Literature DB >> 31248280

Plasticizer Interaction With the Heart: Chemicals Used in Plastic Medical Devices Can Interfere With Cardiac Electrophysiology.

Rafael Jaimes1,2, Damon McCullough1, Bryan Siegel2, Luther Swift1,2, Daniel McInerney1, James Hiebert1, Erick A Perez-Alday3, Beatriz Trenor4, Jiansong Sheng5, Javier Saiz1, Larisa G Tereshchenko3, Nikki Gillum Posnack1,2,6.   

Abstract

BACKGROUND: Phthalates are used as plasticizers in the manufacturing of flexible, plastic medical products. Patients can be subjected to high phthalate exposure through contact with plastic medical devices. We aimed to investigate the cardiac safety and biocompatibility of mono-2-ethylhexyl phthalate (MEHP), a phthalate with documented exposure in intensive care patients.
METHODS: Optical mapping of transmembrane voltage and pacing studies were performed on isolated, Langendorff-perfused rat hearts to assess cardiac electrophysiology after MEHP exposure compared with controls. MEHP dose was chosen based on reported blood concentrations after an exchange transfusion procedure.
RESULTS: Thirty-minute exposure to MEHP increased the atrioventricular node (147 versus 107 ms) and ventricular (117 versus 77.5 ms) effective refractory periods, compared with controls. Optical mapping revealed prolonged action potential duration at slower pacing cycle lengths, akin to reverse use dependence. The plateau phase of the action potential duration restitution curve steepened and became monophasic in MEHP-exposed hearts (0.18 versus 0.06 slope). Action potential duration lengthening occurred during late-phase repolarization resulting in triangulation (70.3 versus 56.6 ms). MEHP exposure also slowed epicardial conduction velocity (35 versus 60 cm/s), which may be partly explained by inhibition of Nav1.5 (874 and 231 µmol/L half-maximal inhibitory concentration, fast and late sodium current).
CONCLUSIONS: This study highlights the impact of acute MEHP exposure, using a clinically relevant dose, on cardiac electrophysiology in the intact heart. Heightened clinical exposure to plasticized medical products may have cardiac safety implications-given that action potential triangulation and electrical restitution modifications are a risk factor for early after depolarizations and cardiac arrhythmias.

Entities:  

Keywords:  action potentials; electrophysiology; heart; plasticizer; plastics

Mesh:

Substances:

Year:  2019        PMID: 31248280      PMCID: PMC6693678          DOI: 10.1161/CIRCEP.119.007294

Source DB:  PubMed          Journal:  Circ Arrhythm Electrophysiol        ISSN: 1941-3084


Phthalate chemical exposure has been reported in patients undergoing invasive medical procedures that use large quantities of plastic materials. We previously reported that phthalate chemical exposure impairs cell coupling and slows conduction velocity in a cardiomyocyte cell model. Acute phthalate exposure slowed atrioventricular conduction and increased atrioventricular node effective refractory period in an intact, whole heart model. Phthalate exposure prolonged action potential duration time, enhanced action potential triangulation, and increased the ventricular effective refractory period. Phthalate exposure slowed epicardial conduction velocity, which may be partly explained by inhibition of Nav1.5. See Editorial by Plastics have revolutionized clinical care. Yet, despite the many advantages, concerns have been raised about the ubiquitous use of plastics in the clinical setting.[1-3] To manufacture flexible plastic products, phthalate esters are added to impart flexibility to otherwise stiff polyvinyl chloride polymers. Di-2-ethylhexyl phthalate (DEHP) is the most commonly used plasticizer in Food and Drug Association–approved medical devices, including blood storage bags, tubing circuits, enteral feeding tubes, endotracheal tubes, and catheters.[4] In the finished product, DEHP can contribute up to 40% by weight in intravenous bags and 80% by weight in medical tubing. Because phthalate esters are hydrophobic and noncovalently bound to the polyvinyl chloride polymer, these additives are highly susceptible to leaching when in contact with blood, plasma, and other lipophilic solutions.[5,6] Consequently, heightened phthalate exposure has been observed in patients undergoing invasive medical procedures that use large quantities of plastic materials—including cardiopulmonary bypass, extracorporeal membrane oxygenation, dialysis, and transfusion procedures (see Table).[7-21] As an example, a neonatal exchange transfusion procedure resulted in plasma concentrations as high as 38 µM DEHP and 54 µM mono-2-ethylhexyl phthalate (MEHP), the primary metabolite of DEHP.[13] Whereas multiple medical procedures can increase a patient’s cumulative phthalate exposure to levels that are 4000× to 160 000× higher than deemed safe.[7] Importantly, phthalate levels can remain elevated in the blood for hours (eg, blood transfusion; 5–24 hour half-lives)[22,23] to weeks (eg, extracorporeal membrane oxygenation support, intensive care unit patient) depending on the course of treatment.[19] Measurements of Clinical Exposure to DEHP and Its Metabolites Including MEHP Incidental plastic chemical exposure during medical procedures is virtually unavoidable, yet the direct impact on patient health remains unclear. Because phthalates structurally resemble natural hormones, they can exert endocrine-disrupting properties and interfere with an array of biological processes.[7,24,25] Moreover, these low molecular weight chemicals have been shown to interact with ion channels, nuclear receptors, and other cellular targets. Accordingly, phthalates are likely to have a global impact on human health that is multifactorial. Indeed, epidemiological studies have reported associations between phthalate exposure and a broad range of health conditions, including metabolic disturbances, reproductive disorders, inflammatory conditions, neurological disorders, and cardiovascular disease.[24-31] The latter is particularly worrisome—since patients who are most susceptible to hospital-based phthalate exposures are often at heightened risk for cardiovascular complications, including those undergoing multiple transfusions, chronic transfusions (sickle cell, thalassemia) or support procedures (extracorporeal membrane oxygenation, cardiopulmonary bypass, dialysis).[4,9,19,32-34] Local and systemic reactions to plastic chemicals may add an extra insult to patients who are prone to secondary complications, including hypoxia (congenital heart defect patients), ischemia-reperfusion injury (cardiac surgery patients), postoperative arrhythmias, cardiac iron overload, pulmonary hypertension, impaired contractility, and diastolic dysfunction. Indeed, animal studies have shown a causal relationship between phthalate exposure and subsequent alterations in cardiovascular function.[35-39] Using a rodent heart-lung preparation, Labow, et al[38] showed that phthalates exert a hypertensive effect on the pulmonary vasculature. This group also showed that phthalate exposure has a negative inotropic effect on isolated human atrial trabeculae, thus highlighting the applicability to humans.[36] Our group has also shown that phthalate exposure alters calcium handling and contractility in vitro, using rodent and human cardiomyocytes.[40,41] Moreover, phthalate exposure was shown to reduce gap junctional connexin-43,[40,41] which decreased intercellular coupling and slowed conduction velocity (CV). Therefore, we hypothesized that phthalate exposure may directly impact cardiac electrophysiology in an intact heart. To test this hypothesis, we comprehensively assessed the electrophysiological effects of MEHP using isolated, intact rodent hearts. Cardiac electrophysiology parameters were measured at baseline and again after treatment with either control or MEHP-supplemented media perfusion (30 minutes). To address clinical relevance, the MEHP dose (60 µM) selected for heart perfusion was comparable to patient exposure after an exchange transfusion procedure[13] or measured levels in stored blood products.[4,42]

Methods

The authors declare that data that supports the findings of this study are available within the article.

Isolated Heart Preparation

Animal protocols were approved by the Institutional Animal Care and Use Committee of the Children’s Research Institute and followed the National Institutes of Health’s Guide for the Care and Use of Laboratory Animals. Experiments were conducted using adult, male Sprague-Dawley rats (>8 weeks old, >280 g, Taconic Biosciences). Animals were housed in conventional rat cages in the Research Animal Facility under standard environmental conditions (12:12 hour light:dark cycle, 64°C–78°C, 30%–70% humidity, free access to reverse osmosis water, corn cob bedding, and food [2918 rodent chow; Envigo]). Animals were anesthetized with 3% to 5% isoflurane; the heart was excised and then transferred to a temperature-controlled (37°C) constant-pressure (70 mm Hg) Langendorff perfusion system for electrophysiology and optical mapping experiments. Excised hearts were perfused with Krebs-Henseleit buffer bubbled with carbogen throughout the duration of the experiment, as previously described (≈1 hour).[43]

General Protocol

MEHP (93+%; Wako Chemicals) stock solution was prepared in 99+% dimethyl sulfoxide (DMSO), and the working concentration was prepared directly in perfusate media with a total DMSO concentration of <0.001%. To avoid bolus exposure, the perfusion system was designed to premix MEHP with perfusate media to the desired concentration before entering the aorta. Cardiac electrophysiology was assessed under baseline conditions (control media perfusion) and again after either control media perfusion or 60 µM MEHP-supplemented media perfusion (30 minutes). This allowed us to control for electrophysiology changes that could be attributed to procedure (control baseline versus control 30 minutes), individual animal differences, and MEHP treatment (control 30 minutes versus MEHP 30 minutes). For clinical relevance, the MEHP dose (60 µM) was selected based on an upper clinically relevant exposure after an exchange transfusion procedure[13] and within the range of measured levels in stored blood products.[4,42]

Electrophysiology Measurements

A reference electrode was positioned proximal to the right atrium and a measuring electrode proximal to the apex to acquire a pseudo-ECG in the lead II configuration (differential amplifier with 1000× gain). During sinus rhythm, ECG signals were collected to analyze heart rate, beat rate variability, atrioventricular conduction (PR interval) depolarization time, and repolarization time (RJpoint, RT).[44] Signals were acquired in iox2 (emka Technologies), and ECG segments were computed in ecgAUTO (emka Technologies). For determination of atrioventricular nodal effective refractive period (AVNERP), Wenckebach cycle length (WBCL), and sinus node recovery time, a coaxial stimulation electrode (Harvard Apparatus, Holliston, MA) was placed on the right atrium. AVNERP was defined as the maximum extrastimulus interval (S1-S2) during atrial pacing that failed to conduct through the atrioventricular node, as indicated by loss of ventricular capture. WBCL was defined as the maximum pacing cycle length (PCL) during atrial pacing that caused the Wenckebach phenomenon. Sinus node recovery time was measured to assess the automaticity of the sinoatrial node and defined as the time delay between the last paced atrial depolarization and the first sinus beat (5 s electrical stimulation at 200 ms PCL). A stimulator (Bloom Electrophysiology, Denver, CO) was set to 1 ms pulse width at 1.5× diastolic current threshold. For ventricular pacing, a 0.25 mm diameter tungsten, unipolar, cathodal electrode was placed on the LV epicardium and centered in the imaging field. A stainless steel indifferent electrode was placed under the heart. To determine the ventricular effective refractory period (VERP) within 10 ms, dynamic pacing (S1-S1) was performed during optical mapping. An arrhythmia score was designated for each individual heart after the dynamic pacing protocol, wherein the most severe electrical disturbance was classified (1=single premature ventricular contraction [polyvinyl chloride], 2=bigeminy/salvos, 3=ventricular tachycardia or fibrillation).[45]

Ventricular Conduction Frequency Composition Assessment

High resolution (2000 Hz) unfiltered digital pseudo-ECG recordings were analyzed using a custom MATLAB (MathWorks, Inc, Natick, MA) software application. Normal sinus median beat was constructed out of 436 normal sinus beats, as previously described.[46] Appropriate selection of beats was confirmed by investigator (Dr Perez-Alday) with the aid of a graphical display. From the median beat obtained, QRS was measured from QRS onset to J point.[47] The power spectral density (PSD) was measured on the median beat of the ECG signal[48] using a nonparametric Fast-Fourier transform algorithm based on the Welch-Bartlett power estimation method,[49] with a Hanning window of length 8 samples and a 50% overlap between segments. From the Nyquist frequency, the available frequency range was from 0 to 1000 Hz; however, the frequency composition was calculated on 5 subsets of 100 Hz frequency intervals from 0 Hz to 500 Hz. The relative contribution of frequencies was normalized over the total power.

Optical Mapping

After establishing a baseline heart rate (10 minutes), the perfusate was supplemented with 10 µmol/L (−)-blebbistatin (Sigma-Aldrich) to reduce motion artifact for imaging experiments.[50] The heart was loaded with a potentiometric dye (62.1 µg RH237, 1-minute staining) through a bubble trap located proximal to the aortic cannula.[51,52] The epicardium was illuminated with 2 broad light-emitting diode spotlights (530 nm, 200 mW; Mightex). RH237 fluorescence was longpass filtered (680+ nm; Chroma Technologies), and optical action potentials were acquired using a fixed focal lens (17 mm, f/0.95; Schneider Optics) attached to an sCMOS camera (Zyla 4.2 PLUS; Andor Technology). The camera sensor was cropped to 384×256 pixels and set to an exposure time of 1.219 ms resulting in the acquisition of 16-bit monochrome images at 814 frames per second for 2 s. Optical maps were acquired during each PCL (S1-S1; 250–80 ms, 10 ms increments). A custom MATLAB (MathWorks, Natick, MA) script was used to calculate the action potential durations (APD) to percent repolarization (APD30, 80, 90) for every beat in the recording, as previously described. APD triangulation was defined as the difference between APD90 and APD30.[53] Restitution curves were generated by plotting APD80 against PCL; curves were then analyzed as either monophasic or biphasic (where applicable) to determine slopes, as indicated by Franz.[54] APD alternans were defined as sequential APD80 measurements that differed by >4 ms.

Image and Signal Processing

Data analysis was performed using a custom MATLAB script. A region of interest (<20 pixel radius) was selected from the raw image, averaged, and plotted against time. Drift removal was performed by subtraction of a polynomial fit if necessary. To remove high-frequency noise, a fifth order Butterworth low-pass filter was applied to the resulting signals with a cutoff frequency adjusted between 50 and 150 Hz. A peak detector was then used to measure the total number of action potentials in the file across time. Characteristics from each event are measured and averaged, including APD30, APD80, and APD90, as described previously. Image processing was performed, and isochrone activation maps were constructed in RHYTHM.[55] The background was removed, convolved with a 15×15 uniform kernel for box blurring and then time signals were low-pass filtered below 100 Hz. The activation time of every pixel on the heart was defined as the maximum derivative of the action potential or transient upstroke. To measure epicardial CV, activation maps were subsequently loaded into ORCA[56] and measured at a minimum of 18 angles separated by 10° each from the pacing stimulus with the median CV reported. Subsequent electrical wave propagation images were then constructed using custom Python scripts and plotted with matplotlib.[57]

Whole-Cell Voltage-Clamp Recordings

Recordings were performed at room temperature (22°C–25°C) using HEK293 (human embryonic kidney 293) cells stably transfected with Nav1.5 cDNA. For Nav1.5 recordings, the extracellular solution (mM) included 137 NaCl, 10 HEPES, 4 KCl, 1 MgCl2, 1 CaCl2, 10 dextrose, and the intracellular solution included 120 aspartic acid, 120 CsOH, 10 CsCl, 10 HEPES, 10 EGTA, 5 MgATP, 0.4 TrisGTP. The voltage protocol (1 s duration) was repeated at 0.1 Hz. Briefly, cells were repolarized from −95 to −120 mV for 200 ms, then depolarized from −120 to −15 mV for 40 ms, and further depolarized to +40 mV for 200 ms, followed by a voltage ramp down phase for 100 ms from +40 to −95 mV. Nav1.5 late current was induced by 20 nmol/L anemone toxin in the extracellular solution, which allowed for the study of late sodium current contribution from Nav1.5 that is separate from fast sodium current, as previously described.[58] Nav1.5 fast current was measured in the absence of anemone toxin. For positive control, 30 µM tetrodotoxin (sodium channel blocker) was directly applied to recordings and resulted in 98.3±0.7% Nav1.5 current block (n=3). To ensure baseline recording was stable enough for drug application, cells were presented with this voltage protocol in control solution until Nav1.5 current amplitudes for 12 consecutively recorded current traces (2-minute duration) exhibit <10% difference. Effects were monitored for 3 minutes before and after application of MEHP (n=4 per dose). To quantify drug potency against Nav1.5 channels, the steady state Nav1.5 current amplitude (averaged value from 5 steady state current traces) in drug solution was divided by the averaged amplitude from the last 5 traces measured in control solution to calculate the fractional block. Then, fractional block was plotted against drug concentration tested, fitted with the Hill Equation to generate an half-maximal inhibitory concentration (IC50) and the Hill coefficient.

In Silico Modeling

Human ventricular action potentials were simulated using previously established models and conductance.[59,60] The block of fast and late sodium, INa and INaL, respectively, were provoked with the computed IC50 from MEHP determined by whole-cell recordings (see previous section) as previously described.[62] The isolated epicardial cellular model was paced with 1000 pulses (−80 µA/µF, 0.5 ms) at PCL=250 ms in the presence of 60 µmol/L MEHP.

Statistical Analysis

Statistical analysis was performed using the R software package and Stata MP 15.1 (StataCorp, College Station, TX). Data normality was confirmed by Shapiro-Wilk test. Datasets were compared using a 2-sample Student independent t test between the 2 treatment groups (30-minute control, 30-minute MEHP treated) or paired t test between baseline and 30-minute time points. Significance was defined as P≤0.05. Results were reported as mean±SE mean, from a minimum of 5 animals per group as indicated. As distribution of PSD variables was skewed, we used Wilcoxon rank-sum test to compare groups (control versus MEHP treated). Wilcoxon matched-pairs signed-ranks test was used for paired comparison of PSD changes before and after MEHP treatment. Variables with skewed distribution are presented as median and interquartile range. Arrhythmia incidence was compared using the χ2 test.

Results

Effects of MEHP Exposure on Cardiac Automaticity and Beat Rate Variability During Sinus Rhythm

ECG characteristics were quantified at 2 time points—first under baseline conditions and again after 30 minutes of control or 60 µM MEHP-supplemented media perfusion during sinus rhythm (Figure 1A). Example ECG traces are shown in Figure 1B. Heart rate at 30 minutes was comparable between experimental groups, 172 beats per minute in control and 178 beats per minute in MEHP-treated hearts (Figure 1C and 1D). Beat rate variability was assessed as a cumulative index of automaticity and sinus node discharge. Although both root means successive square difference and SD of the normal RR intervals trended higher in MEHP-treated hearts (9.4±2.4 and 17.3±4.2 ms, respectively) compared with controls (5.4±1.7 and 11.9±2.5 ms, respectively), neither parameter was significant (root means successive square difference, P=0.19; SD of the normal RR intervals, P=0.28) Figures 1E and 1F. MEHP treatment increased RJpoint and RT intervals by 55% and 30%, respectively (Figure 2), both readily identifiable indicators of depolarization and repolarization in the excised rodent heart[44] (Figures 1G and 1H). Atrioventricular conduction delay was also measured during sinus rhythm. Atrioventricular conduction remained relatively stable in control hearts, as indicated by a PR interval time of 43±2 ms at baseline and 47±2 ms after 30-minute perfusion. In comparison, the PR interval time increased by 44% in MEHP-treated hearts (41±1 ms baseline, 59±3 ms, P<0.05; Figure 1I).
Figure 1.

Effect of mono-2-ethylhexyl phthalate (MEHP) on ECG characteristics during sinus rhythm. A, Experimental protocol timeline; all ECG recordings were performed during sinus rhythm. B, ECG signals collected from excised, intact rat hearts with PR interval denoted. C, Heart rate (HR) decreased with time of perfusion and after application of blebbistatin for mechanical uncoupling, with no difference between treatment groups. D, Similar rate slowing was observed in the RR interval, and coincided with heart rate variability changes in E SD of the normal RR intervals (SDNN) and (F) root means successive square difference (rMSSD) after 30-min perfusion. G and H, Prolonged depolarization and repolarization in MEHP-exposed hearts as measured from R to J inflection point (RJpoint) and R to T-wave (RT) (G) PR interval time was significantly lengthened in MEHP-exposed hearts compared to both baseline and control (Ctrl) hearts after the same duration of perfusion (P<0.05). n≥7 per group. BPM indicates beats per minute; and EP, electrophysiology protocol.

Figure 2.

Sinus node function and atrioventricular conduction altered after mono-2-ethylhexyl phthalate (MEHP) treatment. A, ECG trace corresponding to a 200 ms pacing cycle length (PCL) train (indicated by arrows) on the right atrium in a control (Ctrl) heart. Sinus node recovery time (SNRT) is indicated as the delay between the cessation of pacing and the resumption of sinus node activity. B, SNRT measured in MEHP-treated heart shows marked prolongation compared with Ctrl. C, Aggregate SNRT analysis for MEHP-treated hearts and time-matched Ctrls (P<0.05). D, Wenckebach cycle length (WBCL) measured as the earliest PCL in which Wenckebach phenomenon was visible (P=0.07). E, Atrioventricular nodal refractory period (AVNERP) assessed by S1-S2 pacing on the right atrium was significantly lengthened in MEHP hearts compared with Ctrls (P<0.05). n=5–8 per group.

Effect of mono-2-ethylhexyl phthalate (MEHP) on ECG characteristics during sinus rhythm. A, Experimental protocol timeline; all ECG recordings were performed during sinus rhythm. B, ECG signals collected from excised, intact rat hearts with PR interval denoted. C, Heart rate (HR) decreased with time of perfusion and after application of blebbistatin for mechanical uncoupling, with no difference between treatment groups. D, Similar rate slowing was observed in the RR interval, and coincided with heart rate variability changes in E SD of the normal RR intervals (SDNN) and (F) root means successive square difference (rMSSD) after 30-min perfusion. G and H, Prolonged depolarization and repolarization in MEHP-exposed hearts as measured from R to J inflection point (RJpoint) and R to T-wave (RT) (G) PR interval time was significantly lengthened in MEHP-exposed hearts compared to both baseline and control (Ctrl) hearts after the same duration of perfusion (P<0.05). n≥7 per group. BPM indicates beats per minute; and EP, electrophysiology protocol. Sinus node function and atrioventricular conduction altered after mono-2-ethylhexyl phthalate (MEHP) treatment. A, ECG trace corresponding to a 200 ms pacing cycle length (PCL) train (indicated by arrows) on the right atrium in a control (Ctrl) heart. Sinus node recovery time (SNRT) is indicated as the delay between the cessation of pacing and the resumption of sinus node activity. B, SNRT measured in MEHP-treated heart shows marked prolongation compared with Ctrl. C, Aggregate SNRT analysis for MEHP-treated hearts and time-matched Ctrls (P<0.05). D, Wenckebach cycle length (WBCL) measured as the earliest PCL in which Wenckebach phenomenon was visible (P=0.07). E, Atrioventricular nodal refractory period (AVNERP) assessed by S1-S2 pacing on the right atrium was significantly lengthened in MEHP hearts compared with Ctrls (P<0.05). n=5–8 per group.

Electrophysiological Study

Sinus node function was further investigated by measuring sinus node recovery time, which was delayed by 54% in MEHP-treated hearts (532±57 ms) compared with control hearts (345±10 ms, P=0.01, Figure 2A through 2C). WBCL trended longer in MEHP-treated hearts compared with controls, 155.8±12.4 to 126.6±4.5 (P=0.07, Figure 2D). Atrioventricular node conduction refractoriness was interrogated by implementing an atrial pacing protocol to measure AVNERP, which was significantly longer in MEHP-treated hearts compared with controls, 147±13 ms to 107±4.4 ms (P<0.05, Figure 2E). These results suggest that MEHP may directly increase atrioventricular nodal refractoriness.

Effects of MEHP Exposure on Ventricular Conduction and Refractoriness During Sinus Rhythm

ECG frequency content was also analyzed during sinus rhythm, as high frequencies during depolarization are associated with abnormal or slowed conduction.[47,62] In MEHP-treated hearts, contribution of high frequencies (400–500 Hz) was increased by 0.032% (interquartile range, −0.008% to 0.20%), whereas in control group, contribution of the same frequency band (400–500 Hz) decreased by 0.06% (interquartile range, 0.04%–1.19%), P=0.076. ECG lead placement among isolated hearts proved difficult due to the lack of anatomic markers and consistency in positioning of the heart relative to ECG leads, making experimental variability between individual hearts high for PSD measurements. Follow-up studies were performed using optical mapping, because APD and VERP measurements may more accurately predict alterations in ventricular repolarization in the rodent heart.[63]

Effects of MEHP Exposure on Epicardial Conduction and Refractoriness During Epicardial Pacing

Optical mapping of transmembrane voltage signals was performed to evaluate action potential shape and duration, using an epicardial pacing protocol. APD at 80% repolarization (APD80) was assessed until VERP was reached, and APD80 values were compared between the timed control and MEHP-treatment group giving rise to the electrical restitution curves (Figure 3A). Ventricular refractoriness was determined during our optical mapping protocol which involves ventricular pacing at both baseline and 30-minute conditions. VERP was 51% longer in MEHP-treated hearts (117±9.5 ms) compared with time-matched controls (77.5±5.5 ms, P<0.05; Figure 3B). A longer refractory period suggests that MEHP exposure slows repolarization. MEHP-treated hearts displayed a monophasic restitution curve with a slope of 0.18 throughout (steep portion only, no plateau), as compared to controls which displayed a biphasic shape (steep and plateau phases). The steep slope of the controls was like that of the MEHP group at 0.17, and the slope of the plateau phase was 0.06. Importantly, monophasic restitution curves are associated with an increased risk of electrical alternans.[54,64] With a lengthened VERP in MEHP-treated hearts, loss of capture at PCLs <100 ms limited APD measurements in the MEHP group.
Figure 3.

Mono-2-ethylhexyl phthalate (MEHP) causes action potential morphology changes with corresponding modification of the electrical restitution curve. A, MEHP exposure caused the restitution curve to change from a biphasic form to a monophasic form with no plateau phase. The slope of the plateau on the control (Ctrl) heart was 0.06 compared with 0.18 in the MEHP hearts. B, Despite the similarities in the action potential duration at faster cycle lengths, the ventricular effective refractory period (VERP) was markedly prolonged in the MEHP-treated group (P<0.05). C, With rapid pacing (pacing cycle length [PCL]=140 ms), the action potential duration (APD) appeared similar between the Ctrl and MEHP-treated groups: there were no difference between (D) APD30, (E) APD90, or (F) AP triangulation. G, At slower cycle lengths (PCL=240 ms), the action potential morphology was quite different: while (H) APD30 was similar, (I) lengthening was apparent in the late phase as shown by APD90 (P<0.05). J, The effect on APD90 with no change on APD30 caused an increase in the triangulation shape of the AP after MEHP treatment (P<0.05). N=7 for each group. AP Tri indicates action potential triangulation; and Vm, transmembrane voltage.

Mono-2-ethylhexyl phthalate (MEHP) causes action potential morphology changes with corresponding modification of the electrical restitution curve. A, MEHP exposure caused the restitution curve to change from a biphasic form to a monophasic form with no plateau phase. The slope of the plateau on the control (Ctrl) heart was 0.06 compared with 0.18 in the MEHP hearts. B, Despite the similarities in the action potential duration at faster cycle lengths, the ventricular effective refractory period (VERP) was markedly prolonged in the MEHP-treated group (P<0.05). C, With rapid pacing (pacing cycle length [PCL]=140 ms), the action potential duration (APD) appeared similar between the Ctrl and MEHP-treated groups: there were no difference between (D) APD30, (E) APD90, or (F) AP triangulation. G, At slower cycle lengths (PCL=240 ms), the action potential morphology was quite different: while (H) APD30 was similar, (I) lengthening was apparent in the late phase as shown by APD90 (P<0.05). J, The effect on APD90 with no change on APD30 caused an increase in the triangulation shape of the AP after MEHP treatment (P<0.05). N=7 for each group. AP Tri indicates action potential triangulation; and Vm, transmembrane voltage. APD30 and APD90 were measured to compute the triangulation of the action potential. At a selected fast PCL (140 ms), there were no significant differences between control and MEHP (Figure 3C). APD30 prolonged with time in both groups (Figure 3D), with no difference between APD90 and AP triangulation (Figures 3E and 3F, respectively). At a slower PCL (240 ms), APD90 was significantly prolonged in MEHP-treated hearts (93.5±3.5 ms) compared with controls (75.2±4.2 ms, P=0.02) (Figure 3G). The lengthening of APD90 (Figure 3I) without a significant change in APD30 (Figure 3H) led to an increase in the AP triangulation (APD90–APD30)[53]; AP triangulation was 56.6±4.5 ms in the controls and increased to 70.3±2.9 ms in the MEHP group (Figure 3J). These results suggest that MEHP exposure exerts a reverse use-dependence effect, wherein myocyte refractoriness and action potential prolongation is increased at slower heart rates.[65]

Electrical Alternans and Arrhythmia Susceptibility

AP triangulation, reverse use dependence, and a monophasic electrical restitution curve with no plateau phase have been associated with an increased propensity for electrical alternans and arrhythmias.[53,54,66] APD instability was assessed by quantifying the incidence of electrical alternans or sequential beat-to-beat variations in APD, such as displayed in Figure 4A. The threshold of electrical alternans was observed in MEHP-treated hearts at a cycle length of 122±11 ms compared with 96±6 ms PCL (P=0.06, Figure 4B). Rodent hearts are resistant to fibrillation, due in part to the small tissue mass.[44] Therefore, to assess arrhythmia susceptibility, each individual heart was scored by the most severe electrical disturbance encountered during the electrophysiology study at the end of the experiment. Arrhythmia incidence was significantly increased in MEHP-treated hearts, compared with controls (P<0.05, Figure 4C).
Figure 4.

Action potential alternans and arrhythmia susceptibility from mono-2-ethylhexyl phthalate (MEHP) exposure. A, MEHP-treated hearts displayed increased susceptibility to alternating action potential morphology at the same pacing cycle length (PCL) compared with baseline as predicted by the change in electrical restitution (See Figure 3A), PCL=110 ms this example. B, The longest PCL which resulted in alternating action potentials (alternans threshold) was longer in the MEHP-treated hearts (P=0.06). C, Hearts after MEHP exposure became more susceptible to spontaneous arrhythmia as shown by a higher incidence of both premature ventricular contractions (PVC) and ventricular tachycardia or ventricular fibrillation (VT/VF) following electrophysiological study (P<0.0001, χ2 test). N=8 for each group. AUF indicates arbitrary units of fluorescence; and Ctrl, control.

Action potential alternans and arrhythmia susceptibility from mono-2-ethylhexyl phthalate (MEHP) exposure. A, MEHP-treated hearts displayed increased susceptibility to alternating action potential morphology at the same pacing cycle length (PCL) compared with baseline as predicted by the change in electrical restitution (See Figure 3A), PCL=110 ms this example. B, The longest PCL which resulted in alternating action potentials (alternans threshold) was longer in the MEHP-treated hearts (P=0.06). C, Hearts after MEHP exposure became more susceptible to spontaneous arrhythmia as shown by a higher incidence of both premature ventricular contractions (PVC) and ventricular tachycardia or ventricular fibrillation (VT/VF) following electrophysiological study (P<0.0001, χ2 test). N=8 for each group. AUF indicates arbitrary units of fluorescence; and Ctrl, control.

MEHP Exposure Slows Conduction Velocity

Our previous in vitro studies suggest that phthalate esters uncouple neighboring cardiomyocytes, resulting in slowed conduction and an arrhythmogenic phenotype.[40] To test whether this effect was observed in an intact heart, CV across the epicardial surface was measured using optical mapping and a unipolar, cathodal pacing electrode. The electrode was centered in the field of view, and 10° increments were used to measure ≥180° of coverage for longitudinal and transverse CV measurements. Optical mapping confirmed that paced beats propagated as elliptical wavefronts that emanated from the pacing electrode (Figures 5A through 5D). For depiction, we show 2 selected regions for each heart across the epicardial surface and the conduction time from pacing initiation to the distal site under control (Figure 5E) and MEHP (Figure 5F) conditions. To investigate CV restitution, both a slow (240 ms) and fast (140 ms) PCL were selected for CV measurements. Incidentally, CV increased slightly between baseline and 30-minute timed recordings in control hearts at both pacing rates (Figures 5G and 5H). However, MEHP-treated hearts displayed 47% slower CV at 240 ms PCL (Figure 5G) and 42% slower CV at 140 ms PCL (Figure 5H) compared with time-matched controls.
Figure 5.

Activation maps and measured conduction velocity (CV) after mono-2-ethylhexyl phthalate (MEHP) exposure. A, Each pixel on the heart was assigned an activation time based on the maximum upstroke velocity in the fluorescent signal to create an isochrone map. Wavefronts are visible emanating from the center of the left ventricular epicardium where the cathodal pacing electrode was placed. B, After 30-min control (Ctrl)-media perfusion, the activation maps did not show significant differences. C, In Ctrl hearts, analysis of the pacing site (R1) and a distal site (R2) did not show significant differences in conduction time when compared between baseline and 30-min perfusion. D, Pre-MEHP baseline showed similar CV as the Ctrl baseline. E, After 30-min of MEHP exposure, the activation maps changed in morphology. F, Analysis of the pacing site (R1) and distal site (R2) indicated a decreased upstroke velocity after 30-min of MEHP exposure. G, CV was significantly slower in MEHP-treated hearts compared with Ctrls after 30 min at both pacing cycle length (PCL)=240 ms (P<0.05) and (H) PCL=140 ms (P<0.05), with N=5 per group. I, Sodium current (INa) was measured with HEK293 (human embryonic kidney 293) cells which stably express Nav1.5. Fast sodium current (INa) was measured in the absence of anemone toxin, and half-maximal inhibition concentration (IC50) was determined at 874 µmol/L. Late sodium current (INaL) was activated with anemone toxin, and corresponding IC50 was determined at 231 µmol/L (N=5 for each group).

Activation maps and measured conduction velocity (CV) after mono-2-ethylhexyl phthalate (MEHP) exposure. A, Each pixel on the heart was assigned an activation time based on the maximum upstroke velocity in the fluorescent signal to create an isochrone map. Wavefronts are visible emanating from the center of the left ventricular epicardium where the cathodal pacing electrode was placed. B, After 30-min control (Ctrl)-media perfusion, the activation maps did not show significant differences. C, In Ctrl hearts, analysis of the pacing site (R1) and a distal site (R2) did not show significant differences in conduction time when compared between baseline and 30-min perfusion. D, Pre-MEHP baseline showed similar CV as the Ctrl baseline. E, After 30-min of MEHP exposure, the activation maps changed in morphology. F, Analysis of the pacing site (R1) and distal site (R2) indicated a decreased upstroke velocity after 30-min of MEHP exposure. G, CV was significantly slower in MEHP-treated hearts compared with Ctrls after 30 min at both pacing cycle length (PCL)=240 ms (P<0.05) and (H) PCL=140 ms (P<0.05), with N=5 per group. I, Sodium current (INa) was measured with HEK293 (human embryonic kidney 293) cells which stably express Nav1.5. Fast sodium current (INa) was measured in the absence of anemone toxin, and half-maximal inhibition concentration (IC50) was determined at 874 µmol/L. Late sodium current (INaL) was activated with anemone toxin, and corresponding IC50 was determined at 231 µmol/L (N=5 for each group).

MEHP Inhibits Sodium Channel Current

Slowed myocardial CV can be attributed to a reduction in sodium channel current, anatomic obstacles, or diminished cellular coupling (eg, gap junctions). Because the latter 2 mechanisms may take additional time (exceeding 30-minute chemical exposure), we focused our attention on the direct effects of MEHP on sodium current. The onset of current suppression was concentration-dependent (Figure 5I). MEHP suppressed fast and late sodium current (INa) with an IC50 of 874 µmol/L and 231 µmol/L, respectively. Tetrodotoxin (30 µmol/L), a potent neurotoxin that blocks Nav1.5, served as a positive control resulting in >96% inhibition for all experiments (data not shown). Using in silico modeling combined with IC50 values, 60 µM MEHP is expected to decrease dV/dt by 4.6% (215 to 205 mV/ms), based on a human epicardial action potential at basic cycle length of 250 ms.

Discussion

This study is the first to examine the impact of MEHP on the electrophysiology of the intact heart, using a clinically relevant concentration that was comparable to patient exposure after an exchange transfusion procedure[13] or measured levels in stored blood products.[4,42] Importantly, we show that acute MEHP exposure results in slowed atrioventricular conduction, slowed epicardial CV, and increased atrioventricular and ventricular refractory periods. Optical mapping studies revealed a prolonged APD at slower PCLs, akin to reverse use dependence. MEHP-treated hearts also displayed a triangulated action potential and a steeper electrical restitution curve—both of which have been associated with an increased propensity for electrical alternans and arrhythmias.

Automaticity and Sinus Node Function

Automaticity in the isolated heart is determined by a balance between inward and outward currents (voltage clock)[67-69] and calcium oscillations (calcium clock),[70] and as such, alterations in the spontaneous beating rate can serve as a cumulative index of cardiac excitability. Previous studies have reported bradycardia after phthalate exposure in vivo[39] (>20 mg/kg MEHP), in vitro[71] (10 µM DEHP), and using an isolated heart model[72] (250 µM DEHP). In the current study, we observed comparable heart rates between treatment groups (control, MEHP) at the dose and duration chosen for the study (60 µM MEHP, 30 minutes). Alterations in automaticity may become more pronounced after longer treatment times, with higher concentrations, or exposure to the parent compound versus metabolite (DEHP versus MEHP). We also examined beat rate variability as a cumulative index of automaticity and sinus node discharge. Both root means successive square difference and SD of the normal RR intervals trended higher in MEHP-treated hearts, but neither parameter was statistically significant. Sinus node function was directly measured using a pacing protocol to quantify sinus node recovery time, which was significantly delayed by 54% in MEHP-treated hearts. These results suggest an underlying effect of phthalate esters on potassium channel current or calcium handling in the sinoatrial node[74]; indeed, alterations in calcium cycling and contractility have been reported in cardiac preparations treated with phthalate esters.[36,41]

Atrioventricular Conduction

We assessed the direct effect of MEHP exposure on intrinsic atrioventricular electrical conduction, by quantifying PR interval, AVNERP, and WBCL. Similar to a previous report by Aronson et al,[72] we observed a 44% prolongation of PR interval time in MEHP-treated hearts compared with controls—concomitant with a 37% increase in AVNERP and 23% increase in WBCL. Negative dromotropy through the atrioventricular node is commonly observed with cholinergics, and at least 1 study has shown that the effects of phthalate esters on human atrial trabeculae are partially inhibited by atropine, suggesting an interaction with cholinergic receptors.[36] Alternatively, an interaction between phthalate esters and potassium or calcium channels could also alter nodal conduction. Indeed, a recent whole-cell patch clamp study demonstrated that phthalate exposure (10–100 µM DEHP, parent compound of MEHP) blocks L-type calcium channels in vascular smooth muscle cells.[74]

Ventricular Conduction and Electrical Instability

Using an isolated intact heart preparation, we observed significant effects on ventricular conduction and repolarization. During sinus rhythm, MEHP-treatment prolonged RJpoint and RT intervals, indicating a slowing of the internal conduction system. Notably, recent studies have highlighted the utility of APD and VERP measurements as precise indicators of ventricular repolarization.[63] Using an epicardial pacing protocol, we observed a 51% increase in VERP and 24% increase in APD90 time at slower PCL. On the surface, this result might suggest a beneficial, antiarrhythmic effect of acute MEHP exposure. Though importantly, APD prolongation is a key contributor to electrical instability when the AP is triangulated[53,66]—as is the case with pharmaceuticals that exert off-target effects on potassium channels encoded by hERG, which can set the stage for early afterdepolarizations and Torsades.[75] Moreover, reverse-use dependence can also be proarrhythmic as the APD restitution slope changes shape and becomes steeper. Similar to the effects of sotalol on electrical restitution,[76] we observed significant APD90 prolongation at slower PCL (>190 ms) and a steep, monophasic APD restitution curve in MEHP-treated hearts. A steeper APD restitution slope with lack of plateau phase at longer basic cycle lengths can lead to an increased spatial dispersion of restitution properties, therefore, creating a proarrhythmic milieu and electrical alternans.[64] Due to a significant difference in VERP, we were unable to reach faster pacing rates in MEHP-treated hearts, but electrical alternans were observed at longer PCL in MEHP hearts (122 ms) compared with controls (96 ms). We also observed a 42% to 47% decrease in epicardial CV, which can be partly attributed to an immediate inhibitory effect of MEHP on sodium channels. Indeed, fast and late sodium channel current was inhibited by MEHP (IC50 874 µmol/L and 231 µmol/L MEHP, respectively) using a HEK293 cell line stably expressing Nav1.5. IC50 values were subsequently used for human in silico modeling, predicting a modest slowing of dV/dt. Though the latter was more modest than anticipated, additional voltage-clamp studies using a cardiomyocyte model may aid in the interpretation of these results. Due to the acute nature of the observed effects of MEHP exposure, our study suggests that phthalate esters interfere with ionic currents in the myocardium—and these effects may be multifactorial. Unfortunately, the rodent myocardium is not ideal for assessing arrhythmia incidence because of its small size and robustness to fibrillation.[44] Future electrophysiological studies are needed, in a larger animal model, to definitively draw a conclusion on the arrhythmogenic potential of MEHP and other phthalate esters. Clinical studies showed that abnormal slow conduction in patients with ventricular arrhythmias can manifest by increased power of high frequencies within the ECG signal.[62,77] In this study, we observed a trend towards higher contribution of high frequencies within QRS, but PSD measurements proved difficult due to variability in excised heart positioning relative to ECG lead placement.

Study Limitations

The scope of the study was limited to the effect of acute (30 minutes) MEHP exposure on cardiac electrophysiology, using an isolated heart model. Such a model prevents the investigation of longer exposure time, and ex vivo results may differ from in vivo studies with an intact vascular and autonomic nervous system. The reported results are also limited to the effect of MEHP, and therefore, do not account for additional effects that may be caused by DEHP (the parent compound) or other secondary metabolites. The presented study was also limited to a single dose of MEHP, which was chosen based on the reported blood concentration after an exchange transfusion procedure.

Conclusions

This study reports the electrophysiological effects of MEHP on intact, isolated rodent hearts. The increase in action potential triangulation and altered APD restitution curve are both indicators of increased arrhythmic risk.

Acknowledgments

We gratefully acknowledge Manelle Ramadan, BS, Morgan Burke, BS, and Ashish Doshi, MD, PhD for technical assistance.

Sources of Funding

This work was supported by the National Institutes of Health (R00ES023477 and R01HL139472 to Dr Posnack), Children’s Research Institute and Children’s National Heart Institute. We thank the generosity of the NVIDIA corporation for the graphics processing unit to perform image processing. This work was partially supported by the Dirección General de Política Científica de la Generalitat Valenciana (PROMETEU2016/088) to J.S. and B.T.

Disclosures

None.
Table 1.

Measurements of Clinical Exposure to DEHP and Its Metabolites Including MEHP

  72 in total

1.  Exposure to di(2-ethylhexyl) phthalate in premature neonates in a neonatal intensive care unit in Taiwan.

Authors:  Pen-Hua Su; Yan-Zin Chang; Hua-Pin Chang; Shu-Li Wang; Hsin-I Haung; Po-Chin Huang; Jia-Yuh Chen
Journal:  Pediatr Crit Care Med       Date:  2012-11       Impact factor: 3.624

2.  Feasibility of a semi-automated method for cardiac conduction velocity analysis of high-resolution activation maps.

Authors:  Ashish N Doshi; Richard D Walton; Sébastien P Krul; Joris R de Groot; Olivier Bernus; Igor R Efimov; Bastiaan J Boukens; Ruben Coronel
Journal:  Comput Biol Med       Date:  2015-05-21       Impact factor: 4.589

3.  Human embryonic and induced pluripotent stem cell-derived cardiomyocytes exhibit beat rate variability and power-law behavior.

Authors:  Yael Mandel; Amir Weissman; Revital Schick; Lili Barad; Atara Novak; Gideon Meiry; Stanislav Goldberg; Avraham Lorber; Michael R Rosen; Joseph Itskovitz-Eldor; Ofer Binah
Journal:  Circulation       Date:  2012-01-18       Impact factor: 29.690

4.  Anemone toxin (ATX II)-induced increase in persistent sodium current: effects on the firing properties of rat neocortical pyramidal neurones.

Authors:  M Mantegazza; S Franceschetti; G Avanzini
Journal:  J Physiol       Date:  1998-02-15       Impact factor: 5.182

5.  Kinetics of di(2-ethylhexyl) phthalate (DEHP) and mono(2-ethylhexyl) phthalate in blood and of DEHP metabolites in urine of male volunteers after single ingestion of ring-deuterated DEHP.

Authors:  Winfried Kessler; Wanwiwa Numtip; Wolfgang Völkel; Elcim Seckin; György A Csanády; Christian Pütz; Dominik Klein; Hermann Fromme; Johannes G Filser
Journal:  Toxicol Appl Pharmacol       Date:  2012-08-20       Impact factor: 4.219

Review 6.  Frequency content and characteristics of ventricular conduction.

Authors:  Larisa G Tereshchenko; Mark E Josephson
Journal:  J Electrocardiol       Date:  2015-08-28       Impact factor: 1.438

7.  Dispositions of di- and mono-(2-ethylhexyl) phthalate in newborn infants subjected to exchange transfusions.

Authors:  P Sjöberg; U Bondesson; G Sedin; J Gustafsson
Journal:  Eur J Clin Invest       Date:  1985-12       Impact factor: 4.686

8.  Simulation of the undiseased human cardiac ventricular action potential: model formulation and experimental validation.

Authors:  Thomas O'Hara; László Virág; András Varró; Yoram Rudy
Journal:  PLoS Comput Biol       Date:  2011-05-26       Impact factor: 4.475

9.  Prenatal phthalate exposure and anogenital distance in male infants.

Authors:  Shanna H Swan
Journal:  Environ Health Perspect       Date:  2006-02       Impact factor: 9.031

10.  Diabetic hyperglycaemia activates CaMKII and arrhythmias by O-linked glycosylation.

Authors:  Jeffrey R Erickson; Laetitia Pereira; Lianguo Wang; Guanghui Han; Amanda Ferguson; Khanha Dao; Ronald J Copeland; Florin Despa; Gerald W Hart; Crystal M Ripplinger; Donald M Bers
Journal:  Nature       Date:  2013-09-29       Impact factor: 49.962

View more
  9 in total

1.  Optocardiography and Electrophysiology Studies of Ex Vivo Langendorff-perfused Hearts.

Authors:  Luther M Swift; Rafael Jaimes; Damon McCullough; Morgan Burke; Marissa Reilly; Takuya Maeda; Hanyu Zhang; Nobuyuki Ishibashi; Jack M Rogers; Nikki Gillum Posnack
Journal:  J Vis Exp       Date:  2019-11-07       Impact factor: 1.355

2.  Plastics and cardiovascular disease.

Authors:  Nikki Gillum Posnack
Journal:  Nat Rev Cardiol       Date:  2021-02       Impact factor: 32.419

3.  Age-dependent changes in electrophysiology and calcium handling: implications for pediatric cardiac research.

Authors:  Luther M Swift; Morgan Burke; Devon Guerrelli; Marissa Reilly; Manelle Ramadan; Damon McCullough; Tomas Prudencio; Colm Mulvany; Ashika Chaluvadi; Rafael Jaimes; Nikki Gillum Posnack
Journal:  Am J Physiol Heart Circ Physiol       Date:  2019-12-30       Impact factor: 4.733

4.  Chronic perinatal hypoxia delays cardiac maturation in a mouse model for cyanotic congenital heart disease.

Authors:  Jennifer Romanowicz; Devon Guerrelli; Zaenab Dhari; Colm Mulvany; Marissa Reilly; Luther Swift; Nimisha Vasandani; Manelle Ramadan; Linda Leatherbury; Nobuyuki Ishibashi; Nikki Gillum Posnack
Journal:  Am J Physiol Heart Circ Physiol       Date:  2021-03-19       Impact factor: 4.733

Review 5.  Impacts of Environmental Insults on Cardiovascular Aging.

Authors:  Yang Lan; Shaowei Wu
Journal:  Curr Environ Health Rep       Date:  2022-02-01

6.  Emerging concepts and opportunities for endocrine disruptor screening of the non-EATS modalities.

Authors:  Christopher J Martyniuk; Rubén Martínez; Laia Navarro-Martín; Jorke H Kamstra; Adam Schwendt; Stéphane Reynaud; Lorraine Chalifour
Journal:  Environ Res       Date:  2021-08-19       Impact factor: 6.498

7.  Low-level plasticizer exposure and all-cause and cardiovascular disease mortality in the general population.

Authors:  Guowei Zeng; Qi Zhang; Xiaowei Wang; Kai-Hong Wu
Journal:  Environ Health       Date:  2022-03-09       Impact factor: 7.123

8.  Sex-Specific Programming of Cardiac DNA Methylation by Developmental Phthalate Exposure.

Authors:  Laurie K Svoboda; Kai Wang; Raymond G Cavalcante; Kari Neier; Justin A Colacino; Maureen A Sartor; Dana C Dolinoy
Journal:  Epigenet Insights       Date:  2020-08-05

9.  Potential Consequences of the Red Blood Cell Storage Lesion on Cardiac Electrophysiology.

Authors:  Marissa Reilly; Chantal D Bruno; Tomas M Prudencio; Nina Ciccarelli; Devon Guerrelli; Raj Nair; Manelle Ramadan; Naomi L C Luban; Nikki Gillum Posnack
Journal:  J Am Heart Assoc       Date:  2020-10-22       Impact factor: 5.501

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.