Effect of long-detection interval vs standard-detection interval for implantable cardioverter-defibrillators on antitachycardia pacing and shock delivery: the ADVANCE III randomized clinical trial.

Maurizio Gasparini1, Alessandro Proclemer, Catherine Klersy, Axel Kloppe, Maurizio Lunati, José Bautista Martìnez Ferrer, Ahmad Hersi, Marcin Gulaj, Maurits C E F Wijfels, Elisabetta Santi, Laura Manotta, Angel Arenal.   

Abstract

IMPORTANCE: Using more intervals to detect ventricular tachyarrhythmias has been associated with reducing unnecessary implantable cardioverter-defibrillator (ICD) therapies.
OBJECTIVE: To determine whether using 30 of 40 intervals to detect ventricular arrhythmias (VT) (long detection) during spontaneous fast VT episodes reduces antitachycardia pacing (ATP) and shock delivery more than 18 of 24 intervals (standard detection). DESIGN, SETTING, AND PARTICIPANTS: Randomized, single-blind, parallel-group trial that enrolled 1902 primary and secondary prevention patients (mean [SD] age, 65 [11] years; 84% men; 75% primary prevention ICD) with ischemic and nonischemic etiology undergoing first ICD implant at 1 of 94 international centers (March 2008-December 2010).
INTERVENTIONS: Patients were randomized 1:1 to programming with long- (n = 948) or standard-detection (n = 954) intervals. MAIN OUTCOMES AND MEASURES: Total number of ATPs and shocks delivered for all episodes (primary outcomes) and inappropriate shocks, mortality, and syncopal rate (secondary outcomes).
RESULTS: During a median follow-up of 12 months (interquartile range, 11-13), long-detection group had 346 delivered therapies (42 therapies per 100 person-years, 95% CI, 38-47) vs 557 in the standard-detection group (67 therapies per 100 person-years [95% CI, 62-73]; incident rate ratio [IRR], 0.63 [95% CI, 0.51-0.78]; P < .001). The long- vs the standard-detection group experienced 23 ATPs per 100 person-years (95% CI, 20-27) vs 37 ATPs per 100 person-years (95% CI, 33-41; IRR, 0.58 [95% CI, 0.47-0.72]; P < .001); 19 shocks per 100 person-years (95% CI, 16-22) vs 30 shocks per 100 person-years (95% CI, 26-34; IRR, 0.77 [95% CI, 0.59-1.01]; P = .06), with a significant difference in the probability of therapy occurrence (P < .001); and a reduction in first occurrence of inappropriate shock (5.1 per 100 patient-years [95% CI, 3.7-6.9] vs 11.6 [95% CI, 9.4-14.1]; IRR, 0.55 [95% CI, 0.36-0.85]; P = .008). Mortality (5.5 [95% CI, 4.0-7.2] vs 6.3 [95% CI, 4.8-8.2] per 100 patient-years; HR, 0.87; P = .50) and arrhythmic syncope rates (3.1 [95% CI, 2.6-4.6] vs 1.9 [95% CI, 1.1-3.1] per 100 patient-years; IRR, 1.60 [95% CI, 0.76-3.41]; P = .22) did not differ significantly between groups. CONCLUSIONS AND RELEVANCE: Among patients receiving an ICD, the use of a long- vs standard-detection interval resulted in a lower rate of ATP and shocks, and inappropriate shocks. This programming strategy may be an appropriate alternative. TRIAL REGISTRATION: clinicaltrials.gov Identifier: NCT00617175.

Entities:  

Mesh:

Year:  2013        PMID: 23652522     DOI: 10.1001/jama.2013.4598

Source DB:  PubMed          Journal:  JAMA        ISSN: 0098-7484            Impact factor:   56.272


  89 in total

1.  Device therapy: ADVANCEs in ICD programming.

Authors:  Bryony M Mearns
Journal:  Nat Rev Cardiol       Date:  2013-05-28       Impact factor: 32.419

Review 2.  Roles and indications for use of implantable defibrillator and resynchronization therapy in the prevention of sudden cardiac death in heart failure.

Authors:  Yitschak Biton; Jayson R Baman; Bronislava Polonsky
Journal:  Heart Fail Rev       Date:  2016-07       Impact factor: 4.214

3.  2019 HRS/EHRA/APHRS/LAHRS expert consensus statement on catheter ablation of ventricular arrhythmias.

Authors:  Edmond M Cronin; Frank M Bogun; Philippe Maury; Petr Peichl; Minglong Chen; Narayanan Namboodiri; Luis Aguinaga; Luiz Roberto Leite; Sana M Al-Khatib; Elad Anter; Antonio Berruezo; David J Callans; Mina K Chung; Phillip Cuculich; Andre d'Avila; Barbara J Deal; Paolo Della Bella; Thomas Deneke; Timm-Michael Dickfeld; Claudio Hadid; Haris M Haqqani; G Neal Kay; Rakesh Latchamsetty; Francis Marchlinski; John M Miller; Akihiko Nogami; Akash R Patel; Rajeev Kumar Pathak; Luis C Saenz Morales; Pasquale Santangeli; John L Sapp; Andrea Sarkozy; Kyoko Soejima; William G Stevenson; Usha B Tedrow; Wendy S Tzou; Niraj Varma; Katja Zeppenfeld
Journal:  J Interv Card Electrophysiol       Date:  2020-10       Impact factor: 1.900

4.  Reduction of inappropriate ICD therapies in patients with primary prevention of sudden cardiac death: DECREASE study.

Authors:  Jörg Otto Schwab; Hendrik Bonnemeier; Thomas Kleemann; Johannes Brachmann; Sven Fischer; Frank Birkenhauer; Frank Eberhardt
Journal:  Clin Res Cardiol       Date:  2015-05-23       Impact factor: 5.460

Review 5.  The subcutaneous ICD-current evidence and challenges.

Authors:  Kiran Haresh Kumar Patel; Pier D Lambiase
Journal:  Cardiovasc Diagn Ther       Date:  2014-12

Review 6.  Electrical Storm: Current Evidence, Clinical Implications, and Future Perspectives.

Authors:  Christos Kontogiannis; Konstantinos Tampakis; Georgios Georgiopoulos; Stefano Bartoletti; Christos Papageorgiou; Hector Anninos; Alkistis Kapelouzou; Michael Spartalis; Ioannis Paraskevaidis; Sofia Chatzidou
Journal:  Curr Cardiol Rep       Date:  2019-07-27       Impact factor: 2.931

Review 7.  ICD programming to reduce shocks and improve outcomes.

Authors:  Valentina Kutyifa; Wojciech Zareba; Arthur J Moss
Journal:  Curr Cardiol Rep       Date:  2014       Impact factor: 2.931

Review 8.  Ventricular Tachycardia with ICD Shocks: When to Medicate and When to Ablate.

Authors:  Amir AbdelWahab; John Sapp
Journal:  Curr Cardiol Rep       Date:  2017-09-13       Impact factor: 2.931

Review 9.  Optimal tachycardia programming in ICDs : Recommendations in the post-MADIT-RIT era.

Authors:  Carsten W Israel; Tatsiana Burmistrava
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2016-09

Review 10.  Sensing and detection in Medtronic implantable cardioverter defibrillators.

Authors:  Mark L Brown; Charles D Swerdlow
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2016-09-08
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.