Bone strain index reproducibility and soft tissue thickness influence: a dual x-ray photon absorptiometry phantom study.

C Messina1,2, L P Piodi3, L Rinaudo4, I Emili5, F Porro6, C Buonomenna5, L M Sconfienza1,2, L Vergani7, F M Ulivieri8.   

Abstract

BACKGROUND: Bone strain index (BSI) is a tool measuring bone strain, derived from dual x-ray photon absorptiometry. It is able to characterise an aspect of bone quality that, joined to the quantity and quality parameters of bone mineral density (BMD) and trabecular bone score (TBS), permits an accurate definition of fracture risk. As no data are available about BSI precision, our aim was to assess its in vitro reproducibility.
METHODS: A Hologic spine phantom was used to perform BSI scans with three different scan modes: fast array (FA), array (A), and high definition (HD). Different soft tissue thicknesses (1, 3, 6 cm) of fresh pork rind layers as a surrogate of abdominal fat were interposed. For each scan mode, the phantom was consecutively scanned 25 times without repositioning.
RESULTS: In all scan modes (FA, A, HD) and at every fat thickness, BSI reproducibility was lower than that of BMD. The highest reproducibility was found using HD-mode with 1 cm of pork rind and the lowest one using HD-mode with 6 cm of pork rind. Increasing fat thickness, BSI reproducibility tended to decrease. BSI least significant change appeared to be about three times that of BMD in all modalities and fat thicknesses. Without pork rind superimposition and with 1-cm fat layer, BSI reproducibility was highest with HD-mode; with 3 or 6 cm fat thickness, it was higher with A-mode.
CONCLUSIONS: BSI reproducibility was worse than that of BMD, but it is less sensitive to fat thickness increase, similarly to TBS.

Entities:  

Keywords:  Absorptiometry (photon); Bone density; Bone strain index; Fractures (bone); Reproducibility of results

Mesh:

Year:  2019        PMID: 31410624      PMCID: PMC6692807          DOI: 10.1186/s41747-019-0110-9

Source DB:  PubMed          Journal:  Eur Radiol Exp        ISSN: 2509-9280


Key points

Bone strain index (BSI) is an index of bone quality derived from dual-energy x-ray absorptiometry. BSI reproducibility was tested in a spine phantom study. BSI reproducibility was lower than that of bone mineral density (BMD) and tended to decrease with the increase of interposed swine fat tissue. BSI reproducibility was less sensitive to the increase of interposed fat thickness than BMD. BSI behaviour is similar to that of other bone quality indexes, such as trabecular bone score.

Background

Osteoporosis is defined as a skeletal condition characterised by reduced bone strength, due to an impaired bone mass and compromised microarchitecture that cause an increased risk of non-traumatic fragility fractures [1]. Fragility fractures are important clinical events because they increase the disability and mortality related to this disease, with high social costs [1]. In addition, the presence of a fragility fracture itself predisposes to the recurrence of new fractures [1-3]. Dual x-ray photon absorptiometry (DXA) is considered the gold standard method for the diagnosis of reduced bone mass and for its follow-up [4]. Bone mineral density (BMD) is the most relevant parameter of bone load resistance and is calculated as bone mineral content on area (g/cm2). It is expressed by T-score and Z-score, indicating how the measured data diverge from the mean value of a normal population or from the mean value of a population of the same sex and age, respectively. According to the World Health Organization criteria, a T-score lower than -2.5 defines the cutoff for the diagnosis of osteoporosis in postmenopausal women and men over 50, while a Z-score lower than -2.0 represents the cutoff for the diagnosis of reduced bone mass in premenopausal women and men under 50 [5]. However, while BMD is clearly one of the major determinants of bone strength [6], the assessment of fracture risk based on BMD alone could lack sensitivity. In fact, many fragility fractures occur in osteopenic individuals (T score between -2.5 and -1.0), not only in subjects with osteoporosis (T score below -2.5) [7]. Other factors in addition to BMD account for bone strength and fracture risk, concurring to determine bone quality [8]. They are represented by trabecular architecture, bone geometry, and bone turnover. The last one is evaluated by biochemical assays, while the first two can be directly measured by bone biopsy. Recently, the use of DXA has been enriched with new software that are capable to provide information on bone microarchitecture and turnover, such as the trabecular bone score (TBS) and the hip structural analysis [9]. In this scenario, another DXA-derived software has been recently developed with the aim of further investigating bone quality, a tool called bone strain index (BSI) [10]. This software applies a mathematical model called finite element method (FEM) to DXA lumbar spine scans. FEM derived from quantitative computed tomography scans demonstrated to be predictive of experimental vertebra strength [11]. The BSI parameter uses the format of digital imaging and communications in medicine (DICOM) files to obtain an average strain field from FEM analysis, calculated within each lumbar vertebra. The force acting on the surface of each vertebra has been related to the height and weight of each patient. A classical mathematical matricial approach using a triangular mesh and FEM has been applied defining the stiffness matrix dependent from the local BMD. The result of this process is the distribution of the strain that is defined as the spatial deformation of every single element in which the vertebra was divided before the calculation [12]. In order to make this analysis easier in daily usage, the model described above has been completely automated integrating both the segmentation process and FEM analysis in a single software [10]. A recent analysis of the BSI software showed a good correlation between BSI and the yield strain applied to porcine vertebrae, meaning that BSI can be a predictor of elastic and plastic changes in the mechanical response of the bone. As a consequence, BSI could be applied in the prediction of the possible region of fracture within the vertebrae. When the software was applied to DXA scan of human vertebrae, a good correlation was found with T-score, in patients presenting T-score values lower than -3.0 and high values of BSI. Thus, high BSI values are associated with lower BMD values [10]. Finally, in recent clinical studies, BSI appeared to be useful to identify the osteoporotic patients’ subgroup particularly prone to fragility fractures [13], and to characterise young patients affected by secondary osteoporosis [14, 15]. Bone densitometry precision is one of the most important factors in clinical practice, especially for patient’s monitoring. Precision of DXA-related measurements such as BMD and TBS has been widely assessed [16-19]. Currently, no data are available about BSI precision. Thus, the aim of this study was to evaluate the reproducibility of BSI measurements on a phantom and to compare it with BMD reproducibility.

Methods

This phantom study was carried out at the IRCCS Istituto Ortopedico Galeazzi of Milan, Italy, in 2017–2018. For the in vitro study, the Institutional Review Board approval was not requested, because no patients were involved. A spine phantom (Model DPA/QDR-1, S/N: 13129, fan beam BMD = 0.985 g/cm2) provided by Hologic (Hologic Inc, Marlborough, Massachusetts, USA) for daily quality control was used. All measurements were performed using a QDR-Discovery W densitometer, software version n. 13.6.0.2 able to perform exams with three different scan modes: fast array (FA), array (A), and high definition (HD). To reproduce the human soft tissues, 5-mm layers of fresh pork rind were interposed as a surrogate of subcutaneous and visceral abdominal fat, given its high lipids’ content (80%). For each scan mode, the phantom was consecutively scanned 25 times without repositioning, with the specific purpose to test the best reproducibility for the software. The set of the three scan modes was performed without any layer of pork rind and then with layers creating an increasing thickness of 1 cm, 3 cm, and 6 cm. Thus, a total of 300 acquisitions (3 modes x 25 scans x 4 different pork rind thicknesses) were acquired. The region of interest (ROI) of the first scan was automatically set by the computer. At this stage, the operator could make image corrections if needed, for example in case of phantom ROI contour inaccuracies or intervertebral lines misplacement. After delineating the first ROI, the software automatically used it for the following acquisitions, thus avoiding any variability due to manual correction. Once DXA analysis was completed, data were obtained from the same ROI used for phantom BMD. For each scan, a standard measure of 190 cm and 80 kg was used in order to reproduce the same normal human body mass index (BMI) value of 22 kg/m2 for all acquisitions.

Statistical analysis

Normality of the data was assessed using Shapiro-Wilk test. Mean ± standard deviation (SD) of BMD and BSI were calculated. Precision was obtained in accordance to the suggestions of the International Society for Clinical Densitometry official positions, and the coefficient of variation (CoV) was calculated as the ratio between standard deviation and mean. Least significant change percentage (LSC%) was calculated as 2.77 × CoV. Reproducibility was calculated as the complement to 100% of LSC% [20]. We first evaluated BMD and BSI mean values differences with increasing thickness and between the three scan modes using one-way analysis of variance (ANOVA) with Tukey post-hoc test, after verification of variance homogeneity according to Levene test. As the BSI HD subset showed variance in homogeneity, we used the Welch ANOVA with the Games-Howell post-hoc test. The comparison between BMD and BSI reproducibility at different thicknesses was performed by calculating the SD distribution of all measurements. The resulting distributions were again tested using one-way ANOVA. A p value lower than 0.05 was considered statistically significant.

Results

The general overview of the results is reported in Table 1, with a direct comparison of BMD and BSI reproducibility for each modality with different thicknesses of interposed soft tissue. In general, without soft tissue superimposition, the LSC% of BSI was about three times higher than that of BMD, with the greatest difference for A-mode and FA-mode. This difference was somewhat lower at increasing thickness, with BSI LSC reaching the lowest value of 1.7% with 1 cm thickness of interposed soft tissue using HD-mode (versus BMD LSC = 0.7%). Overall, the reproducibility of BSI was lower compared to that of BMD.
Table 1

BMD and BSI mean values and their SD, LSC%, and reproducibility. Data are presented according to the specific scan mode and to the degree of soft tissue thickness

FA (0 cm)A (0 cm)HD (0 cm)
BMDBSIBMDBSIBMDBSI
Mean0.9872.1680.9872.1211.0052.077
SD0.0040.0220.0030.0190.0020.013
CoV0.4%1.0%0.3%0.9%0.2%0.6%
LSC1.0%2.8%0.7%2.5%0.6%1.8%
Reproducibility99.0%97.2%99.3%97.5%99.4%98.2%
FA (1 cm)A (1 cm)HD (1 cm)
BMDBSIBMDBSIBMDBSI
Mean0.9902.1460.9902.1421.0032.084
SD0.0030.0180.0040.0230.0030.012
CoV0.3%0.8%0.4%1.1%0.3%0.6%
LSC0.9%2.3%1.2%3.0%0.7%1.7%
Reproducibility99.1%97.7%98.8%97.0%99.3%98.3%
FA (3 cm)A (3 cm)HD (3 cm)
BMDBSIBMDBSIBMDBSI
Mean0.9802.1780.9812.1591.0012.081
SD0.0050.0290.0040.0170.0040.022
CoV0.5%1.3%0.4%0.8%0.4%1.1%
LSC1.3%3.7%1.1%2.2%1.1%3.0%
Reproducibility98.7%96.3%98.9%97.8%98.9%97.0%
FA (6 cm)A (6 cm)HD (6 cm)
BMDBSIBMDBSIBMDBSI
Mean0.9942.1340.9952.0991.0202.045
SD0.0050.0250.0050.0230.0060.029
CoV0.5%1.2%0.5%1.1%0.6%1.4%
LSC1.4%3.3%1.3%3.0%1.6%3.9%
Reproducibility98.6%96.7%98.7%97.0%98.4%96.1%

BMD Bone mineral density, BSI Bone strain index, FA Fast array, A Array, HD High definition, CoV Coefficient of variation, SD Standard deviation, LSCLeast significant change percentage, 0 cm Absence of interposed soft tissue

BMD and BSI mean values and their SD, LSC%, and reproducibility. Data are presented according to the specific scan mode and to the degree of soft tissue thickness BMD Bone mineral density, BSI Bone strain index, FA Fast array, A Array, HD High definition, CoV Coefficient of variation, SD Standard deviation, LSCLeast significant change percentage, 0 cm Absence of interposed soft tissue Table 2 shows a direct comparison between BMD and BSI reproducibility values with different thicknesses of interposed soft tissue using the three modalities. Regarding BMD, the highest reproducibility was 99.4% (0 cm soft tissue thickness, HD-mode), while the lowest was 98.4% (6 cm soft tissue thickness, HD-mode). For BMD, the increase in soft tissue thickness was associated with a decrease in reproducibility, which showed to be significant only for HD-mode (p < 0.001). On the other hand, the highest value of BSI reproducibility was 98.3% (1-cm soft tissue thickness, HD-mode), whereas the lowest was 96.1% (6 cm soft tissue thickness, HD-mode).
Table 2

Comparison between BMD and BSI reproducibility at different thicknesses

BMD (Fast array)BSI (Fast array)
Pork rind thickness0 cm1 cm3 cm6 cmp value0 cm1 cm3 cm6 cmp value
SD0.0040.0030.0050.005n.s.0.0220.0180.0290.025n.s.
CoV0.4%0.3%0.5%0.5%1.0%0.8%1.3%1.2%
LSC1.0%0.9%1.3%1.4%2.8%2.3%3.7%3.3%
Reproducibility99.0%99.1%98.7%98.6%97.2%97.7%96.3%96.7%
BMD (Array)BSI (Array)
Pork rind thickness0 cm1 cm3 cm6 cm0 cm1 cm3 cm6 cm
SD0.0030.0040.0040.005n.s.0.0190.0230.0170.023n.s.
CoV0.3%0.4%0.4%0.5%0.9%1.1%0.8%1.1%
LSC0.7%1.2%1.1%1.3%2.5%3.0%2.2%3.0%
Reproducibility99.3%98.8%98.9%98.7%97.5%97.0%97.8%97.0%
BMD (HD)BSI (HD)
Pork rind thickness0 cm1 cm3 cm6 cm0 cm1 cm3 cm6 cm
SD0.0020.0030.0040.006< 0.0010.0130.0120.0220.029< 0.001
CoV0.2%0.3%0.4%0.6%0.6%0.6%1.1%1.4%
LSC0.6%0.7%1.1%1.6%1.8%1.7%3.0%3.9%
Reproducibility99.4%99.3%98.9%98.4%98.2%98.3%97.0%96.1%

BMD Bone mineral density, BSI Bone strain index, HD High definition, CoV Coefficient of variation, SD Standard deviation, LSCLeast significant change percentage, 0 cm Absence of interposed soft tissue, n.s. Not statistically significant

Comparison between BMD and BSI reproducibility at different thicknesses BMD Bone mineral density, BSI Bone strain index, HD High definition, CoV Coefficient of variation, SD Standard deviation, LSCLeast significant change percentage, 0 cm Absence of interposed soft tissue, n.s. Not statistically significant Table 3 shows the comparisons between the mean values of BMD and BSI at 0 cm and 6 cm soft tissue thickness for each scan mode. There was a significant increase in BMD values in each scan mode, with the highest difference of 1.46% for HD-mode and the lowest difference of 0.76% for FA-mode. Differently from BMD, BSI values significantly decreased from 0 to 6 cm soft tissue thickness in all scan modes, and the variation was more pronounced than for BMD. In fact, the highest negative variation was 1.57% for FA-mode, while the lowest negative variation was 1.03% in A-mode.
Table 3

Comparison of BMD and BSI mean values at 0 and 6 cm of pork rind layers for each scan modality

FAArrayHD
BMDBSIBMDBSIBMDBSI
0 cm0.9872.1680.9872.1211.0052.077
6 cm0.994*2.134*0.995*2.099*1.020*2.045*
Difference 0–6 cm+ 0.007-0.034+ 0.008-0.022+ 0.015-0.032
% difference 0–6 cm+ 0.76%-1.57%+ 0.83%-1.03%+ 1.46%-1.55%
Highest LSC% (§)1.4% (6 cm)3.7% (3 cm)1.3% (6 cm)3.0% (both 1 and 6 cm)1.6% (6 cm)3.9% (6 cm)

BMD Bone mineral density, BSI Bone strain index, FA Fast array, HD High definition, LSC% Least significant change percentage, 0 cm No pork rind layers. § indicates at which soft tissue thickness there was the highest value of LSC% in each scan modality

*p < 0.05 when compared with no soft tissue superimposition (0 cm)

Comparison of BMD and BSI mean values at 0 and 6 cm of pork rind layers for each scan modality BMD Bone mineral density, BSI Bone strain index, FA Fast array, HD High definition, LSC% Least significant change percentage, 0 cm No pork rind layers. § indicates at which soft tissue thickness there was the highest value of LSC% in each scan modality *p < 0.05 when compared with no soft tissue superimposition (0 cm)

Discussion

This is the first study in which the reproducibility of BSI is tested on a phantom, showing that this new software has higher CoV and LSC% compared to BMD, being about two to three times higher. As a consequence, BSI reproducibility was lower than that of BMD, and it ranged from 96.1% (6 cm soft tissue thickness, HD-mode) to 98.3% (1 cm soft tissue thickness, HD-mode). BMD reproducibility confirmed to be very good, but was affected by the increase of soft tissue thickness. This is not the first phantom study aimed to evaluate the reproducibility of DXA measurements. The important aspect of these studies relies on the fact that phantoms are usually less affected by external factors, and this represents a methodological advantage to better understand the working principle of new techniques, such as the BSI. Previous studies have been published with data regarding BMD and TBS reproducibility [16–18, 21]. Concerning TBS, phantom studies showed that its reproducibility is somewhat lower compared to that of BMD, being between 96.4–98.3% [21] and 97.7–98.3% [18]. Usually, BMD reproducibility is known to be very good and typically represents the standard of reference for other DXA-based measurements. This was confirmed by our study, as BMD showed high values of reproducibility being around 99% in all the three scan modalities. On the other hand, the reproducibility of BSI was lower than that of BMD, and this finding has implication in clinical practice especially for patient’s monitoring. From a clinical point of view, the lower is the reproducibility the longer has to be the time of follow-up to observe a clinically significant variation. From a practical point of view, the increase in soft tissue thickness is a condition that is frequently found in clinical DXA routine, in patients with high values of BMI and waist circumference. In our study, we tried to understand the behaviour of this new technology by simulating this frequent clinical scenario, showing that BSI, despite having lower reproducibility, was equally affected as BMD. The increase in soft tissue thickness was associated to a reduction of precision, but this was statistically significant for both BMD and BSI only for the HD-mode. Thus, we found a detrimental effect of increasing soft tissue on BSI reproducibility, similar to that of BMD. When considering BMD and BSI mean values difference between 0 and 6 cm, we notice again a similar behaviour of these two parameters, as a statistically significant variation was found for both for BMD and BSI, despite the first increased its value and the latter showed a decrease. This divergence is expected, as higher BSI values are associated to lower BMD values and vice versa. Nevertheless, the amplitude of variation between 0 and 6 cm was slightly higher for BSI compared to BMD, a difference that was more evident for the FA-mode. This study presents some limitations. The first one is that we evaluated BSI reproducibility on a phantom that was designed for BMD quality control. Thus, the intrinsic value of BSI measurement may not be directly transferred to clinical practice, despite the evaluation of reproducibility remains independent from BSI significance. Another limitation is that we performed this study on a single phantom with a single densitometer. Thus, these results may present variations when applied to other settings, and the precision of this software may vary when a different phantom or densitometer is used, and of course, the precision may vary among patients with different degrees of bone quality. Lastly, we simulated fat tissue using a material (pork rind) which is similar but not equal to human fat; therefore, we may not directly apply these results to real BMI variations. In conclusion, this phantom study assessed the reproducibility of BSI compared to that of BMD in different scan modes and interposed fat thicknesses, showing that BSI reproducibility was overall lower than that of BMD and only slightly negatively influenced by interposed fat thickness increase respect to BMD.
  19 in total

1.  Fracture Risk Prediction by Non-BMD DXA Measures: the 2015 ISCD Official Positions Part 2: Trabecular Bone Score.

Authors:  Barbara C Silva; Susan B Broy; Stephanie Boutroy; John T Schousboe; John A Shepherd; William D Leslie
Journal:  J Clin Densitom       Date:  2015 Jul-Sep       Impact factor: 2.617

2.  Short-term precision assessment of trabecular bone score and bone mineral density using dual-energy X-ray absorptiometry with different scan modes: an in vivo study.

Authors:  Michele Bandirali; Alessandro Poloni; Luca Maria Sconfienza; Carmelo Messina; Giacomo Davide Edoardo Papini; Marcello Petrini; Fabio Massimo Ulivieri; Giovanni Di Leo; Francesco Sardanelli
Journal:  Eur Radiol       Date:  2015-02-08       Impact factor: 5.315

3.  Correlations between trabecular bone score, measured using anteroposterior dual-energy X-ray absorptiometry acquisition, and 3-dimensional parameters of bone microarchitecture: an experimental study on human cadaver vertebrae.

Authors:  Didier Hans; Nicole Barthe; Stephanie Boutroy; Laurent Pothuaud; Renaud Winzenrieth; Marc-Antoine Krieg
Journal:  J Clin Densitom       Date:  2011-07-02       Impact factor: 2.617

4.  Differences among array, fast array, and high-definition scan modes in bone mineral density measurement at dual-energy x-ray absorptiometry on a phantom.

Authors:  A Delnevo; M Bandirali; G Di Leo; C Messina; L M Sconfienza; A Aliprandi; F M Ulivieri; F Sardanelli
Journal:  Clin Radiol       Date:  2013-01-23       Impact factor: 2.350

5.  Reproducibility of trabecular bone score with different scan modes using dual-energy X-ray absorptiometry: a phantom study.

Authors:  Michele Bandirali; Giovanni Di Leo; Carmelo Messina; Maria Juana Pastor Lopez; Alessandro Mai; Fabio M Ulivieri; Francesco Sardanelli
Journal:  Skeletal Radiol       Date:  2014-08-12       Impact factor: 2.199

6.  Predictive value of BMD for hip and other fractures.

Authors:  Olof Johnell; John A Kanis; Anders Oden; Helena Johansson; Chris De Laet; Pierre Delmas; John A Eisman; Seiko Fujiwara; Heikki Kroger; Dan Mellstrom; Pierre J Meunier; L Joseph Melton; Terry O'Neill; Huibert Pols; Jonathan Reeve; Alan Silman; Alan Tenenhouse
Journal:  J Bone Miner Res       Date:  2005-03-07       Impact factor: 6.741

Review 7.  Osteoporosis prevention, diagnosis, and therapy.

Authors: 
Journal:  JAMA       Date:  2001-02-14       Impact factor: 56.272

8.  Risk of new vertebral fracture in the year following a fracture.

Authors:  R Lindsay; S L Silverman; C Cooper; D A Hanley; I Barton; S B Broy; A Licata; L Benhamou; P Geusens; K Flowers; H Stracke; E Seeman
Journal:  JAMA       Date:  2001-01-17       Impact factor: 56.272

Review 9.  Microarchitecture, the key to bone quality.

Authors:  Maria Luisa Brandi
Journal:  Rheumatology (Oxford)       Date:  2009-10       Impact factor: 7.580

Review 10.  Executive summary of the 2013 International Society for Clinical Densitometry Position Development Conference on bone densitometry.

Authors:  John T Schousboe; John A Shepherd; John P Bilezikian; Sanford Baim
Journal:  J Clin Densitom       Date:  2013 Oct-Dec       Impact factor: 2.617

View more
  6 in total

1.  Bone Strain Index: preliminary distributional characteristics in a population of women with normal bone mass, osteopenia and osteoporosis.

Authors:  Fabio Massimo Ulivieri; Luca Rinaudo; Carmelo Messina; Alberto Aliprandi; Luca Maria Sconfienza; Francesco Sardanelli; Bruno Mario Cesana
Journal:  Radiol Med       Date:  2022-09-04       Impact factor: 6.313

Review 2.  The Bone Strain Index: An Innovative Dual X-ray Absorptiometry Bone Strength Index and Its Helpfulness in Clinical Medicine.

Authors:  Fabio Massimo Ulivieri; Luca Rinaudo
Journal:  J Clin Med       Date:  2022-04-20       Impact factor: 4.964

3.  Reproducibility of DXA-based bone strain index and the influence of body mass: an in vivo study.

Authors:  Carmelo Messina; Luca Petruccio Piodi; Luca Rinaudo; Ciriaco Buonomenna; Luca Maria Sconfienza; Laura Vergani; Fabio Massimo Ulivieri
Journal:  Radiol Med       Date:  2019-12-27       Impact factor: 3.469

4.  Bone strain index in the prediction of vertebral fragility refracture.

Authors:  Fabio Massimo Ulivieri; Luca Petruccio Piodi; Luca Rinaudo; Paolo Scanagatta; Bruno Mario Cesana
Journal:  Eur Radiol Exp       Date:  2020-04-09

5.  Bone microarchitectural analysis using ultra-high-resolution CT in tiger vertebra and human tibia.

Authors:  Ryota Inai; Ryuichi Nakahara; Yusuke Morimitsu; Noriaki Akagi; Youhei Marukawa; Toshi Matsushita; Takashi Tanaka; Akihiro Tada; Takao Hiraki; Yoshihisa Nasu; Keiichiro Nishida; Toshifumi Ozaki; Susumu Kanazawa
Journal:  Eur Radiol Exp       Date:  2020-01-28

Review 6.  Image-based biomechanical models of the musculoskeletal system.

Authors:  Fabio Galbusera; Andrea Cina; Matteo Panico; Domenico Albano; Carmelo Messina
Journal:  Eur Radiol Exp       Date:  2020-08-13
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.