Literature DB >> 3119784

A small bypass mixing chamber for monitoring metabolic rate and anesthetic uptake: the bymixer.

Y Sanjo1, K Ikeda.   

Abstract

We developed a miniature mixing chamber, the bymixer, that aids in the measurement of respiratory gas metabolism and inhalation anesthetic uptake. A small fraction of total respiratory gas flow is bypassed from the inspiratory or expiratory limb of the breathing circuit to the bymixer. We tested the relationship between total flow and bypass flow. To analyze the error and response time of the system, we compared the mixed expired carbon dioxide from the bymixer (capacity, 0.3 L, ratio of bypass flow to total flow, 1/4.5) with that from a conventional 2.0-L mixing chamber in 15 volunteers and 12 anesthetized patients during spontaneous or controlled ventilation. Bypass flow correlated well with total flow (r2 = 0.999 to 1.000) when total flow ranged from 0 to 70 L/min and the ratio of bypass flow to total flow ranged from 1/3 to 1/20. The difference between the values of mixed expired carbon dioxide from the two mixing chambers was small, ranging from -0.03 to 0.01 vol% in both of the ventilatory modes. The relative error was within 2.3% of the carbon dioxide values obtained from the conventional chamber. We observed the error at the lowest minute volume (4 L/min). The 90% response time of the bymixer (24.8 seconds) was similar to that of the conventional chamber (19.1 seconds) at a minute volume of 7.5 L/min. For clinical use, we combined a conventional breathing circuit with two bymixers, one for mixing inspired gas and the other for mixing expired gas. We found this device accurate and easy to use.

Entities:  

Mesh:

Substances:

Year:  1987        PMID: 3119784     DOI: 10.1007/bf03337377

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  7 in total

1.  Frequency response of long mass-spectrometer sampling catheters.

Authors:  F L Scamman; J K Fishbaugh
Journal:  Anesthesiology       Date:  1986-10       Impact factor: 7.892

2.  The measurement of metabolic gas exchange and minute volume by mass spectrometry alone.

Authors:  N J Davies; D M Denison
Journal:  Respir Physiol       Date:  1979-02

3.  An oxygen replenishment technique for the continuous monitoring of oxygen uptake rate.

Authors:  W J Jones
Journal:  J Assoc Adv Med Instrum       Date:  1971 Sep-Oct

4.  [Soft mixing chamber technique for continuous monitoring of Vo2, Vco2, and RQ (author's transl)].

Authors:  Y Sanjo; K Ikeda; I Matsuda
Journal:  Kokyu To Junkan       Date:  1980-02

5.  The uses of long sampling probes in respiratory mass spectrometry.

Authors:  N J Davies; D M Denison
Journal:  Respir Physiol       Date:  1979-08

6.  A systematic method for validation of gas exchange measurements.

Authors:  M C Damask; C Weissman; J Askanazi; A I Hyman; S H Rosenbaum; J M Kinney
Journal:  Anesthesiology       Date:  1982-09       Impact factor: 7.892

7.  Capnography and the Bain circuit I: A computer model.

Authors:  J E Beneken; N Gravenstein; J S Gravenstein; J J van der Aa; S Lampotang
Journal:  J Clin Monit       Date:  1985-04
  7 in total
  8 in total

1.  The relationship between anesthetic potency (minimum alveolar concentration) and molecular shape; structural studies on conventional inhalational anesthetics.

Authors:  Y Shiraishi; Y Fujise; K Ikeda; Y Takahashi; T Akagi; S Sasaki; Y Nozue
Journal:  J Anesth       Date:  1991-10       Impact factor: 2.078

2.  Oxygen uptake and carbon dioxide elimination during epinephrine-induced arrhythmias in humans.

Authors:  Y Shiraishi; Y Nakajima; J Yokoyama; K Ikeda
Journal:  J Anesth       Date:  1992-04       Impact factor: 2.078

3.  Frequently repeated Fick cardiac output measurements during anesthesia.

Authors:  M Doi; K Morita; K Ikeda
Journal:  J Clin Monit       Date:  1990-04

Review 4.  Non-steady state monitoring by respiratory gas exchange.

Authors:  P H Breen; S A Isserles; U Z Taitelman
Journal:  J Clin Monit Comput       Date:  2000       Impact factor: 2.502

5.  New metabolic lung simulator: development, description, and validation.

Authors:  Abraham Rosenbaum; Christopher Kirby; Peter H Breen
Journal:  J Clin Monit Comput       Date:  2007-03-01       Impact factor: 2.502

Review 6.  Open-circuit respirometry: real-time, laboratory-based systems.

Authors:  Susan A Ward
Journal:  Eur J Appl Physiol       Date:  2018-05-04       Impact factor: 3.078

Review 7.  Degenerative cervical myelopathy - update and future directions.

Authors:  Jetan H Badhiwala; Christopher S Ahuja; Muhammad A Akbar; Christopher D Witiw; Farshad Nassiri; Julio C Furlan; Armin Curt; Jefferson R Wilson; Michael G Fehlings
Journal:  Nat Rev Neurol       Date:  2020-01-23       Impact factor: 42.937

8.  Comparative carbon dioxide output through injured and noninjured peritoneum during laparoscopic procedures.

Authors:  T Kazama; K Ikeda; Y Sanjo
Journal:  J Clin Monit Comput       Date:  1998-04       Impact factor: 2.502

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.