| Literature DB >> 31286912 |
Kwan-Sik Sung1, You Gyoung Yi2, Hyung-Ik Shin3.
Abstract
BACKGROUND: Muscle strength measurements using hand-held dynamometry (HHD) can be affected by the inadequate strength of the tester and lack of stabilization of the participants and the device. A portable HHD anchoring system was designed that enabled the measurement of isometric knee extensor muscle strength in a supine position. This can be used with individuals who are unable to assume the sitting position required for the measurement of knee extensor strength in conventional isokinetic dynamometry (IKD). The aim of this study was to evaluate the reliability and validity of knee extensor strength measurements using this device.Entities:
Keywords: Hand-held dynamometry; Portable dynamometer anchoring system; Supine
Year: 2019 PMID: 31286912 PMCID: PMC6615264 DOI: 10.1186/s12891-019-2703-0
Source DB: PubMed Journal: BMC Musculoskelet Disord ISSN: 1471-2474 Impact factor: 2.362
Fig. 1The portable dynamometer anchoring system. (a) The portable dynamometer anchoring system for measuring supine isometric knee extensor strength. A: Movable frame that can adjust the placement of the hand-held dynamometer (HHD) according to the size of the lower leg. B: Movable frame that can adjust the HHD placement according to leg length. C: Belt to fix the anchoring system to the bed. (b) Position of the examiner to prevent pelvic rotation of the participant
Maximal isometric knee extensor strength measurements for each trial and each rater
| Measurement | Mean (N) | SD (N) | Range (N) |
|---|---|---|---|
| Trial 1 – rater 1 | 492.00 | 157.1 | 213.3–813.8 |
| Trial 2 – rater 1 | 486.87 | 153.0 | 239.3–761.2 |
| Trial 3 – rater 2 | 499.32 | 159.2 | 238.0–799.0 |
SD: standard deviation; N: Newtons
Intra- and inter-rater reliability of maximal isometric knee extensor strength in each trial
| ICC Mean (95% CI) | SEM (%) Mean (SD) | MDC (%) Mean (SD) | |
|---|---|---|---|
| Intra-rater | 0.98 (0.96–0.98) | 21.8 (4.5) | 60.4 (12.3) |
| Inter-rater | 0.98 (0.96–0.99) | 22.2 (4.5) | 61.6 (12.4) |
ICC: intraclass correlation coefficient, CI: confidence interval, SEM: standard error of measurement, MDC: minimal detectable change
Fig. 2Bland and Altman plot for intra-rater measurements. (a) The mean difference and limits of agreement (LOA) between the maximum knee extensor strength measurements, (b) The mean difference (%) and LOA (%) between the maximum knee extensor strength measurements
Fig. 3Bland and Altman plot for inter-rater measurements. (a) The mean difference and limits of agreement (LOA) between the maximum knee extensor strength measurements, (b) The mean difference (%) and LOA (%) between the maximum knee extensor strength measurements
Fig. 4Correlation between maximal voluntary knee extensor torque obtained using the portable anchoring system and isokinetic dynamometer. r: Pearson correlation coefficient