| Literature DB >> 31208342 |
François Regoli1, Denis Graf2, Beat Schaer3, Firat Duru4, Peter Ammann5, Lorenza Mangoni di S Stefano6, Barbara Naegli7, Haran Burri8, Rainer Zbinden9, Nazmi Krasniqi10, Martin Fromer11.
Abstract
BACKGROUND: Little is known about the ICD performance using enhanced detection algorithms in unselected, non-trial patients. Performance of recent generation ICD equipped with SmartShock™ technology (SST) for detection and conversion of ventricular tachyarrhythmias (VTA) was investigated.Entities:
Keywords: Antitachycardia pacing; Implantable cardioverter-defibrillator; Shock; Ventricular tachyarrhythmia
Year: 2019 PMID: 31208342 PMCID: PMC6580638 DOI: 10.1186/s12872-019-1121-4
Source DB: PubMed Journal: BMC Cardiovasc Disord ISSN: 1471-2261 Impact factor: 2.298
Fig. 1Patient-flow diagram
Baseline characteristics in the 4P-study
| SC-ICD ( | DC-ICD ( | CRT-D ( | All patients ( | ||
|---|---|---|---|---|---|
| Patient demographics and clinical presentation | |||||
| Male (%) | 81 | 88 | 84 | 84 | 0.645 |
| Age (years ± SD) | 59 ± 13 | 63 ± 14 | 68 ± 11 | 64 ± 13 | 0.0012 |
| LVEF (% ± SD) | 44 ± 14 | 42 ± 14 | 31 ± 10 | 36 ± 13 | < 0.0011,2 |
| QRS (ms ± SD) | 107 ± 26 | 112 ± 30 | 143 ± 35 | 126 ± 36 | < 0.0011,2 |
| Secondary prevention (%) | 58 | 48 | 17 | 35 | < 0.0011,2 |
| Device replacement or upgrade (%) | 63 | 41 | 61 | 56 | 0.039 |
| NYHA functional class (%) | |||||
| I | 17 | 18 | 10 | 14 | < 0.0011,2 |
| II | 19 | 24 | 50 | 34 | |
| III | 6 | 8 | 39 | 21 | |
| IV | 2 | 0 | 1 | 1 | |
| Underlying cardiac disease (%) | |||||
| Ischemic heart disease, leading diagnosis | 60 | 66 | 58 | 61 | 0.652 |
| CABG | 21 | 18 | 30 | 25 | 0.239 |
| PCI | 27 | 44 | 37 | 36 | 0.217 |
| Cardiomyopathy, other | 25 | 26 | 27 | 26 | 0.668 |
| Arrhythmia and conduction defects (%) | |||||
| Non permanent atrial fibrillation | 15 | 24 | 31 | 25 | 0.108 |
| Atrial tachycardia | 4 | 14 | 10 | 10 | 0.251 |
| Ventricular fibrillation | 27 | 12 | 13 | 16 | 0.066 |
| Ventricular tachycardia (sustained) | 25 | 34 | 8 | 19 | 0.0031,2 |
| Ventricular tachycardia (non-sustained) | 21 | 26 | 16 | 20 | 0.372 |
| AV-block | 10 | 10 | 28 | 19 | 0.0082 |
| Medication (%) | |||||
| Beta-blocker | 91 | 85 | 90 | 89 | 0.333 |
| Calcium channel blocker | 18 | 11 | 6 | 10 | 0.129 |
| Digoxin | 0 | 2 | 13 | 7 | 0.0042 |
| Anti-arrhythmics | 25 | 34 | 29 | 29 | 0.599 |
| of which amiodarone | 14 | 28 | 25 | 23 | 0.187 |
| Implantation type (%) | |||||
| New | 38 | 59 | 39 | 44 | < 0.0011,2 |
| Replacement | 62 | 37 | 33 | 41 | |
| Upgrade | 0 | 4 | 28 | 15 | |
Significant post-hoc comparisons are indicated as: 1 SC-ICD vs CRT-D; 2 DC-ICD vs CRT-D
Fig. 2Sequential management of ventricular tachyarrhythmia (VTAs) by ICD devices, from the detection to the electrical therapy of categorized VTAs
Device-based treatment of ventricular tachyarrhythmia episodes, as number of episodes (number of patients) having experienced one or more episode
| Delivered therapy (%) | VT (91.1%) | FVT (1.3%) | VF (7.6%) | Total |
|---|---|---|---|---|
| ATP only (94.8%) | 1146 (23) | 11 (4) | 51 (11) | 1208 (31) |
| SC-ICD | 327 (11) | 7 | 22 | 356 |
| DC-ICD | 739 (4) | 2 | 1 | 742 |
| CRT-D | 80 (8) | 2 | 28 | 110 |
| ATP + shock (0.3%) | 0 (0) | 0 (0) | 4 (1) | 4 (1) |
| SC-ICD | 0 (0) | 0 (0) | 0 (0) | 0 (0) |
| DC-ICD | 0 (0) | 0 (0) | 0 (0) | 0 (0) |
| CRT-D | 0 (0) | 0 (0) | 4 (1) | 4 (1) |
| Shock only (4.9%) | 15 (7) | 5 (4) | 42 (19) | 62 (27) |
| SC-ICD | 6 (3) | 0 (0) | 8 (5) | 14 (8) |
| DC-ICD | 1 (1) | 2 (1) | 9 (5) | 12 (7) |
| CRT-D | 8 (5) | 3 (2) | 25 (9) | 36 (12) |
| Total | 1161 (26) | 16 (6) | 97 (25) | 1274 (44) |
Death and Hospitalization rates per 100 patient-years (95%CI) by cause and device type
| SC-ICD | DC-ICD | CRT-D | Overall | |
|---|---|---|---|---|
| Death | 2.6 (0.8–8.3) | 3.5 (1.3–9.4) | 5.3 (3.0–9.4) | 4.2 (2.6–6.6) |
| Any hospitalization | 16.0 (10.1–25.4) | 36.1* (26.6–49.1) | 28.3 (22.2–36.2) | 27.2 (22.8–32.5) |
| Non-cardiac hospitalization | 7.1 (3.6–14.2) | 17.6 † (11.4–27.3) | 14.2 †† (10.0–20.0) | 13.3 (10.3–17.1) |
| Cardiac hospitalization | 8.9 (4.8–16.5) | 18.5 ** (12.1–28.4) | 14.2 (10.0–20.0) | 13.9 (10.9–17.8) |
| Hospitalization for arrhythmia | 1.8 (0.4–7.1) | 7.0 # (3.5–14.1) | 5.8 ## (3.3–9.9) | 5.1 (3.4–7.7) |
| Ventricular pacing % | 6.9 (0–29.2) | 11.3 (0–41.1) | 96.8 (92.8–100.0) | / |
Mean values with corresponding standard deviation between parenthesis
* p = 0.010 vs. SC-ICD
† p = 0.03 vs. SC-ICD; †† p = 0.045 vs. SC-ICD
** p = 0.04 vs. SC-ICD
# p = 0.017 vs. SC-ICD; ## p = 0.043 vs. SC-ICD
Fig. 3Kaplan-Meier curves for survival (a) and freedom of hospitalization (b)
Fig. 4Last generation ICDs are primarily ATP devices with defibrillation backup (rescue shock)
Details on the ICD programming divergences from the recommended settings (presented as Figures 5 and 6 in Appendix)
| TOTAL | SC-ICD | DC-ICD | CRT-D | |
|---|---|---|---|---|
| Primary prevention (pts) | 121 | 19 | 25 | 77 |
| FVT detection- ON [n, (%)] | 19 (10.2) | 1 (2.3) | 4 (8.3) | 14 (15.1%) |
| VT detection- ON [ | 90 (74.4) | 15 (78.9) | 18 (72.0) | 57 (64.0) |
| PR Logic/ Wavelet- OFF or Monitor [ | 18 (14.9) | 3 (15.8) | 2 (8.0) | 13 (16.9) |
| Onset ON or OFF [ | 49 (40.5) | 12 (53.1) | 9 (36.0) | 28 (36.4) |
| T-Wave- OFF [ | 6 (5) | 3 (15.8) | 0 | 3 (3.9) |
| RV-Lead Noise [ | 3 (2.5) | 2 (10.5) | 0 | 1 (1.3) |
| ATP before charging | 8 (6.6) | 1 (5.3) | 2 (8.0) | 5 (6.5) |
| FVT- ATP & shocks | 121 (100) | 19 (100) | 25 (100) | 77 (100) |
| VT- ATP & shocks | 121 (100) | 19 (100) | 25 (100) | 77 (100) |
| Secondary prevention (pts) | 64 | 25 | 23 | 16 |
| FVT detection- ON [ | 19 (29.7) | 8 (32.0) | 9 (39.1) | 2 (12.6) |
| VT detection- OFF [ | 11 (17.2) | 6 (24.0) | 5 (21.7) | 0 |
| PR Logic/ Wavelet- OFF or Monitor [ | 4 (6.3) | 3 (12.0) | 0 | 1 (6.3) |
| Onset ON or OFF [ | 32 (50.0) | 10 (40.0) | 14 (60.8) | 8 (50.0) |
| ATP before charging | 8 (12.5) | 3 (12.0) | 2 (8.7) | 3 (18.8) |
| FVT- ATP & shocks | 121 (100) | 19 (100) | 25 (100) | 77 (100) |
Classification of 176 symptomatic VTA episodes in 39 patients
| VT ( | FVT ( | VF ( | Total | |
|---|---|---|---|---|
| ATP | 125 | 0 | 0 | 125 |
| ATP and shock | 11 | 1 | 1 | 13 |
| Shock only | 11 | 1 | 9 | 21 |
| Self-terminating | 10 | 5 | 2 | 17 |
| Total episodes | 157 | 7 | 12 | 176 |