| Literature DB >> 441578 |
Abstract
Metabolic gas exchange and respired minute volume were measured on 37 occasions in 9 normal adults on exercise and at rest, using the concentration of an injected gas, argon, as a measure of flow. These measurements were compared with simultaneous Douglas bag collections in which volumes were determined with a dry gas meter. The two types of measurements agreed within +/- 5%. It was concluded that this technique, designed for the remote monitoring of unconscious patients was suitable for a wide variety of clinical and experimental situations.Entities:
Mesh:
Substances:
Year: 1979 PMID: 441578 DOI: 10.1016/0034-5687(79)90029-x
Source DB: PubMed Journal: Respir Physiol ISSN: 0034-5687