Improving SVT discrimination in single-chamber ICDs: a new electrogram morphology-based algorithm.

George J Klein1, Jeffrey M Gillberg, Anthony Tang, Shmuel Inbar, Arjun Sharma, Christina Unterberg-Buchwald, Paul Dorian, Hans Moore, Firat Duru, Ethan Rooney, Daniel Becker, Katie Schaaf, David Benditt.   

Abstract

INTRODUCTION: Wide-spread adoption of ICD therapy has focused efforts on improving the quality of life for patients by reducing "inappropriate" shock therapies. To this end, distinguishing supraventricular tachycardia from ventricular tachycardia remains a major challenge for ICDs. More sophisticated discrimination algorithms based on ventricular electrogram morphology have been made practicable by the increased computational ability of modern ICDs. METHODS AND
RESULTS: We report results from a large prospective study (1,122 pts) of a new ventricular electrogram morphology tachycardia discrimination algorithm (Wavelet Dynamic Discrimination, Medtronic, Minneapolis, MN, USA) operating at minimal algorithm setting (RV coil-can electrogram, match threshold of 70%). This is a nonrandomized cohort study of ICD patients using the morphology discrimination of the Wavelet algorithm to distinguish SVT and VT/VF. The Wavelet criterion was required ON in all patients and all other supraventricular tachycardia discriminators were required to be OFF. Spontaneous episodes (N = 2,235) eligible for ICD therapy were adjudicated for detection algorithm performance. The generalized estimating equations method was used to remove bias introduced when an individual patient contributes multiple episodes. Inappropriate therapies for supraventricular tachycardia were reduced by 78% (90% CI: 72.8-82.9%) for episodes within the range of rates where Wavelet was programmed to discriminate. Sensitivity for sustained ventricular tachycardia was 98.6% (90% CI: 97-99.3%) without the use of high-rate time out.
CONCLUSIONS: Results from this prospective study of the Wavelet electrogram morphology discrimination algorithm operating as the sole discriminator in the ON mode demonstrate that inappropriate therapy for supraventricular tachycardia in a single-chamber ICD can be dramatically reduced compared to rate detection alone.

Entities:  

Mesh:

Year:  2006        PMID: 17096661     DOI: 10.1111/j.1540-8167.2006.00643.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  11 in total

Review 1.  Strategic choices to reduce implantable cardioverter-defibrillator-related morbidity.

Authors:  Oussama Wazni; Bruce L Wilkoff
Journal:  Nat Rev Cardiol       Date:  2010-04-20       Impact factor: 32.419

2.  Long-term incidence of inappropriate shocks in patients with implantable cardioverter defibrillators in clinical practice-an underestimated complication?

Authors:  Daniel Hofer; Jan Steffel; David Hürlimann; Laurent Haegeli; Thomas F Lüscher; Firat Duru; Urs Eriksson; Nazmi Krasniqi
Journal:  J Interv Card Electrophysiol       Date:  2017-11-25       Impact factor: 1.900

3.  Reduction of the inappropriate ICD therapies by implementing a new fuzzy logic-based diagnostic algorithm.

Authors:  Michał Lewandowski; Andrzej Przybylski; Wiesław Kuźmicz; Hanna Szwed
Journal:  Ann Noninvasive Electrocardiol       Date:  2013-09       Impact factor: 1.468

4.  Optimal Programming of ICDs for Prevention of Appropriate and Inappropriate Shocks.

Authors:  Ronald Lo; Amin Al-Ahmad; Henry Hsia; Paul C Zei; Paul J Wang
Journal:  Curr Treat Options Cardiovasc Med       Date:  2008-09

Review 5.  Sensing and detection in Medtronic implantable cardioverter defibrillators.

Authors:  Mark L Brown; Charles D Swerdlow
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2016-09-08

6.  Dynamic analysis of cardiac rhythms for discriminating atrial fibrillation from lethal ventricular arrhythmias.

Authors:  Deeptankar DeMazumder; Douglas E Lake; Alan Cheng; Travis J Moss; Eliseo Guallar; Robert G Weiss; Steven R Jones; Gordon F Tomaselli; J Randall Moorman
Journal:  Circ Arrhythm Electrophysiol       Date:  2013-05-16

7.  Inappropriate Detection of a Supraventricular Tachycardia as Dual Tachycardia by the PR Logic™ Algorithm.

Authors:  Ajit Thachil; Sridevi Chennapragada; Narasimhan Calambur
Journal:  Indian Pacing Electrophysiol J       Date:  2014-05-25

8.  2015 HRS/EHRA/APHRS/SOLAECE expert consensus statement on optimal implantable cardioverter-defibrillator programming and testing.

Authors:  Bruce L Wilkoff; Laurent Fauchier; Martin K Stiles; Carlos A Morillo; Sana M Al-Khatib; Jesœs Almendral; Luis Aguinaga; Ronald D Berger; Alejandro Cuesta; James P Daubert; Sergio Dubner; Kenneth A Ellenbogen; N A Mark Estes; Guilherme Fenelon; Fermin C Garcia; Maurizio Gasparini; David E Haines; Jeff S Healey; Jodie L Hurtwitz; Roberto Keegan; Christof Kolb; Karl-Heinz Kuck; Germanas Marinskis; Martino Martinelli; Mark McGuire; Luis G Molina; Ken Okumura; Alessandro Proclemer; Andrea M Russo; Jagmeet P Singh; Charles D Swerdlow; Wee Siong Teo; William Uribe; Sami Viskin; Chun-Chieh Wang; Shu Zhang
Journal:  J Arrhythm       Date:  2016-02-01

9.  Influence of myopotential interference on the Wavelet discrimination algorithm in implantable cardioverter-defibrillator.

Authors:  Kazuya Mizukami; Hisashi Yokoshiki; Hirofumi Mitsuyama; Masaya Watanabe; Taro Tenma; Rui Kamada; Masayuki Takahashi; Ryo Sasaki; Motoki Maeno; Hiroyuki Tsutsui
Journal:  J Arrhythm       Date:  2016-10-01

10.  Arrhythmic episodes in patients implanted with a cardioverter-defibrillator - results from the Prospective Study on Predictive Quality with Preferencing PainFree ATP therapies (4P).

Authors:  François Regoli; Denis Graf; Beat Schaer; Firat Duru; Peter Ammann; Lorenza Mangoni di S Stefano; Barbara Naegli; Haran Burri; Rainer Zbinden; Nazmi Krasniqi; Martin Fromer
Journal:  BMC Cardiovasc Disord       Date:  2019-06-17       Impact factor: 2.298

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.