Abstract
INTRODUCTION: This study investigates basic dosimetric properties of unflattened 6 MV photon beam shaped by multileaf collimator and compares them with those of flattened beams. MATERIALS AND METHODS: Monte Carlo simulation model using BEAM code was developed for a 6MV photon beam based on Varian Clinic 600 unique performance linac operated with and without a flattening filter in beam line. Dosimetric features including lateral profiles, central axis depth dose, photon and electron spectra were calculated for flattened and unflattened cases, separately. RESULTS: An increase in absolute depth dose with a factor of more than 2.4 was observed for unflattened beam which was dependent on depth. PDDs values were found to be lower for unflattened beam for all field sizes. Significant decrease in calculated mlc leakage was observed when the flattening filter was removed from the beam line. The total scatter factor, SCP was found to show less variation with field sizes for unflattened beam indicating a decrease in head scatter. The beam profiles for unflattened case are found to have lower relative dose value in comparison with flattened beam near the field edge, and it falls off faster with distance. CONCLUSION: Our study showed that increase in the dose rate and lower peripheral dose could be considered as realistic advantages for unflattened 6MV photon beams.
INTRODUCTION: This study investigates basic dosimetric properties of unflattened 6 MV photon beam shaped by multileaf collimator and compares them with those of flattened beams. MATERIALS AND METHODS: Monte Carlo simulation model using BEAM code was developed for a 6MV photon beam based on Varian Clinic 600 unique performance linac operated with and without a flattening filter in beam line. Dosimetric features including lateral profiles, central axis depth dose, photon and electron spectra were calculated for flattened and unflattened cases, separately. RESULTS: An increase in absolute depth dose with a factor of more than 2.4 was observed for unflattened beam which was dependent on depth. PDDs values were found to be lower for unflattened beam for all field sizes. Significant decrease in calculated mlc leakage was observed when the flattening filter was removed from the beam line. The total scatter factor, SCP was found to show less variation with field sizes for unflattened beam indicating a decrease in head scatter. The beam profiles for unflattened case are found to have lower relative dose value in comparison with flattened beam near the field edge, and it falls off faster with distance. CONCLUSION: Our study showed that increase in the dose rate and lower peripheral dose could be considered as realistic advantages for unflattened 6MV photon beams.
Entities:
Keywords:
Flattening Filter Free Beams ; Multileaf Collimator; Monte Carlo Simulation
Medical linear accelerators are equipped with a flattening filter (FF) as their integral part which is primarily designed to produce a flat beam profile by compensating for the non-uniformity of photon fluence across the field. Flattening filter is a major source of quality changes within the primary beam due to scattering and absorbing of primary photons, and it also decreases beam output noticeably. Introduction of new technologies such as intensity-modulated radiation therapy (IMRT) or intensity-modulated arc therapy have removed the necessity to have a flattened beam profile. In principle, the flattening filter can be removed, and the leaf sequences of mlc can be adjusted accordingly to produce desired fluence distributions similar to those of a beam with a flattening filter. Substantial reduction in head scatter as well as significant increases in dose rate can be achieved by the removal of flattening filter with its associated attenuation from the x-ray beam path [1]. Thus flattening filter free (FFF) beams in radiotherapy have the advantage of shorter treatment delivery time and lower peripheral dose compared to conventional flattened beams. This is particularly vital for the treatment where large doses per fraction are advised, e.g., stereotactic ablative body radiotherapy [2,3] or where patient’s motion might affect the efficacy of the delivery or both [4]. Many studies have been conducted using Monte Carlo (MC) method for analyzing the influence of linac head components on beam characteristics [5,6,7]. Thus, the effect of flattening filter on absolute absorbed dose, beam profiles and various energy spectra could be studied by this method [8]. In an MC study on flattening filter free beams, dose rates increase by a factor of 2.31(6 MV) and 5.45(18 MV) and out-of-field dose reductions were reported [9]. In a similar study, a significant improvement in out-of-field dose was reported for small field sizes [10]. The effects of mlc on beam characteristics for flattened and unflattened beam were also investigated in previous studies coating clear advantage of mlc over jaw to define treatment field [11]. Above-mentioned studies outline the possible profit of removing the flattening filter. It is therefore important to examine these properties for a typical modern accelerator such as Varian Clinic 600 unique performance. Our study reports on depth-dose dependencies, lateral profiles, total scatter factors, mlc leakage and various energy spectra in a conventional accelerator and a flattening filter-free system.
Material and Methods
Beam-line Simulation Model for 6 MV Varian Linac
Monte Carlo code system BEAMnrc [12,13] is used to model Varian
Clinic 600 unique performance in our study. To derive the best estimates for the mean energy and full width at half maximum (FWHM) of the electron
beam incident on the target, Monte Carlo simulations for monoenergetic beams ranging from 5.5 to 6.2 MeV with FWHM varied from 0.15 to 0.25 cm were
performed to find the best match with percentage depth dose (PDD) and profile measurements. A monoenergetic source with kinetic energy of the beam
5.7 MeV and FWHM for X and Y directions of 0.2 cm was found to give the best agreement with measured data. Geometry and materials used to build Monte
Carlo simulation model of the linear accelerator were based on machine specifications as provided by the manufacturer Varian Medical Systems. The linac
was structured in the following order: a target slab of tungsten and copper, primary collimator (tungsten), flattening filter, ion chamber, mirror,
jaws (tungsten) and finally the option for 120- leaf Varian MillenniumTM Multileaf Collimator. All materials used in the Monte Carlo (MC) simulation
were extracted from the 700 ICRU PEGS4 (pre-processor for Electron Gamma Shower) cross section data available in BEAMnrc, and met the specifications
for the linac as provided by the manufacturer. Different stages of simulation for 6 MV photon beam produced by Varian Linac using principal features
of the BEAMnrc-DOSXYZnrc code [14,15] are shown in Figure 1.
In the simulation of the full accelerator unit, we have split the calculation into three steps in order to save time. In the first step which
takes the most computing time, 1.5×108 initial histories are initiated and a monoenergetic electron beam source of kinetic energy of 5.7 MeV with FWHM for
the X and Y directions of 0.2 cm was incident on the target. The primary collimator, flattening filter and ion chamber are included in this step. The output
of this step is a phase space file at plain one as show in Figure 1, having information of energy, position, direction,
charge and history variable for every particle exiting downstream from the end of ion chamber. Since source and primary collimator have fixed openings,
it is possible to use this phase space data for the simulation of different field sizes. Figure 1 lists the component
module of BEAMnrc code used for modeling of fixed opening part of treatment head in the first step. This large set of particles produced in first step is used
repeatedly as the input to the next step of simulation. The second step of calculation simulates the passage of the particles through the mirror, adjustable
collimator, mlc and air slab to a plane at source to surface distance (SSD) 100 cm from target. We simulate different openings of jaw as well as mlc to get
field sizes from 5×5 to 20×20 cm2 at an SSD equal to 100 cm. For the latter case in mlc defined field sizes the projected jaw setting was 5 cm larger than
that of mlc. Figure 1 also lists the component module of BEAMnrc code used for modeling of variable opening part
of treatment head in second step of simulation. The output of this step is a phase space file at plain two as shown in Figure 1,
having information of energy, position, direction, charge and history variable for every particle reaching the plain at SSD 100 cm from target. The data analysis program
BEAMDP [16] is used to analyze the phase space data files to extract various types of spectra of all particles reaching
the plane at SSD 100 cm. In the third step of the simulation, the phase space files for field sizes of 5×5 to 20×20 cm2 at an SSD of 100 cm which are obtained
at the end of second step are reused by the DOSXYZnrc code as an input for dose calculations in a water phantom as shown in Figure 1.
We transport particles through a water phantom of dimension 30×30×30 cm3 with voxels size of 0.25×0.25×0.25 cm3. In the simulation of “unfiltered” 6MV photon beam,
all three steps of simulation are the same expect in first step where the flattening filter is being removed from the beam line. A comprehensive set of dosimetric
data for 6 MV filtered photon beams where acquired using a three-dimensional (3D) phantom, Blue phontom2 IBA Dosimetry GmbH and OmniPro-Accept 7 data acquisition
software. All the measurements were performed with a Scanditronix/ Wellhofer compact ionization chamber CC13, in the water phantom.
Figure1
6 MV Varian linac simulation model separated into three parts, Treatment head fixed and variable opening part representing first and second step of simulation modelled using component module of BEAMnrc code and Dose Calculation inside water phantom using DOSXYZnrc code in third step.
6 MV Varian linac simulation model separated into three parts, Treatment head fixed and variable opening part representing first and second step of simulation modelled using component module of BEAMnrc code and Dose Calculation inside water phantom using DOSXYZnrc code in third step.
Results
Validation of Monte Carlo Simulation Model
Depth-dose curves for filtered 6MV photon beam for field size 5×5 to 20×20 cm2 were calculated in an on axis cylinder of radius 1 cm using Monte Carlo simulation (MC)
and compared with measured data for the validation of simulation model. The calculated central axis depth-dose curves were normalized to unity at the depth, dmax,
of the maximum dose deposition and the Dmax. Both results measured and calculated, could then be compared with respect to the relative value of the maximum
dose Dmax and the corresponding depth dmax. Figure 2 shows the comparison between the calculated
depth-dose distributions and measurements of three different field sizes. The comparison shows that the calculated and measured data agree within 1% of local relative dose,
and 1 mm in depth at all depths and field sizes which are summarized in Table 1.
Figure2
Comparison of measured and calculated depth doses curves of 6MV photon beam for 20×20, 10×10 and 5×5 cm2 field sizes.
Table 1
Comparison of calculated and measured central-axis depth-dose profiles at various Field sizes. A denotes the field size, dmax (cm) denotes the location
of the maximum Dose, and ΔDmax is the relative dose difference between the measurement and the Calculations at dmax
A/cm2
dmax (simulated)
dmax (measured)
ΔDmax
5×5
1.5
1.56
0.2
10×10
1.5
1.52
0.17
15×15
1.48
1.5
0.13
20×20
1.38
1.40
0.10
Comparison of measured and calculated depth doses curves of 6MV photon beam for 20×20, 10×10 and 5×5 cm2 field sizes.Comparison of calculated and measured central-axis depth-dose profiles at various Field sizes. A denotes the field size, dmax (cm) denotes the location
of the maximum Dose, and ΔDmax is the relative dose difference between the measurement and the Calculations at dmaxLateral beam profiles for the filtered 6MV photon beam were also simulated for 5×5 to 20×20 cm2 field sizes at 1.5,5 and 10 cm depths.
The measured and calculated lateral dose profiles were normalized to unity on the central axis for comparison. Figure 3
shows the comparison of Monte Carlo calculations to measured data for a field size of 20×20, 10×10 and 5×5 cm2 at depth of 10 cm. The lateral
field size at the 50% dose level(X50) and penumbra widths, P90−10 and P80−20 (calculated from the 90% level to the 10% level and from 80% to 20%) where
calculated using Monte Carlo simulation and the results of the comparisons are summarized in Table 2. The differences
between the measurement and the simulations results in lateral field size at the 50% dose level, X50, was found to be less than 1 mm.
Figure3
Comparison of measured and calculated beam profiles of the 6MV photon beam at a Depth of 10 cm for 20×20, 10×10 and 5×5 cm2 field sizes.
Table 2
Comparison of measured and calculated lateral dose profiles at 10 cm depth. A Denotes the field Size, ΔX50 (mm) is the lateral
difference measured at the 50% Dose point in the Penumbra, and ΔP90-10(mm) as well as ΔP80-20(mm) describe the Difference
in width of the penumbra measured from the 90% point to 10% dose Point and from 80% to 20% Dose point respectively.
A/cm2
ΔX50
Δ P80-20
Δ P90-10
5×5
0.10
1.5
0.8
10×10
0.50
1.52
1.0
15×15
0.40
1.2
2.0
20×20
0.50
1.0
2.2
Comparison of measured and calculated beam profiles of the 6MV photon beam at a Depth of 10 cm for 20×20, 10×10 and 5×5 cm2 field sizes.Comparison of measured and calculated lateral dose profiles at 10 cm depth. A Denotes the field Size, ΔX50 (mm) is the lateral
difference measured at the 50% Dose point in the Penumbra, and ΔP90-10(mm) as well as ΔP80-20(mm) describe the Difference
in width of the penumbra measured from the 90% point to 10% dose Point and from 80% to 20% Dose point respectively.
Comparison of Flattened and Unflattened Beam Characteristics
Profile Comparison
Beam profiles for different field sizes were calculated at 1.5,5 and 10 cm depth for both cases with/without flattening filter in a water phantom.
The lateral profile for 20×20 cm2 and 10×10 cm2 field sizes at a depth of 10 cm are compared for two cases as shown
in Figures 4 and 5. For this comparison
of unflattened beam with flattened one, the flat profile is normalized to 1 on the central axis, and the nonflat profile is normalized by dose (D) which is calculated using this formula:
Figure4
Comparison of lateral profile for 6MV photon beam at 10 cm depth for field size 20×20 cm2. FF (MLC) denotes flattened beam shaped by mlc, FFF (MLC) and FFF (JAW) denotes unflattened beam shaped by mlc and jaw, respectively.
Figure5
Comparison of lateral profile for 6MV photon beam with/without a flattening filter in Beam line Shaped by mlc at depth 10 cm for field size 10×10 cm2. FF and FFF denotes calculation with/without a flattening filter in beam line.
Comparison of lateral profile for 6MV photon beam at 10 cm depth for field size 20×20 cm2. FF (MLC) denotes flattened beam shaped by mlc, FFF (MLC) and FFF (JAW) denotes unflattened beam shaped by mlc and jaw, respectively.Comparison of lateral profile for 6MV photon beam with/without a flattening filter in Beam line Shaped by mlc at depth 10 cm for field size 10×10 cm2. FF and FFF denotes calculation with/without a flattening filter in beam line.Where D is the dose at the inflection point of penumbra region of the unflattened beam, D is the dose at the inflection point of the flattened
profile and D is the dose on the central axis of the flattened beam [11]. It is observed in our study that
the beam profiles for unflattened case to have relative dose value lower than the flattened beam are close to the measured field size edge. However,
the amount of reduction in dose was less for mlc shaped in comparison to jaw shaped unflattened beam. For field size 20×20 cm2 measured
at 9 cm off axis distance reduction in relative dose for mlc and jaw shaped unflattened beam was found to be 12% and 15%, respectively. For field size
10×10 cm2 measured at 4 cm off axis distance reduction in relative dose for mlc shaped unflattened beam was found to be 8%, respectively.The dose in near field edge region for small field size shaped by mlc was investigated in our study for unflattened beam and compared with that of the
flattened beam. Figure 6 shows the calculated flattened and unflattened beam profiles for a small field size
(5×5 cm2) at a depth of 5 cm. The dose at 2 cm off-axis distance is lower in unflattened beams by 5% and it tends to decrease faster with
increasing off axis distance than in flattened beams. Faster lateral dose fall-off outside the treatment field will result in lower doses to surround normal tissues.
Figure6
Comparison of lateral dose profiles Shaped by mlc for a 5×5 cm2 field size at a depth of 5 cm. FF and FFF denotes calculation with/without a flattening filter in beam line.
Comparison of lateral dose profiles Shaped by mlc for a 5×5 cm2 field size at a depth of 5 cm. FF and FFF denotes calculation with/without a flattening filter in beam line.
Analysis of Spectra
Photon Energy Fluences Spectra
Figure 7 shows photon energy fluences as a function of off axis distance calculated for 20×20 cm2 field size. Photon emerging from target passes through the components of the collimating system on their way to the scoring plain at an SSD 100 cm. Scoring plain is an annular region around the central axis with radius 0 < r < 15 cm. The annular region has been divided into equal interval (bin) of 0.5 cm. The number of photon within each bin crossing the scoring plain is being recorded for with/without flattening filter case separately. The precision of calculated photon fluences spectra for all field sizes used in the dose calculations is high and uncertainty in each 0.5 cm wide bin is usually between 1 to 5%, except for the high-energy end of the spectra. There is a noticeable increase observed in the photon energy fluence when the flattening filter is removed from the beam line.
Figure7
Photon energy fluences per initial electron on the target, at the top of the water phantom as a function of off axis distance for 20×20 cm2 field size calculated for with/without a flattening filter in beam line.
Photon energy fluences per initial electron on the target, at the top of the water phantom as a function of off axis distance for 20×20 cm2 field size calculated for with/without a flattening filter in beam line.
Average Energy Distribution
Figure 8 shows the calculated photon average energies distribution at 100 cm SSD for 20×20 cm2 field
size as a function of off axis distance for with/without flattening filter case. From the above distribution beam hardening effect produced by
the flattening filter [8] for the filtered beam can be verified as we find that the mean photon energy for flattened beam to have a value at central axis 1.5 MeV and decreases to 1.12 MeV at off axis distance of 20 cm. In cases of unflattened beam, the mean energy of spectra did not change significantly with increasing off axis distance, and it respectively decreased from 1.2 MeV on central axis to 1.10 MeV at 20 cm off axis distance.
Figure8
Photon average energy distribution of the filtered and unfiltered 6-MV beams as a function of the off axis distance for 20×20 cm2 field size.
Photon average energy distribution of the filtered and unfiltered 6-MV beams as a function of the off axis distance for 20×20 cm2 field size.
Contaminant Electron Energy Fluence Spectra
Figure 9 shows the calculated electron energy fluence spectra
for contaminant electrons calculated as a function of off axis distance for 20×20 cm2 field size at 100 cm SSD for with/without
flattening filter case separately. Number of electron reaching the phantom surface is found to increase with removing the flattening filter
from the beam line. However, the difference at near center is higher and tends to decrease as the off axis distances increase between the two cases.
Monte Carlo calculation demonstrated that the electron fluence at the center for a 6 MV flattening filter free beam with the same electron beam hitting
the target was found to be 1.2 times greater than its value for with flattening filter case for field size 20×20 cm2. This increase in electron
energy fluence indicates a potential risk of delivering a high skin dose to the patient and also the risk of placing ion chamber used for the measurement
outside the range of its reliable operation.
Figure9
Electron energy fluences per initial electron on target, at the top of the water phantom as a function of off axis distance for 20×20 cm2 field size calculated for with/without a flattening filter in beam line.
Electron energy fluences per initial electron on target, at the top of the water phantom as a function of off axis distance for 20×20 cm2 field size calculated for with/without a flattening filter in beam line.
Depth Dose Analysis
Absolute Dose: For comparison purposes, absolute absorbed doses per initial electron on target were calculated for flattened and unflattened beam at two reference depths of 1.5 and 10 cm. The ratio of absolute depth doses for flattening filter free to standard flattened beams were calculated and are presented in Table 3. A significant increase in absorbed dose was observe by removing flattening filter, indicating an increased in dose rate for unflattened beam. However, the increase in dose rate decreases with increasing depth.
Table 3
Ratios of absolute depth doses for flattening filter free to flattened beams at two reference depths for different field sizes.
A denotes the field size; d denotes the depth inside water phantom. Absorbed dose calculated without the flattening filter in the beam
line is denoted as DFFF (flattening filter free) and with filter in beam line is denoted as DFF.
A/cm2
(DFFFDFF) At d=1.5
(DFFFDFF) At d=10
5×5
2.49
2.42
10×10
2.47
2.45
20×20
2.44
2.40
Ratios of absolute depth doses for flattening filter free to flattened beams at two reference depths for different field sizes.
A denotes the field size; d denotes the depth inside water phantom. Absorbed dose calculated without the flattening filter in the beam
line is denoted as DFFF (flattening filter free) and with filter in beam line is denoted as DFF.Percentage Depth-Dose Characteristics: Absolute depth dose values were used to calculate percentage depth-dose characteristic (PDD)
curves. It can be seen from Figure 10 that unflattened beam shows slightly lower PDDs values in comparison
with the standard beam for all field sizes. Difference in PDDs between the two cases is evident at deeper depths and increasedswith depth
for 10×10 and 20×20 cm2 field sizes. This difference is validated by calculating two parameters which are reported in Table 4,
namely, the relative dose at a depth of 10 and 20 cm (D10, D20)
Figure10
Comparison of relative depth dose curves calculated for with and without flattening filter for 6MV photon beams for different field sizes: (a)10×10 cm2 (b) 20×20 cm2
Table 4
Comparison of relative depth doses for flattening filter free to standard flattened beams at two reference depths for different Field sizes.
A denotes the field size; D10 and D20 denote relative depth dose at 10 and 20 cm depth.
A/cm2
D10
D20
With FF
Without FF
With FF
Without FF
5×5
62.43
59.40
33.26
30.84
10×10
66.70
63.80
37.80
34.34
20×20
71.65
68.96
40.98
37.55
Comparison of relative depth dose curves calculated for with and without flattening filter for 6MV photon beams for different field sizes: (a)10×10 cm2 (b) 20×20 cm2Comparison of relative depth doses for flattening filter free to standard flattened beams at two reference depths for different Field sizes.
A denotes the field size; D10 and D20 denote relative depth dose at 10 and 20 cm depth.
MLC Linkage
MLC leakage is an important parameter needed for the commissioning of a treatment-planning system. We calculated the MLC leakage as a function
of field size for the unflattened profile in our study as presented in Table 5. MLC leakage represents the dose on the central beam axis with MLC blocked fields normalized by the dose of open fields of the same field size at 1.5 cm depth for SSD 100 cm. Open MLC fields are defined as MLC leaves are withdrawn underneath the jaws as not to intercept the beam, the field size is defined by the treatment jaws. MLC blocked fields define a field in which the MLC leaves are configured to fully block the open field produced by the jaw. To ensure that the jaws blocked the rounded tips of the leaves completely in MLC blocked fields the leaves of MLC were positioned asymmetrically with respect to the central axis, and there projected offset at isocenter was 8.0 cm.
Table 5
Calculated MLC leakage for 6 MV photon beam delivered with or without flattening Filter in beam line for different field sizes. Calculations were made at 1.5 cm depth and SSD 100 cm.
Field size
MLC LINKAGE
With flattening filter
Without flattening filter
5×5
-
1.10
10×10
1.40
1.23
20×20
-
1.32
Calculated MLC leakage for 6 MV photon beam delivered with or without flattening Filter in beam line for different field sizes. Calculations were made at 1.5 cm depth and SSD 100 cm.
Scatter Function
The total scatter factor, SCP is defined as ‘the dose rate at a reference depth for a given field size divided by the dose rate at the
same point and depth for the reference field size (10×10 cm2). It was measured at SSD = 100 cm and a depth equal to dmax of a 10×10 cm2 field
for different field sizes. The data for with/without flattening filter case are presented in Table 6.
Table 6
Total scatter factor Scp of 6 MV photon beams measured for with / without a flattening filter cases. The Scp was measured at SSD = 100 cm, and at the depth of maximum dose dmax of a 10×10 cm2 field size.
Field size
Scp (MLC shaped with flattening filter)
Scp (JAW shaped without flattening filter)
Scp (MLC shaped without flattening filter)
5×5
0.967
0.97
0.98
10×10
1
1
1
15×15
1.021
1.012
1.010
20×20
1.054
1.027
1.018
Total scatter factor Scp of 6 MV photon beams measured for with / without a flattening filter cases. The Scp was measured at SSD = 100 cm, and at the depth of maximum dose dmax of a 10×10 cm2 field size.The unflattened beams are found to have less value of SCP for lager field sizes in comparison wuth flattened beams indicating a reduced head scatter in unflattened
beams compared with the standard flattened beam.
Discussion
Flattening filter is a main source of absorption of majority portion of primary photons particularly of those which are close to central axis of beam, thus removing filter from the beam line should result in considerable increase in dose rate, and therefore a decrease in beam-on time should be achieved when radiation treatment is delivered. The ratios of absolute depth doses calculated in our study for flattening filter free to standard flattened beams for field size 10×10 cm2, at 10 cm depth for an SSD equal to 100 cm was found to be 2.45 signifying the potential higher dose rate delivered by the unflattened beam. Unflattened beam is found to have slightly lower PDDs value compared to the standard beam for all field sizes. Difference in the PDDs of flattened and unflattened beams are apparent at deeper depths and increase with depth for all field sizes. The photon energy fluence and average energy distribution as a function of off axis distance for flattened and unflattened beams are being calculated in our study. It was observed that the fluence of photon on central axis increased nearly two times with removing flattening filter, and with increasing off axis distance this difference reduced. The average energy of photon spectrum on central axis decreased from 1.5 to 1.2 MeV for 20×20 cm2 field size at 100 cm SSD with removing flattening filter. The average energy of photon for flattened beam was found to decrease with increasing off axis distance from 1.5 MeV on central axis to 1.12 MeV at 20 cm off axis distance for 20×20 cm2 field size. Whereas for flattening filter free beam, the mean energy of spectra did not change significantly with increasing off axis distance, and it respectively decreased from 1.2 MeV on central axis to 1.1 MeV at 20 cm off axis distance for 20×20 cm2 field size. The differential attenuation produced by flattening filter with increasing off axis distance from central axis of beam is the possible explanation for this behavior. As the central thick part of flattening filter attenuates more low energy photons, and as off axis distance increases, more low energy photons are allowed to penetrate the thin lateral part of the flattening filter contributing to the photon energy fluence; thus the mean energy of spectra decreases significantly with increasing distance for the flattened beam. In our study, we have calculated the mlc leakage for both cases with or without flattening filter in the beam line. It was observed that there is a considerable decrease in mlc leakage when the flattening filter was removed from the beam line, as for 10×10 cm2 field size, its value was 1.4 which decreases to 1.23 with filter removed from the beam line. The average energy difference on the central axis is considered to be the major reason for this decrease. As the filter is removed from the beam line, the average energy of photon beam decreases causing more attenuation of the photon beam by mlc. The total scatter factor, SCP for the unflattened and flattened beam has been investigated in our study. It can be seen from the data that the flattening filter free beam SCP increases more slowly with increasing field size than that of the flattened beam. In addition, the amount of variation in SCP was even less for mlc shaped in comparison with jaw shaped unflattened beam; this is due to the forward-peaked profile of unflattened beam, which produces less SCP because of the reduced off-axis intensity. The flattening filter free beam has greatly reduced fluence off axis; hence, less secondary head scatter is created, which is directed towards the central axis. Due to this reason, as the measured field size increases, the expected increase in SCP for flattening filter free beams is not seen which is found with the flattened ones. Our study showed that flattened and unflattened beams are comparable within a few centimetres from the central axis; therefore, unflattened beams are unlikely to present a problem for treatments with small fields. Moreover, the treatments can also profit from an increased dose rate. The beam profiles for unflattened cases are found to have lower relative dose value in comparison with flattened beam close to the field edge. The key reason for this behavior is that the flattening filter elevates relative fluence of primary photons propagating off-axis and reduces head scatter present in unflattened beams. The near field edge dose calculated without the flattening filter is found to be smaller for small field sizes when compared with the flattened beams; it falls off faster with distance. This means that an improved sparing of normal tissues close to small treatment fields can be achieved.
Conclusion
We developed a Monte Carlo simulation model for a 6MV photon beam based on Varian Clinic 600 unique performance linac and calculated basic dosimetric features which were benchmarked against measurement. It was then used to investigate the basic dosimetric properties of photon beams generated by the accelerator with the flattening filter removed. Comparison of dosimetric features of flattened and unflattened beams revealed that beam profiles for unflattened cases were found to have faster rate of decline and lower dose value. Mlc leakage calculated for both cases showed a considerable decrease for unflattened beam which is due to the differences in average energy on the central axis for the two cases. The total scatter factor, SCP less variation with field sizes indicate that removing the filter from the beam line could significantly reduce the amount of head scatter photons and therefore doses to normal tissues and organs.