Recommended by Rajabi MT |
| Guideline #6 | BOZR=Km (mm) - 0.20 Recommended to fit RoseK2 GP lens (Menicon, Co., Ltd., North Billeric, MA, USA)[20] Recommend 3-touch and apical clearance fitting philosophy. |
| Guideline #7 | BOZR=Kf (mm) - [1/3 astigmatism (mm)] Proposed by Bausch & Lomb[21] to fit their keratoconus lens design or OP8 GP lens (Soflex, Israel).[22] Recommend 3-touch fitting philosophy (with slight central touch). |
| Guideline #8 | BOZR=Kmean (mm) Recommended to fit ACL KERA lens (Australian Contact Lenses, Australia), FlexCone (SwissLens, Switzerland), Keracon (Gelflex, Australia), McGuire lens (Ultravision, United Kingdom) and Nissel K2 lens (Cantor+Nissel, United Kingdom).[18] Recommend 3-touch fitting philosophy. |
| Guideline #9 | BOZR=Ks (mm) Proposed to fit Comfort Kone lens (MetroOptics, USA)[23] or iKone lens (Valley Contax, USA).[24] Recommend slight apical clearance fitting philosophy. |
| Guideline #10 | Calculens.com Algorithm developed to select the first diagnostic lens in keratoconus.[14] Recommend 3-touch fitting philosophy. |
Kmean=Mean corneal curvature; Kf=Flattest corneal meridian; Ks=Steepest corneal meridian
Summary of the means and standard deviation (SD) of the BOZR proposed by each guideline. The correlation, arithmetic, and absolute mean and SD of the BOZR differences between each guideline and the BOZR that was fitted are shown
| Guideline BOZR proposed | Mean difference* and correlation between BOZR proposed and BOZR fitted | Absolute difference between proposed and fitted BOZR (mm) | Success rate† (95% CI) | |
|---|---|---|---|---|
| Guideline #1 | 7.34±0.39 | -0.14±0.14 ( | 0.16±0.12 | 26.3% (16.5-36.0%) |
| Guideline #2 | 7.23±0.34 | -0.03±0.14 ( | 0.10±0.10 | 41.3% (30.4-52.1%) |
| Guideline #3 | 7.05±0.48 | +0.15±0.26 ( | 0.23±0.18 | 19.8% (11.0-28.5%) |
| Guideline #4 | 7.16±0.40 | +0.03±0.17 ( | 0.14±0.11 | 26.3% (16.5-36.0%) |
| Guideline #5 | 7.58±0.23 | -0.38±0.22 ( | 0.39±0.21 | 3.8% (0.2-9.7%) |
| Guideline #6 | 6.96±0.47 | +0.23±0.18 ( | 0.24±0.16 | 12.3% (5.1-19.6%) |
| Guideline #7 | 7.31±0.41 | -0.11±0.20 ( | 0.16±0.16 | 23.4% (15.2-34.1%) |
| Guideline #8 | 7.16±0.47 | +0.03±0.18 ( | 0.13±0.12 | 34.6% (24.1-45.0%) |
| Guideline #9 | 7.04±0.44 | +0.15±0.22 ( | 0.20±0.18 | 29.6% (19.6-39.6%) |
| Guideline #10 | 7.19±0.40 | 0.00±0.12 ( | 0.09±0.08 | 50.6% (39.7-61.6%) |
*Paired t-Test (P<0.05 statistically significant). †Percentage of cases with a difference ≤0.05 mm with definitive BOZR fitted
Figure 1Agreement between the BOZR proposed by each guideline and the final BOZR fitted. Guideline #10 exhibiting better agreement with lower LoA
Figure 2Cumulative percentages of the differences between the BOZR proposed by each guideline and the final BOZR fitted
Figure 3Success rate of the GP guideline fitting (difference between the BOZR of the diagnostic lens proposed with the final BOZR prescribed was ≤0.05 mm) according to Amsler–Krumeich classifications