Validation of intracardiac shunt using thoracic bioimpedance and inert gas rebreathing in adults before and after percutaneous closure of atrial septal defect in a cardiology research unit: study protocol.

Laura Filaire1, Aurelie Chalard2, Hélène Perrault3,4, Romain Trésorier2, Jean-René Lusson2, Bruno Pereira5, Frederic Costes6,7, Claire Dauphin2, Ruddy Richard4,6,7.   

Abstract

INTRODUCTION: Intrathoracic shunt quantification is a major factor for appropriate clinical management of heart and pulmonary diseases. Intracardiac shunts quantified by pulmonary to systemic output ratio (Qp/Qs) are generally assessed by Doppler echocardiography, MRI or catheterisation. Recently, some authors have suggested the concomitant use of thoracic bioimpedance (TB) and inert gas rebreathing (IGR) techniques for shunt quantification. The purpose of this study is to validate the use of this approach under conditions where shunt fraction is directly quantified such as in patients with isolated atrial septal defect (ASD). METHODS AND ANALYSIS: This trial is a prospective, observational single-centre, non-blinded study of adults seen for percutaneous closure of ASD. Qp/Qs ratio will be directly measured by Doppler echocardiography and direct Fick. IGR and TB will be used simultaneously to measure the cardiac output before and after closure: the ratio of outputs measured by IGR and TB reflecting the shunt fraction. The primary outcome will be the comparison of shunt values measured by TB-IGR and Doppler echocardiography. ETHICS AND DISSEMINATION: The study has been approved by an independent Research Ethics Committee (2017-A03149-44 Fr) and registered as an official clinical trial. The results will be published in a peer-reviewed journal. TRIAL REGISTRATION NUMBER: NCT03437148; Pre-results. © Author(s) (or their employer(s)) 2019. Re-use permitted under CC BY-NC. No commercial re-use. See rights and permissions. Published by BMJ.

Entities:  

Keywords:  atrial septal defect (ASD); cardiac output; inert gas rebreathing; shunt quantification; thoracic bioimpedance

Year:  2019        PMID: 31133575      PMCID: PMC6538205          DOI: 10.1136/bmjopen-2018-024389

Source DB:  PubMed          Journal:  BMJ Open        ISSN: 2044-6055            Impact factor:   2.692


The validation of a double non-invasive approach for assessing the extent of cardiac shunt will provide an alternative to catheterisation. The double non-invasive approach is used concomitantly to an arteriovenous catheterisation. The advantage of the bioimpedance method for stroke volume determination is that it is continuous and does not require patient participation. A limitation of inert gas rebreathing resides in the lack of control on patient rebreathing manoeuvres.

Introduction

Intrathoracic shunt quantification is a major parameter for the clinical management of cardiac and pulmonary diseases. Several techniques are currently available for quantification but present some limitations: the direct Fick method is invasive and therefore not first indication, easy access to MRI remains an issue in several clinical centres while the precision and repeatability of Doppler echocardiography measurements is highly dependent on technical ability and experience. Recently, Perrault et al 1 suggested a new approach of intrathoracic shunt quantification by using concomitantly thoracic bioimpedance (TB) and inert gas rebreathing (IGR). TB allows the determination of the systemic cardiac output () and IGR enables the assessment of the pulmonary blood flow () such that the difference between measured by TB and determined from the IGR measurement may be taken to reflect the intrathoracic shunt. The extent to which the use of these combined non-invasive techniques provides quantification that is sensitive enough to be of use in the clinical management of patients with intrathoracic shunts remains to be validated. Isolated atrial septal defect (ASD) is a congenital heart disease leading to intracardiac left-to-right shunts of varying degrees depending on size of defect and left ventricular compliance.2 Clinical management generally entails assessment of ratio since the clinical decision for percutaneous closure is highly dependent on the magnitude of the left-to-right shunt.3 4 The comparison of shunt quantification using the combined TB and IGR techniques with that obtained through routine measurements in isolated ASD provides an opportunity to validate the usefulness of this approach in clinical settings. The aim of this study is thus to validate the combined non-invasive approach of ratio measurement using TB and IGR in adults with secundum ASD by comparing results obtained with those from usual methods of Doppler echocardiography and direct Fick.

Methods and analysis

Setting

The present trial is a prospective, single-centre, observational, non-blinded study that compares the measurement of the pulmonary to systemic blood flow ratio () by the concomitant use of IGR and TB first to the non-invasive Doppler echocardiography before and after defect closure and, second, to the direct Fick method during the defect closure procedure.

Study objectives

The aim of the study is to validate the combined non-invasive ratio measurement in adults with ASD both before and after interventional closure in comparison to the Doppler echocardiography measure and during the closure procedure to the direct Fick method.

Methods

Methodology and design

The study will take place in the cardiology unit of the University Hospital Center, with the view to include 30 patients over a period of 2 years from inclusion of first patient. Patients will be enrolled and followed by cardiologists who will perform right heart catheterisation and Doppler echocardiography. The non-invasive measurement of using IGR and TB will be made before and after closure by trained clinical investigators without knowing the results obtained by cardiologists at the time of data acquisition.

Patients and public involvement

This research was designed by clinicians with the aim of simplifying patient’s management. Patients were not involved in the recruitment to or conduct of the study. According to the law in force in France, participants will be individually informed of the results at the end of the study.

Inclusion criteria

Patients with the following conditions will be included: Adult patients (≥18 years old) with confirmed secundum ASD with indication for interventional closure.5 ASD regardless of the size, with suspicion of paradoxical embolism. Patients with significant intracardiac shunt (right ventricular volume overload) and pulmonary vascular resistance (PVR) <5 Wood units (WU). Patients with PVR ≥5 WU but <2/3 systemic vascular resistance or pulmonary arterial pressure <2/3 systemic pressure and evidence of net left-to-right shunt ( >1.5).

Exclusion criteria

Patients will be excluded for the study if they present with one of the following criteria: Pregnant and breastfeeding women. Patient under assisted ventilation. Chronic respiratory disease ventilation/perfusion mismatch abnormalities. Complex congenital heart disease. Patients with coagulation function abnormalities. Patients who have not provided written consent. Patients in whom the use of TB or IGR is contraindicated.

ASD closure

Patients with secundum ASD eligible for percutaneous closure will be invited to participate in the study and to provide their signed informed consent. The percutaneous closure will be performed during a 3-day hospitalisation as per the usual institutional protocol. Resting haemodynamic measurements of shunt flow (), systemic cardiac output () and right and left ventricular functions will be performed using Doppler echocardiography the day before as well as the day after percutaneous closure. Concomitantly, pulmonary blood flow by IGR () and the systemic cardiac output by TB () will be measured. Right heart catheterisation will also be performed to directly measure and through a direct Fick computation (). An ‘Amplatzer Septal Occluder (Saint Jude)’ will be used for ASD closure. The sequence of procedures is shown in figure 1.
Figure 1

Consort diagram: flow chart. Patients indicated to ASD percutaneous closure will be proposed to the study. Three assessments of the ratio with the concomitant use of IGR and TB are planned during a 3-days hospitalization: the day before and after intervention in comparison to the Doppler Cardiac Ultrasound measure and one the day of intervention in comparison to the direct Fick method. ASD: atrial septal defect.

Consort diagram: flow chart. Patients indicated to ASD percutaneous closure will be proposed to the study. Three assessments of the ratio with the concomitant use of IGR and TB are planned during a 3-days hospitalization: the day before and after intervention in comparison to the Doppler Cardiac Ultrasound measure and one the day of intervention in comparison to the direct Fick method. ASD: atrial septal defect.

Study group

In this study, subjects serve as their own control for all relevant measures.

Approaches for cardiac output determination and shunt fraction quantification

Doppler echocardiography

Cardiac output measurement will be obtained using standard cardiac ultrasound (Ge-Vid S9, Philips-IE 33 and Philips-EPIC). Images will be captured in parasternal (short and long axes) and apical four-chamber views with 2D cardiac ultrasound with patient in left lateral decubitus. Cross-sectional area of the pulmonary annulus (Ap) and aortic annulus (Aa) will be measured.6 Velocity of blood will be assessed by pulsed Doppler. Velocity at the pulmonary annulus and aortic annulus is plotted on time to obtain the pulmonary and systemic velocity time integral (VTIp and VTIa, respectively). Pulmonary and systemic cardiac output measurements are based on the following formulae.7 The echocardiographic shunt fraction is estimated from ratio with values from three non-continuous cardiac cycles being used for computation. In turn, the computed values will serve as reference for statistical analyses.8

Fick quantification of shunt

Under local anaesthesia and after pressure and ECG monitoring, a catheter will be inserted through the right femoral vein allowing blood gas samples at various sites and then pushed through the ASD for pulmonary vein sampling to enable oxygen saturation comparisons and left-to-right shunt contribution. The standard Fick equation will be applied for computation of cardiac output to include or exclude shunted blood fractions. Systemic cardiac output will be taken as: =whole body oxygen consumption. = arterial oxygen content in the femoral artery. =mixed venous oxygen content obtained from the oxygen saturation and haemoglobin concentration in inferior vena cava (IVC) and superior vena cava (SVC) especially for patients with ASD.9 Shunt contribution will be obtained through a series of Fick computations using successive venous content values of samples in the IVC, the SVC, the innominate vein, the right atrium and ventricle and the pulmonary artery. The oxygen content measured at each of these sites will be introduced into the Fick equation to determine the corresponding cardiac output, reflecting the absence or presence of blood shunt contribution.

Double non-invasive cardiac output measurement

This approach will be based on the use of two cardiac output determination methods simultaneously, leading to intrathoracic shunt quantification: TB measurement reflects  while IGR reflects (figure 2).
Figure 2

Illustration of the thoracic cardiovascular circulation in Atrial septal defect (arrow). The blue frame represents the pulmonary circulation measured with the IGR (Innocor®) and the red frame the systemic circulation measured with the TB (Physioflow®). 1. Right side of the heart; 2. Vena cava; 3. Pulmonary artery; 4. Trachea; 5. Lung; 6. Pulmonary Veins; 7. Left side of the heart; 8. Aorta.

Illustration of the thoracic cardiovascular circulation in Atrial septal defect (arrow). The blue frame represents the pulmonary circulation measured with the IGR (Innocor®) and the red frame the systemic circulation measured with the TB (Physioflow®). 1. Right side of the heart; 2. Vena cava; 3. Pulmonary artery; 4. Trachea; 5. Lung; 6. Pulmonary Veins; 7. Left side of the heart; 8. Aorta.

Thoracic bioimpedance

The methodology of PhysioFlow (Manatec, France) has been well described in the study of Charloux et al.10 TB was validated against the Fick method in healthy subjects11 and in patients with chronic heart and respiratory diseases.10 12–14 Two sets of cutaneous electrodes are placed on the thorax of the subject to reflect changes in bioimpedance. The changes in thoracic impedance (ΔZ) occurring as a result of cardiac ejection are continuously recorded and reflect the stroke volume (SV). Systemic cardiac output is then determined from the multiplication of SV and heart rate. The value is obtained as the average of continuous data over the last 30 s (approximately 25–30 heartbeats) immediately prior to the rebreathing manoeuvres.

Inert gas rebreathing

IGR is validated against invasive method Fick and thermodilution for cardiac output measurement in both healthy subjects and patients with heart disease.15–18 The Innocor (Innovision, Odense, Denmark) device, which will be used in this study, measures the  using the rebreathing technique with a mixture of gases (sulfur hexafluoride (SF6), nitrous oxide (N2O) and oxygen (O2)). SF6 is an insoluble inert gas which is used as a marker of gas mixture distribution homogeneity. N2O is a blood soluble gas which serves as gas tracer. Its disappearance from the lung is proportional to the rate of pulmonary blood flow.19 20 Rebreathing procedures will be achieved at a breathing rate of 20/min, each separated by 3 min. Each rebreathing manoeuvre will last 15–20 s which represents a total of 25–30 cardiac cycles and will result in a cardiac output value. The average of the three cardiac output measurements will be used as mean for statistical analyses. Under normal physiological conditions,  is equivalent to . The absolute difference between  and  quantifies the intrathoracic shunt. The qualitative aspect of the intrathoracic shunt with a shunt fraction ratio such that ratio >1.0 reflects a left-to-right shunt. The severity of the shunt is expressed by the value of the ratio, such that higher shunt ratios reflect greater severity.

Study outcomes

Primary outcome

The main outcome of the study is the difference between ratio measured by the double non-invasive technique () when compared with the Doppler echocardiography ratio taken before interventional closure of secundum ASD. The working hypothesis is that there will be no significant difference between these ratios.

Secondary outcomes

The secondary outcomes are: The ratio measured by the double non-invasive technique () in comparison to the invasive direct Fick method. The ratio measurements obtained after ASD closure by the double non-invasive technique () to those measured using Doppler echocardiography (). Relationship between extent of shunt assessed by Fick, echocardiography and double approach and cardiac output measurements.

Study limitations

The study presents some methodological limitations: First, there is a lack of control on patient rebreathing manoeuvre with the IGR. Second, some patient’s factors such as preload or anxiety might influence physiological left-to-right shunt volumes before, during or after catheterisation to varying degrees. This bias becomes irrelevant after shunt closure, if shunt closure is completed. Third, the difference of time periods of data acquisition between the methods of cardiac output measurement (lower number of heart cycles captured by echocardiography, as compared with the other monitoring methods averaging more heart and respiratory cycles) might introduce a bias in shunt determination.

Statistical considerations

Estimation of sample size

To evaluate the agreement between the double non-invasive approach and the gold standard technique, a sample size of 30 subjects would provide a power of 90% to highlight an agreement of 0.8 (Lin’s concordance coefficient) for a two-sided type I error at 5%. A sequential exploratory analysis will be proposed each 10 patients without correction of type I error.

Statistical analysis

All statistical analyses will be performed using Stata software (V.13, StataCorp, College Station, USA). A two-sided type I error will be set at 0.05 to indicate statistical significance. Continuous data will be presented as mean±SD or as median (IQR), according to the statistical distribution. The assumption of normality will be studied with the Shapiro-Wilk test. The primary analysis will be performed using correlation coefficient (Pearson or Spearman according to the statistical distribution), Lin’s concordance coefficient and Bland and Altman plot in order to study the accuracy between non-invasive measurement and Doppler echocardiography in adults with ASD. The correlation and concordance coefficients will be presented with 95% CIs.

Secondary analysis

In addition, the  obtained using each method will be examined regardless of the size of ASD using correlation coefficient and analysis of variance (ANOVA) or Kruskal-Wallis test (if the assumptions of ANOVA are not met: normality and homoscedasticity studied by Bartlett test). The same statistical analyses will be carried out with respect to direct Fick measured shunt ratios. Finally, the intraindividual and interindividual reproducibility of the double non-invasive approach will be evaluated by intraclass correlation coefficient (ICC estimated from a mixed model with patient as random effect). ICC will be presented with a 95% CI.

Ethics and dissemination

Approval

According to the French regulation on clinical trials, the study has been submitted to the ‘Comité de Protection des Personnes Ouest V’ (Reference 2017-A03149-44) and to the ‘Agence Nationale de Sécurité du Medicament (ANSM)’ (the French regulatory authority for research). Approval from the Ethics Review Board is dated 13 February 2018 and from the ANSM 30 January 2018. Any modification in the protocol or informed consent during the study will be presented to the reference authority. The study is currently registered on the clinical website under the number NCT03437148.

Patient informed consent

All patients will receive verbal and written information on the aim of the study and the protocol. Written informed consent will be obtained prior to their participation in the study. During the study, the patients will have the possibility to ask all questions concerning the protocol from the cardiologist or investigator. They will be informed that they are free to stop the study at any time at their own discretion.

Data collection and quality management

Data will be collected by the principal investigator and the trained clinical research assistant. Data will be registered in written notebooks at each assessment point for each patient. Data capture will be achieved using Research Electronic Data Capture. A clinical research assistant will be commissioned to ensure the progress of the study, and the data capture according to the standard operating procedures implemented at the University Hospital of Clermont-Ferrand in accordance with the Good Clinical Practice in current French laws.

Access to data

The data set will be the property of the institution. However, the principal investigator and the project manager will have full access to the final data set. The results will be communicated in a peer-reviewed journal, will be presented at an international conference and will appear in ClinicalTrials.gov.

Discussion

To the best of our knowledge, our study is the first to assess the ratio using a combined non-invasive approach of ratio measurement using TB and IGR. Cardiac output is a valuable physiological measurement to provide insight into integrated cardiocirculatory systemic functions at play to refine diagnosis and capability for several chronic disease states.18 To date, accurate measurement of cardiac output remains challenging in everyday clinical environments. Over the last decade, TB has been introduced which provides an interesting alternative as it allows to monitor values of continuous periods.10 11 This approach, however, may be limiting with respect to identifying right and left heart contributions. Rebreathing methodologies for cardiac output determinations provide such an opportunity. Peyton et al demonstrated the value of IGR to calculate the ratio Qp/Qs by combining the shunt equation and Fick principle. The limits of their approach stand in the impossibility to extend the application to intracardiac shunt.21 The aim of this study is to demonstrate the non-inferiority of the combined non-invasive approach of  ratio measurement using TB and IGR in adults with secundum ASD by comparing the results obtained with those from usual methods (Doppler echocardiography and direct Fick). In this context, Perrault et al 1 used TB and IGR simultaneously to measure the cardiac output either by TB, IGR or CO2 rebreathing (with or without correction by PaCO2) in patients with chronic obstructive pulmonary disease at rest and exercise. The ability of TB and IGR to correctively measure Qs and Qp, respectively, was confirmed. Interesting results lie on the possibility to estimate the shunt effect due to underestimation of cardiac output by IGR in comparison to TB. The added value of the double non-invasive approach lies in the possibility to quantify and qualify the intrathoracic shunt from physiological haemodynamic parameters with relatively minor discomfort to the patient. The simultaneous use of IGR and TB is non-operator dependent, totally non-invasive and easy to use contrary to the Doppler echocardiography or the Fick methodology. In addition, the combined technique can be successfully used during clinical cardiopulmonary exercise testing enabling early or refined diagnoses or for follow-up and clinical management.5 22 23 Such an approach enables to quantify or qualify the shunt on exercise or shunt reversal and cyanosis which may first appear under exercise.24 Thus, the first step of this process is to evaluate the concomitant use of TB and IGR at rest in intracardiac shunt in comparison to standard methods.
  23 in total

1.  Noninvasive assessment of cardiac output in children using impedance cardiography.

Authors:  D S Braden; L Leatherbury; F A Treiber; W B Strong
Journal:  Am Heart J       Date:  1990-11       Impact factor: 4.749

2.  Noninvasive measurement of cardiac output during exercise by inert gas rebreathing technique: a new tool for heart failure evaluation.

Authors:  Piergiuseppe Agostoni; Gaia Cattadori; Anna Apostolo; Mauro Contini; Pietro Palermo; Giancarlo Marenzi; Karlman Wasserman
Journal:  J Am Coll Cardiol       Date:  2005-10-10       Impact factor: 24.094

3.  Effects of ventilation on cardiac output determined by inert gas rebreathing.

Authors:  Morten Damgaard; Peter Norsk
Journal:  Clin Physiol Funct Imaging       Date:  2005-05       Impact factor: 2.273

4.  Validation of the use of foreign gas rebreathing method for non-invasive determination of cardiac output in heart disease patients.

Authors:  Liang Dong; Jian-an Wang; Chen-yang Jiang
Journal:  J Zhejiang Univ Sci B       Date:  2005-12       Impact factor: 3.066

5.  Reliability of measurements with Innocor during exercise.

Authors:  P Fontana; U Boutellier; M Toigo
Journal:  Int J Sports Med       Date:  2009-07-29       Impact factor: 3.118

6.  Influence of sampling site and flow area on cardiac output measurements by Doppler echocardiography.

Authors:  H Dittmann; W Voelker; K R Karsch; L Seipel
Journal:  J Am Coll Cardiol       Date:  1987-10       Impact factor: 24.094

7.  Accuracy of electrical impedance cardiography for measuring cardiac output in children with congenital heart defects.

Authors:  D S Miles; R W Gotshall; J C Golden; D T Tuuri; R H Beekman; T Dillon
Journal:  Am J Cardiol       Date:  1988-03-01       Impact factor: 2.778

8.  Noninvasive measurement of intrapulmonary shunting.

Authors:  Philip J Peyton; Gavin J B Robinson; Peter R McCall; Bruce Thompson
Journal:  J Cardiothorac Vasc Anesth       Date:  2004-02       Impact factor: 2.628

9.  Addressing Assumptions for the Use of Non-invasive Cardiac Output Measurement Techniques During Exercise in COPD.

Authors:  Hélène Perrault; Ruddy Richard; Sophia Kapchinsky; Jacinthe Baril; Jean Bourbeau; Tanja Taivassalo
Journal:  COPD       Date:  2015-09-25       Impact factor: 2.409

10.  The benefit of atrial septal defect closure in elderly patients.

Authors:  Monika Komar; Tadeusz Przewlocki; Maria Olszowska; Bartosz Sobien; Piotr Podolec
Journal:  Clin Interv Aging       Date:  2014-07-16       Impact factor: 4.458

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.