Sasithorn Amnuaywattakorn1, Chanika Sritara2, Chirawat Utamakul3, Wichana Chamroonrat4, Arpakorn Kositwattanarerk5, Kanungnij Thamnirat6, Boonsong Ongphiphadhanakul7.
Abstract
BACKGROUND: Trabecular bone score (TBS), which has been proposed to be used in complementary with bone mineral density (BMD) to improve the assessment of fracture risk, is negatively associated with body mass index (BMI). The effect of soft tissue, which is expected to be thicker in subjects with high BMI, on TBS was studied using three scan types: Hologic with fast array mode (Hfa), Hologic with high definition mode (Hhd), and GE-Lunar iDXA.Entities:
Mesh:
Substances:
Year: 2016 PMID: 26757709 PMCID: PMC4710977 DOI: 10.1186/s12891-016-0886-1
Source DB: PubMed Journal: BMC Musculoskelet Disord ISSN: 1471-2474 Impact factor: 2.362
Fig. 1Lumbar spine phantom setup. A Hologic BMD spine phantom, for routine quality control of BMD assessment, with a soft tissue equivalent material placed on top
TBS of a Lumbar Spine Phantom with Various Thicknesses of Soft Tissue Equivalent Material
| Soft | Hologic Fast Array (Hfa) | Hologic HD (Hhd) | iDXA | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Tissue | 95%CI | 95%CI | 95 % CI | |||||||||
| (cm) | mean | SD | Lower | Upper | mean | SD | Lower | Upper | mean | SD | Lower | Upper |
| 0 | 1.379 | 0.018 | 1.372 | 1.386 | 1.430 | 0.009 | 1.427 | 1.433 | 1.423 | 0.005 | 1.421 | 1.424 |
| 1 | 1.355* | 0.018 | 1.348 | 1.361 | 1.398* | 0.008 | 1.395 | 1.401 | 1.367* | 0.004 | 1.366 | 1.369 |
| 2.5 | 1.287* | 0.014 | 1.282 | 1.293 | 1.335* | 0.010 | 1.332 | 1.339 | 1.248* | 0.005 | 1.246 | 1.249 |
| 3.5 | 1.243* | 0.015 | 1.238 | 1.249 | 1.286* | 0.016 | 1.280 | 1.292 | 1.170* | 0.006 | 1.167 | 1.172 |
| 5 | 1.161* | 0.023 | 1.153 | 1.170 | 1.120* | 0.017 | 1.194 | 1.207 | 1.037* | 0.007 | 1.035 | 1.040 |
| 7.5 | 0.968* | 0.013 | 0.963 | 0.973 | 1.015* | 0.019 | 1.008 | 1.022 | 0.819* | 0.006 | 0.816 | 0.821 |
* P < 0.05 as compared with soft tissue = 0 cm
BMD of a Lumbar Spine Phantom with Various Thicknesses of Soft Tissue Equivalent Material
| Soft | Hologic Fast Array (Hfa) | Hologic HD (Hhd) | iDXA | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Tissue | 95%CI | 95%CI | 95%CI | |||||||||
| (cm) | mean | SD | lower | upper | mean | SD | lower | upper | mean | SD | lower | upper |
| 0 | 1.006 | 0.004 | 1.004 | 1.007 | 1.002 | 0.004 | 1.001 | 1.003 | 1.139 | 0.001 | 1.138 | 1.139 |
| 1 | 1.009 | 0.003 | 1.008 | 1.010 | 1.005* | 0.004 | 1.004 | 1.007 | 1.136* | 0.001 | 1.136 | 1.137 |
| 2.5 | 1.009 | 0.003 | 1.008 | 1.010 | 1.005* | 0.003 | 1.004 | 1.006 | 1.132* | 0.001 | 1.131 | 1.132 |
| 3.5 | 1.008 | 0.004 | 1.007 | 1.010 | 1.005* | 0.002 | 1.004 | 1.006 | 1.129* | 0.001 | 1.129 | 1.129 |
| 5 | 1.011* | 0.004 | 1.009 | 1.012 | 1.008* | 0.004 | 1.006 | 1.010 | 1.127* | 0.001 | 1.127 | 1.128 |
| 7.5 | 1.019* | 0.008 | 1.016 | 1.022 | 1.009* | 0.005 | 1.007 | 1.011 | 1.126* | 0.001 | 1.125 | 1.126 |
* P < 0.05 as compared with soft tissue = 0 cm
Least significant change (LSC) of TBS and BMD
| Mode | Hfa | Hhd | iDXA |
|---|---|---|---|
| TBS LSC (%) | 3.62 | 1.74 | 0.97 |
| TBS Reproducibility (%) | 96.4 | 98.3 | 99.0 |
| BMD LSC (%) | 1.10 | 1.11 | 0.24 |
| BMD Reproducibility (%) | 98.9 | 98.9 | 99.8 |
Hfa Hologic fast array; Hhd Hologic high definition
Fig. 2Phantom TBS images. TBS images and values of a Hologic BMD spine phantom by Hologic fast array (Hfa), Hologic high definition (Hhd), and Lunar iDXA with overlying soft tissue material of 0, 1, 2.5, and 7.5 cm
Fig. 3Phantom BMD images. BMD images of a Hologic BMD spine phantom by Hfa (a), Hhd (b), and iDXA (c) without and with overlying soft tissue equivalent materials. * Denotes the least thickness of soft tissue equivalent material that caused a statistically significant difference in BMD. Despite statistically significant changes, the absolute changes in BMD are smaller than LSC values; hence, unlikely to be a problem in BMD monitoring
Fig. 4Box-and-whisker plots of TBS and BMD. Box-and-whisker plots of TBS (a) and BMD (b) with various thicknesses of overlying soft tissue equivalent materials using three scan types: Hfa (blue dashed line), Hhd (red dotted line), and iDXA (green solid line)