| Literature DB >> 33505769 |
Debananda Padhy1,2, Shrikant R Bharadwaj2, Suryasmita Nayak1,2, Suryasnata Rath3, Taraprasad Das4.
Abstract
Purpose: The purpose of this study was to determine the accuracy and repeatability of refractive errors obtained using three autorefractors based on different measurement principles, vis-à-vis, gold-standard retinoscopy. Methodology: Accuracy of noncycloplegic, sphero-cylindrical refractive error of 234 eyes was obtained using the rotary prism-based RM-8900 closed-field autorefractor, photorefraction based Spot vision screener, wavefront aberrometry based E-see, and streak retinoscopy by four different examiners, masked to the results of each other. Intersession repeatability of autorefractors was determined by repeat measurements in a subset of 40 subjects.Entities:
Keywords: accuracy; astigmatism; autorefractor; emmetropia; myopia; photorefraction; repeatability; screening; wavefront aberrations
Year: 2021 PMID: 33505769 PMCID: PMC7794271 DOI: 10.1167/tvst.10.1.2
Source DB: PubMed Journal: Transl Vis Sci Technol ISSN: 2164-2591 Impact factor: 3.283
Detailed Characteristics of the Four Refractive Error Measurement Techniques Evaluated in this Study
| Characteristics | Retinoscope | Tabletop Autorefractor | Photoscreener | E-see |
|---|---|---|---|---|
| Working principle | Foucault knife-edge | Image size assessment using rotary prisms | Eccentric IR photorefraction | Wavefront aberrometry |
| Operating range | No specific operating range | +22 D to -25 D for sphere and ±10.00 D for cylinder | ±7.50 D for sphere and ±3.50 D for cylinder | ±10.00 D for sphere and ±6.00 D for cylinder |
| Invasiveness of measurement | Noninvasive. Measurements performed from a distance. | Noninvasive. Participant stabilized on forehead rest. | Noninvasive. Measurements are performed from a distance. | Device touches the face for measurement |
| Portability | Portable | Non-portable | Portable | Portable |
| Measurement time | Consumes time for measurement | ∼5 measurements per second | <1 s per measurement | ∼10 measurements per second |
| Fixation distance | Distant target | Near target with simulated distance viewing | Distant target | Distant target |
| Pupil size dependence | Measurements become challenging with pupil miosis | Stable measurements for pupil diameter up to 2 mm | Measurement accuracy decreases with pupil miosis | Measurement accuracy may vary with pupil diameter |
| Binocular viewing / measurements | Viewing can be binocular, but the measurement is monocular | Only monocular viewing and measurements | Viewing and measurement can be binocular | Only monocular viewing and measurements |
| Participant cooperation | Needs limited cooperation. Useful in challenging cases. | Participant cooperation is essential for reliable measurements | Needs limited cooperation. Useful in challenging cases. | Participant cooperation is critical for reliable measurements |
| Examiner training | Examiner training is time-consuming | Examiner can be trained within a short time | Examiner can be trained within a short period of time | Examiner can be trained within a short period of time |
| Near-triad measurements | Only measurement of accommodation is possible | Only measurement of accommodation is possible | Near-triad can be measured in sync with each other | Only measurement of accommodation is possible |
Data on operating ranges and measurement time are obtained from the manufacturer prescribed user manual of each instrument.
Demographic Details of the Study Participants
| Attributes | Range, [Mean (±1 SD)] |
|---|---|
| Sample size | 234 20 children (≤16 y) | 214 adults |
| Age, y | 5–58 [29.3 (±11.1)] 5–16 [12.1 (±3.4)] for children | 17– 58 [30.9 (±10.2)] for adults |
| Male / Female | 151/ 83 |
| Refractive error diagnosis (D) | Emmetropia: n = 35 (15%) Simple myopia: n = 39 (16.45%) | −8 to −0.2 [−1.8 (±2.1)] Simple hyperopia: n = 26 (11%) | 0.2–8 [1.3 (±1.7)] Simple myopic astigmatism: n = 38 (17%) | −3.0 to −0.2 [−0.9 (±0.6)] Compound myopic astigmatism: n = 74 (31%) | −0.5 to 10.2 [−3.7 (±2.2)] Simple hyperopic astigmatism: n = 4 (2%) | 0.2–0.7 [0.4 (±0.2)] Compound hyperopic astigmatism: n = 8 (4%) | 0.5–3.0 [1.4 (±0.8)] Mixed astigmatism: n = 10 (4%) | −0.6–0.25 [−0.6 (±0.2)] |
| Retinoscopy refraction (D) | M: −10.25 to 8 [−1.2 (±2.3)] J0: −1.4 to 1.8 [0.02 (±0.4)] J45: −0.9 to 1.2 [0.01 (±0.1)] |
Wherever applicable, both the range and the mean ± 1 standard deviation (within square parenthesis) of the data are shown in the table.
Figure 1.Bland-Altman type plots of the agreement in M power vector values between the three autorefractor and retinoscopy evaluated in this study. The mean difference (MD) and the 95% limits of agreement (LOA) are indicated by the solid, dashed lines in each panel, respectively. The numerical values of these parameters are also indicated in each panel. Negative values along the abscissa of each panel indicate myopic refraction. Positive and negative values along the ordinate of each panel indicate an overestimation and underestimation bias by a given autorefractor, relative to retinoscopy.
Figure 2.Bland-Altman type plots of the agreement in J0 power vector values among the three autorefractor and retinoscopy evaluated in this study. All other details of the plots are the same as Figure 1.
Figure 3.Scatter diagram of the difference in M (gray filled symbols) and J0 (black open symbols) power vector components of refraction obtained using retinoscopy and table-top autorefractor (panel A), retinoscopy and Photoscreener (panel B) and retinoscopy and E-see (panel C) plotted as a function of the subject's age.
Figure 4.Bland-Altman type plots for the repeatability of the M power vector across the three autorefractor evaluated in this study. All other details of the plots are the same as Figure 1.
Figure 5.Cumulative frequency distribution plots indicating the percentage of M (panel A) and J0 (panel B) power vector values within a given dioptric difference from gold-standard retinoscopy. Panel C provides data of repeatability of the M power vector for all three techniques in the same format as panels A and B. The dioptric difference shown along the abscissa of all three panels is unsigned absolute values, unlike the Bland-Altman plots in Figures 1 to 3. The abscissa range in the three panels is different to demonstrate the data trends.